The popularity of argon gas cluster ion beams (Ar-GCIB) as primary ion beams in time-of-flight secondary ion mass spectrometry (TOF-SIMS) has increased because the molecular ions of large organic- and biomolecules can be detected with less damage to the sample surfaces. However, Ar-GCIB is limited by poor mass resolution as well as poor mass accuracy. The inferior quality of the mass resolution in a TOF-SIMS spectrum obtained by using Ar-GCIB compared to the one obtained by a bismuth liquid metal cluster ion beam and others makes it difficult to identify unknown peaks because of the mass interference from the neighboring peaks. However, in this study, the authors demonstrate improved mass resolution in TOF-SIMS using Ar-GCIB through the delayed extraction of secondary ions, a method typically used in TOF mass spectrometry to increase mass resolution. As for poor mass accuracy, although mass calibration using internal peaks with low mass such as hydrogen and carbon is a common approach in TOF-SIMS, it is unsuited to the present study because of the disappearance of the low-mass peaks in the delayed extraction mode. To resolve this issue, external mass calibration, another regularly used method in TOF-MS, was adapted to enhance mass accuracy in the spectrum and image generated by TOF-SIMS using Ar-GCIB in the delayed extraction mode. By producing spectra analyses of a peptide mixture and bovine serum albumin protein digested with trypsin, along with image analyses of rat brain samples, the authors demonstrate for the first time the enhancement of mass resolution and mass accuracy for the purpose of analyzing large biomolecules in TOF-SIMS using Ar-GCIB through the use of delayed extraction and external mass calibration.

Several types of cluster ion beams have been examined and applied to TOF-SIMS analyses to measure large organic- and biomolecules with high mass because these beams inflict less damage to the sample surface. According to computer simulations1 and experimental2–6 reports, the ejection of high mass secondary ions is enhanced with the nonlinear effects of multiple collisions when large-sized cluster ion beams bombard the organic surfaces. There are some reports that a cluster ion beam as the sputtering ion source can successfully accomplish depth profiling of organic7–10 and amino acid films11 without ion damage. For example, Mochiji and Oshima et al. used argon gas cluster ion beams (Ar-GCIB) to measure several peptides and proteins with molecular weights of up to 15 kDa such as insulin, cytochrome C, lysozyme, and chymotrypsin, which were previously undetected in TOF-SIMS using atomic ion beams.4,6 Recently, a number of studies have examined the performance of cluster ion beams with respect to tissue imaging on the brains of rodents.12–16 In these studies, various biological molecules such as cholesterols, fatty acids, and sulfatides were detected in the brain tissue, with the intensity of the secondary ions boosted significantly when using Ar-GCIB than when using a bismuth cluster ion beam.

Understandably, GCIB is gaining in popularity as the primary ion source for measuring organic- and biomaterials in TOF-SIMS analyses. Of these cluster ion beams, Ar-GCIB is the preferred analysis beam because argon gas is relatively cheap and is easily formed into various sizes of clusters. Unfortunately, Ar-GCIB has poor lateral resolution as well as poor mass resolution caused by difficulties in both spatial and time focusing of large argon clusters. Poor lateral resolution results in low image quality for small sample areas such as cells or hair in the SIMS imaging analysis. One method to overcome this has been to increase the energy of the primary ion beam, which facilitates the focusing of the Ar-GCIB. This approach was utilized by Angerer et al. who demonstrated that a 40 keV Ar4000+ beam could be focused to approximately 2–3 μm at the sample surface to successfully produce a chemical image of sphingomyelin on human hair.12 Ar-GCIB is also limited by poor mass resolution, which prevents the isotope peaks from separating, thereby causing them to become overlapped in the spectrum. The resultant broadness of the peaks is problematic when trying to assign accurate mass for mass calibration; by the same token, the isotope distribution cannot be utilized to determine completely unknown peaks.

In this study, we are concerned with increasing the mass resolution and mass accuracy of the spectrum and image in the Ar-GCIB-TOF-SIMS analysis by using delayed extraction and external mass calibration. Delayed extraction17 has been widely used in matrix-assisted laser desorption ionization time-of-flight (MALDI-TOF) mass spectrometry18,19 and TOF-SIMS (Refs. 20–28) to achieve high mass resolution. This technique achieves time-of-flight compensation in ion energy spread by applying additional accelerating energy onto the sample plate with a specific time delay after the laser irradiates the sample surface. Thus, ions having the same mass-to-charge ratio (m/z) can arrive at the detector simultaneously, to generate peaks with narrow widths. We demonstrate here that the isotope peaks can be separated enough to distinguish them from the lowest to the highest after applying the delayed extraction technique to TOF-SIMS. In TOF-SIMS, delayed extraction achieves time-of-flight compensation in ion energy spread by applying extracting pulse energy onto the extractor of the secondary ions with a specific time delay after the pulsed primary ion beam irradiates the sample surface. Unfortunately, the spectrum acquired by using the delayed extraction method has lost the peaks of hydrogen and carbon ions, which causes difficulties in internal mass calibration—the typical method of calibrating mass in TOF-SIMS analysis. Furthermore, the asymmetric peaks in the low mass range make it difficult to determine their specific m/z values, which often leads to inaccurate internal mass calibration. As a result, we have used the external mass calibration method together with delayed extraction in our TOF-SIMS analysis. As with the delayed extraction method, the external mass calibration method is also widely used in various mass spectrometry techniques.29–32 Although these two methods, delayed extraction and external mass calibration, are not new, they will be useful in identifying unknown materials when newly applied to an Ar-GCIB-TOF-SIMS analysis.

Irganox® 1010 was purchased from Sigma Aldrich (St. Louis, MO, USA) and dissolved in ethanol. Tryptic digest of bovine serum albumin (BSA) and peptide calibration standard II were purchased from Bruker Daltonics (Bremen, Germany). Tryptic digest of BSA and the peptide calibration standard II samples were each dissolved in 125 μl of distilled water, according to the instructions provided. The silicon wafers were cleaned in a super-piranha solution33 and rinsed sufficiently with distilled water. Each sample of 10 μl were loaded onto a clean silicon wafer and air dried. After preparing the samples, Irganox® 1010 was added adjacent to the dried sample spots. For the rat brain sampling, all experimental procedures were conducted using protocols approved by the Institutional Animal Care and Use Committee of Konkuk University (Seoul, Korea). Male Sprague–Dawley rats (aged 10 weeks and weighing 300 g) were purchased from Orient Co., Ltd., a branch of Charles River Laboratories (Seoul, Korea). The animals were anesthetized using xylazine/ketamine (30 mg/75 mg/kg, ip) and maintained at 37 ± 1 °C during all surgical procedures. The dissected brains were immediately transferred to liquid nitrogen and then transferred to −80 °C. For TOF-SIMS analysis, the frozen brain tissues were sectioned at 10 μm-thickness in −20 °C using a cryostat (Leica CM 3050S, Leica Microsystems Inc., IL). The tissue sections were deposited onto a stainless steel substrate and then stored at −80 °C before the TOF-SIMS analysis.

TOF-SIMS spectra were obtained using a TOF-SIMS V instrument (ION-TOF GmbH). Ar1000+ with 10 keV was used as the analysis ion beam and the primary ion current was approximately 0.01 pA. During data acquisition, a pulsed electron flood gun was not used since charge compensation was unnecessary for Ar-GCIB. For the spectrum analysis, the analysis area was 300 μm × 300 μm and the primary ion dose density was 2.2 × 1011 ions/cm2 to ensure static limits. Five spectra were acquired for each sample in the conventional static and delayed extraction modes, whose timing diagrams are shown in Scheme 1. For the imaging analysis, we generated images with a wider field of view of 17.4 × 10.8 or 15 × 10 mm2 using a “stage-scan” technique in a saw-tooth mode that scanned the entire surface area with a spatial resolution of 10 μm. The primary ion current was 0.03 or 0.01 pA and the primary ion dose density was 8.4 × 1010 or 4.5 × 1010 ions/cm2 to ensure static limits. In the stage-scan mode, the sample was rastered by moving the sample stage. A delayed extraction time of 1.95 μs was used in the delayed extraction mode. The positive ion spectra were internally mass calibrated using C3H5+, C3H7+, C4H7+, C4H9+, and C5H11+ peaks for the tryp-BSA and C3H5+, C3H7+, C4H9+, and C5H11+ peaks for the peptide standard calibration II and rat brain samples. For external calibration, four peaks of Irganox® 1010 were selected and used as external calibrants, as explained in Sec. III.

Scheme 1.

Timing diagrams of (a) conventional static extraction mode and (b) delayed extraction mode.

Scheme 1.

Timing diagrams of (a) conventional static extraction mode and (b) delayed extraction mode.

Close modal

The molecular ion peaks of Irganox® 1010, recorded using Ar1000+ in the conventional extraction mode, are broad and show poor mass resolution, as shown in Fig. 1(a). With a mass resolution of approximately 260, the isotope peaks of the molecular ion are not separated and cannot be distinguished from each other. One attempt to improve mass resolution in TOF-SIMS using Ar-GCIB involves applying delayed extraction, an essential technique in TOF mass spectrometry.17–28 

Fig. 1.

TOF-SIMS spectra of Irganox® 1010 {[M]+ = C73H108O12+, m/z (calc.) = 1176.78}, obtained by using Ar1000+ in (a) conventional static extraction mode and (b) delayed extraction mode.

Fig. 1.

TOF-SIMS spectra of Irganox® 1010 {[M]+ = C73H108O12+, m/z (calc.) = 1176.78}, obtained by using Ar1000+ in (a) conventional static extraction mode and (b) delayed extraction mode.

Close modal

In the conventional static extraction mode (standard mode) of TOF-SIMS analysis, the width of the primary ion pulse directly affects the start accuracy of the TOF measurement, regardless of whether the analysis is performed with Ar-GCIB or with a liquid metal ion gun. This is because the pulsed primary ions desorb the secondary ions while the extraction field is already switched on, as shown in Scheme 1(a). The difficulty of Ar-GCIB is that the pulse duration, even in a bunched mode, is quite long (more than 20 ns) due to the extreme mass distribution of the large argon cluster primary ions. This results in a limited time and, hence, mass resolution. In the delayed extraction mode, however, the pulsed primary ions desorb the secondary ions while the extraction field is still switched off so that the secondary ions drift in field-free space above the sample surface. After a specific delay time, the extraction field is switched on shortly after the end of the primary ions pulse, so the start of the TOF measurement is no longer spread out over the primary ions pulse width. In other words, the delayed extraction mode separates the pulsing of the secondary ion extraction from that of the primary, and with the right delayed extraction time, the secondary ions that are located between the sample surface and the secondary ion extractor are able to enter the analyzer at the same time. As a result, mass resolution is dramatically improved from m/z ∼ 30 to ∼1200. Figure 1(b) shows the molecular peaks of Irganox® 1010 obtained with Ar-GCIB in the delayed ion extraction mode. The isotopic peaks of Irganox® 1010 are clearly separated, and mass resolution is visibly improved to 4972 calculated at m/z 1176.78.

Mass accuracy is contingent on accurate peak assignments that are reliant on mass resolution. Without these peak assignments, mass calibration is difficult. In a TOF-SIMS analysis, internal mass calibration is typically carried out by choosing peaks generated from small molecules that already exist in the sample. Thus, the mass error between the measured and theoretical values increases with molecular weight. Consequently, identifying unknown peaks obtained by Ar-GCIB is difficult when using the internal mass calibration method, and is further hampered when using the delayed extraction mode, as described in the Introduction.

Irganox® 1010 was selected for external mass calibration29–32 because this organic material is inexpensive and commonly used to evaluate the performance of cluster ion beams through depth profiling.7–9 Further, positive and negative spectra that contain well-known multiple characteristic and molecular peaks from the low- to the high-mass spectral regions (up to m/z 1200) can be obtained from this material in a reasonable time (less than 10 s). Figures 2(a) and 2(b) are positive and negative spectra obtained by using Ar-GCIB in the delayed ion extraction mode. The peaks selected as the external mass calibrants are indicated with arrows in Figs. 2(a) and 2(b). The Irganox® 1010 solution was deposited adjacent to the analytes, and a spectrum was acquired before the analytes were analyzed. Mass calibration was performed by using the four selected peaks shown in Fig. 2(a) for the positive spectrum and Fig. 2(b) for the negative spectrum, and the channel values or flight times of the selected peaks were measured from the mass calibrated spectrum. Then, the spectra of the analytes were externally calibrated using these channel values or flight times of the four specific m/zs. It is important to note that the analytes should be analyzed without z-movement.

Fig. 2.

TOF-SIMS spectra of Irganox® 1010 recorded in (a) positive and (b) negative ion modes, respectively, using Ar1000+ with delayed extraction.

Fig. 2.

TOF-SIMS spectra of Irganox® 1010 recorded in (a) positive and (b) negative ion modes, respectively, using Ar1000+ with delayed extraction.

Close modal

To evaluate the effectiveness of the external mass calibration method, trypsin-digested bovine serum albumin (tryp-BSA) and peptide calibration standard II were used. Figure 3 shows positive spectra of tryp-BSA and peptide calibration standard II obtained with Ar-GCIB using the delayed extraction mode and external mass calibration. Internal mass calibration was first performed using the C3H5+, C3H7+, C4H7+, C4H9+, and C5H11+ peaks for the tryp-BSA, and the C3H5+, C3H7+, C4H9+ and C5H11+ peaks for the peptide standard calibration II. External mass calibration was then performed using the four peaks of Irganox® 1010. Here, it should be noted that in the delayed extraction mode, the setting values of the delay time can be crucial—particularly for the flight time of the low-mass ions because the setting condition of the delayed extraction mode is typically optimized to separate the high-mass ion peaks (m/z ∼ 1000 in our study). This is because low-mass ions drift faster than ions with higher mass, meaning they experience a smaller fraction of the extraction field compared to the slower (high-mass) ions. As a consequence, the deviation of the mass calibration curve for low-mass ions is generally larger than that of the high mass ions. For internal calibration, then, it is advisable to use ions with higher mass—the higher the better—preferably above m/z 40 in order to minimize this effect.34 

Fig. 3.

TOF-SIMS spectra of (a) trypsin-digested BSA and (b) peptide calibration standard II obtained using delayed extraction and external mass calibration.

Fig. 3.

TOF-SIMS spectra of (a) trypsin-digested BSA and (b) peptide calibration standard II obtained using delayed extraction and external mass calibration.

Close modal

Figure 4 shows the mass differences of the internally mass calibrated and externally mass calibrated spectra. The average values and deviations calculated from the five spectra for each sample are also indicated in the graph. For the internally mass calibrated spectra, the mass differences are in the range of 0.7–2.0 Da, with the values tending to increase as a function of m/z. Here, in the case of tryp-BSA, the average mass difference of the eight peaks is approximately 1.2 Da. However, after external mass calibration, the average mass difference rapidly decreases to 0.03 Da. Since these values can be also expressed by means of mass error in parts per million (ppm), they were converted to 827.14 ppm for the internal mass calibration and 23.64 ppm for external mass calibration. For the peptide calibration standard II, the average mass difference of the seven peaks is approximately 1.3 Da in the internally mass calibrated spectrum and is reduced to 0.002 Da in the externally mass calibrated spectrum. These values correspond to 925.86 and 1.65 ppm, respectively. From these results, we are confident that the external mass calibration method is a useful calibration method for Ar-GCIB-TOF-SIMS analysis.

Fig. 4.

Mass difference for each peak in the spectra of (a) trypsin-digested BSA and (b) peptide calibration standard II after internal and external mass calibrations.

Fig. 4.

Mass difference for each peak in the spectra of (a) trypsin-digested BSA and (b) peptide calibration standard II after internal and external mass calibrations.

Close modal

Rather than using external mass calibration, well-known higher m/z peaks could be included to improve mass accuracy in the high mass range. For example, the average mass difference of the eight peaks can be reduced from 1.2 Da (827.14 ppm) to 0.013 Da (8.78 ppm) for tryp-BSA with the use of m/z 1283.71 and 1479.80 as internal calibrants. Similarly, the average mass difference of the seven peaks can be reduced from 1.3 Da (925.86 ppm) to 0.007 Da (6.49 ppm) for the peptide calibration standard II with the use of m/z 1347.74 and 1758.93 as internal calibrants. However, in general, it is difficult to determine the exact chemical formula of a high molecular peak for the internal calibrant unless peak identification with the MS/MS measurement can be successfully performed for each peak.

In addition to a spectrum analysis, the delayed extraction and external calibration methods were also applied to an image analysis of a rat brain sample obtained by using Ar-GCIB. Figure 5 shows the images and spectra of protonated cholesterol trimer15 [(C27H46O)3H+, m/z (calc.) = 1160.07] obtained by conventional [(a) and (c)] and delayed extraction modes [(b) and (d)], as explained above. As with the spectrum analysis case, mass error was significantly reduced from 215.5 to 60.34 ppm by using delayed extraction and external mass calibration, although it was higher than in the spectrum analysis due to the large scanning area for image acquisition.

Fig. 5.

Spectra and images of cholesterol trimers [(C27H46O)3H+, m/z (calc.) = 1160.07) obtained from the surface of a rat brain using (a), (c) the conventional mode and (b), (d) the delayed extraction mode.

Fig. 5.

Spectra and images of cholesterol trimers [(C27H46O)3H+, m/z (calc.) = 1160.07) obtained from the surface of a rat brain using (a), (c) the conventional mode and (b), (d) the delayed extraction mode.

Close modal

The use of Ar-GCIB as analysis beams in the TOF-SIMS technique has rapidly increased because they produce molecular ions of biomolecules with high mass. However, their poor mass resolution results in insufficient separation of the isotope peaks, thus making it difficult to distinguish them from each other. This affects mass calibration by aggravating the mass error up to several thousands of ppm. To circumvent these difficulties, two common techniques in MALDI-TOF mass spectrometry, delayed extraction and external mass calibration, were applied to improve mass resolution and mass accuracy, respectively. Mass resolution was improved to 4972 for the molecular peak (m/z 1176.78) of Irganox® 1010, and mass accuracy was enhanced up to 1.65 ppm in the spectrum analysis and up to 60.34 ppm in the image analysis. When Ar-GCIB is used as the analysis beam in TOF-SIMS, the delayed extraction technique and external mass calibration can be valuable tools to identify unknown peaks.

This study was supported by the Development of Platform Technology for Innovative Medical Measurements Program (GP2016-0022) from the Korea Research Institute of Standards and Science, the Pioneer Research Center Program (NRF-2012-0009541), the Bio & Medical Technology Development Program (NRF-2015M3A9D7029894), and Global Frontier Project (H-GUARD_2013M3A6B2078962) of the National Research Foundation (NRF) funded by the Ministry of Science, ICT & Future Planning. The authors are also grateful to Derk Rading (ION-TOF GmbH) for his valuable comments and suggestions.

1.
Z.
Postawa
,
R.
Paruch
,
L.
Rzenzik
, and
B. J.
Garrison
,
Surf. Interface Anal.
45
,
35
(
2013
).
2.
E.
Niehuis
,
R.
Möllers
,
D.
Rading
,
H.-G.
Cramer
, and
R.
Kersting
,
Surf. Interface Anal.
45
,
158
(
2013
).
3.
K.
Moritani
,
M.
Tanaka
,
N.
Inui
, and
K.
Mochiji
,
Surf. Interface Anal.
45
,
143
(
2013
).
4.
K.
Mochiji
,
J. Anal. Bioanal. Tech.
S2:001
(
2011
).
5.
D.
Weibel
,
S.
Wong
,
L.
Lockyer
,
P.
Blenkinsopp
,
R.
Hill
, and
J. C.
Vickerman
,
Anal. Chem.
75
,
1754
(
2003
).
6.
O.
Oshima
,
I.
Kashihara
,
K.
Moritani
,
N.
Inui
, and
K.
Mochiji
,
Rapid Commun. Mass Spectrom.
25
,
1070
(
2011
).
7.
A. G.
Shard
 et al.,
Anal. Chem.
84
,
7865
(
2012
).
8.
A. G.
Shard
,
S.
Ray
,
M. P.
Seah
, and
L.
Yang
,
Surf. Interface Anal.
43
,
1240
(
2011
).
9.
A. G.
Shard
 et al.,
J. Phys. Chem. B
119
,
10784
(
2015
).
10.
S.
Muramoto
,
D.
Rading
,
B.
Bush
,
G.
Gillen
, and
D. G.
Castner
,
Rapid Commun. Mass Spectrom.
28
,
1971
(
2014
).
11.
N.
Wehbe
,
T.
Tabarrant
,
J.
Brison
,
T.
Mouhib
,
A.
Delcorte
,
P.
Bertrand
,
R.
Moellers
,
E.
Niehuis
, and
L.
Houssiau
,
Surf. Interface Anal.
45
,
178
(
2013
).
12.
T. B.
Angerer
,
P.
Blenkisopp
, and
J. S.
Fletcher
,
Int. J. Mass Spectrom.
377
,
591
(
2015
).
13.
A.
Brunelle
,
D.
Touboul
, and
O.
Laprévote
,
J. Mass Spectrom.
40
,
985
(
2005
).
14.
J. S.
Fletcher
,
N. P.
Lockyer
, and
J. C.
Vickerman
,
Surf. Interface Anal.
43
,
253
(
2011
).
15.
T. B.
Angerer
,
M. D.
Pour
,
P.
Malmberg
, and
J. S.
Fletcher
,
Anal. Chem.
87
,
4305
(
2015
).
16.
J. S.
Fletcher
and
J. C.
Vickerman
,
Anal. Bioanal. Chem.
396
,
85
(
2010
).
17.
W. C.
Willy
and
I. H.
McLaren
,
Rev. Sci. Instrum.
26
,
1150
(
1955
).
18.
M. L.
Vestal
,
P.
Juhasz
, and
S. A.
Martin
,
Rapid Commun. Mass Spectrom.
9
,
1044
(
1995
).
19.
P.
Juhasz
,
M. L.
Vestal
, and
S. A.
Martin
,
J. Am. Soc. Mass Spectrom.
8
,
209
(
1997
).
20.
C.
Cersoy
,
P.
Richardin
,
P.
Walker
, and
A.
Brunelle
,
J. Mass Spectrom.
47
,
338
(
2012
).
21.
L. L.
Hanay
and
D. E.
Riederer
,
Anal. Chim. Acta
397
,
225
(
1999
).
22.
A. G.
Sostarecz
,
C. M.
McQuaw
,
A.
Wucher
, and
N.
Winograd
,
Anal. Chem.
76
,
6651
(
2004
).
23.
J.
Cheng
and
N.
Winograd
,
Anal. Chem.
77
,
3651
(
2005
).
24.
S.
Parry
and
N.
Winograd
,
Anal. Chem.
77
,
7950
(
2005
).
25.
S. V.
Baryshev
,
A. V.
Zinovev
,
C. E.
Tripa
,
M. J.
Pellin
,
Q.
Peng
,
J. W.
Elam
, and
I. V.
Veryovkin
,
Rapid Commun. Mass Spectrom.
26
,
2224
(
2012
).
26.
A.
King
,
T.
Henkel
,
D.
Rost
, and
I. C.
Lyon
,
Rapid Commun. Mass Spectrom.
24
,
15
(
2010
).
27.
J. L S.
Lee
,
I. S.
Gilmore
,
M. P.
Seah
, and
I. W.
Fletcher
,
J. Am. Soc. Mass Spectrom.
22
,
1718
(
2011
).
28.
Q. P.
Vanbellingen
,
N.
Elie
,
M. J.
Eller
,
S.
Della-Negra
,
D.
Touboul
, and
A.
Brunelle
,
Rapid Commun. Mass Spectrom.
29
,
1187
(
2015
).
29.
Goborn
,
M.
,
Mueller
,
V.
,
Egelhofer
,
D.
,
Theiss
,
H.
Lehrach, and
E.
Nordhoff
,
Anal. Chem.
74
,
3915
(
2002
).
30.
R. L.
Wong
and
I. J.
Amster
,
Am. Soc. Mass Spectrom.
17
,
1681
(
2006
).
31.
K.
Sládková
,
J.
Houška
, and
J.
Havel
,
Rapid Commun. Mass Spectrom.
23
,
3114
(
2009
).
32.
J.
Lyczko
,
D. G.
Beach
, and
W.
Gabryelski
,
J. Mass Spectrom.
50
,
463
(
2015
).
33.
H.
Min
,
J. W.
Park
,
H. K.
Shon
,
D. W.
Moon
, and
T. G.
Lee
,
Appl. Surf. Sci.
255
,
1025
(
2008
).
34.
ION-TOF GmbH Technical Engineer, personal communication (12 January 2016).