In microfluidic impedance cytometry, the change in impedance is recorded as an individual cell passes through a channel between electrodes deposited on its walls, and the particle size is inferred from the amplitude of the impedance signal using calibration. However, because the current density is nonuniform between electrodes of finite width, there could be an error in the particle size measurement because of uncertainty about the location of the particle in the channel cross section. Here, a correlation is developed relating the particle size to the signal amplitude and the velocity of the particle through the channel. The latter is inferred from the time interval between the two extrema in the impedance curve as the particle passes through a channel with cross-sectional dimensions of 50 μm (width) × 30 μm (height) with two pairs of parallel facing electrodes. The change in impedance is predicted using 3D COMSOL finite-element simulations, and a theoretical correlation that is independent of particle size is formulated to correct the particle diameter for variations in the cross-sectional location. With this correlation, the standard deviation in the experimental data is reduced by a factor of two to close to the standard deviation reported in the manufacturer specifications.
HIGHLIGHTS
Measurement of particle size through microfluidic-based impedance cytometry using parallel facing electrodes.
Development of an analytical formula for a rectangular channel via 3D COMSOL simulation.
Implementation of the analytical formula for accurate determination of particle size irrespective of value.
I. INTRODUCTION
Single-cell impedance cytometry is becoming a widely used noninvasive method for determining the properties of cells, such as their sizes, shapes, and dielectric properties.1 The size of a single cell was first determined by Coulter2 via DC resistance measurement. More recently, AC methods have been used to determine both the size and properties of the cell membrane, such as cytoplasm and subcellular components.3–8 In general, impedance cytometry of a cell is carried out in a microfluidic channel based on the impedance signal when the cell—suspended in a conductive solution—passes through microelectrodes fabricated on the inner walls of the channel, and accurate determination of particle size has been explored with different microelectrode configurations.9–12
The literature contains two different microelectrode configurations, i.e., (i) coplanar, where the electrodes are etched on one wall of the channel with rectangular cross section, and (ii) parallel facing, where the electrodes are located on opposite walls of the channel. A microfluidic chip with a coplanar configuration is a relatively cheap device that is easy to fabricate. However, as the cross-stream distance of the particle from the nearest electrode surface increases, the measured signal amplitude decreases, and this limits the range of particle size for a particular channel configuration; in a channel with a height of tens of micrometers, the minimum particle size that can be detected is ∼2 μm. By contrast, with parallel facing electrodes, the impedance depends on only the perpendicular distance of the particle from the nearest electrode surface, increasing as that distance decreases. However, the fabrication procedure is more complicated because it requires accurate alignment of the electrodes on opposite walls.13
Recently, the coplanar configuration was used to measure the diameters of submicrometer particles.1,14–16 De Ninno et al.14 demonstrated accurate size determination of particles with diameters of 5.2 μm and above by using five electrodes in a microchannel with cross-sectional dimensions of 40 μm (width) × 21 μm (height) and with the addition of floating electrodes; the impedance signal involved an additional parameter known as relative prominence, which was used to correct the particle diameter with a resolution of 0.3 μm, where the resolution is defined as the difference in particle size corresponding to twice the standard deviation of the distribution. Using double differential electrodes in the coplanar configuration, Zhong et al.15 measured particle size with a resolution of 0.1 μm, but the channel dimensions were restricted to 8 μm (width) × 10 μm (height), and such small channels are susceptible to clogging. Several studies have used parallel electrodes to determine particle diameters,1,17–19 and in most cases particle diameters of 3–10 μm were studied with a resolution of 1 μm.
Irrespective of the electrode configuration in the microchannel, the differential signal amplitude generated by a particle as it passes the electrodes depends on its perpendicular distance from them. Known as position dependency, this phenomenon results in significant errors in diameter measurements, and various approaches have been used to understand the position dependency of the signal amplitude and correct for it.14,16,17 Spencer et al.17 measured both transverse and oblique signal amplitudes simultaneously by using multiple pairs of electrodes to determine the transit times of particles, and the relative values of these transit times were used to correct for the position dependency in the direction perpendicular to the electrode surface. The signals measured using three pairs of coplanar electrodes were fitted using a bipolar Gaussian equation, and shape parameters obtained from this fit varied with both particle position and diameter, allowing correction for position dependency. However, because the coplanar configuration is more sensitive to measurement inaccuracy due to position dependency, using it to quantify smaller particle diameters is considered infeasible.
Herein, we propose a method involving three pairs of parallel facing electrodes to improve the estimation of particle diameter. By extracting both peak amplitude and travel time from the measured signal traces and the theoretically simulated curves, this method provides a simple strategy for correcting for particle position dependency. The novelty of the present work stems from the choice of travel time, for which a general strategy is introduced.17,20
II. EXPERIMENTAL METHODS
The experimental setup is shown schematically in Fig. 1(a). A microfluidic chip with dimensions of 5.4 mm × 5.4 mm with parallel facing electrodes was fabricated on a two-inch wafer using our in-house facilities. The top and bottom walls for the electrodes were made of Si/SiO2 and glass, respectively. As shown in Fig. 1(b), the configuration comprised a microfluidic channel with cross-sectional dimensions of 50 μm (width) × 30 μm (height). On the top and bottom walls of the channel were three pairs of electrodes, each electrode with a width of 30 μm and separated by an edge-to-edge distance of 30 μm. The electrical circuit is shown in Fig. 1(a), and the center electrodes on the top and bottom walls were grounded.
Complete setup for impedance measurements of beads suspended in phosphate-buffered saline solution.
Complete setup for impedance measurements of beads suspended in phosphate-buffered saline solution.
The Cr/Pt/Au electrodes (20 nm/150 nm/100 nm) were deposited on both the Si/SiO2 and glass substrates, and the desired electrode pattern as shown in Fig. 1(b) was fabricated using optical photolithography and ion milling. The channel fabrication and bonding were done using a photoimageable resist (Perminex 2015, USA) and a bonding chuck as per the manufacturer data sheet. See our previous work21 for the details of fabricating the sensor. Using an existing semiconductor facility, the electrical contacts were wire-bonded to a printed circuit board and connected to male header pins for direct coupling with the female sockets; this was possible by an innovative change in our design, with the top electrode contacts accessible through the bottom electrodes by filling the gap with fine solder paste (Chip Quik-SMD291SNL T7, USA). Side holes for fluid access were injected with transparent fluid connectors that were glued using UV glue (Bondic, USA) directly on the chip. A photograph of the fabricated lab-on-a-chip is shown in Fig. 1(c). The inlet of the fluid connector was then connected to a syringe pump via a Teflon tube (inner diameter: 0.3 mm; outer diameter: 0.8 mm).
Polystyrene beads with a density of 1.05 g/cm3, a DC conductivity of 6.6 × 10−4 S/m, and diameters of 2 μm, 3 μm, 4 μm, 5 μm, and 6 μm (Lab261, 1% solid) were used for the measurements. They were suspended in phosphate-buffered saline (PBS) with a DC conductivity of 1.6 S/m and diluted to a concentration of ∼200 beads per microliter. The carrier signal applied to the top electrodes had a frequency of 1.8 MHz and a peak-to-peak amplitude of 7.5 V. The currents from the bottom left and right electrodes were converted to a voltage difference using transimpedance operational amplifiers (THS 4303) in which the inverting inputs were grounded. The outputs of the two transimpedance amplifiers were connected to the input of a differential amplifier (ADA4927), which provided the voltage difference between the two electrodes. At constant voltage, the current is inversely proportional to the impedance, and the results herein are expressed in terms of the impedance. As a cell passes sequentially past the left and right electrodes, the difference in impedance is positive when the particle is between the upstream pair of electrodes and then becomes negative when it passes through the downstream pair.
The impedance measurements of the fast-moving cells suspended in the PBS solution required a high sampling rate. The fastest-moving cells in the channel with cross-sectional dimensions of 50 μm × 30 μm at a flow rate of 30 μl/min travelled at 0.7 m/s, and the time to cross the electrode zone of length 150 μm was ∼0.2 ms. Therefore, a sampling rate of 350 000 samples per second provided ∼75 measurements within the time taken for a cell to cross the electrode zone, which was sufficient to capture the signal with no distortion. The current lock-in amplifier designed by MicroX could generate three different frequencies in a single channel ranging from 10 kHz to 15 MHz and also had the in-built feature of real-time data processing for event counting by a custom-made algorithm programmed in the digital signal processor integrated in the lock-in board. The demodulated data from the lock-in amplifier were transferred to computer memory through a universal serial bus and were recorded in binary format. The frequency generation, low-pass-filter cut-off frequency, and sampling rate were controlled using an application programming interface made for a Linux-based system. The front-end electronics shown in Fig. 1 had precision resistors of 1 kΩ with a tolerance of less than 0.05%, and these were used to set the required gain of the differential amplifier.
III. THEORETICAL ANALYSIS
Electric-field simulations were carried out using the COMSOL 5.6 finite-element solver, and the analytical solution for a single pair of electrode was reported earlier.22 The potential V in the fluid and particle is governed by the Laplace equation ∇2V = 0. The boundary condition of zero potential gradient was applied at the top and bottom surfaces of the channel, i.e., n · ∇V = 0, because these surfaces are considered to be insulating; here, n is the unit normal to the surface. The boundary condition of constant potential was applied at the electrode surfaces. Because the boundary condition changes discontinuously at the junction between a bare channel wall and an electrode deposited thereon, the field lines have large curvature near the electrodes, and because of this, it was necessary to refine the mesh near the electrodes as shown in Fig. 2(b). In total, 2 × 106 mesh points were used in a typical simulation configuration of the type shown in Fig. 2(b). Distributions of potential and current density across the channel with an applied voltage of 1 V are shown in Fig. 2(c).
(a) Channel configuration showing (i) particle traveling in channel with cross-sectional dimensions of 30 μm × 50 μm, (ii) top and bottom electrodes, (iii) applied voltage, and (iv) output current. (b) Mesh used in typical 3D COMSOL simulations for a particle with a diameter of 3 μm, highlighting the higher meshing density at the electrodes. (c) and (d) Potential and current density distributions in xz plane of channel with an applied voltage of 1 V.
(a) Channel configuration showing (i) particle traveling in channel with cross-sectional dimensions of 30 μm × 50 μm, (ii) top and bottom electrodes, (iii) applied voltage, and (iv) output current. (b) Mesh used in typical 3D COMSOL simulations for a particle with a diameter of 3 μm, highlighting the higher meshing density at the electrodes. (c) and (d) Potential and current density distributions in xz plane of channel with an applied voltage of 1 V.
Simulation results for the change in current due to a particle with a diameter of 3 μm are shown in Fig. 3, where the impedance is shown as a function of downstream distance at different z locations in the direction perpendicular to the top and bottom walls. Even for the same particle size, the impedance curves differ significantly in shape depending on the z position of the particle. In each case, there are at least four extrema as the particle passes through the channel. When it passes through the center of the channel, the current changes relatively smoothly and can be fitted well by Gaussian curves. When the particle is close to a wall, there are two maxima in the current when the particle crosses the left and right edges of each electrode, and the amplitude deviates from single Gaussian behavior with two sharp peaks near the edges of the electrodes.17
Differential current ΔI versus particle position along x direction of microchannel for different z positions perpendicular to electrodes. The inset shows the maximum impedance at different z locations scaled by the impedance at the center of the channel at z = 0.
Differential current ΔI versus particle position along x direction of microchannel for different z positions perpendicular to electrodes. The inset shows the maximum impedance at different z locations scaled by the impedance at the center of the channel at z = 0.
To understand the position dependency of the peak amplitude, the maximum from each set of curves for different z values was extracted and then normalized by the impedance at the center of the channel at z = 0 [Z/Z0 ∝ ΔImax(z)/ΔImax(0)]. Here, the maximum of the time trace of the impedance is used for bimodal curves, and Z0 is the maximum impedance measured at z = 0. The fitted curve in the inset of Fig. 3(b) shows that the maximum amplitude changes exponentially as a function of vertical position.
A series solution can be obtained for the fluid velocity profile for the flow in a rectangular channel by using separation of variables. The velocity depends on the pressure gradient dp/dx along the channel:23
where −h/2 ≤ z ≤ h/2 is the direction perpendicular to the electrodes, and −w/2 ≤ y ≤ w/2 is that perpendicular to the flow direction and parallel to the electrodes. The maximum velocity at the center of the channel at y = z = 0 is
and the average velocity vav is
Although the velocity profile in 3D space is an inverted paraboloid (not shown here), for y = 0 it becomes a parabola as shown in Fig. 4(a), which can be fitted using
At the experimental flow rate of 30 μl/min, the value of vmax at the center was 0.68 m/s, and for this maximum velocity, the coefficients a1, a2, and a3 for the parabolic fit in Fig. 4(a) are 0.68 m/s, −1.13 × 10−5 s−1, and −0.003 m−1 s−1, respectively.
(a) Velocity profile for a 30 μm (height) × 50 μm (width) channel obtained from COMSOL simulation (verified using analytical solution). (b) Normalized impedance as a function of velocity normalized with maximum velocity at y = z = 0.
(a) Velocity profile for a 30 μm (height) × 50 μm (width) channel obtained from COMSOL simulation (verified using analytical solution). (b) Normalized impedance as a function of velocity normalized with maximum velocity at y = z = 0.
Because the particle position and size are unknown, the velocity is used as a parameter for correcting the particle size. The velocity is inferred from the time taken by the particle to travel from one electrode pair to the next. Although the maximum impedance is independent of the y position, the maximum velocity in a horizontal plane at a fixed z location does depend on the horizontal cross-stream coordinate y. Here, the maximum velocity vmax at the channel center at y = z = 0 is used as the normalizing factor.
The variation of Z/Z0 versus vx/vmax at y = 0 is shown in Fig. 4(b), and the following provides a good fit for this curve:
where d is the corrected diameter, D is the electrical diameter, and c1, c2, and c3 are the fitting constants. The curve fitting was done using cftool in MATLAB, using the linear least-squares method known as the Levenberg–Marquardt algorithm to obtain a minimum R2 = 99%.
For the present experimental system, the electrical diameter D is defined as
where G is the amplification factor to account for the current-to-voltage converter board and other electronic peripherals to be calibrated from experimental data. The value of G was estimated as 73.5 μm (nA)(−1/3) using the assumption that the true diameter and the electrical diameter are equal for particles moving at the center of the channel with maximum velocity. Equation (5) allows the corrected diameter to be determined once the values of D and the particle velocity vx are known.
IV. RESULTS AND DISCUSSION
Figure 5 shows time traces of impedance measurements with sampling at intervals of 3.3 μs for particles with a diameter of 3 μm. Figure 5(a) shows that the peak amplitude of Z/Z0 is in the range of 1%–1.75%, which corresponds to variation of the electrical diameter D by 20%. This agrees well with the theoretical prediction as shown in the inset of Fig. 3. Figures 5(b)–5(e) show expanded time traces of the impedance as a single particle travels through the channel. The velocities inferred from the time interval between two extrema are 0.54 m/s, 0.49 m/s, 0.36 m/s, and 0.28 m/s, respectively, and the values of the electrical diameter D derived using Eq. (5) are 3.05 μm, 3.35 μm, 3.45 μm, and 3.71 μm, respectively. From these figures, it is clear that the shapes of the impedance trajectories of particles moving with different velocities corresponding to different z positions agree qualitatively with the theoretical predictions shown in Fig. 3. As the velocity decreases, the particles are located closer to the electrodes, and the impedance trajectory has a bimodal Gaussian shape.
(a) Normalized impedance Z/Z0 versus number of events. Here, the value of Z0 is taken as 1.23 × 106, which corresponds to the maximum velocity of the particle, vmax = 0.68 m/s. Circles (positive maxima) and stars (negative maxima) show the events selected for further data processing. (b)–(e) Enlarged traces of single events obtained for a particle moving with different velocities.
(a) Normalized impedance Z/Z0 versus number of events. Here, the value of Z0 is taken as 1.23 × 106, which corresponds to the maximum velocity of the particle, vmax = 0.68 m/s. Circles (positive maxima) and stars (negative maxima) show the events selected for further data processing. (b)–(e) Enlarged traces of single events obtained for a particle moving with different velocities.
Figure 6(a) is a scatter plot of velocity versus electrical diameter D for particles with a diameter of 2 μm. This shows clearly that the instrument resolution was ∼1.7 μm, and there was electrical noise for electrical diameters below this value. The maximum measured particle velocity vmax was ∼0.67 m/s, which is very close to the theoretically predicted value of ∼0.68 m/s for the aforementioned channel dimensions and flow rate. The increase of D signifies the position dependency of the particle in the z direction as shown in Fig. 6(c). Using Eq. (5), the corrected particle diameter d was derived for the whole population of data; the corresponding histogram plot is shown in Fig. 6(b), and a color density plot of velocity versus d is shown in Fig. 6(d). These show a significant narrowing of the distribution and a reduction in the coefficient of variation, with the particle size now independent of velocity. The values of the coefficients c1, c2, and c3 used in Eq. (5) were obtained from the theoretical results given in Table I.
(a) Scatter plot of velocity versus electrical diameter for 2-μm particles from separate runs, showing clearly the instrument resolution of up to 1.7 μm. (b) Histogram of corrected diameter of 2-μm particles. (c) and (d) Color density plots of velocity versus electrical and corrected diameter, respectively.
(a) Scatter plot of velocity versus electrical diameter for 2-μm particles from separate runs, showing clearly the instrument resolution of up to 1.7 μm. (b) Histogram of corrected diameter of 2-μm particles. (c) and (d) Color density plots of velocity versus electrical and corrected diameter, respectively.
Particle diameter (μm) | c1 | c2 | c3 |
Theoretical values | 0.99 | 2.67 | 0.14 |
2 (1.98) | 0.99 | 1.9 | 0.145 |
3 (3.098) | 0.99 | 2.89 | 0.145 |
4 (3.97) | 0.99 | 2.67 | 0.14 |
5 (4.77) | 0.99 | 2.67 | 0.14 |
6 (6.05) | 0.99 | 2.67 | 0.14 |
Particle diameter (μm) | c1 | c2 | c3 |
Theoretical values | 0.99 | 2.67 | 0.14 |
2 (1.98) | 0.99 | 1.9 | 0.145 |
3 (3.098) | 0.99 | 2.89 | 0.145 |
4 (3.97) | 0.99 | 2.67 | 0.14 |
5 (4.77) | 0.99 | 2.67 | 0.14 |
6 (6.05) | 0.99 | 2.67 | 0.14 |
The density of sampling points is shown in the electrical-diameter–velocity plane in Fig. 7 for particles with average diameters of ∼3.0 μm, 4.0 μm, 5.0 μm, and 6.0 μm. For all diameters, the density plot in particle-diameter–velocity space [Fig. 7(a)] shows a systematic variation of the electrical diameter with velocity, which is a consequence of the position dependence of the impedance in the z direction. Applying Eq. (5) to all the individual populations reduces the standard deviation in the distributions for all sizes, and the particle size is now independent of velocity. Figure 7(b) shows that the histogram of the uncorrected diameter is an asymmetric Gaussian distribution with a relatively large standard deviation. When Eq. (5) is applied, the distribution shown in Fig. 7(d) is closer to a symmetric Gaussian with a smaller standard deviation. From these plots, it is inferred that the correlation used to correct the electrical diameter works effectively for all particle sizes with the same fitting constants.
Separate runs with different particle sizes: (a) and (b) electrical-diameter histogram for particles of different sizes, and corresponding correction in histogram shape; (c) and (d) scatter plot of velocity versus electrical diameter for different particle sizes, and corresponding change in plot due to size correction.
Separate runs with different particle sizes: (a) and (b) electrical-diameter histogram for particles of different sizes, and corresponding correction in histogram shape; (c) and (d) scatter plot of velocity versus electrical diameter for different particle sizes, and corresponding change in plot due to size correction.
Figure 8 shows similar plots for experiments with a mixture of beads. Figure 8(a) shows that regardless of particle size, the ratio D/d varies in a similar manner with velocity. Figure 8(b) shows that the histogram of the uncorrected diameter is an asymmetric Gaussian distribution and for each particle diameter, the distribution width is large. The electrical diameter was corrected using Eq. (5) with the same values for c1, c2, and c3 as those in the simulations for all particle sizes. There is a significant reduction in the spread of the histograms for particles of different sizes as shown in Figs. 8(c) and 8(d), and the comparative results are provided in Table II for different particle sizes. From Table II, for the separate runs it is clear that there was a reduction in the standard deviation by a factor of two irrespective of particle diameter, which indicates the validity of the empirical relationship in Eq. (5). The experimental data from the mixed runs show again the applicability of Eq. (5), where both the standard deviation and coefficient of variation (CV) were found to be of the same orders of magnitude as those estimated from the separate runs. The estimated CV for each particle diameter was found to be very close to that from the manufacturer; the exception was that for the diameter of 3 μm, for which the CV provided by the manufacturer is very low compared to those for the other diameters.
Mixed runs with different particle sizes: (a) and (b) electric-diameter histogram, and corresponding correction in histogram shape; (c) and (d) scatter plot of velocity versus electrical diameter for mixed runs, and corresponding plot with correction.
Mixed runs with different particle sizes: (a) and (b) electric-diameter histogram, and corresponding correction in histogram shape; (c) and (d) scatter plot of velocity versus electrical diameter for mixed runs, and corresponding plot with correction.
Comparison between statistical data provided by manufacturer and those obtained before and after corrections were applied to the experimental data.
Manufacturer data . | Experimental data . | ||||||||
---|---|---|---|---|---|---|---|---|---|
Separate runs . | Mixed runs . | ||||||||
Particle diameter (μm) . | Mean . | SD . | CV . | Mean (before) . | SD (before) . | CV . | Mean . | SD . | CV . |
2 | 1.98 | 0.03 | 1.5 | 1.98 (2.0) | 0.065 (0.13) | 3.2 | |||
3 | 3.098 | 0.01 | 0.32 | 3.11 (3.22) | 0.083 (0.14) | 2.7 | 3.11 | 0.075 | 2.3 |
4 | 3.97 | 0.06 | 1.51 | 3.96 (4.07) | 0.089 (0.15) | 2.2 | 3.90 | 0.073 | 1.8 |
5 | 4.77 | 0.04 | 0.78 | 4.75 (4.81) | 0.088 (0.21) | 1.85 | 4.79 | 0.079 | 1.6 |
6 | 6.05 | 0.10 | 1.65 | 5.98 (6.08) | 0.11 (0.25) | 1.88 | 5.98 | 0.12 | 2.0 |
Manufacturer data . | Experimental data . | ||||||||
---|---|---|---|---|---|---|---|---|---|
Separate runs . | Mixed runs . | ||||||||
Particle diameter (μm) . | Mean . | SD . | CV . | Mean (before) . | SD (before) . | CV . | Mean . | SD . | CV . |
2 | 1.98 | 0.03 | 1.5 | 1.98 (2.0) | 0.065 (0.13) | 3.2 | |||
3 | 3.098 | 0.01 | 0.32 | 3.11 (3.22) | 0.083 (0.14) | 2.7 | 3.11 | 0.075 | 2.3 |
4 | 3.97 | 0.06 | 1.51 | 3.96 (4.07) | 0.089 (0.15) | 2.2 | 3.90 | 0.073 | 1.8 |
5 | 4.77 | 0.04 | 0.78 | 4.75 (4.81) | 0.088 (0.21) | 1.85 | 4.79 | 0.079 | 1.6 |
6 | 6.05 | 0.10 | 1.65 | 5.98 (6.08) | 0.11 (0.25) | 1.88 | 5.98 | 0.12 | 2.0 |
V. CONCLUSION
An important difficulty in determining the diameter of a particle of micrometer size by using impedance measurements in a microchannel is the uncertainty about its location in the channel cross section. For a particle of given size, the measured impedance is higher if the particle is closer to the electrodes and lower when it is in the center of the channel. Because of this, the electrical diameter inferred from impedance measurements has a significant error when compared to the true diameter. Herein, a procedure was formulated for correcting the electrical diameter to obtain the true diameter. The particle velocity is used as a single parameter to represent the variation in position in the cross section, and the ratio of the electrical and true diameters is expressed as a three-parameter fit [Eq. (5)]. Although the velocity depends on the two cross-stream coordinates whereas the impedance correction depends on only the coordinate perpendicular to the electrodes, we found a very good correlation between the velocity and the correction to the diameter using the three-parameter fit. There is a reduction by a factor of two in the standard deviation of the measured particle diameter by using this empirical correlation.
Surprisingly, we also found that the coefficients in Eq. (5) are independent of the particle diameter. The same coefficients were found to minimize the standard deviation in particle size for particles with diameters of 2–8 μm. This indicates that the coefficients are functions of only the channel and electrode configurations and do not depend on the size of the particle in the channel. Therefore, once the coefficients have been calibrated for one particle size for a given channel configuration, they can be used for other particle sizes. This provides a robust method for correcting the particle size for variations in cross-sectional location for samples in which the particle size distribution is discrete, so that the spread in the size distribution is smaller than the difference in the discrete size ranges. A more sophisticated procedure is being developed for size correction when there is a continuous size distribution.
ACKNOWLEDGMENTS
The authors would like to thank (i) the Director of CSIR–NAL for supporting this activity and (ii) the Polish grant committee for funding the “Bridge Alpha” grant for this project.
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
DATA AVAILABILITY
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
N. Priyadarshi graduated in Chemical Engineering from the Maulana Abul Kalam Azad University of Technology in West Bengal in 2020 and then completed an M.Tech. degree in Chemical Engineering at the Indian Institute of Science in Bangalore in 2022.
Usama Ahmed Abbasi graduated in Chemical Engineering from the Indian Institute of Technology (IIT) in Varanasi in 2009 and then completed a Master’s degree in Engineering at the Indian Institute of Science in Bangalore. Currently, he is the chief technology officer of MicroX Labs. His current activities include developing sensors, developing disposable plastic cartridges for automated sample preparation, and designing and developing point-of-care medical instruments.
Prof. V. Kumaran received a B.Tech. degree from the IIT in Madras and a Ph.D. degree from Cornell University in the United States. He is a professor in the Department of Chemical Engineering of the Indian Institute of Science in Bangalore. His research interests are fluid mechanics, statistical mechanics, and the dynamics of complex fluids.
Dr. P. Chowdhury graduated in Physics from the University of Burdwan in 1992 and completed postgraduate studies at the same university in 1995. He obtained a doctorate degree in Physics from the IIT in Bombay in 2000. He worked in South Korea as a postdoctoral fellow. Currently, he is working as a scientist at the National Aerospace Laboratories in India. His current activities include developing magnetic sensors for industrial applications and impedance-based sensors for biomedical applications. He is the author of over 65 international journal publications.