We present a novel, interferometric, two-color, high-order harmonic generation setup based on a turn-key Ytterbium-doped femtosecond laser source and its second harmonic. Each interferometer arm contains a spatial light modulator with individual capabilities to manipulate the spatial beam profiles and to stabilize the relative delay between the fundamental and the second harmonic. In addition, separate control of the relative power and focusing geometries of the two color beams is implemented to conveniently perform automated scans of multiple parameters. A live diagnostics system gives continuous information during ongoing measurements.
I. INTRODUCTION
Sources of extreme ultraviolet (XUV) attosecond pulses, derived from high-order harmonic generation (HHG) in gases, hold considerable importance for basic research, including time-resolved spectroscopy with attosecond precision,1 and serve practical uses such as metrology tools in the semiconductor sector.2,3 However, due to its high nonlinearity, HHG is an inefficient process with typical conversion efficiencies on the order of 10−6 and lower.4 To optimize the HHG flux, the microscopic interaction between the laser and a single atom5,6 as well as macroscopic propagation effects in the nonlinear interaction medium7 have to be taken into account. The single-atom response is mainly governed by laser parameters, such as peak intensity, wavelength, and pulse duration. It is also strongly affected by the use of multiple frequencies, such as the second or third harmonic in addition to the fundamental.
The first two-color HHG experiments were performed already in 1993,8,9 following theoretical predictions.10 Watanabe et al. obtained an increase in the harmonic yield by several orders of magnitude by adding the third harmonic of the driving field, thus opening an extremely promising route to optimize HHG.9 In fact, it was later predicted that the optimum electric field shape, which gives not only a high cutoff but also a high yield, can be achieved by a multi-color synthesis.11 Multi-color HHG has been studied for many purposes ranging from a fundamental understanding of the HHG process12–16 to yield optimization,17–21 orbital-angular momentum studies,22,23 and spectral24–27 and temporal28 shaping.
Experimentally, multi-color HHG is not easy to realize. Even in the simplest case with only two different colors in the driving field, one needs to obtain precise control over the relative field intensities, their relative delay with sub-cycle stability, and their individual focusing geometries. An additional, recurring question for flux optimization is how to rigorously compare single-color and multi-color HHG.
In this work, we describe an experimental setup to generate high-order harmonics with a combination of the fundamental field and a fraction of its second harmonic. The setup is designed to obtain precise parameter control with the help of live diagnostics. First, we introduce the principles of multi-color HHG physics to identify the most important parameters for the experiment. A small review of possible experimental realizations is presented, where we explain the advantages and drawbacks of certain design choices to motivate the relevance of our setup. Second, we describe an experimental setup that uses a commercial laser with a pulse duration of 180 fs at a central wavelength of 1030 nm, together with its second harmonic. The integration of spatial light modulators (SLMs) gives us full control over the focusing conditions of the individual two color beams. Furthermore, one SLM is used to actively stabilize the phase delay between the two colors with a stability of <150 mrad. Finally, we present a selection of HHG results, highlighting the versatility of our setup.
II. DESIGN OF TWO-COLOR HIGH-ORDER HARMONIC GENERATION
A. Introduction to two-color high-harmonic generation
HHG is a laser-based source of coherent secondary radiation, discovered three decades ago. When intense ultrashort pulses are focused on a gas target, discrete spectral lines corresponding to odd high-order harmonics of the driving laser frequency ω are generated.29,30 This process can be understood physically by the three-step model.5,6 In step one, due to a strong distortion of the atomic potential by the laser field, an electron can tunnel-ionize to the continuum. It is then driven away by the laser field. When the laser field changes sign, the electron is driven back to the parent atom (step two). It may then recombine back to the ground state, and its excess energy is emitted as a photon, typically in the XUV spectral range (step three). This process is repeated every half-cycle of the driving field. The spectrum of odd-order harmonics, typically associated with HHG, follows from the spectral interference of the mutually coherent emissions from every half-cycle. In the presence of the second harmonic, the symmetry between consecutive half-cycles is broken, leading to the generation of even-order harmonics. The predictions of the three-step model for one and two-color HHG with the same linear polarization are presented in Fig. 1. The driving field is shown in (a)–(c) as a black solid line. In (b) and (c), the contributions of the fundamental (F, red) and second harmonic (SH, green) are indicated.
The distribution of return energies is distorted by the addition of the second harmonic and strongly depends on the values for R and ϕ. This can lead to a cut-off energy higher than in the single-color case [Eq. (3)] at the same total intensity [Fig. 1(h)] as well as to the appearance of flat regions in the return energy vs time, where the yield is expected to be high.
B. Possible experimental realizations
Experimentally, two-color HHG can be very difficult to realize, as one needs to first perform waveform-synthesis and then obtain control over the relative phase, the relative field amplitudes, and the individual focusing geometries. For this reason, most studies have focused on a narrow parameter space. We will review and comment on the design choices of each of the abovementioned experimental challenges to motivate our own setup implementation.
1. Two-color field synthesis
In Fig. 2, different schemes for two-color field synthesis are shown. The easiest to implement and the most compact solution experimentally is the inline option (a), where a nonlinear crystal is inserted into the beam path. This method is widely used.8,12,16,17,19,20,25,26,31–43 The desired intensity ratio between the fundamental and the harmonic can be obtained by choosing the crystal thickness and phase-matching angle. Limitations might occur when using very short pulses, where the phase-matching bandwidth of the material will limit the spectrum and, therefore, the pulse duration of the harmonic radiation. Another drawback is the impossibility of controlling the beam size, the field intensities, and the waveform shapes individually.
This individual control can be achieved by an interferometric approach instead, as illustrated in Figs. 2(b) and 2(c). The most energy efficient one at first glance is obtained by placing the nonlinear crystal into the beam and then splitting the colors with dichroic optics, as depicted in (b). A slightly different version is shown in Fig. 2(c). Here, the beam is split first with a fixed ratio, and the nonlinear crystal is placed in one of the arms. In this setting, the crystal can be optimized to yield the best harmonic output in terms of efficiency and beam quality. Also, impairment of the fundamental beam, for example, depletion in the center, is avoided. In both interferometer versions, each arm contains a single color and allows individual beam manipulation. However, due to longer beam paths and a larger number of optomechanical elements, interferometric setups are more susceptible to mechanical instabilities than inline setups.
When the second harmonic is generated in a nonlinear crystal co-linearly, as shown in all examples in Fig. 2, it will be perpendicularly polarized relative to the fundamental. If parallel relative polarization is desired, a half-wave (λ/2) plate designed for one color can be added to the beam path.
2. Focusing geometry
3. Relative phase control
To control the relative phase between the two fields in an inline setup, one has to rely on changing the amount of material the light goes through. Most works use a thin plate of fused silica at the Brewster angle, where small deviations from this angle will delay the components differently. While many publications do not comment at all on the introduced spatial beam displacement by transmitting through an angled plate, Brugnera et al. concluded that as long as the displacement is significantly smaller than the medium width and length, it will not have an influence on the experiment.35 Eliminating the displacement issue can be done using a wedge-pair with variable insertion instead.33,40,43,44 In an interferometric setup, the relative phase control can be easily realized by adding a delay stage to one arm, which then does not affect the spatial properties.45,46
4. Relative intensity control
Inline-setups are implemented most straight-forwardly but provide very limited options to carefully control the relative intensities, except for turning the nonlinear crystal to vary the conversion efficiency. In an interferometer, the options for attenuating pulse energies to change R or the total intensity must be carefully considered: when the pulses are sufficiently long (>30 fs), an attenuator consisting of a half wave plate and a polarizer can work very well. For shorter pulses, when high reflectivity thin film polarizers do not support a sufficiently large bandwidth, an iris is often used. This, however, affects not only the transmitted energy but also the beam size before the final focusing optics and, consequently, the focus waist sizes. In experiments where exploring different ratios of the two colors is not the main goal, aperturing one of the beams can be used to match the focus sizes of the individual colors.13
5. Design choices for a setup prioritizing flexibility and parameter control
In summary, two-color setup design and diagnostic capabilities highly depend on the purpose for which the complications mentioned above can be neglected. In our work, we prioritize flexibility and parameter control, thus choosing a setup with a configuration as shown in Fig. 2(c). In the next section, we will present our interferometric setup with the possibility of shaping the focusing geometry, intensity, and phase relation between both colors by integrating SLMs.
III. FLEXIBLE INTERFEROMETRIC ω/2ω HIGH-ORDER HARMONIC GENERATION
A. Experimental setup
Our experimental setup is depicted in Fig. 3. The output of a compact, turn-key Ytterbium-doped laser system with a pulse duration of 180 fs and pulse energy of 700 µJ at a repetition rate of 10 kHz (Pharos, Light Conversion) is divided by a beam splitter. 80% of the power is reflected into the 1030 nm (black) interferometer arm and 20% is transmitted. A 1 mm thick BBO crystal converts the transmitted light to 515 nm (green). A dichroic mirror reflects the residual infrared (IR) into a beam dump. In both interferometer arms, a combination of a motorized λ/2 plate and a thin-film polarizer acts as an attenuator to continuously tune the transmitted power. After the attenuators, both colors are horizontally polarized. To use the whole area of the SLMs (9.6 × 15.36 mm2, Santec SLM300) and to decrease the intensity to avoid damage, the beams are enlarged via lens telescopes with a 1/e2 diameter of 4–8 mm. Both colors are recombined through another dichroic mirror and are focused with an f = 250 mm lens into a gas jet (backing pressure of 2–4 bars and gas nozzle diameter of 42 µm), which can be moved in three directions. The generated XUV radiation is sent via an imaging flat-field grating onto a microchannel plate (MCP) where the phosphor screen behind is recorded with a camera, denoted by IV in Fig. 3. A very thin glass plate close to Brewster angle before the vacuum chamber reflects ∼1% of the power into the delay stabilization and focus diagnostic parts. Just in front of the vacuum chamber, the maximum power for 1030 nm is 4.5 W, corresponding to a pulse energy of 450 µJ, and for 515 nm 0.35 W, corresponding to 35 µJ. The split of the maximum pulse energy values can be varied by simply replacing the beam splitter at the beginning of the interferometer.
B. Focus control employing SLMs
The two-color beam is focused into the HHG chamber using a plano-convex lens, which introduces chromatic aberrations resulting in different focus positions for the two colors along the propagation direction. In addition, a non-perfect re-collimation after the enlargement of the lens telescopes in the two interferometer arms influences focus position and size, as well as aberrations introduced by the large number of transmissive optics.
For diagnostics, a part of the beam is reflected out by the thin plate directly in front of the vacuum chamber and focused onto the chip of a camera, denoted by I in Fig. 3. This camera is mounted on a motorized translation stage, allowing to take images along the propagation direction. Images captured a few millimeters before and after the focal plane can be used to obtain the beam quality factor M2 as well as to retrieve the wavefront using a modified version of the Gerchberg–Saxton algorithm.47,48 With the help of numerical beam propagation, the wavefront can be obtained in any plane of interest, and aberrations can be corrected by displaying the inverse of the retrieved wavefront in the plane of the SLM. In Fig. 4, our generation foci are shown without active wavefront shaping for 1030 nm (a) and 515 nm (b). The results of this correction for both colors are shown in Figs. 4(c) and 4(d). The images before and after correction share the same color scale, highlighting the significant improvement of the peak intensity in the focus. The wavefront needs to be corrected, with the method described above, only when significant modifications to the setup are made, such as exchanging optical elements in combination with an optical realignment procedure.
We measured focus sizes equal to w0,ω = 27 µm and w0,2ω = 14 µm, in excellent agreement with the predicted beam sizes, see Eq. (5). Adding a tilt to the wavefront in the x and y directions, perpendicular to the propagation axis, allowed for small alignment corrections. To move the beams independently of each other along the propagation direction, a parabolic phase front is applied, resulting in minimal beam focusing or defocusing after the SLM without influencing w0 significantly. Scanning camera I in Fig. 3 on the translation stage allows the determination of both foci positions relative to each other as well as their size and the overall beam quality at all times.
To estimate the focus position relative to the gas target inside the vacuum chamber, two cameras are used to image the proximity of the gas target through chamber windows from the side and from the top, denoted in Fig. 3 by II and III, respectively. In Fig. 5, the side (a) and top views (b) are shown. These views serve the purpose of visualizing the alignment of the gas nozzle relative to the laser focus. In Fig. 5(a), an example is shown where the pressure in the vacuum chamber is ≈0.1 mbar, which is sufficient to ionize the remaining gas. Using this method, one can calibrate with high precision the position of the focus waist relative to the gas target. The longitudinal focus position can then be changed either by moving the focusing lens or by applying a parabolic phase to the SLMs.
C. Phase control
The stabilization of the delay between the 1030 and the 515 nm arm uses the beam transmitted through the second thin plate in the diagnostic part of the setup. The box marked “Delay stabilization” in Fig. 3 is presented explicitly in Fig. 6(a). Both colors first go through a zero-order λ/2 plate designed for 1030 nm, which rotates the polarization of the 1030 nm radiation by 90°, but leaves the 515 nm radiation almost untouched. Transmission through 9 mm of BK-7 glass introduces a delay between the two-color components. The second harmonic of the fundamental pulse is generated in a BBO. Both second harmonic components have the same polarization and spectrally interfere. The thickness of the glass plate was chosen to have enough well-resolved spectral fringes. The spectral fringes are recorded by a commercial fiber spectrometer with an integration time of 50 ms and are shown in Fig. 6(b). The recorded spectrum is Fourier transformed (c), and the phase is extracted at the position of the peak marked by an arrow. This phase is plotted in Fig. 6(d) over a span of 50 s. It is used as a feedback to the 515 nm SLM. A proportional–integral–derivative (PID) control calculates the phase correction that is necessary for active stabilization. The SLM allows a phase variation of 2π, with a response time of ≈200 ms.
In Fig. 6(d), the free-running, unstabilized case (red) is compared to the actively stabilized case (black) over the same time span. The free-running phase-drift is slow compared to the stabilization response time. The standard deviation of the stabilized phase has a typical value of <150 mrad. During each measurement, the phase is logged to allow finer data sorting later for even higher accuracy. This is especially useful if the setup is operated in conditions where the phase is changed very rapidly. Such a case is presented in Fig. 6(d) in blue, where the intensity ratio of fundamental to green was changed from below 1% to 30% for different fast phase scans consecutively. Depending on the experimental conditions, the fringe contrast can be optimized by placing selective filters for IR or green in front of the spectrometer, as shown in Fig. 6(a). Alternatively, the BBO crystal can be moved along the beam focus to produce more or less second harmonic. There were no issues with phase stability as long as the Fourier peak signal was above ≈−25 dB. The phase measured this way is purely relative. To obtain an absolute phase reference for the experiment, one can calibrate it by using, for example, the minimum in harmonic yield.21
D. ω/2ω ratio control
To control the total field intensity in the gas target as well as the ratio between the two colors, each interferometer arm contains an attenuator with a half-wave plate on a motorized rotation mount and a 45° thin film polarizer (TFP). The beam size of the fundamental and second harmonic is measured using the camera I in Fig. 3 for each experimental configuration, and an M2 fit is performed. With this knowledge, automated scans are possible at a constant total peak intensity, Itot = ISH + IIR, while changing the contribution of the individual fields.
R = 0 and R = 1 refer to only 1030 and 515 nm, respectively, corresponding to single-color HHG in either case. The total intensity Itot defines the values of R that can be taken. Itot depends on the focusing geometries, pulse durations, and pulse energies. In our experiment, the pulse duration for both colors is fixed, and the pulse energies can be controlled by the attenuators. Since the beam sizes before the focusing lens are comparable, the focus size of the second harmonic is a factor of 2 smaller than for the fundamental [Eq. (5)]. With the current configuration of parameters presented in Sec. III A, the possible values for R are shown as the shaded gray area under the black line in Fig. 7. If another telescope were inserted in, for example, the 515 nm arm with the aim of having a comparable beam size in the focus instead, the maximum value for R would decrease rapidly (red). For Itot < 0.8 × 1014 W/cm2, the signal to noise ratio was not sufficient anymore to evaluate the relative yield of the high harmonics in our experiment (blue dashed line). To still be able to access a large range for R, having a smaller second harmonic focus compared to the fundamental was deemed necessary. To the best of our knowledge, the impact of the relative focal sizes in two or more color HHG has not been studied yet.
E. Integrated measurement platform
All motorized components (waveplates for the attenuators, linear translation stages for the focusing lens, and camera I) are controlled via a home-built control and measurement platform in Python. The cameras imaging the gas nozzle and the MCP as well as the SLMs can be accessed within the same environment. For each image of the XUV spectrum, a variety of information is logged together with it. This includes the measured two-color phase, its standard deviation, the settings of the waveplates defining the intensity ratio, and the settings of the SLMs and focusing lenses. With the help of this measurement platform, different scan options can be selected that are then being conducted and saved in a completely automated way, for example:
relative phase over a 2π range;
relative intensity ratio while keeping the same total intensity;
foci position along the propagation direction;
foci position relative to each other.
IV. RESULTS
Our experimental setup allows us to perform systematic studies of two-color HHG as a function of several parameters independently, such as R, ϕ, and Itot, with the possibility of bringing improved insights into the underlying physics.
In Fig. 8, we present results that demonstrate the capabilities of our setup. In this example, the total intensity of Itot = 0.8 × 1014 W/cm2 is kept constant while varying the intensity ratio between 1030 and 515 nm. In Fig. 8(a), the harmonic spectrum in argon at a backing pressure of 3 bars is shown for only the fundamental wavelength (R = 0, black). As expected, only odd harmonic orders are present. For the results shown in the red and blue curves, we keep the same total intensity but change the ratio to R = 0.32, thus lowering the contribution of 1030 nm and adding 515 nm. We consider two relative phases of ϕ = π (red) and ϕ = π/2 (blue). We use the same parameters for R and ϕ, as shown in Fig. 1; see Sec. II A. Even harmonic orders appear in addition to the odd ones, with comparable intensities. By changing the phase, one can either move the cut-off to higher energies (red) or maximize the lower and middle harmonic yields by more than one order of magnitude (blue) compared to the single-color case (black).
In Fig. 8(b), the evolution of the yield of two different harmonics is shown for a full 2π phase scan. Harmonic 25 (black) belongs to the HHG plateau, while harmonic 31 (red) is near the cut-off. Their oscillations are out of phase, reproducing previous studies.15,27,39,49
Examples of a full scan of R and ϕ at the same total intensity can be seen in Fig. 8(c) (harmonic 25) and (d) (harmonic 31). The color scale represents the XUV yield enhancement compared to R = 0. While for harmonic 25 in the plateau, there is a distinct maximum in R and ϕ, and the structure becomes richer, with two (R, ϕ) maxima for the 31st harmonic near the cutoff.
We noticed a continuous degradation of the 515 nm SLM after operating it on a daily basis over the course of 1 year, exposing it to a peak intensity of W/cm2 corresponding to an average intensity of 2 W/cm2. The slow deterioration resulted in a decrease in the reflection efficiency from 80% initially to 70% after 1 year. Further investigation of damage thresholds for SLMs using ultrashort laser pulses is therefore deemed necessary in the future.
V. CONCLUSION AND OUTLOOK
In this work, we present a novel experimental setup designed for studying two-color HHG. A two-color interferometer is combined with SLMs for convenient spatial beam control but also for delay stabilization with a standard deviation of <150 mrad and a delay range corresponding to 2π of the second harmonic. Cameras for the beam profile and gas target allow the precise positioning of the two beams with respect to each other as well as the position of the geometrical focus with respect to the generation gas nozzle. A summarizing overview of the diagnostic and control capabilities of the respective laser parameters that influence the XUV properties is shown in Fig. 9. The strength of our setup lies in the extensive parameter control, which allows us to tune the relative intensity ratio R and the phase difference between the two fields continuously, while keeping the total intensity fixed thanks to variable attenuators. This enables us to rigorously investigate properties such as yield enhancement as a function of independent parameters and disentangle the contributions of different effects. Apart from the described two-color HHG, the setup additionally features unique possibilities for spatial beam shaping, for example, top-hat beams and angular orbital momentum, with virtually unlimited possibilities to study details of single- and two-color HHG.
ACKNOWLEDGMENTS
The authors acknowledge support from the Swedish Research Council (Grant Nos. 2013-8185, 2021-04691, 2022-03519, and 2023-04603), the European Research Council (advanced grant QPAP, Grant No. 884900), the Crafoord Foundation, and the Knut and Alice Wallenberg Foundation. A.L'H. is partly supported by the Wallenberg Center for Quantum Technology (WACQT), funded by the Knut and Alice Wallenberg Foundation.
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
Author Contributions
A.-K. Raab: Conceptualization (equal); Data curation (lead); Formal analysis (lead); Investigation (lead); Methodology (lead); Software (equal); Validation (lead); Visualization (lead); Writing – original draft (lead); Writing – review & editing (lead). M. Schmoll: Investigation (supporting); Methodology (equal); Software (equal); Writing – review & editing (supporting). E. R. Simpson: Conceptualization (supporting); Funding acquisition (equal); Investigation (supporting); Methodology (equal); Project administration (equal); Resources (equal); Software (supporting); Supervision (supporting); Writing – review & editing (supporting). M. Redon: Conceptualization (supporting); Data curation (equal); Formal analysis (supporting); Investigation (equal); Methodology (equal); Software (equal); Validation (equal); Visualization (supporting); Writing – original draft (supporting); Writing – review & editing (supporting). Y. Fang: Data curation (supporting); Investigation (supporting); Methodology (supporting); Writing – review & editing (supporting). C. Guo: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Investigation (equal); Methodology (equal); Resources (equal); Supervision (equal); Validation (equal); Writing – review & editing (supporting). A.-L. Viotti: Conceptualization (supporting); Formal analysis (supporting); Funding acquisition (supporting); Methodology (supporting); Resources (supporting); Supervision (equal); Visualization (supporting); Writing – review & editing (supporting). C. L. Arnold: Conceptualization (equal); Funding acquisition (equal); Methodology (equal); Project administration (equal); Resources (equal); Supervision (equal); Writing – review & editing (supporting). A. L’Huillier: Conceptualization (equal); Formal analysis (supporting); Funding acquisition (lead); Investigation (equal); Methodology (equal); Project administration (equal); Resources (equal); Supervision (equal); Validation (supporting); Visualization (supporting); Writing – original draft (supporting); Writing – review & editing (supporting). J. Mauritsson: Conceptualization (supporting); Funding acquisition (equal); Investigation (supporting); Methodology (supporting); Project administration (equal); Resources (equal); Supervision (supporting); Visualization (supporting); Writing – review & editing (supporting).
DATA AVAILABILITY
The data that support the findings of this study are available from the corresponding author upon reasonable request.