We have modeled, designed, built, and tested a novel reflectron time-of-flight (TOF) analyzer, which is capable of performing surface analysis using both secondary ion mass spectroscopy (SIMS) and mass spectroscopy of recoiled ions (MSRI). All elements (including H and He) can be identified, with isotopic resolution, using both MSRI and SIMS. For ions of a given mass, the higher energy ions penetrate further into the reflectron before being turned around while the lower energy ions do not penetrate as deeply. By properly adjusting both the experimental geometry and the reflectron voltages, all ions of a given mass arrive at the detector simultaneously—resulting in enhanced mass resolution compared with simple TOF detection. SIM spectra are complicated by molecular fragments in addition to elemental ions. In MSRI only elemental ions are detected. As a result, data analysis in MSRI is less complicated than in SIMS. Being able to use a single analyzer to selectively obtain SIMS or MSRI data provides complementary surface information. MSRI has a number of unique capabilities for surface studies. In situ, real-time surface analysis can be performed during film growth at pressures of ∼2.5 mTorr at the substrate by differentially pumping both the ion source and the reflectron analyzer region. It has been demonstrated that the ratio of the positive to negative ion yield is phase specific; for example, one can clearly distinguish the different forms of carbon (diamond versus graphite versus amorphous carbon) during film growth. MSRI analysis of poorly conducting surfaces is possible.
Skip Nav Destination
Article navigation
September 1999
Research Article|
September 01 1999
Novel reflectron time of flight analyzer for surface analysis using secondary ion mass spectroscopy and mass spectroscopy of recoiled ions
V. S. Smentkowski;
V. S. Smentkowski
Chemistry and Materials Science Divisions, Argonne National Laboratory, Argonne, Illinois 60439
Search for other works by this author on:
A. R. Krauss;
A. R. Krauss
Chemistry and Materials Science Divisions, Argonne National Laboratory, Argonne, Illinois 60439
Search for other works by this author on:
D. M. Gruen;
D. M. Gruen
Chemistry and Materials Science Divisions, Argonne National Laboratory, Argonne, Illinois 60439
Search for other works by this author on:
J. C. Holecek;
J. C. Holecek
Ionwerks, Houston, Texas 77005
Search for other works by this author on:
J. A. Schultz
J. A. Schultz
Ionwerks, Houston, Texas 77005
Search for other works by this author on:
J. Vac. Sci. Technol. A 17, 2634–2641 (1999)
Article history
Received:
January 14 1999
Accepted:
March 26 1999
Citation
V. S. Smentkowski, A. R. Krauss, D. M. Gruen, J. C. Holecek, J. A. Schultz; Novel reflectron time of flight analyzer for surface analysis using secondary ion mass spectroscopy and mass spectroscopy of recoiled ions. J. Vac. Sci. Technol. A 1 September 1999; 17 (5): 2634–2641. https://doi.org/10.1116/1.581923
Download citation file:
Sign in
Don't already have an account? Register
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Sign in via your Institution
Sign in via your InstitutionPay-Per-View Access
$40.00
Citing articles via
Related Content
In situ surface composition and structure of InGaN and GaN thin films by time-of-flight mass spectroscopy of recoiled ions
J. Vac. Sci. Technol. B (May 1999)
Characterization of TiAlN thin film annealed under O 2 by in situ time of flight direct recoil spectroscopy/mass spectroscopy of recoiled ions and ex situ x-ray photoelectron spectroscopy
Journal of Vacuum Science & Technology A (June 2002)
A study of HfO 2 film interfaces with Si and SiO 2
J. Appl. Phys. (January 2006)
Investigation of GaN deposition on Si, Al2O3, and GaAs using in situ mass spectroscopy of recoiled ions and reflection high‐energy electron diffraction
J. Vac. Sci. Technol. B (May 1996)
Reflectron velocity map ion imaging
Rev. Sci. Instrum. (October 2005)