The range of combined intersubject and cushion‐fit variation in the response of the TDH‐39 audiometer earphone in an MX41/AR cushion, reported in the literature and confirmed in these measurements, is of the order of 25 dB at 100 Hz, 6 dB at 1 kHz, and 15 dB at 5 kHz. The performance of this earphone system is further impaired by cushion‐induced physiological noise that masks low‐frequency thresholds, typically by 6 dB in normal‐hearing subjects. An experimental mounting for the TDH‐39 is described and analyzed; it is a circumaural device designed to keep the earphone‐ear acoustical system as near constant as possible with successive fittings and different subjects, and to reduce physiological noise. The performance of the TDH‐39 in this mounting, compared to performance in an MX41/AR cushion, showed a substantial reduction in cushion‐fit and intersubject response variation, and masking by physiological noise was largely eliminated in the subjects tested. The mounting has an optional cup for reducing the transmission of external noise, but this reduction is at the expense of a partial return of physiological noise masking. Average midrange sensitivity was reduced 5–6 dB. Response measurements were made with a free‐field reference method, using a probe microphone at the subject's eardrum.
Skip Nav Destination
Article navigation
December 1970
December 01 1970
Audiometer‐Earphone Mounting to Improve Intersubject and Cushion‐Fit Reliability Available to Purchase
Edgar Villchur
Edgar Villchur
Foundation for Hearing Aid Research, Woodstock, New York 12498
Search for other works by this author on:
Edgar Villchur
Foundation for Hearing Aid Research, Woodstock, New York 12498
J. Acoust. Soc. Am. 48, 1387–1396 (1970)
Article history
Received:
July 16 1970
Citation
Edgar Villchur; Audiometer‐Earphone Mounting to Improve Intersubject and Cushion‐Fit Reliability. J. Acoust. Soc. Am. 1 December 1970; 48 (6B): 1387–1396. https://doi.org/10.1121/1.1912297
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Citing articles via
Focality of sound source placement by higher (ninth) order ambisonics and perceptual effects of spectral reproduction errors
Nima Zargarnezhad, Bruno Mesquita, et al.
A survey of sound source localization with deep learning methods
Pierre-Amaury Grumiaux, Srđan Kitić, et al.
Drawer-like tunable ventilated sound barrier
Yong Ge, Yi-jun Guan, et al.
Related Content
MX41/AR earphone cushions versus a new circumaural mounting
J. Acoust. Soc. Am. (February 1981)
Large‐scale check on the consistency of coupler threshold standards
J. Acoust. Soc. Am. (August 2005)
Influence of earphone‐cushion center‐hole diameter on the acoustic output of audiometric earphones
J. Acoust. Soc. Am. (January 1979)
Effect of Circumaural Earphones and Earphone Cushions on Auditory Threshold
J. Acoust. Soc. Am. (November 1963)
Residual hearing in deafness
J. Acoust. Soc. Am. (August 2005)