To ensure the continued safety of diagnostic ultrasound, accurate measurements of the acoustic output from clinical devices are necessary. The effects of nonlinear propagation can cause the current derating method to substantially underestimate relevant values in tissue. One suggested method to overcome this is to measure acoustic output parameters at low levels and extrapolate to values at full output. Tissue mimicking material were produced from concentrated milk. The samples had low scatter but speed of sound, attenuation, and nonlinear parameter B/A consistent with nonfatty soft tissues. Measurements of acoustic output from single element transducers were conducted at a wide variety of transmit levels in both water and through a path consisting mostly of tissue mimicking material. Results of peak rarefactional pressure and MI/MIE using tissue mimicking material at high pressure levels were consistent with extrapolated derated values from low level measurements in water. This agreement was true for the six measured transducers, which spanned a range of center frequencies from 2.0 MHz to 15.0 MHz. These results indicate that it may be possible to use low level transmit pulses in water, derate the measured values, and extrapolate those results to the higher pressure levels typically used in diagnostic ultrasonography.
Skip Nav Destination
,
Article navigation
October 2020
Meeting abstract. No PDF available.
October 01 2020
Experimental test of extrapolated acoustic output using single-element transducers Free
Timothy A. Stiles;
Timothy A. Stiles
Phys., Kettering Univ., 1700 University Ave., Flint, MI 48504, [email protected]
Search for other works by this author on:
Clayton Sparks
Clayton Sparks
Phys., Kettering Univ., Flint, MI
Search for other works by this author on:
Timothy A. Stiles
Clayton Sparks
Phys., Kettering Univ., 1700 University Ave., Flint, MI 48504, [email protected]
J. Acoust. Soc. Am. 148, 2520 (2020)
Citation
Timothy A. Stiles, Clayton Sparks; Experimental test of extrapolated acoustic output using single-element transducers. J. Acoust. Soc. Am. 1 October 2020; 148 (4_Supplement): 2520. https://doi.org/10.1121/1.5147002
Download citation file:
78
Views
Citing articles via
Focality of sound source placement by higher (ninth) order ambisonics and perceptual effects of spectral reproduction errors
Nima Zargarnezhad, Bruno Mesquita, et al.
A survey of sound source localization with deep learning methods
Pierre-Amaury Grumiaux, Srđan Kitić, et al.
Variation in global and intonational pitch settings among black and white speakers of Southern American English
Aini Li, Ruaridh Purse, et al.
Related Content
Pulse-echo sound speed estimation using second order speckle statistics
AIP Conf. Proc. (October 2012)
A new perspective for lung ultrasonography, preliminary results
J. Acoust. Soc. Am. (September 2015)
Analysis of ultrasonography distance accuracy and maximum depth of visualization based-on silicone rubber phantom
AIP Conf. Proc. (December 2019)
Hybrid-array-based optoacoustic and ultrasound (OPUS) imaging of biological tissues
Appl. Phys. Lett. (May 2017)
Ultrasonography education in the first medical-engineering based college
J. Acoust. Soc. Am. (March 2019)