This article describes time-resolved optical measurements of H2S partial pressure and mass flow in a pulsed gas delivery system approximating injection conditions encountered during atomic layer deposition. A high-speed nondispersive ultraviolet (NDUV) gas analyzer design is employed for in-line H2S detection in a gas delivery line with flowing carrier gas. An in-place analyzer calibration performed in a reference cell yields an H2S detection limit of ≈1.4 Pa (at 22 °C) at a sampling rate of 1 kHz. Flow measurements performed on the delivery line are used to evaluate the effects of adjustable delivery parameters on the time-dependent injection system output. Short pulse widths exhibit partial pressure transients attributed to flow development within the different volumes of the delivery system. After ≈1.0 s of injection, steady-state flow is established across flow elements. A partial pressure of H2S in the delivery line is found to vary linearly with upstream H2S pressure, consistent with choked flow. A stronger scaling of partial pressure is evident when the flow coefficient of the downstream metering valve is adjusted. Estimated steady-state H2S flow rates in the range of 0.05–0.21 mg/s are observed within a limited range of valve flow coefficients. However, further increases in the flow coefficient do not result in increased flow, likely due to conductance limitations in downstream flow system components. The utility of NDUV absorption measurements for high-pressure pulsed gas delivery systems is discussed.

Hydrogen sulfide (H2S) is used as a sulfur source in the vapor phase deposition of sulfur-containing thin films such as transition metal dichalcogenides1–4 and chalcogenide glasses,5,6 some of which are candidate materials for future transistor channels and nonvolatile phase change memory. Unlike the low-volatility metal precursors used in chalcogenide film deposition processes, the chalcogen source (e.g., H2S and H2Se) is a compressed gas requiring pressure reduction and flow control components downstream of the source. In a typical gas delivery system, H2S is supplied from a high-pressure cylinder equipped with a pressure regulator, a downstream metering valve or critical orifice, and an injection valve. Operating conditions such as upstream pressure, metering valve conductance, and injection pulse width all contribute to the time-varying output of the H2S delivery system. For continuous deposition processes where a steady-state flow of H2S can be established, H2S concentration and flow rate may be conveniently measured employing a mass flow meter using a sensing element based on heat loss or Coriolis forces. In processes that involve pulsed injections of H2S, for instance, pulsed chemical vapor deposition and atomic layer deposition (ALD), measurement of H2S delivery system output requires time resolution significantly shorter than the injection pulse width. Pulse widths on the order of 0.1–1.0 s are common for ALD processes, giving rise to pressure and flow transients during each injection. Commercial thermal mass flow meters achieve sampling rates of 10 Hz or lower, making it difficult to characterize the pulse profiles of H2S under typical injection conditions. Further, the pressure drop introduced by the flow meter itself can alter the injection profile of a pulse. Therefore, a noninvasive, fast, and direct measurement of H2S concentration and flow rates is needed for the characterization and monitoring of pulsed processes such as ALD. Optical absorption spectroscopy meets these requirements and is well-suited for in situ gas sensing in deposition processes. Our prior work has focused on the delivery monitoring of low-volatility organometallic precursors using nondispersive infrared (NDIR) measurements.7,8 This work demonstrates the application of a nondispersive ultraviolet (NDUV) gas analyzer for the characterization of pulsed H2S delivery from a high-pressure source.

Optical absorption measurements are commonly used in the monitoring of environmental pollutants, toxic gases, and indoor air quality.9 While NDIR is the more commonly used technique due to its high selectivity to gases of interest, NDUV instruments have also found use in gas detection.10–13 In environmental applications, H2S is challenging to measure using NDIR and NDUV due to interference from species such as SF6 in the infrared and CS2, SO2, and mercaptans in the ultraviolet region.14,15 Various methods have been developed to prevent interference in gas mixtures, such as wavelength modulation measurements or the use of multivariate optical elements.15,16 Interference is less of a concern in semiconductor gas delivery applications where ultrahigh purity gases are delivered in dedicated gas lines. In this case, a direct absorption measurement using NDUV can provide low-cost and high-speed monitoring of gas pulsing. A typical NDUV measurement setup consists of a light source, an optical cell containing the sampled gas, and a detector. Selectivity to the target gas is provided by matching either the light source output or the detected wavelengths to characteristic absorption bands of the analyte. Quantitative measurements are obtained by calibrating sensor response against a known concentration or partial pressure of analyte. The use of nondispersive gas sensors in semiconductor processes has largely focused on steady-state flow measurements of metalorganics during continuous processes such as vapor phase epitaxy.17 Instruments making use of NDUV sensing have also been demonstrated for the steady-state monitoring of trimethyl aluminum delivery18 and NO2 flow visualization.19 We have taken a similar approach and employed a high-speed NDUV absorption measurement to characterize H2S concentration and flow rates under fast pulsing conditions used in ALD. The sensor is based on the H2S absorption band centered at ≈195 nm and a bandpass filter that selects for this feature.20 The NDUV gas sensor is installed on an H2S delivery line consisting of pressure reduction and flow control elements representative of ALD reactors incorporating high-pressure gas sources.4,21 We evaluate sensor response linearity and detection limit using an in-place calibration procedure. Using the calibrated sensor, we demonstrate the use of the sensor during pulsing to characterize H2S partial pressure and mass flow rate as a function of delivery system parameters such as injection timing, upstream pressure, and the flow coefficient of a metering valve. The uses and limitations of the measurement approach are discussed with respect to dynamic measurements of gas delivery for pulsed processes.

A dilute mixture of research grade H2S in argon with an H2S mole fraction ( x H 2 S ) of (3.01 ± 0.06)% was used as the H2S source. The H2S component of the gas mixture had a specified purity of 99.9% with x O C S 0.06 %, x C S 2 0.02 %, x C O 2 0.01 %, x N 2 0.01 %, and x hydrocarbon 0.01 %. A point-of-use purifier installed downstream of the H2S cylinder was used to remove moisture, metals, and particles. Ultrahigh purity Ar was used as a carrier gas during all measurements and was passed through a purifier to remove moisture, H2, CO2, O2, CO, organics, acids, bases, and metals.

A simplified schematic of the flow system used in this work is shown in Fig. 1.

FIG. 1.

Schematic of the H2S delivery system and nondispersive gas analyzer. Flow control elements include a MFC, pneumatic valves (PV1–PV4), pressure regulator (RV), NV, and a TV. The optical measurement consists of a deuterium lamp, optical fibers, lenses, a RC, a FC, optical filters, and an APD. Reference pressure is measured at CDG1 during calibration and the downstream pressure gauge CDG2 is used for flow experiments.

FIG. 1.

Schematic of the H2S delivery system and nondispersive gas analyzer. Flow control elements include a MFC, pneumatic valves (PV1–PV4), pressure regulator (RV), NV, and a TV. The optical measurement consists of a deuterium lamp, optical fibers, lenses, a RC, a FC, optical filters, and an APD. Reference pressure is measured at CDG1 during calibration and the downstream pressure gauge CDG2 is used for flow experiments.

Close modal

The upstream portion of the gas delivery system consisted of 4.572 mm inner-diameter electropolished 316L stainless-steel tubing. The carrier gas flow rate was 100 cm3/min at standard temperature and pressure (STP), defined as 0 °C and 101.33 kPa. A pneumatic shutoff valve (PV1) was located downstream of the Ar mass flow controller (MFC). The H2S delivery system including the gas cylinder was housed in a ventilated cabinet (dashed lines in Fig. 1). A pressure regulator (RV) was used to set the H2S delivery pressure (PR) upstream of the injection valving. Electronically controlled pneumatic diaphragm valves PV2 and PV3 were used to define injection, pressure equilibration, and purge timings. An all-metal metering needle valve (NV) installed between PV2 and PV3 was used as an adjustable flow restrictor. During injection, the H2S mixture flowed into the delivery line by way of a tee junction. Past the tee, the combined gas flow was directed through a three-port pneumatic valve PV4 and the optical flow cell (FC). H2S could be directed into the reference cell (RC) by opening PV4. A 2.6 kPa full-scale capacitance diaphragm gauge (CDG1) was directly connected to the RC to record cell pressure during sensor calibration. Gas lines downstream of the FC were transitioned to 25 mm inner-diameter stainless-steel tubing where a 13.3 kPa full-scale gauge (CDG2) was installed on a tee to measure downstream pressure. A throttle valve (TV) positioned downstream of the gauge was used to increase the delivery line pressure to ≈0.5 kPa, simulating ALD delivery line pressures under carrier gas flow. Gases exiting the flow system were released through an H2S scrubber containing high surface area Fe2O3 granules and were pumped out using a dry pump.

Gas injection timings were controlled by opening and closing solenoid-controlled valves PV2 and PV3 driven by a digital delay generator. Valve actuation states corresponding to purging, H2S injection, and delivery line pressure equilibration are summarized in Table I.

TABLE I.

Flow system state and corresponding pneumatic valve states.

StatePV1PV2PV3
Purge OPEN CLOSED CLOSED 
Inject OPEN OPEN OPEN 
Equilibrate OPEN CLOSED OPEN 
StatePV1PV2PV3
Purge OPEN CLOSED CLOSED 
Inject OPEN OPEN OPEN 
Equilibrate OPEN CLOSED OPEN 

When idle, the system is in a purge state with PV1 open and carrier gas flowing. Injection valves PV2 and PV3 are closed during purging and the volume between the two valves contains the source H2S mixture. After a time delay ( t d ) of 1.0 s, H2S is injected for a pulse time of t p by simultaneously opening PV2 and PV3, allowing the H2S mixture to flow through NV into the tee. At time t d + t p, the upstream injection valve PV2 is closed, allowing the volume between PV2 and PV3 to equilibrate with the delivery line pressure. After an equilibration time ( t e ) of 1.0 s, downstream injection valve PV3 is closed at time t d + t p + t e to return the system to a purge state. The delayed closing of PV3 is intended to reduce the buildup of pressure between the injection valves during purges, which would cause a large pressure transient in the delivery system upon subsequent injection. An alternative design incorporating a vent line between the flow restricting valve and the injection valve was described by Ahmido.22 For all experiments, t d and t e were set to 1.0 s, while t p varied. The cycle time between injections was 10.0 s.

Two separate optical cells were used for gas analyzer calibration and flow measurements. The cells were positioned such that their optical paths were collinear, allowing in-place calibration of the NDUV system. Both cells were made from stainless steel and had a central bore that matched the inner diameter of the upstream gas delivery lines. Pairs of recessed windows made of UV-grade fused silica were sealed to each cell using perfluoroelastomer O-rings. The optical path length between the windows within each cell was 9.2 mm, assuming 25% O-ring compression. For the FC, the optical axis was perpendicular to the gas flow direction. The RC was used only during sensor calibration and was evacuated prior to flow measurements. During measurement, light from a deuterium (D2) lamp was focused onto a 450 μm core-diameter, 2 m-long, solarization-resistant silica fiber. The output of the fiber was coupled into the optical cells using a pair of f/2 plano convex singlet lenses made of fused silica. Light was focused to a spot between the two optical cells. A matching pair of lenses was used to direct the light exiting the flow cell through a bandpass filter, through another f/2 lens, which focused the light onto a UV-enhanced silicon avalanche photodiode (APD). The detector had an active area of 1.0 mm diameter and a transimpedance gain of 25 MV/A. The bandpass filter had a nominal center wavelength of 214 nm and a full width at half maximum of 25 nm. The measured transmittance of the filter and the molar absorption coefficient of H2S at 22 °C are shown in Fig. 2. The transmittance peak near ≈185 nm in the filter spectrum may be an artifact of the charge-coupled device used to record the spectrum. Both detector and pressure signals were digitized at 1 kHz using a 24-bit analog-to-digital converter. All measurements were done at an ambient temperature of 22 ± 1 °C.

FIG. 2.

Molar absorption coefficient of H2S at 22 °C between 170 and 250 nm adapted from Wu and Chen (Ref. 23) (left axis) and the transmittance of the bandpass filter used in the NDUV gas analyzer (right axis).

FIG. 2.

Molar absorption coefficient of H2S at 22 °C between 170 and 250 nm adapted from Wu and Chen (Ref. 23) (left axis) and the transmittance of the bandpass filter used in the NDUV gas analyzer (right axis).

Close modal
Analyzer response was calibrated by measuring detector signal at a series of H2S partial pressures in the reference cell. To perform a calibration, the delivery line and the optical cells were first evacuated by closing all pneumatic valves, except for PV4. Next, the vacuum system was isolated from the pump and the two cells were charged with dilute H2S by opening PV2 and PV3. Last, the reference cell was isolated by closing PV4 and the delivery line was evacuated, including the flow cell. To achieve different H2S fill levels, the cell was successively evacuated by briefly opening and closing PV4 to the pump. Detector signal and reference cell pressure were recorded at each fill level. The molar density of H2S in the cell at the i th fill level was calculated assuming ideal gas behavior,
(1)
where x H 2 S , i n is the H2S mole fraction in the source container, P i is the average pressure for the i th fill level, P i 1 is the pressure at the previous fill level, R is the gas constant, and T is the gas temperature. Decadic absorbance in the reference cell between successive fills is then calculated as
(2)
where I i is the average detector signal at the i th fill level, I i 1 is the signal at the previous fill level, and S is the analyzer sensitivity factor given in m3/mol. For sensor calibration, S was calculated from the slope of a series of A i and n ^ i measurements. Each data point in the curve represents an absorbance due to a molar density change between consecutive fill levels in the reference cell. Rather than referencing absorbance to the initial APD signal, the calculation of absorbance between fill levels minimizes contributions to measured absorbance from residual adsorption on the cell windows. Another reason for employing this method was the observation of APD signal loss after each calibration. This could be a result of deposition on windows due to UV photodissociation of trace CS2 in the H2S mixture.24 Signal loss was especially noticeable during calibration where the cell windows were exposed to the gas for an extended period of time. The gain setting of the APD was adjusted after calibration to compensate for signal loss.
Signals from the APD and the downstream CDG were collected during H2S injections to calculate partial pressure and mass flow rate. The reference cell was evacuated and isolated during these measurements. Time-dependent absorbance in the flow cell was calculated as
(3)
where I ( t ) is the APD signal at time t, I 0 is the average reference signal recorded during the purge state prior to injection (td = 1.0 s), and I d is the dark signal recorded with the light source blocked. Absorbance was converted to a partial pressure ( p H 2 S , out ) using the calibration factor S and the ideal gas law.
Hydrogen sulfide flow rate was estimated using a set of mole balances written for the control volume around the tee injection point. Inputs into the control volume are carrier gas flows from the MFC and the dilute H2S/Ar mixture from the injection system. The output of the tee has a total gas flow rate consisting of an unknown mole fraction of H2S. Treating the gases as incompressible fluids and using the continuity equation, we obtain
(4)
where F M F C and F i n are the volumetric flow rates entering the tee from the MFC and the injection system, respectively. F o u t is the total flow coming out of the tee. Using this relationship, component mole balances can be written. Because no H2S is present in the carrier gas, the mole balance for H2S is
(5)
where x H 2 S , i n and x H 2 S , o u t are the H2S mole fractions at the injection side and the outlet side of the tee. Similarly, an Ar mole balance can be written as
(6)
where F A r , M F C is the Ar gas flow rate controlled by the MFC. Combining Eqs. (4)–(6), the flow rate of H2S exiting the tee can be expressed in terms of the carrier gas flow rate and H2S mole fractions at the tee inlet and outlet,
(7)

The carrier gas flow rate and the input mole fraction of H2S are known and x H 2 S , o u t can be estimated using the H2S partial pressure measurement at the optical cell and the total pressure ( P ) measurement at the downstream CDG. A steady-state mass flow rate for H2S was calculated using Eq. (7) and converting the resulting volumetric flow (at STP) to mass flow ( m ˙ H 2 S ) at the system temperature and pressure. Time-dependent mass flow rates were not calculated since the absorbance and pressure measurements were located at different points within the flow system. Fast pressure transients that could be observed optically in the flow cell were significantly distorted by the time the injection front reached the downstream CDG. This resulted in an underestimation of instantaneous system pressure at the start of each injection, which translated to unphysically large x H 2 S , o u t values. However, at time points sufficiently far from the pressure transient (t > 2.5 s), steady-state values for p H 2 S , o u t and P o u t could be used to calculate a steady-state x H 2 S , o u t. This was the approach taken in this paper. Furthermore, to obtain a steady-state pressure estimate closer to the flow cell, the pressure drop between the downstream CDG and the reference cell CDG was measured at representative pressures and a linear relationship was developed. A correction factor obtained from this analysis was applied to the downstream pressure measurement to approximate P o u t near the flow cell.

The gas analyzer response in the reference cell during a single calibration run is plotted in Fig. 3. The results of linear regression (dashed line) give a calibration factor of (0.454 ± 0.003) m3/mol and an absorbance intercept of 8.09 × 10−6 ± 1.36 × 10−5. When the calibration sequence is repeated 16 times, the average value of S is (0.450 ± 0.003) m3/mol. Given in terms of H2S partial pressure, the analyzer sensitivity is (1.83 × 10−4 ± 1.22 × 10−6) Pa−1. Blank absorbance measurements were additionally performed in the absence of H2S and referenced against an average I 0 taken during the first 1.0 s of measurement. The absorbance standard deviation (σ) for a 1.0 s long blank is ≈8.2 × 10−5. The detection limit of the NDUV analyzer, calculated as 3σ of the blank measurement, corresponds to a H2S partial pressure of ≈1.4 Pa at a sampling rate of 1 kHz. Analyzer sensitivity and time resolution are both sufficient to characterize H2S delivery under conditions relevant for pulsed processes such as ALD. For the remainder of this manuscript, we report on the results of a parametric study of delivery conditions and the corresponding H2S output measured using our NDUV gas analyzer. Adjustable delivery parameters examined here are injection pulse width, H2S cylinder outlet pressure, and the flow coefficient of the limiting constriction at the delivery system outlet.

FIG. 3.

Absorbance plotted as a function of H2S density change in the reference cell. The dashed lines show the result of a linear fit to the data points.

FIG. 3.

Absorbance plotted as a function of H2S density change in the reference cell. The dashed lines show the result of a linear fit to the data points.

Close modal

Time-resolved H2S partial pressure measurements and the corresponding total pressure are plotted in Fig. 4. Each trace is an average of eight consecutive injections. Injection pulse widths vary from 0.2 to 3.2 s with the H2S cylinder regulator pressure (PR) set to 68.9 kPa (10 psig) and the flow restricting needle valve at the 1.0 turn open position. Injection delay, pulse width, and postinjection equilibration durations are indicated by arrows for the longest injection. The total pressure measurement [Fig. 4(a)] at the downstream gauge shows a smooth increase in pressure ≈0.1 s after the start of the injection, reaching a steady-state value after ≈1.6 s of injection. Pressure starts to decay during the 1.0 s long equilibration as the volume between the injection valves equilibrates with the delivery line pressure. After both injection valves close, pressure continues to decrease as the unswept volume between the downstream gauge and the flow line is evacuated. The H2S partial pressure measurement shown in Fig. 4(b) differs from the total pressure measurement in several ways. The rate of rise of p H 2 S is sharper than the pressure gauge measurement and reaches a value of ≈65 Pa approximately 0.25 s after the start of the injection. After ≈0.5 s into the injection, p H 2 S sharply drops to a new steady-state value of ≈55 Pa. The first plateau is due to the pressurized volume between the needle valve NV and the downstream injection valve PV3. Evidently, this volume takes ≈0.25 s to drain, after which point the pressure upstream of the needle valve drives flow, resulting in the second plateau. At this point, flow is fully established within the injection system, with the needle valve acting as a critical orifice under choked flow conditions ( P upstream 2 P downstream ). Upon closing the upstream injection valve, p H 2 S decreases much more rapidly than the downstream pressure measurement would indicate. At the end of the injection, ≈1 Pa of H2S is measured immediately before the downstream injection valve closes.

FIG. 4.

(a) Total pressure and (b) H2S partial pressure as a function of time for injection pulse widths from 0.2 to 3.2 s. Each curve is an average of eight injections. Injection delay, pulse width, and equilibration time are shown as dashed lines for the longest injection.

FIG. 4.

(a) Total pressure and (b) H2S partial pressure as a function of time for injection pulse widths from 0.2 to 3.2 s. Each curve is an average of eight injections. Injection delay, pulse width, and equilibration time are shown as dashed lines for the longest injection.

Close modal

Next, we varied the flow coefficient (Cv) of the needle valve at the exit of the delivery system. This valve had an approximately linear Cv adjustment range over 0.1–2.0 turns of a vernier micrometer handle. Measurements from experiments spanning this Cv range are shown in Fig. 5.

FIG. 5.

(a) Total pressure and (b) H2S partial pressure traces from injections performed under various needle valve positions (larger values represent higher flow coefficients). The inset in (c) shows the beginning of each injection in more detail. Each curve is an average of eight injections.

FIG. 5.

(a) Total pressure and (b) H2S partial pressure traces from injections performed under various needle valve positions (larger values represent higher flow coefficients). The inset in (c) shows the beginning of each injection in more detail. Each curve is an average of eight injections.

Close modal

Upstream regulator pressure and injection pulse width for these measurements were fixed at 68.9 kPa and 1.6 s, respectively. The downstream pressure traces in Fig. 5(a) show a slow rise followed by a plateau after ≈1.0 s of injection. The amplitude of the pressure signal increases by a factor of ≈9.7 as the needle valve is opened from 0.1 turns to 2.0 turns. Fast transients associated with the developing flow are not apparent in the total pressure data as the gauge is too far downstream of the injection point. However, the optical measurement of p H 2 S in Fig. 5(b) reveals transients associated with flow development. As before, two regimes are apparent in the data. A first maximum in p H 2 S is reached shortly after the start of the injection driven by the pressurized volume between the needle valve and the downstream injection valve. Once this volume equilibrates with the delivery line, p H 2 S drops to its steady-state value where flow is established across the needle valve. As the needle valve is incrementally opened, the point at which steady-state flow is established shifts to the right. This trend is shown by the dashed arrow in the inset [Fig. 5(c)]. A larger flow coefficient (or conductance) permits higher pressures to develop at the point of injection. For a fixed equilibration time (te = 1.0 s), the trapped volume between PV2 and PV3 contains a higher density of gas at the end of each injection. Therefore, at the start of the subsequent injection, the time required to equilibrate this volume with the delivery line is longer. This could explain the positive shift in the initial pressure transient as a function of needle valve position.

Similar trends are evident in experiments where the H2S pressure regulator output is varied. Figure 6 shows total pressure and partial pressure data for injections with an upstream cylinder outlet pressure ( P R ) range of 17.2–96.5 kPa. For these injections, the pulse width and needle valve opening were fixed at 1.6 s and 0.4 turns, respectively.

FIG. 6.

(a) Total pressure and (b) H2S partial pressure traces from injections performed with upstream cylinder outlet pressure varying from 17.2 to 96.5 kPa. The inset in (c) shows the beginning of each injection in more detail. Each curve is an average of eight injections.

FIG. 6.

(a) Total pressure and (b) H2S partial pressure traces from injections performed with upstream cylinder outlet pressure varying from 17.2 to 96.5 kPa. The inset in (c) shows the beginning of each injection in more detail. Each curve is an average of eight injections.

Close modal

The total pressure profiles in Fig. 6(a) show similar time-varying features as the series of experiments where the needle valve opening was varied. The amplitude of the pressure traces scale with upstream pressure, but the gain associated with upstream pressure is weaker than that observed for flow coefficient adjustment. Here, the maximum total pressure increases by ≈1.6× when the upstream pressure is raised by a factor of ≈5.6. This scaling is readily discerned in the total pressure measurement because H2S is delivered from a dilute source, with Ar constituting most of the injected gas. An equivalent H2S injection from a 100% source would result in a small deflection in steady-state P in the delivery line, which would be challenging to detect using a pressure gauge alone. The optical measurement in Fig. 6(b) shows p H 2 S scale as a function of P R, similar to the total pressure measurement. Sensitivity to P R is weaker when compared with the scaling observed with Cv. The positive shift in the time it takes to reach a steady-state value for p H 2 S is shown in the inset [Fig. 6(c)]. As was the case when the needle valve opening was adjusted, increasing P R results in a longer duration transient that occurs at the start of the injection. This could be explained by a higher initial pressure (number of moles) between the injection valves at larger P R values. A variation in initial pressure occurs because the postinjection equilibration time was constant for all experiments.

Using total pressure and partial pressure data presented thus far, the impact of adjustable delivery parameters on H2S partial pressure and mass flow rate can be evaluated. Figure 7 shows the average values for p H 2 S and m ˙ H 2 S calculated at a steady state between t = 2.50 and 2.65 s.

FIG. 7.

Steady-state H2S partial pressure and mass flow rate plotted as a function of upstream regulator pressure [(a) and (b)] and needle valve flow coefficient [(c) and (d)]. The dashed lines in (a) and (b) are linear fits to the data. Each point represents an average of 150 time points (0.15 s). The error bars representing 1σ are too small to display. Conventional units for flow rates ( Q H 2 S ) are given in cm3/min at STP (SCCM).

FIG. 7.

Steady-state H2S partial pressure and mass flow rate plotted as a function of upstream regulator pressure [(a) and (b)] and needle valve flow coefficient [(c) and (d)]. The dashed lines in (a) and (b) are linear fits to the data. Each point represents an average of 150 time points (0.15 s). The error bars representing 1σ are too small to display. Conventional units for flow rates ( Q H 2 S ) are given in cm3/min at STP (SCCM).

Close modal

The first adjustable parameter is the upstream regulator pressure. Panels (a) and (b) show that both p H 2 S and m ˙ H 2 S vary linearly with P R. This is consistent with sonic injection (choked flow) where the flow rate is expected to vary linearly as a function of pressure upstream of the flow-limiting element (needle valve). The degree of adjustment afforded by P R is notably small for both partial pressure and flow rate. For a sixfold increase in upstream pressure, we observe an increase of ≈1.9× in p H 2 S and of ≈1.6× in m ˙ H 2 S. The second parameter of interest is the flow coefficient of the needle valve. The numerical value of Cv was provided by the valve manufacturer and given in volumetric flow rate of water, as defined in NFPA/T3.21.3 R1-2008 (R2018).25 However, relative changes in Cv are expected to map to changes in vacuum conductance. Therefore, we show changes in p H 2 S and m ˙ H 2 S as a function of a normalized Cv value and the corresponding valve position (top axis, number of turns) in panels (c) and (d). At a fixed upstream pressure of 68.9 kPa, increasing normalized Cv by a factor of ≈6.5 results in an increase in p H 2 S of ≈9.7×. Partial pressure increases nonlinearly at first but becomes linear as the valve is opened past 0.5 turns. In contrast, m ˙ H 2 S shows a similar response up to 0.5 turns but stops increasing thereafter (the shaded region in the figure). Therefore, after 0.5 turns, the H2S mole fraction stops increasing in the delivery line, while p H 2 S continues to rise. With the needle valve position between 0.1 and 0.5 turns open, measured flow rates of 0.05–0.21 mg/s are within a factor of ≈2.5 of the estimated flow rates (dashed lines) based on choked flow calculations and manufacturer-provided Cv values.26 This is reasonable since the flow does not account for any downstream flow-reducing components and reported Cv values have a typical error of >10%.25 A lack of change in mass flow in response to increasing Cv suggests that beyond a certain needle valve position, a downstream flow element becomes the limiting constriction in the delivery system. This could be the injection valve (PV3) or the throttle valve downstream of the pressure gauge.

The NDUV analyzer and the optical measurements described here provide insights into delivery system design and operation for pulsed processes. A designer may choose to optimize a delivery system for p H 2 S, m ˙ H 2 S, or both. Control over the partial pressure reaching the wafer surface is important for reaction kinetics, which could be a source of variability in process performance. On the other hand, mass delivery may be a consideration in reactor scaling or changes in wafer load. Adjustable delivery parameters examined here show varying degrees of controllability over p H 2 S and m ˙ H 2 S. First, we examine injection pulse width. For a dilute H2S source, as is the case here, injections lasting several seconds are needed to deliver sufficient mass to the reactor. As shown in Fig. 4, changes in p H 2 S and P stabilize after 1.0 s of injection, after which point total mass delivery ( m H 2 S ) should scale linearly with injection time. If supplied from a pure source, rather than a dilute mixture, the same mass of H2S could be delivered with pulse widths that are ≈33× shorter. This would be preferable for high-volume process equipment, where shorter cycle times are desired for tool productivity. However, as the NDUV data show, stable flow in the delivery system is not fully established at tp < 1.0 s as evidenced by the transients. Therefore, p H 2 S, m ˙ H 2 S, and m H 2 S are likely to vary nonlinearly with t p for short pulse widths. This, in turn, would make injection system outputs sensitive to uncertainties in injection timing.

Next, we consider the upstream pressure and injection system flow coefficient in terms of delivery system performance. Both parameters provide control over p H 2 S and m ˙ H 2 S, as shown in Fig. 7. Adjusting the delivery system conductance (needle valve opening), when compared with upstream pressure, provides a greater degree of control over p H 2 S. However, m ˙ H 2 S and m H 2 S do not scale past a certain Cv value in our flow system. Note that a gas delivery system design employing higher conductance downstream flow elements may not exhibit this limitation. Upstream pressure regulation, on the other hand, enables linear control over both p H 2 S and m ˙ H 2 S but provides comparatively weak scaling. The set of delivery parameter choices that a process designer makes will depend upon the desired system output. High-speed and high-sensitivity measurements discussed here provide one means of evaluating parameter response for pulsed delivery applications. Although dilute H2S is used as a test case for the optical measurements in this work, more common reactants such as NH3, H2, and O2 are also delivered from high-pressure sources in pulsed processes. These gases all require similar pressure reduction and control elements to achieve controlled pulsed injections. Insights gained here from optical measurements are expected to be applicable to pulsed processes employing other high-pressure gas sources.

We described an NDUV gas analyzer for H2S detection and demonstrated its use in a gas delivery line characteristic of a pulsed process such as ALD. The minimum detectable partial pressure was ≈1.4 Pa (at 22 °C) with a time resolution of 1 kHz. The analyzer was used to monitor hydrogen sulfide partial pressure and flow rate as dilute H2S was injected into a flowing carrier gas. Flow transients that could not be resolved with a downstream pressure gauge were readily detected by the optical measurement. Steady-state flow development was observed to occur ≈1.0 s after the start of the H2S injection. Adjustment ranges for H2S partial pressure and steady-state flow rates were characterized as a function of upstream pressure and the flow coefficient of a downstream metering valve. The partial pressure of H2S in the delivery line could be tuned by changing upstream pressure or the flow restricting valve opening. The mass flow rate could be controlled using the same parameters, but only a limited range of flow rates was accessible due to flow restrictions downstream of the gas delivery system. For long injections, total mass delivery could be further adjusted by changing the injection pulse width. The optical measurements demonstrated here could aid in the design of process equipment or in process parameter selection for pulsed gas delivery applications.

Certain commercial equipment, instruments, or materials are identified in this paper in order to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by NIST, nor is it intended to imply that the materials or equipment identified are necessarily the best available for the purpose.

The authors have no conflicts to disclose.

Berc Kalanyan: Conceptualization (lead); Data curation (lead); Formal analysis (lead); Investigation (lead); Writing – original draft (lead); Writing – review & editing (lead). Evan P. Jahrman: Conceptualization (supporting); Data curation (supporting); Formal analysis (supporting); Investigation (supporting); Writing – review & editing (supporting). James E. Maslar: Conceptualization (supporting); Data curation (supporting); Formal analysis (supporting); Investigation (supporting); Writing – review & editing (supporting).

The data that support the findings of this study are available from the corresponding author upon reasonable request.

1.
T. W.
Scharf
,
S. V.
Prasad
,
M. T.
Dugger
,
P. G.
Kotula
,
R. S.
Goeke
, and
R. K.
Grubbs
,
Acta Mater.
54
,
4731
(
2006
).
2.
V.
Pore
,
M.
Ritala
, and
M.
Leskelä
,
Chem. Vap. Deposition
13
,
163
(
2007
).
3.
L. K.
Tan
,
B.
Liu
,
J. H.
Teng
,
S.
Guo
,
H. Y.
Low
, and
K. P.
Loh
,
Nanoscale
6
,
10584
(
2014
).
4.
M.
Mattinen
,
G.
Popov
,
M.
Vehkamäki
,
P. J.
King
,
K.
Mizohata
,
P.
Jalkanen
,
J.
Räisänen
,
M.
Leskelä
, and
M.
Ritala
,
Chem. Mater.
31
,
5713
(
2019
).
5.
M.
Leskelä
,
L.
Niinistö
,
P.
Niemela
,
E.
Nykänen
,
P.
Soininen
,
M.
Tiitta
, and
J.
Vähäkangas
,
Vacuum
41
,
1457
(
1990
).
6.
E.
Färm
,
M. J.
Heikkilä
,
M.
Vehkamäki
,
K.
Mizohata
,
M.
Ritala
,
M.
Leskelä
, and
M.
Kemell
,
J. Vac. Sci. Technol. A
35
,
01B114
(
2017
).
7.
J. E.
Maslar
,
W. A.
Kimes
,
B. A.
Sperling
, and
R. K.
Kanjolia
,
Appl. Spectrosc.
74
,
1219
(
2020
).
8.
J. E.
Maslar
,
W. A.
Kimes
,
B. A.
Sperling
, and
R. K.
Kanjolia
,
J. Vac. Sci. Technol. A
39
,
012403
(
2021
).
9.
J.
Hodgkinson
and
R. P.
Tatam
,
Meas. Sci. Technol.
24
,
012004
(
2013
).
10.
G.
Dooly
,
C.
Fitzpatrick
, and
E.
Lewis
,
Sens. Actuators B
134
,
317
(
2008
).
11.
J. J.
Davenport
,
J.
Hodgkinson
,
J. R.
Saffell
, and
R. P.
Tatam
,
IEEE Sens. J.
18
,
2218
(
2018
).
12.
J.-L.
Tirpitz
,
D.
Pöhler
,
N.
Bobrowski
,
B.
Christenson
,
J.
Rüdiger
,
S.
Schmitt
, and
U.
Platt
,
Front. Earth Sci.
7
(published online 2019).
13.
Y.
Guo
et al,
Infrared Phys. Technol.
105
,
103153
(
2020
).
14.
Y.
Zhang
,
Y.
Wang
,
Y.
Liu
,
X.
Dong
,
H.
Xia
,
Z.
Zhang
, and
J.
Li
,
Spectrochim. Acta Part A
210
,
120
(
2019
).
15.
B.
Dai
,
C. M.
Jones
,
M.
Pearl
,
M.
Pelletier
, and
M.
Myrick
,
Sensors
18
,
2006
(
2018
).
16.
M.
Nikodem
,
K.
Krzempek
,
D.
Stachowiak
, and
G.
Wysocki
,
Opt. Eng.
57
(1),
011019
(
2017
).
17.
D.
Hayashi
,
Y.
Sato
,
M.
Minami
,
N.
Muramatsu
, and
M.
Iwaya
,
Jpn. J. Appl. Phys.
58
,
SC1013
(
2019
).
18.
H.
Ishii
,
M.
Nagase
,
N.
Ikeda
,
Y.
Shiba
,
Y.
Shirai
,
R.
Kuroda
, and
S.
Sugawa
,
Jpn. J. Appl. Phys.
58
,
SBBL04
(
2019
).
19.
Y.
Saka
,
Y.
Shiba
,
T.
Inada
,
T.
Goto
,
T.
Suwa
, and
T.
Oikawa
,
IEEE Trans. Semicond. Manuf.
36
, 515–519 (
2023
).
20.
C. Y. R.
Wu
and
F. Z.
Chen
,
J. Quant. Spectrosc. Radiat. Transfer
60
,
17
(
1998
).
21.
S.
Letourneau
,
M. J.
Young
,
N. M.
Bedford
,
Y.
Ren
,
A.
Yanguas-Gil
,
A. U.
Mane
,
J. W.
Elam
, and
E.
Graugnard
,
ACS Appl. Nano Mater.
1
,
4028
(
2018
).
22.
T.
Ahmido
,
W. A.
Kimes
,
B. A.
Sperling
,
J. T.
Hodges
, and
J. E.
Maslar
,
J. Vac. Sci. Technol. A
34
,
031512
(
2016
).
23.
H.
Keller-Rudek
,
G. K.
Moortgat
,
R.
Sander
, and
R.
Sörensen
,
Earth Syst. Sci. Data
5
,
365
(
2013
).
24.
W. S.
McGivern
,
O.
Sorkhabi
,
A. H.
Rizvi
,
A. G.
Suits
, and
S. W.
North
,
J. Chem. Phys.
112
,
5301
(
2000
).
25.
Pneumatic Fluid Power—Flow Rating Test Procedure and Reporting Method—For Fixed Orifice Components
(
National Fluid Power Association
,
Milwaukee
,
WI
,
2008
).
26.
Flow Equations for Sizing Control Valves
(
The International Society of Automation
,
Research Triangle Park
,
NC
,
2002
).