Microspheres can break the diffraction limit and magnify nano-structure imaging, and with its advantages of low cost and label-free operation, microsphere-assisted imaging has become an irreplaceable tool in the life sciences and for precision measurements. However, the tiny size and limited imaging field of traditional solid microspheres cause difficulties when imaging large sample areas. Alternatively, droplets have similar properties to those of microspheres, with large surface curvature and refractive-index difference from the surrounding environment, and they can also serve as lenses to focus light for observation and imaging. Previous work has shown that droplets with controllable size can be generated using an optical tweezer system and can be driven by optical traps to move precisely like solid microspheres. Here, a novel microdroplet-assisted imaging technology based on optical tweezers is proposed that better integrates the generation, manipulation, and utilization of droplets.
ARTICLE HIGHLIGHTS
•Controllable droplet generation in an optical tweezer system by using photogenerated droplets.
•The generated droplets can aid imaging in the same way as can solid microspheres but with larger field of view and higher imaging quality.
•Droplets can be cleaned easily from flexible samples without causing mechanical damage.
I. INTRODUCTION
Microsphere nanoscopy emerged in 2011, and as a simple technology, microsphere imaging is still based on optical principles that allow direct magnification and imaging of nanostructures at scales below the diffraction limit using a conventional optical microscope.1,2 With intensive research, several teams have developed various microsphere-based methods for optical super-resolution. By simply placing the transparent microspheres on the surface of the sample, the minimum resolution of the microscope can reach 50 nm.3 To overcome their tiny field of view (FOV) and inconvenient maneuverability, the microspheres can scan a large area of the sample via the high-precision manipulation capability of an atomic force microscope (AFM)4,5 or an optical tweezer (OT) system,6,7 thus enabling imaging of a larger area of the sample. However, these approaches still require the scanning and stitching of microsphere images, and influenced by spherical distortion, microsphere imaging shows obvious distortion and brightness unevenness, which makes image stitching significantly more difficult.8
In previous studies, we have shown a precisely controllable way to generate photogenerated droplets in a specific mixed solution by using an OT system.9 The generated droplets have a higher refractive index than the surrounding solution and thus achieve similar properties to those of microspheres.10 Consequently, the droplets can also be manipulated by optical traps to move randomly.11,12 Through further research, we found that the essence of droplet generation is the action of laser traps in multiple physical fields:13 the “trapping” process dominated by gradient forces and the “convergence” process caused by photothermal and thermal acceleration effects.14,15 This indicates that droplet photogeneration can occur in not only a mixture of isopropyl alcohol (IPA) and phosphate-buffered saline but also other solutions with similar properties. Also, the generated droplets have ideal light transmission and can gather light as can solid microspheres but while being softer and more plastic, and they will not scratch the sample surface, similar to a cell biomagnifier.16
Indeed, some researchers have already used OTs to manipulate lipid droplets in order to detect the fluorescence signals of other structures in the whole cellular environment,17 leading to the expectation that liquid droplets can give rise to harmless and compatible observation technology with nanoscale spatial resolution. However, because lipid droplets can only be used in biological environments and it is difficult to change their size, thereby preventing taking advantage of the mutual solubility of droplets, herein we use an OT system to generate controlled droplets efficiently and liberally in a mixture of water, IPA, and silicone oil (40 cs, water-soluble), and we use them to assist in the magnification of samples. In this way, we integrate droplet generation, transport, and utilization in the OT system, offering a preliminary verification of the feasibility of microdroplet-assisted imaging. Furthermore, photogeneration makes the preparation and use of droplets more convenient, being simpler and faster than the cantilever structure and other non-contact microsphere imaging solutions. With excellent controllability and wide versatility, this method has great potential for applications such as harmless detection and nanostructure measurement.
II. EXPERIMENTAL WORK
A conventional transmission-based OT system uses objective and condenser lenses opposed along the axis, which requires high-precision focal alignment of the illumination with the optical trap. Consequently, the overall thickness of the sample chamber is limited by the short working distances of the objective and condenser lenses, and the observed sample and the sample chamber must have high transmittance to ensure sufficient illumination for imaging. Therefore, in the present work, a reflective OT system using only one objective lens is chosen to observe non-transparent samples and to achieve a simplified system structure as shown schematically in Fig. 1.
To produce a stable and reliable optical trap and obtain a clear resolution, the chosen laser is a single-frequency linearly polarized ytterbium fiber laser from IPG Photonics (YLR-5-LP, 1064 nm, 5 W), and the objective lens is a high numerical aperture (NA) objective (ZEISS, NA:1.2, 63×). The main video camera (Manta G-507, monochrome) is mounted on a linear guide to separate the imaging of the sample from the optical trap. Kohler illumination is used to generate a bright and extremely even illumination of the sample chamber to avoid interferences such as lamp filament and shadows, in the resulting image. Besides, several 450-nm bandpass filters are used to prevent chromatic dispersion resulting in reduced imaging quality.
For stable droplet generation, a custom microfluidic chip and an air pressure-driven injection system are used. The custom microfluidic chip uses a metal substrate and is capped with a cover slip and UV-curable adhesive to ensure maximum light transmission for the optical trap manipulation and the sample observation.18 The height of the microfluidic channel can be adjusted in the range of 200–1000 μm, ensuring the ability to load samples with different thicknesses. The air pressure can be adjusted with an accuracy of 400 Pa, and the injection rate of droplet samples can be controlled between 10 μL/min and 500 mL/min under constant air pressure. Because of the high concentration of the original silicone oil, it must be diluted with pure water to disperse into nano-droplets in IPA and distribute evenly. Therefore, the original solution is mixed with silicone oil, water, and IPA at a volume ratio of 1:5:10. See the supplementary material for the detailed photogeneration process and the droplet characteristics.
Because droplet photogeneration involves the aggregation and fusion of nano-scale droplets, the generated droplets have a higher refractive index than the surrounding solution.19 This also means that the OT system can manipulate the droplets to achieve precise displacement, as with conventional microspheres. In addition, with the ideal light transmission and refractive-index differences, droplets can also concentrate light and magnify the sample for imaging. Compared with solid microspheres, the size of droplets can be adjusted by changing the optical trap power, droplet generation time, or original solution concentration during the generation process. This means that it is possible to change the FOV, depth of field (DOF), contrast, or other imaging characteristics by generating droplets with different ingredients or shapes.20,21
For instance, Fig. 2 shows a typical droplet generation process. In step 1, the laser is turned on and the optical trap is placed above the sample surface. The injector is driven by the air pressure, and the injection rate is 200 μl/min. Within half a minute, the droplet diameter grows to greater than 5 μm under the optical trap, which is sufficient to replace solid microspheres for assisted imaging. Then in step 2, the height of the sample chamber is raised slightly so the droplet can be brought closer to the sample with nanostructures. The focusing surface of the camera is adjusted at the same time to allow precise focusing of the droplet image. Benefiting from the affinity of liquids to solids, the droplet is attached stably to the sample surface, but the contact surface with the surrounding solution still maintains a certain curvature under tension, determined by the properties of the original solution.22 Besides, the imaging FOV and DOF are also influenced by the droplet size and its surface curvature after attachment.23,24 See the supplementary material for a comparison of the imaging characteristics with solid microspheres.
In addition, liquid droplets can penetrate and fill tiny pits on the sample surface to a certain extent, obtaining stronger light reflection of the sample and effectively improving the imaging quality of the camera. Compared with solid microspheres, the generated droplets can prevent scratches on the sample surface and are easily cleaned off with pure water, making it possible to observe samples without damage or contamination.
III. THEORY AND DISCUSSION
It has been demonstrated that microspheres can be treated as micro-lenses and can be used to identify details of objects with sizes below the conventional diffraction limit.25 If microspheres are half immersed in solution, then the difference in the refractive index also leads to the convergence of light and thus magnifies the sample.26 Geometrically, a droplet adhering to a sample surface is propped up under the action of liquid tension, similar to a semi-submerged microsphere or a plano-convex solid immersion lens,27–29 and as described previously, with the high curvature of the contact surface and the refractive-index difference with the surrounding solution, the droplet can refract light like a microsphere with amplification ability.26 This indicates that the research methods used for solid microspheres are also applicable to the imaging properties of liquid droplets.
However, because of its tiny size and all-liquid environment, the thickness of the droplet after adhering to the sample surface is unknown and cannot be measured directly. However, given that (i) the droplet volume does not change during adhesion and (ii) only its shape changes under the balance of tension and sample affinity, the droplet thickness can be inferred from the droplet diameters as seen from the imaging camera before and after adhering to the sample surface.30
Based on these parameters, we simulated the focusing characteristics of the droplet in the COMSOL Multiphysics software. With irradiance being the main factor affecting the imaging of a microscale lens,32 we began by using wave optics simulation to find the focal point of the micro-droplet; in this simulation, the refractive indices of the silicone oil and surrounding solution were set to 1.5 and 1.3, respectively, and the results are shown in Fig. 4(a). The maximum light intensity is at 15 μm along the light propagation axis, which corresponds to 12.45 μm from the optical center. Because the surface curvature of the droplet after attachment is not as large as that of a microsphere, there is no need to consider the change in reflectivity of the light when it passes through the droplet, so this point of maximum intensity can be regarded as being the optical imaging focus of the droplet.33
IV. RESULTS
As typical observation samples for verifying super-resolution imaging, Blu-ray discs have 100-nm stripe-like track structures with constant and uniform periodicity, making them ideal samples for studying the imaging properties of droplets. However, under monochromatic illumination, the multi-stripe sample can produce artifacts parallel to the stripes under the effect of optical interference or diffraction, which interferes with the focusing process during camera imaging and affects the measurement accuracy. Therefore, to prevent artifact interference in the imaging of the data tracks, the Blu-ray disc was pre-burned before the experiment to generate tiny data points on the disc, effectively enhancing the accuracy and precision in the focusing process. With the data points being distributed non-uniformly on the tracks, dimension reduction processing was carried out along the tracks for all disc images to extract the data track period accurately. Furthermore, a silicon wafer with nanostructures was also used to observe droplet-assisted imaging.
According to the simulation and calibration measurements, the maximum imaging magnification of a single droplet is only 1.14×, which is less than the magnification of a solid polystyrene (PS) microsphere in water as shown in Fig. 5(b-ii) (see the supplementary material for more comparisons). This is due to the affinity of the droplet with the sample as it approaches the sample surface, resulting in the entire droplet adhering to the sample and the droplet structure changing from spherical to semi-parabolic, which greatly reduces the object distance u for the magnification imaging process. Also, the hydrophobicity of the sample surface greatly influences the microdroplet morphology during attachment. For example, the droplet shown in Fig. 5(a) did not exhibit strong affinity with the Blu-ray disc, and its shape did not spread much after attachment, maintaining a thickness equal to more than half of its diameter. However, after replacing the sample with a silicon wafer, the affinity between the droplet and the sample surface improved significantly, so that the droplet spread significantly after attachment, and both its thickness and surface curvature decreased significantly, resulting in the lower magnification of 1.05×.
The droplet–sample affinity could be changed by adjusting the concentration of the surrounding solution or the viscosity of the dimethyl silicone oil, thus increasing the droplet surface curvature. However, the concentration and viscosity have wide ranges for adjustment, thereby imposing a huge workload for detailed correlation research. Therefore, more studies and experiments are still needed to determine the internal mechanism and summarize the correlations, and these will be explored in future research.
V. CONCLUSIONS
Herein, we described a method for microdroplet-assisted imaging in an OT system. By utilizing the properties of photogeneration, the OT system can generate arbitrarily sized droplets and achieve different sizes of FOV for magnification imaging. Although droplets are yet to provide the same super-resolution imaging capability as solid microspheres, their controlled generation technology, non-mechanical damage, convenient cleaning process, and larger imaging FOV may offer new ideas for developing microsphere-assisted imaging technology. In addition, the magnification of droplet imaging can be improved further by changing the droplet viscosity or original solution, which is left for future exploration.
SUPPLEMENTARY MATERIAL
See the supplementary material for the detailed photogeneration process and the droplet characteristics.
ACKNOWLEDGMENTS
This work was supported by the National Natural Science Foundation of China (Grant Nos. 52075383 and 61927808).
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
DATA AVAILABILITY
The data that support the findings of this study are available within the article and its supplementary material.
REFERENCES
Cong Zhai is a Ph.D. student in the School of Precision Instrument and Opto-Electronics Engineering at Tianjin University. His research interests include optical tweezers and the development of nano/micro-scale sensors and instruments.
Chunguang Hu received a Ph.D. degree in 2007 and is currently a Professor in the School of Precision Instrument and Opto-Electronics Engineering at Tianjin University. His research interests include optical tweezers, optical characterization of nano/micro-structures on surfaces, nano/micro-scale sensors, and single-molecule technology for biology.