Determining real-time changes in the local atomistic order is important for a mechanistic understanding of shock wave induced structural and chemical changes. However, the single event and short duration (nanosecond times) nature of shock experiments pose challenges in obtaining Extended X-ray Absorption Fine Structure (EXAFS) measurements—typically used for monitoring local order changes. Here, we report on a new single pulse (∼100 ps duration) transmission geometry EXAFS capability for use in laser shock-compression experiments at the Dynamic Compression Sector (DCS), Advanced Photon Source. We used a flat plate of highly oriented pyrolytic graphite (HOPG) as the spectrometer element to energy disperse x rays transmitted through the sample. It provided high efficiency with ∼15% of the x rays incident on the HOPG reaching an x-ray area detector with high quantum efficiency. This combination resulted in a good signal-to-noise ratio (∼103), an energy resolution of ∼10 eV at 10 keV, EXAFS spectra covering 100 s of eV, and a good pulse to pulse reproducibility of our single pulse measurements. Ambient EXAFS spectra for Cu and Au are compared to the reference spectra, validating our measurement system. Comparison of single pulse EXAFS results for ambient and laser shocked Ge(100) shows large changes in the local structure of the short lived state of shocked Ge. The current DCS EXAFS capability can be used to perform single pulse measurements in laser shocked materials from ∼9 keV to 13 keV. These EXAFS developments will be available to all users of the DCS.

Planar shock wave compression—generated via rapid deposition of energy on a planar surface by methods such as plate impact loading or ablation from a high-energy laser—results in near discontinuous changes in stress and density and can bring materials to extreme pressure–temperature conditions on nanosecond timescales.1,2 As such, shock wave compression has been used to examine materials under extreme conditions such as iron at earth core pressures3 and phase transformations in carbon at terapascal pressures.4 By measuring both shock wave velocity and particle velocity behind the shock wave, stress, density, and internal energy can be calculated using the Rankine–Hugoniot jump conditions,1,2,4 which provides valuable equation of state information. However, such continuum measurements lack detailed atomic level information. Using brilliant x rays from modern light sources, in situ x-ray diffraction measurements can now directly investigate shock-induced solid–solid phase transformations,5,6 defects,7,8 and melting9 on nanosecond timescales.

X-ray absorption fine structure spectroscopy (XAFS) is another x-ray diagnostic that can be useful for examining atomic level information in shocked solids.10–18 It is a powerful experimental tool to determine the local atomic structure for selected atomic species.19–22 It can probe a variety of systems including liquids, crystalline and non-crystalline solids, and disordered materials. Thus, XAFS is complementary to the standard x-ray diffraction (XRD) technique, which probes long range atomic order in crystalline materials. XAFS is based on the x-ray photoelectric effect, where a photo-electron is ejected from a core atomic orbital after the absorption of an incident x-ray photon by an atom. The photo-electron wave scatters from the atoms surrounding the absorbing atom, creating interferences between the scattered and outgoing parts of the photo-electron wavefunction, resulting in an energy-dependent XAFS spectrum.21,22 This spectrum consists of two regions—x-ray absorption near edge spectroscopy (XANES) within ∼30 eV of the absorption edge and extended x-ray absorption fine structure spectroscopy (EXAFS) from ∼30 eV above the edge to 100s of eV above the edge. XANES generally requires high energy resolution to resolve the detailed structure around the absorption edge and provides information on the oxidation state and local coordination chemistry. EXAFS refers to the oscillatory component of the absorption spectrum well above the absorption edge, and a large energy bandwidth above the edge is generally needed to get accurate structural and thermal information. The periods of the EXAFS modulations provide information on atomic distances and coordination numbers, while the decay of the modulation is dependent on the Debye–Waller factors, from which the material temperature can be determined.19–22 

Over the last decade, developments have been made in measuring in situ EXAFS under dynamic compression at the Omega Laser (University of Rochester)10–17 using a laser driven spherical implosion x-ray source.23 The x rays transmitted through the target are energy-dispersed by a multi-angle flat crystal spectrometer.15 Based on the same principle, an EXAFS capability is also being developed at the National Ignition Facility (NIF).24,25 Single shot EXAFS measurements under dynamic compression have been reported in a series of materials for energies below 8 keV, including V (K-edge 4.96 keV),10,11,14 Ti (K-edge 5.46 keV),10,11 and Fe (K-edge 7.11 keV).12,13,15,16 However, extending to higher energies has been a challenge due to the loss of x-ray flux by a factor of ∼8 to 10 between 6.5 keV and 10 keV17,25 for laser driven implosion x-ray sources.

An alternate approach to the studies above is to have a real time EXAFS capability under dynamic compression using a synchrotron source. A synchrotron x-ray source can typically provide greater spectral resolution, more energy tunability, and better pulse-to-pulse reproducibility than the laser-based x-ray sources. Recently, EXAFS measurements under ambient and laser shock conditions have been performed at beamline ID2426 of the European Synchrotron Radiation Facility (ESRF) using a 35 J laser.27,28 In this case, a divergent x-ray beam is focused and energy-dispersed by a bent crystal monochromator. The x-ray focal spot is at the target, and the transmitted x rays diverge onto a position sensitive detector. The bent crystal monochromator provides high resolution (ΔE/E ∼ 10−4), necessary for probing the XANES region,26–28 but provides limited energy bandwidth for EXAFS measurements27,28 (for example, ∼300 eV bandwidth near the Fe K-edge energy28).

In this paper, we present a newly developed single synchrotron x-ray pulse (∼100 ps) EXAFS capability at the Dynamic Compression Sector (DCS) located at the Advanced Photon Source (APS). This EXAFS capability is developed as a diagnostic to examine dynamically compressed materials using the DCS 100 J laser as the driver.29 To date, the laser drive capabilities at the DCS have focused on single shock compression, and as such, the temperature is governed by the material Hugoniot. The laser was designed to also achieve ramp compression (5 ns–10 ns rise times), resulting in lower temperatures, and these experiments will be undertaken in the future.

Using a flat plate of highly oriented pyrolytic graphite (HOPG) as the spectrometer element in combination with a high quantum efficiency detector (>90%), we have been able to obtain EXAFS spectra with a good signal-to-noise ratio (SNR) ∼103, an x-ray flux on the sample ∼106 photons/eV per x-ray pulse (∼100 ps), a resolution of ∼10 eV at 10 keV (ΔE/E ∼ 0.1%), an energy bandwidth of over 500 eV for the Ge K-edge, and a good pulse to pulse reproducibility (<2.0% SD) of our single pulse EXAFS measurements. Here, we also report, for the first time, EXAFS measurements under dynamic compression above 8 keV at 11.1036 keV (Ge K-edge30). Using our system, we have also obtained single pulse ambient EXAFS spectra for other materials with absorption edges ranging from ∼9 keV to 13 keV. Our setup, however, does not have the resolution for XANES, which requires a resolution of ∼1 eV (ΔE/E ∼ 10−4).

In Sec. II, we describe our new approach for single shot EXAFS measurements using a low divergence broadband x-ray source and a flat HOPG mosaic crystal as a spectrometer. We demonstrate the spectrometer efficiency and the corresponding energy resolution and bandwidths, with Ge as a primary example. In Sec. III, we first show ambient EXAFS spectra for Cu (K-edge 8.98 keV)30 and Au (L3-edge 11.92 keV)30 and compare them to the reference data obtained from the APS XAFS library.31 Next, we present our EXAFS results on ambient and shock compressed Ge samples, along with a fit of a simulated EXAFS spectrum for the ambient Ge. The excellent match between the measured and simulated ambient Ge EXAFS spectra, as well as the comparison of ambient and reference spectra for Cu and Au, demonstrates the quantitative capability of our EXAFS technique. Concluding remarks are given in Sec. IV.

Figure 1 shows the configuration for single pulse EXAFS measurements in the C-station (laser-shock station) of the DCS. As shown in Fig. 1(a), a combination of two rotating choppers [High Heat Load (HHL) chopper and Jülich chopper] and a millisecond x-ray shutter (ms-shutter) are used to isolate a single x-ray pulse (∼100 ps duration) incident on the laser shock target chamber.29 A single x-ray pulse can be isolated in either the APS 24-bunch mode operation, which has periodic x-ray pulses every 153.4 ns, or in the APS hybrid mode operation, which has a single x-ray pulse separated in time from other x-ray pulses by 1.59 µs. Single pulse EXAFS measurements have been obtained at the DCS using the configuration described in this section using both APS operating modes. However, the hybrid mode is preferable because the x-ray flux from the hybrid singlet is approximately four times larger than that of a single pulse in the 24-bunch mode, resulting in a better SNR. Selected parameters for the APS operating modes are provided in Table I.

FIG. 1.

(a) Layout of the beamline. The front-end station (A-station) consists of the white beam (WB) slits, high heat load (HHL) chopper, millisecond shutter (ms-shutter), and a pair of horizontal and vertical focusing mirrors (A-HFM and A-VFM). The laser-shock station (C-station) houses a horizontal focusing mirror (C-HFM), the Jülich chopper, and the laser target chamber. Relative distances between components are also shown. (b) Experimental configuration for single pulse EXAFS measurements within the C-station. The 2.5 m distance between the graphite spectrometer and the detector is an upper limit due to the size of the hutch and the laser target chamber location. The details of the laser target chamber are shown in Fig. 13.

FIG. 1.

(a) Layout of the beamline. The front-end station (A-station) consists of the white beam (WB) slits, high heat load (HHL) chopper, millisecond shutter (ms-shutter), and a pair of horizontal and vertical focusing mirrors (A-HFM and A-VFM). The laser-shock station (C-station) houses a horizontal focusing mirror (C-HFM), the Jülich chopper, and the laser target chamber. Relative distances between components are also shown. (b) Experimental configuration for single pulse EXAFS measurements within the C-station. The 2.5 m distance between the graphite spectrometer and the detector is an upper limit due to the size of the hutch and the laser target chamber location. The details of the laser target chamber are shown in Fig. 13.

Close modal
TABLE I.

APS x-ray operating modes.

Fill patternDescriptionX-ray spacingBunch current (102 mA total)
24 bunch mode 24 single bunches equally distributed around the storage ring 153.4 ns 4.25 
Hybrid mode A single bunch with 16 mA separated from adjacent septuplets by 1.59 µ1.59 µs, 28 ns 16, 1.5 
324 bunch mode 324 single bunches equally distributed 11.4 ns 0.31 
Fill patternDescriptionX-ray spacingBunch current (102 mA total)
24 bunch mode 24 single bunches equally distributed around the storage ring 153.4 ns 4.25 
Hybrid mode A single bunch with 16 mA separated from adjacent septuplets by 1.59 µ1.59 µs, 28 ns 16, 1.5 
324 bunch mode 324 single bunches equally distributed 11.4 ns 0.31 

The chopped x-ray pulse passes through a target located at the laser target chamber center (TCC) and is then incident on a flat plate of highly oriented-pyrolytic-graphite (HOPG) located downstream of the target chamber [see Fig. 1(b)]. HOPG has the graphite c-axis nominally normal to the plate face with a full-width-at-half-maximum (FWHM) mosaic spread of the order of one degree, with the actual mosaic spread value depending on the HOPG grade. The HOPG acts as an efficient energy dispersive spectrometer with diffraction of incident x rays from the (002) graphite plane.32,33

As shown in Fig. 2, the HOPG is oriented to have the diffraction vector nominally in the yz plane. A pair of horizontal and vertical x-ray slits centered on the incident x-ray beam were placed just upstream of the HOPG in order to have a well defined spectrometer element. The x rays diffracted from the HOPG have an angle–energy correlation described by Bragg’s law,34 

λ=2d(002)sinθ.
(1)

In Eq. (1), λ is the x-ray wavelength related to x-ray energy E via E = hc/λ, where h is Planck’s constant and c is the speed of light, d(002) is the d-spacing for the (002) planes of graphite, and θ is the Bragg angle.

FIG. 2.

(a) Representative energy dispersed x-ray signal on the Rayonix SX165 detector from a single hybrid singlet x-ray pulse in the APS hybrid mode. X rays are angle dispersed with energy (E) increasing with decreasing 2θ scattering angle. (b) Image of the x-ray signal on the Rayonix detector. (c) Integrated x-ray spectrum of the signal in (b). The spectrum was recorded without a sample placed in the beam. For the recorded spectrum, the U27 undulator, HOPG/x-ray beam angle, and x-ray detector position were set to examine the EXAFS signal around the Ge K-edge (∼11.1 keV).

FIG. 2.

(a) Representative energy dispersed x-ray signal on the Rayonix SX165 detector from a single hybrid singlet x-ray pulse in the APS hybrid mode. X rays are angle dispersed with energy (E) increasing with decreasing 2θ scattering angle. (b) Image of the x-ray signal on the Rayonix detector. (c) Integrated x-ray spectrum of the signal in (b). The spectrum was recorded without a sample placed in the beam. For the recorded spectrum, the U27 undulator, HOPG/x-ray beam angle, and x-ray detector position were set to examine the EXAFS signal around the Ge K-edge (∼11.1 keV).

Close modal

The energy dispersed x rays pass through a helium filled tube (∼2.5 m in length, with 50 μm thick Kapton windows on both ends) to limit x-ray scattering and absorption (only ∼1% scattering loss at 11 keV). The x rays are then detected using a Rayonix SX165 x-ray area detector. The detector has 2048 × 2048 pixels with 79 μm pixel size. The x-ray detector vertical plane (xz) is located at a distance Rd (∼2.5 m) along the incident x-ray beam direction from the intersection of the x-ray beam and the HOPG. Figure 2 shows a typical single pulse spectrum on the Rayonix detector without a target. X-ray energy is related to height z above the direct x-ray beam in the detector plane,

E=hc/(2d(002)sin[atan(z/Rd)/2]).
(2)

At DCS, x rays are generated from one of the two undulators—U27 or U17 with periods of 2.7 cm or 1.72 cm, respectively. Figure 3 compares the tuning curves for each of the undulators, showing the coverage of the energy range. The on-axis brilliance was calculated using the XOP software package35 with parameters for U27 (period length: 2.7 cm and number of periods: 88) and U17 (period length: 1.72 cm and number of periods: 137) and with an APS beam current of 100 mA and 7 GeV energy. The x rays from the first harmonic of the U27 undulator are used for the EXAFS measurements giving a peak energy range of 7.2 keV–16.8 keV. However, the intensity falls off drastically above 15 keV. On the lower energy side, below 8 keV, there is an additional issue of higher harmonic contamination, as discussed later. Figure 4 shows the calculated number of x-ray photons/eV emitted from a U27 undulator with the first harmonic peaked at ∼11.5 keV (using XOP35). This energy spectrum is used for Ge K-edge (11.1036 keV)30 EXAFS measurements. The bandwidth of the U27 undulator output is 560 eV FWHM and 1380 eV wide at 10% maximum, wide enough for EXAFS measurements. Figure 5 shows how the FWHM of the U27 first harmonic varies as a function of energy. Tapering the undulator gap can give a broader incident x-ray spectrum if desired, but at the expense of the decreased x-ray flux at a given energy.

FIG. 3.

Undulator tuning curves. Both the first and third harmonics are shown for the U27 undulator, while only the first harmonic is shown for the U17 undulator. The green vertical dashed lines bracket the energy range (9 keV–13 keV) over which our new EXAFS configuration was tested.

FIG. 3.

Undulator tuning curves. Both the first and third harmonics are shown for the U27 undulator, while only the first harmonic is shown for the U17 undulator. The green vertical dashed lines bracket the energy range (9 keV–13 keV) over which our new EXAFS configuration was tested.

Close modal
FIG. 4.

Calculated x-ray flux for a 2.4 m long, 27 mm period U27 undulator at the Advanced Photon Source. The undulator magnet gap is set to give a peak flux for the first harmonic at ∼11.5 keV (black solid line). X-ray numbers correspond to a single x-ray pulse (hybrid singlet in the APS hybrid operation mode). Red dashed line is the calculated x-ray flux after filtering the x rays by reflection from two horizontal silicon KB mirrors (2.1 mrad and 2.4 mrad pitches) and one vertical silicon KB mirror (2.1 mrad pitch). The total number of x rays per pulse in the filtered first harmonic is ∼4 × 109.

FIG. 4.

Calculated x-ray flux for a 2.4 m long, 27 mm period U27 undulator at the Advanced Photon Source. The undulator magnet gap is set to give a peak flux for the first harmonic at ∼11.5 keV (black solid line). X-ray numbers correspond to a single x-ray pulse (hybrid singlet in the APS hybrid operation mode). Red dashed line is the calculated x-ray flux after filtering the x rays by reflection from two horizontal silicon KB mirrors (2.1 mrad and 2.4 mrad pitches) and one vertical silicon KB mirror (2.1 mrad pitch). The total number of x rays per pulse in the filtered first harmonic is ∼4 × 109.

Close modal
FIG. 5.

Energy full-width-at-half-maximum (FWHM) diffracted by various grades of SPI Supplies HOPG and the FWHM of the U27 undulator first harmonic as a function of x-ray energy.

FIG. 5.

Energy full-width-at-half-maximum (FWHM) diffracted by various grades of SPI Supplies HOPG and the FWHM of the U27 undulator first harmonic as a function of x-ray energy.

Close modal

The x-ray beam sizes and divergences are different in the horizontal and vertical directions and are defined by the front-end beam apertures [white-beam slits, 1.2 mm (V) × 1.5 mm (H)] and several x-ray focusing mirrors. The details of the x-ray beamline and the relative distances between the x-ray focusing components are shown in Fig. 1(a). The beam is focused on the sample at the target chamber with a beam spot size of ∼30 μm (V) × ∼50 μm (H). The vertical focus is achieved by a single vertical focusing Kirkpatrick–Baez (KB) mirror in the A-station (A-VFM) with a pitch of 2.1 mrad. The horizontal focus is achieved by a combination of two horizontal focusing KB mirrors: one in the A-station (A-HFM) with a pitch of 2.1 mrad and the other in the C-station (C-HFM) with a pitch range of 2.1 mrad–2.7 mrad. The beam spot size at the HOPG spectrometer is ∼40 μm (V) × ∼250 μm (H). The EXAFS configuration shown in Figs. 1 and 2 is well matched to the DCS x-ray source and optics, which provide a low vertical divergence x-ray beam (∼13 μrad FWHM) and a small vertical beam size (tens of micrometers tall) near the sample and HOPG.

In addition to focusing the x-ray beam, the KB mirrors also act as low pass filters, removing unwanted higher order harmonics from the x-ray beam incident on the sample. Removing higher harmonics from the incident x-ray beam is important because the second and third order harmonics will be diffracted from (004) and (006) graphite planes at the same scattering angle as the first harmonic from (002) graphite planes. The Si stripe on the KB mirrors is used for these EXAFS measurements since it has the proper cutoff energy at a given pitch. The two A-station mirrors (A-HFM and A-VFM) with Si stripes at 2.1 mrad pitches have a low-pass cutoff energy of ∼15.1 keV and the C-station mirror (C-HFM) with the Si stripe at 2.4 mrad pitch has a cutoff energy of ∼13.2 keV. Figure 4 shows that the KB mirrors effectively filter out the higher harmonics when the undulator gap is set to examine the Ge K-edge. However, for energies below 8 keV, the contribution from the second harmonic will gradually become more significant as energy decreases. With a maximum possible C-HFM pitch of 2.7 mrad, the second harmonic peak intensity is ∼0.5% of that of the first harmonic at 7.9 keV, which increases to ∼1.3% at 7.2 keV.

HOPG with different degrees of mosaic spread are commercially available. For example, SPI Supplies offers grades 1, 2, and 3 HOPG with mosaic spreads of 0.4° ± 0.1°, 0.8° ± 0.2°, and 3.5° ± 1.5°, respectively. We can calculate the energy FWHM for each grade from Bragg’s law using dE = (E/tan θ), where is the mosaic spread and θ is the Bragg angle for the graphite (002) reflection at energy E. Figure 5 shows the FWHM energy ranges that are diffracted from the various grades of HOPG as a function of center energy. It also shows the FWHM of the flux spectrum from the first harmonic of undulator U27. The U27 parameters were 2.7 cm period length, number of periods, N = 88, and calculations were performed using the standard APS simulation code TCAP with a ring current of 100 mA and 7 GeV energy. The FWHM of the flux spectrum from the U27 first harmonic is bracketed by the FWHMs diffracted by SPI grades 1 and 2 HOPG, making either grade a candidate for the spectrometer element. The SPI grade 2 HOPG was chosen as the spectrometer element because the U27 undulator flux spectrum has a large tail on the low energy side (see Fig. 4), and to capture the entire U27 undulator first harmonic with a reasonable reflectivity (>10%) requires a larger FWHM for the x rays diffracted from the HOPG than the U27 undulator first harmonic FWHM.

Theoretical x-ray reflectivities were calculated for HOPG as a function of thickness dHOPG and mosaic spread using XOP software.35 The calculations were done for graphite (002) reflection with the energy centered at 11.3 keV. Figures 6(a)–6(c) show the calculated HOPG reflectivities for the three HOPG grades at several HOPG thicknesses as a function of energy over an energy range suitable for Ge K-edge EXAFS measurements. Figure 6(d) shows the peak reflectivity as a function of HOPG thickness for the three different grades. The HOPG plate/x-ray beam angle in the yz plane was chosen such that the (002) graphite planes, without misorientation, diffract 11.3 keV x rays. The SPI grade 2 HOPG provides a good combination between the high peak reflectivity and a fairly flat reflectivity curve (less than a factor of 2 variation in reflectivity over 1000 eV). Although the thicker HOPG has a higher overall reflectivity, as discussed next, energy resolution becomes poorer as the HOPG thickness increases. Therefore, an HOPG thickness of 100 μm was chosen as a compromise between reflectivity and energy resolution considerations. The corresponding average calculated HOPG reflectivity between 10.8 keV and 11.8 keV is 17.3%.

FIG. 6.

Calculated reflectivities of HOPG with various mosaic spreads as a function of x-ray energy near the Ge K-edge: (a) SPI grade 1, (b) SPI grade 2, and (c) SPI grade 3. Calculations are for HOPG thicknesses of 20 μm (black lines), 50 μm (red lines), 100 μm (blue lines), 200 μm (green lines), and 500 μm (dark red lines). In these calculations, the HOPG/incident beam angle was set to diffract 11.3 keV x rays for the nominal c-axis orientation of (002) planes. (d) Peak reflectivity as a function of HOPG thickness for the three different grades.

FIG. 6.

Calculated reflectivities of HOPG with various mosaic spreads as a function of x-ray energy near the Ge K-edge: (a) SPI grade 1, (b) SPI grade 2, and (c) SPI grade 3. Calculations are for HOPG thicknesses of 20 μm (black lines), 50 μm (red lines), 100 μm (blue lines), 200 μm (green lines), and 500 μm (dark red lines). In these calculations, the HOPG/incident beam angle was set to diffract 11.3 keV x rays for the nominal c-axis orientation of (002) planes. (d) Peak reflectivity as a function of HOPG thickness for the three different grades.

Close modal

The actual reflectivity of a ∼100 μm thick SPI grade 2 HOPG was measured for x-ray spectra peaked near 12 keV and near 13 keV. A diode was first used to record a signal proportional to the flux of the direct x-ray beam at the location of the HOPG spectrometer. The same diode was then moved to record a signal proportional to the x-ray flux diffracted by the HOPG. The x-ray beam/HOPG angle was scanned to find the angle at which the reflectivity was maximized, resulting in ∼17% peak reflectivity for x rays near 12 keV and ∼16% peak reflectivity for x rays near 13 keV. The HOPG reflectivity calculations using XOP software35 along with these measurements demonstrate that the HOPG acts as an efficient spectrometer suitable for the x-ray flux in single event experiments.

Energy resolution in EXAFS measurements is an important consideration as poor resolution will wash out sharp absorption features. Several parameters affect the achievable energy resolution.32,33,36,37 These include the thickness of the HOPG, dHOPG, the vertical size of the x-ray beam h at the HOPG, the vertical divergence δθv of the x-ray beam incident on the HOPG, the detector point spread function (DPSF), and the intrinsic diffraction broadening of the HOPG. The intrinsic diffraction broadening is due to a combination of the finite size of graphite crystallites along the c-axis in the HOPG (size broadening34) and the c-axis d-spacing distribution δd(002) in the graphite (strain broadening32,36). The effects of finite dHOPG, h, and DPSF on the energy resolution are mitigated by increasing the HOPG to the detector plane distance Rd, but the DCS C-station configuration limits Rd to a maximum value of ∼2.5 m.

Before presenting measurements for the energy resolution of the EXAFS system, we estimate the energy resolution due to several parameters, presently ignoring any component of the HOPG c-axis misorientation transverse to the nominal diffraction plane; later, we comment on the effects of transverse misorientation on the energy resolution. The effective energy broadening δE1 for purely monochromatic incident x rays due to finite dHOPG and h can be calculated ignoring x-ray extinction in the HOPG,

δE1=E*atantan(2θ)+wd/(2Rd)atantan(2θ)wd/(2Rd)/(2tanθ).
(3)

In Eq. (3), wd = [h + 2dHOPG cos θ]/cos(2θ) is the height of the detected signal on the detector plane for monochromatic x rays solely due to dHOPG and h. The energy spreads δE2 solely due to the vertical divergence δθv of the x-ray beam incident on the HOPG, δE3 solely due to DPSF, δE4 solely due to crystallite size Lc of the HOPG grains along the c-axis,34 and δE5 solely due to c-axis lattice spacing distributions are given by Eqs. (4)–(7), respectively,

δE2=E*δθv/(2tanθ),
(4)
δE3=E*atantan(2θ)+DPSF/(2Rd)atantan(2θ)DPSF/(2Rd)/(2tanθ),
(5)
δE4=E*[0.94λ]/2Lcsinθ,
(6)
δE5=E*δd(002)/d(002).
(7)

The actual energy resolution was measured as a function of energy using the Si(111) double crystal monochromator in the DCS A-station to produce an x-ray beam of bandwidth ∼1 eV incident on the HOPG spectrometer. The x rays dispersed by the HOPG were detected by a Rayonix SX165 x-ray area detector. The measured spectra were fitted to a pseudo-Voigt function to extract the FWHM. For these measurements, Rd = 1860 mm, dHOPG = 100 μm, h ≈ 34 μm, the detector point spread function is ∼150 μm, and θv is ∼13 μrad. Figure 7(a) shows the FWHM of the measured energy spectra diffracted from the HOPG as a function of energy. The blue dashed line shows the energy resolution calculated by adding in quadrature to the various sources described above, other than intrinsic broadening of the HOPG (δE4 and δE5). The calculations also include the Darwin width of the Si(111) double crystal monochromator.38 The energy resolution due to the Darwin width varies linearly with energy (from ∼0.9 eV at 7 keV to ∼1.8 eV at 14 keV). The large difference between the calculated and measured energy resolutions indicates that intrinsic broadening from the HOPG crystal contributes substantially to the overall energy resolution. Using reasonable values of Lc = 3 μm and δd(002) = 0.0025 Å to calculate the total energy resolution results in a good match between the calculated (red solid line in Fig. 7) and the measured energy resolutions, although independently determining precise values for Lc and δd(002) from the (002) graphite peak diffraction is not possible. In any case, the measured energy resolution is a better indicator of the actual energy resolution than the calculated energy resolution. Near the Ge K-edge at 11.1 keV, the energy resolution is ∼12 eV. Figure 7(b) shows the same data and calculations as in Fig. 7(a), except that EE is plotted rather than ΔE. The measured values of EE range from ∼800 to 1000, similar to the other reported EE values when the (002) graphite planes of HOPG are used to energy disperse x rays.32,36 As we will show later, this resolution is good enough for EXAFS measurements, but not sufficient to resolve the fine structures near the edge in the XANES region.

FIG. 7.

(a) Energy resolution ΔE of EXAFS measurements using a HOPG (SPI Supplies grade 2) spectrometer with Rd = 1860 mm. Black circles represent the measured FWHM energy resolution. The blue dashed lines are the calculated energy resolution for h = 34 μm, dHOPG = 100 μm, θv = 13 μrad, and DPSF = 150 μm, excluding the HOPG intrinsic broadening terms (δE4 and δE5). The individual contributions to the calculated energy resolution, δE1 (green), δE2 (pink), and δE3 (brown) are also shown. The red solid lines are the calculated energy resolutions, including the HOPG intrinsic broadening terms with Lc = 3 μm and δd(002) = 0.0025 Å. (b) The same results, but plotted as EE vs E. All calculations also included the Darwin width of the Si(111) double crystal monochromator.

FIG. 7.

(a) Energy resolution ΔE of EXAFS measurements using a HOPG (SPI Supplies grade 2) spectrometer with Rd = 1860 mm. Black circles represent the measured FWHM energy resolution. The blue dashed lines are the calculated energy resolution for h = 34 μm, dHOPG = 100 μm, θv = 13 μrad, and DPSF = 150 μm, excluding the HOPG intrinsic broadening terms (δE4 and δE5). The individual contributions to the calculated energy resolution, δE1 (green), δE2 (pink), and δE3 (brown) are also shown. The red solid lines are the calculated energy resolutions, including the HOPG intrinsic broadening terms with Lc = 3 μm and δd(002) = 0.0025 Å. (b) The same results, but plotted as EE vs E. All calculations also included the Darwin width of the Si(111) double crystal monochromator.

Close modal

Beer’s law21 describes the transmission of x rays through the sample as

Tsample(E)=Itrans(E)I0(E)=eμ(E)t.
(8)

In Eq. (8), I0(E) is the x-ray flux incident on the sample, Itrans(E) is the x-ray flux transmitted through the sample, μ(E) is the energy-dependent absorption coefficient, and t is the effective sample thickness (thickness of the sample material along the x-ray beam path).

The two-dimensional image shown in Fig. 2, obtained without a sample in the x-ray beam, is integrated horizontally (along x^) to obtain the monitor spectrum Imonitor(E),

Imonitor(E)=Isource(E)×RHOPG(E)×SDet(E).
(9)

In Eq. (9), Isource(E) is the spectral flux from the undulator source modified by the mirror reflectivities and any absorption along the beamline (air gaps and Kapton windows), excluding sample absorption. RHOPG(E) is the energy dependent reflectivity of the HOPG via diffraction from (002) graphite planes. SDet(E) is the energy dependent quantum efficiency of the Rayonix detector. Figure 8 shows a typical single pulse monitor spectrum Imonitor(E) with a configuration setup to examine Ge K-edge EXAFS. The width and shape of the monitor spectrum are largely due to the undulator source spectrum.

FIG. 8.

Intensity vs energy using the HOPG spectrometer. Each x-ray photon corresponds to ∼1.45 counts. The black solid line is the monitor signal Imonitor(E) obtained without a sample in the x-ray beam, and the red solid line is the signal I(E) obtained with a Ge sample placed in the x-ray beam. The Ge K-edge is clearly visible, and EXAFS oscillations at energies larger than the Ge K-edge are also noticeable.

FIG. 8.

Intensity vs energy using the HOPG spectrometer. Each x-ray photon corresponds to ∼1.45 counts. The black solid line is the monitor signal Imonitor(E) obtained without a sample in the x-ray beam, and the red solid line is the signal I(E) obtained with a Ge sample placed in the x-ray beam. The Ge K-edge is clearly visible, and EXAFS oscillations at energies larger than the Ge K-edge are also noticeable.

Close modal

The two-dimensional x-ray image recorded with a sample in the x-ray beam path is also integrated horizontally to obtain I(E),

I(E)=Tsample(E)×Isource(E)×RHOPG(E)×SDet(E).
(10)

The horizontal pixel range used in the integration is the same as used to determine Imonitor(E). A representative single pulse spectrum I(E) for a Ge sample is shown in Fig. 8. In Eq. (10), Tsample(E) is the transmission of x rays through the sample as a function of energy. All factors common to Eqs. (9) and (10) are nominally identical during the measurements of Imonitor(E) and I(E), and the product of the absorption coefficient times the effective sample thickness is given by

μ(E)t=lnImonitor(E)I(E).
(11)

The effective sample thickness t = dsample/sin ϕ is known from the metrology of the actual sample thickness dsample and the x-ray beam/sample angle ϕ. Thus, with the x-ray absorption spectroscopy configuration used at the DCS, an absolute measure of μ(E) can be directly determined from the measurements of Imonitor(E) and I(E), provided the product of Isource(E), RHOPG(E), and SDet(E) does not change between single x-ray pulse measurements. The stability of this product or Imonitor(E) is demonstrated in Sec. III. The product μ(E)t vs energy calculated from the single pulse monitor and sample spectra (see Fig. 8) is shown in Fig. 9(a).

FIG. 9.

(a) Measured product of the absorption coefficient and the sample thickness μ(E)t vs energy from monitor Imonitor(E) and I(E) spectra shown in Fig. 8. (b) Measured (red) and calculated (black) signal to noise ratio for μ(E)t. Measurement and calculation correspond to a single pulse (hybrid singlet) with energy bins corresponding to a single row of detector pixels.

FIG. 9.

(a) Measured product of the absorption coefficient and the sample thickness μ(E)t vs energy from monitor Imonitor(E) and I(E) spectra shown in Fig. 8. (b) Measured (red) and calculated (black) signal to noise ratio for μ(E)t. Measurement and calculation correspond to a single pulse (hybrid singlet) with energy bins corresponding to a single row of detector pixels.

Close modal

To determine the pixel range over which the horizontal integration is performed when calculating Imonitor(E) and I(E), the image is first integrated vertically (along z^ in Fig. 2) and the horizontal pixel indices at the peak value and at the two half peak values are determined. The horizontal integration range extends three times from the difference between the peak-value pixel index and the half-peak-value pixel indices. A background is also subtracted from Imonitor(E) and I(E) to account for the small offset in counts in each pixel.

The reason that the recorded x-ray signal is broad in the x^ direction is due to the HOPG c-axis mosaic spread. The HOPG mosaic spread component in the yz plane provides the energy dispersion vertically on the detector, and the transverse mosaic spread component causes the broadening in the x^ direction. Because the diffracted x rays at a fixed energy form a diffraction cone, integrating horizontally along a row of pixels on the detector introduces an additional contribution, reducing energy resolution. For a typical configuration, this additional broadening in a single row of pixels is less than 1 eV and is a small fraction of the overall energy resolution.

Due to the high single pulse x-ray flux (∼4 × 109 x rays per pulse), high HOPG integrated reflectivity (∼15%), and high detector quantum efficiency (>90% over the 9 keV–13 keV range for the 80 μm thick phosphor used in the Rayonix SX165 detector), a high signal to noise ratio (SNR) can be achieved in EXAFS measurements at the DCS. As an example, we consider the Ge K-edge EXAFS measurement shown in Fig. 9(a). To obtain the incident number of photons/eV on the detector, the red dashed curve in Fig. 4 is multiplied by (1) the transmission through windows and air gaps (calculated to be 0.5), (2) transmission through the Ge sample given by the measured ratio of I(E)/Imonitor(E), where the two spectra are shown in Fig. 8, (3) the HOPG reflectivity [blue curve in Fig. 6(b)], and (4) the detector quantum efficiency. The calculated SNR for μ(E)t was based on a Poisson distribution for the number of x-ray photons reaching the detector per row of pixels. Readout noise in each pixel (9e) was also included in the SNR calculation and corresponds to approximately one x-ray photon/pixel. The noise is dominated by the photon statistics. The calculated SNR and the estimated SNR based on the measurements are shown in Fig. 9(b). To estimate the SNR for the measurement, a smoothing function was generated from the data, which was then used as a baseline to calculate the average variance within an energy window. The measured SNR is ∼1000, while the calculated SNR for μ(E)t is ∼500 for each row of pixels (∼1 eV energy range/pixel) over the energy range used for Ge EXAFS measurements. The calculation seems to underestimate the actual SNR by a factor of 2. Because the energy resolution solely due to the height of a single pixel (∼1 eV resolution) is significantly smaller than the total energy resolution (∼12 eV), it is possible to average the signal over several rows of pixels to improve the calculated SNR without significant additional broadening of the absorption spectra. For example, averaging over seven rows of pixels increases the calculated SNR to over 1000.

As mentioned in Sec. II, the stability of Imonitor(E) [see Eq. (9)] is important for the accuracy and reproducibility of the extracted EXAFS oscillations. Assuming that the quantum efficiency of the Rayonix detector SDet(E) is constant, the stability of Imonitor(E) depends on the HOPG reflectivity RHOPG(E) and the source spectrum Isource(E). While Isource(E) should not change, except for the minor variation in the flux, RHOPG(E) can vary depending on the mosaic distribution of the HOPG crystal, which can vary spatially in the HOPG. Therefore, in order to maintain a defined spectrometer element on the HOPG, two important steps were taken. First, a back-end x-ray camera was placed downstream of the HOPG to establish the half-cut and the zero angle (i.e., 2θ = 0° in Fig. 2) of the HOPG spectrometer. Using this as a fiducial, we can then adjust for any vertical drift in the x-ray beam during the experiment. Second, a pair of horizontal and vertical x-ray slits centered on the incident x-ray beam were placed just upstream of the HOPG to define the size of the x-ray beam incident on the HOPG. If we did not establish a stringent horizontal window of the x-ray beam, we found distinct changes in the monitor x-ray spectrum that drifted in time. Therefore, the horizontal x-ray slit had an opening width ∼100 μm narrower than the horizontal width of the incident x-ray beam on the HOPG. The final width of the horizontal x-ray slit was ∼170 μm. The vertical x-ray slit was kept relaxed. The vertical position was monitored with the back-end camera and adjusted as needed by changing the A-station vertical mirror pitch. These steps established a well defined stable region of the HOPG, acting as the spectrometer element.

Figure 10(a) shows a series of five monitor spectra from five single x-ray pulses, showing a minor pulse-to-pulse variation in maximum intensity. When normalized to the maximum intensity, all the scans fall almost on the same curve, as shown in Fig. 10(b). The percentage differences of the individual scans from their mean curve are within ±2% [Fig. 10(c)], and the standard deviation is ∼1.0% to 2.0% over the scan range [Fig. 10(d)], confirming the stability and reproducibility of the monitor spectra.

FIG. 10.

(a) Monitor spectra, Imonitor(E), for a series of five single x-ray pulse measurements, showing a small pulse-to-pulse variation in maximum intensity. (b) Same monitor spectra normalized by the maximum intensity. (c) Percentage difference of each scan from the average of the five scans. (d) Standard deviation of the percentage difference for the five scans.

FIG. 10.

(a) Monitor spectra, Imonitor(E), for a series of five single x-ray pulse measurements, showing a small pulse-to-pulse variation in maximum intensity. (b) Same monitor spectra normalized by the maximum intensity. (c) Percentage difference of each scan from the average of the five scans. (d) Standard deviation of the percentage difference for the five scans.

Close modal

Figure 11 shows the normalized absorption spectra for Cu (K-edge 8.98 keV)30 and Au (L3-edge 11.92 keV).30 The single pulse x-ray measurements were performed on 5 μm thick Cu and Au foils under ambient conditions. Our data are compared with the reference data available from the APS XAFS spectra library.31 As pointed out previously, our measurements do not have high enough resolution to resolve the fine structures associated with XANES (the region about 10 eV below the absorption edge and ∼20 eV above the edge). Instead, our system has been designed and optimized for the EXAFS region. For both Cu and Au, clear EXAFS features can be observed up to ∼400 eV above the absorption edge, which are in good agreement with the reference spectra. The EXAFS signal χ(k) can be calculated by

χ(k)=[μ(k)μ0(k)]/Δμ0(k),
(12)

where k is the wave number of the photo-electron given by k=2m2(EE0), m is the electron mass, is the reduced Planck’s constant, μ(k) is the measured absorption coefficient, μ0(k) is the background absorption coefficient ignoring photoelectron scattering from neighboring atoms, and Δμ0(k) is the jump in the absorption coefficient at the threshold energy E0. Figure 12 shows the k2 weighted χ(k) for Cu and Au, calculated using the ATHENA software package.39 The results are compared to the k2χ(k) obtained for the reference data. The usable χ(k) is in the range k = 3 Å−1–10 Å−1 above which it becomes increasingly noisy. Over this range, the main oscillations and their amplitudes are in good agreement with the reference data for both Cu and Au. However, as expected from a single x-ray pulse measurement, our data are slightly noisier than the reference data. Below ∼3 Å−1 our observed EXAFS oscillation amplitudes, particularly for Au, are also significantly smaller than the reference data likely due to the lower spectral resolution in our single pulse configuration.40 

FIG. 11.

Normalized absorption spectra under ambient conditions for (a) Cu (K-edge 8.98 keV) and (b) Au (L3-edge 11.92 keV). For both plots, the reference data are shifted vertically for clarity.

FIG. 11.

Normalized absorption spectra under ambient conditions for (a) Cu (K-edge 8.98 keV) and (b) Au (L3-edge 11.92 keV). For both plots, the reference data are shifted vertically for clarity.

Close modal
FIG. 12.

Ambient EXAFS signal χ(k) with a k-weight of 2 for (a) Cu K-edge and (b) Au L3-edge.

FIG. 12.

Ambient EXAFS signal χ(k) with a k-weight of 2 for (a) Cu K-edge and (b) Au L3-edge.

Close modal

Figure 13(a) shows the experimental configuration for laser shock compression of a Ge(100) sample. The target consisted of a 50 μm thick aluminized Kapton ablator epoxy bonded to a ∼15 μm thick Ge(100) single crystal, which was mounted on an aluminum bar, as shown in Fig. 13(b). The Al bar had a countersunk through-hole located beneath the Ge/Kapton allowing the laser drive and x rays to directly reach the Kapton/Ge without reflection or absorption from the Al bar. The aluminum bar was attached to a rotating sample holder wheel, which itself was attached to a kinematic stage.29 Velocity interferometry measurements were performed using a central VISAR (Velocity Interferometer System for Any Reflector)41 probe with ∼600 ps time resolution and four Photon Doppler Velocimetry (PDV) probes.42 One PDV probe was reflecting from the center of the sample (overlapping with VISAR), and three PDV probes were reflecting from a circle around it on a 125 μm radius at 120° spacing around the circle. The VISAR spot size was 90 μm FWHM, and the PDV spot sizes were 20 μm. The PDV data were analyzed using a Fast Fourier Transform (FFT) window-size of 1.28 ns and a Hamming window-function of 0.67 ns FWHM. The x-ray monitor signal was measured with a calibration target consisting of 50 μm aluminized Kapton while the ambient Ge measurements were performed just before the sample was shocked. For shock compression, the aluminized side of the Kapton was ablated using a 10 ns duration laser pulse with ∼16 J of energy. The laser pulse is incident at 7° from the sample-normal in order to minimize the laser reflection from the ablator front surface back to the upstream optics and laser system. The temporal profile of the 10 ns duration laser pulse with 500 μm Distributed Phase Plate (DPP) is shown in the inset of Fig. 13(c). Laser ablation of the Kapton results in the formation of a 500 μm diameter plane shock wave in the Kapton, which then propagates into the Ge sample.

FIG. 13.

(a) Experimental configuration used for laser shock EXAFS measurements. The x-ray pulse was incident at an angle ϕ = 52° from the target plane. The aluminized side of the Kapton was ablated using a 10 ns duration laser pulse incident at 7° from the sample-normal, providing a peak stress of 29(2) GPa in the Ge sample. (b) Picture of Ge target mounted over a 2.26 mm diameter countersunk through-hole in a 5 mm wide by 41.4 mm long Al bar. The target consisted of a 50 μm thick aluminized Kapton ablator epoxy bonded to a ∼15 μm thick Ge(100) single crystal. The aluminized Kapton side of the Ge(100)/Kapton target package is glued to the Al bar covering the entire through hole. The Ge free surface side of the target is visible in (b). (c) The measured free surface velocity history of the Ge(100) sample was recorded using a velocity interferometer (VISAR)41 and four Photon Doppler Velocimetry (PDV) probes.42 The temporal profile of the 10 ns duration laser pulse with 500 μm The Distributed Phase Plate (DPP) is shown in the inset of Fig. 13(c).

FIG. 13.

(a) Experimental configuration used for laser shock EXAFS measurements. The x-ray pulse was incident at an angle ϕ = 52° from the target plane. The aluminized side of the Kapton was ablated using a 10 ns duration laser pulse incident at 7° from the sample-normal, providing a peak stress of 29(2) GPa in the Ge sample. (b) Picture of Ge target mounted over a 2.26 mm diameter countersunk through-hole in a 5 mm wide by 41.4 mm long Al bar. The target consisted of a 50 μm thick aluminized Kapton ablator epoxy bonded to a ∼15 μm thick Ge(100) single crystal. The aluminized Kapton side of the Ge(100)/Kapton target package is glued to the Al bar covering the entire through hole. The Ge free surface side of the target is visible in (b). (c) The measured free surface velocity history of the Ge(100) sample was recorded using a velocity interferometer (VISAR)41 and four Photon Doppler Velocimetry (PDV) probes.42 The temporal profile of the 10 ns duration laser pulse with 500 μm The Distributed Phase Plate (DPP) is shown in the inset of Fig. 13(c).

Close modal

The free surface velocity histories of the Ge(100) sample recorded using the VISAR41 and the four PDV probes42 are shown in Fig. 13(c). They show an elastic wave followed by a slower phase transformation wave. All four of the measured shock breakouts using the PDV probes were within 100 ps. This is close to the 40 ps/point PDV digitization rate and also close to the ±40 ps relative timing accuracy between PDV probes. We cannot resolve any spatial variations in shock breakout timing due to the possible target thickness variations over the 250 μm PDV probe diameter. However, we note that the spatial shock stress variations are expected to be less than 1% over the 250 μm PDV probe diameter since the average peak state free surface velocities from each PDV probe agree within ±1%. The peak state free surface velocities and phase change shock arrival times are also consistent between all four PDV probes and the VISAR probe. By comparing the free surface velocity after the phase transformation wave arrival with the previous results from Gust and Royce,43 the peak state stress is determined to be 29(2) GPa. For this peak stress, continuum results suggest transformation to a high-pressure structure,43 which was recently identified as the β-Sn structure using in situ XRD measurements in Ge(100) subjected to plate impact loading.44 

Figure 14 shows the single pulse x-ray absorption spectra from a Ge(100) single crystal prior to shock compression and during laser shock-compression, plotted as a function of photon energy transfer, EEk, where E is the incident energy and Ek = 11.1036 keV is the Ge K-edge. The ambient spectrum shows distinct EXAFS features up to 500 eV above the K-edge. The small peak around ∼570 eV energy transfer is due to extinction via Laue diffraction from the single crystal Ge. The laser shocked Ge spectrum at 29(2) GPa shows EXAFS features that are distinctly different from the ambient data. The absorption spectrum from the shocked Ge was obtained shortly before the phase transformation shock wave reached the rear surface of the Ge sample, so the majority of the Ge is expected to be in the β-Sn structure44 during the EXAFS measurement. Detailed analysis and fits to the EXAFS data on laser shocked Ge will be presented elsewhere. Here, we present our analysis of the ambient Ge spectrum.

FIG. 14.

Ambient (blue line) and laser-shocked (in red) absorption spectra measured at the Ge K-edge (Ek) from a single x-ray pulse. The laser-shock data were offset vertically by 0.2. Laser-shocked Ge with a peak stress of 29(2) GPa shows distinct changes from the ambient spectrum.

FIG. 14.

Ambient (blue line) and laser-shocked (in red) absorption spectra measured at the Ge K-edge (Ek) from a single x-ray pulse. The laser-shock data were offset vertically by 0.2. Laser-shocked Ge with a peak stress of 29(2) GPa shows distinct changes from the ambient spectrum.

Close modal

Quantitative analysis of the ambient Ge absorption spectrum is performed using the ATHENA and ARTEMIS software packages.39 The EXAFS signal χ(k) was obtained after removing the background and normalizing the data with the edge-step using ATHENA. The k-weighted EXAFS signal χ(k) is shown in Fig. 15(a), while the modulus of the Fourier transform of χ(k) is shown in Fig. 15(b). The simulated EXAFS signal was then fitted to the data using ARTEMIS. Ge is in the cubic diamond phase under ambient conditions.45 Back-scattering amplitudes and phases were thus calculated using the cubic diamond structural model, and the first three neighboring shell Ge–Ge paths were considered for the modeling. Assuming only volumetric lattice expansion, only one parameter is needed for varying the Ge–Ge distances during the fit. Additionally, the amplitude reduction factor S02, energy shift E0, and the Debye–Waller factors for the three paths σi2 (i = 1, 2, 3) were also allowed to vary. The coordination number, N, for each scattering path was fixed during the fitting. The Fourier transform from χ(k) to χ(R) was performed within a Hanning window of k-range 3.0 Å−1–11.5 Å−1. The fit to the model was performed in χ(R) space within a R-range of 1.25 Å–4.7 Å, as shown in Fig. 15(b). From the fit, S02 was found to be 0.93 ± 0.07, while E0 was 3.0 ± 0.8 eV. The fitted values of bond lengths (Rfit) and σ2 values are shown in Table II. The Debye–Waller factors, σi2, correspond to the mean-square relative displacement along the different bond directions, and the fitted values are within the reported range.46–48 The fitted bond lengths (Rfit) are in agreement with the known structural parameters (Reff) and published results.45,46 This is also represented in Fig. 15(b), where the three distinct measured peaks are contributions from the first three neighboring shell Ge–Ge paths; the simulated peaks provide an excellent match to the measured peaks.

FIG. 15.

(a) Measured Ge EXAFS signal χ(k) with a k-weight of 2 (in blue). (b) χ(R), modulus of the Fourier transform of χ(k), plotted in real space (in blue). Fitting of the simulated EXAFS to the measured EXAFS was performed in R-space. χ(k)model and χ(R)model are shown in red in (a) and (b), respectively.

FIG. 15.

(a) Measured Ge EXAFS signal χ(k) with a k-weight of 2 (in blue). (b) χ(R), modulus of the Fourier transform of χ(k), plotted in real space (in blue). Fitting of the simulated EXAFS to the measured EXAFS was performed in R-space. χ(k)model and χ(R)model are shown in red in (a) and (b), respectively.

Close modal
TABLE II.

Parameters used for fitting the EXAFS signal using the first three neighboring shell Ge–Ge paths within the cubic diamond structural model of ambient Ge. N is the coordination number, σ2 is the Debye–Waller factor, Reff are the known structural parameters,45 and Rfit are the fitted values.

Scattering pathsNσ22)46–48 σ22)Reff (Å)Rfit (Å)
Ge–Ge1 3–7 (10−33.5(5) (10−32.4494 2.444(5) 
Ge–Ge2 12 10–20 (10−314(1) (10−33.9999 3.992(8) 
Ge–Ge3 12 10–20 (10−317(3) (10−34.6903 4.68(1) 
Scattering pathsNσ22)46–48 σ22)Reff (Å)Rfit (Å)
Ge–Ge1 3–7 (10−33.5(5) (10−32.4494 2.444(5) 
Ge–Ge2 12 10–20 (10−314(1) (10−33.9999 3.992(8) 
Ge–Ge3 12 10–20 (10−317(3) (10−34.6903 4.68(1) 

We have demonstrated that high quality real-time single x-ray pulse EXAFS measurements on dynamically compressed materials can be performed at the DCS for materials with absorption edges ranging from ∼9 keV to 13 keV. Using a flat plate of HOPG as the spectrometer element to energy disperse x rays, in combination with a high quantum efficiency detector (>90%), we have obtained good SNRs (∼103), an energy resolution of ∼10 eV at 10 keV, and pulse to pulse reproducibility (<2.0% SD) of our single pulse EXAFS measurements. This can be confirmed from the comparison of ambient and reference spectra of Cu and Au. The energy resolution of ∼10 eV is good enough for the EXAFS experiment, but not sufficient for XANES. The lower spectral resolution also results in the reduced amplitude of EXAFS oscillations in the low-k region as observed in the Au spectrum, where the amplitude of k2χ(k) agrees poorly in the low-k region but matches well in the high-k region.40 This will not affect the evaluation of the atomic bond distances but may affect the coordination numbers and Debye–Waller factors extracted from the fitting process.40 However, we note that for the ambient Ge EXAFS data presented here, the Debye–Waller factors obtained from fitting are within the ranges of the previously published values.46–48 

Based on our EXAFS measurements using the new DCS system on a variety of elements with absorption edges between 9 keV and 13 keV, clear single pulse EXAFS oscillations were observed up to ∼350 eV to 550 eV beyond the absorption edge, depending on the material. For example, for Cu, Ge, and Au, clear EXAFS oscillations were observed up to 400 eV, 550 eV, and 350 eV beyond the edge, respectively. The useable ranges of EXAFS oscillations were all smaller than expected based on the FWHM of the HOPG reflectivity (red curve in Fig. 5). This is, in part, because the FWHM of the undulator flux spectrum is narrower than the HOPG reflectivity FWHM. In addition, we had less than optimal matching between undulator energies and HOPG spectrometer angles for some of our measurements (for example, Au). For the Ge measurements, the useable EXAFS energy range was ∼400 eV smaller than expected based on the FWHM of the HOPG reflectivity curve at 11.5 keV. A potential option for expanding the useable EXAFS energy range is to taper the U27 undulator gap. This will increase the bandwidth of incident x rays to better match the width of the HOPG reflectivity, provided the HOPG center angle (energy) and the undulator energies are well matched. However, tapering the undulator will reduce the number of incident x rays/eV and will lower the SNR. Because tapering the undulator to extend the useable EXAFS range has not yet been explored experimentally at the DCS, future DCS users can safely expect ∼350 eV to 550 eV of useable EXAFS oscillations for absorption edges between 9 keV and 13 keV.

We have performed EXAFS measurements on a shock-compressed material for energies above 8 keV for the first time at the Ge K-edge of 11.1036 keV. Fitting a simulated ambient Ge EXAFS spectrum to the ambient Ge data gives an excellent match using the known structural parameters. The EXAFS spectrum obtained for the shock compressed Ge exhibits substantial changes from the ambient Ge spectra, indicating significant changes in the local structure of the shock compressed Ge.

For reference, the DCS 100 J laser can currently provide peak stresses between ∼30 GPa and 400 GPa, depending on the sample material, through single shock loading. As demonstrated in the present work for Ge, at the lower end of the accessible stress range for shocked Ge, the increased temperature due to shock compression does not diminish the EXAFS oscillation amplitudes significantly. The EXAFS Debye–Waller factor decreases with compression and increases with temperature, and for a single material phase, EXAFS oscillation amplitudes will increase up to a certain shock stress, followed by a rapid reduction with shock stress when the material becomes stiffer and temperature increases rapidly with shock stress. For high enough shock stress, the material will melt, resulting in even weaker EXAFS oscillations. The 100 J laser was also designed for ramp compression, which results in lower temperature for a given peak stress. Future laser compression experiments with the new EXAFS capability at the DCS will elucidate the degree of thermal disorder that can be quantitatively examined with the current system. Finally, the EXAFS capability presented here—combined with the DCS laser shock capability—is ready for user experiments.

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Dr. J. Hawreliak and Dr. S. Sharma are thanked for useful discussions regarding this work. This publication is based on the work performed at the Dynamic Compression Sector, which is operated by Washington State University under the U.S. Department of Energy (DOE), National Nuclear Security Administration (NNSA) Award No. DE-NA0002442. This research used resources of the Advanced Photon Source, a DOE Office of Science User Facility operated for the DOE Office of Science by Argonne National Laboratory under Contract No. DE-AC02-06CH11357. This publication is also based on work supported by the DOE NNSA under Award No. DE-NA0002007.

1.
G. E.
Duvall
and
R. A.
Graham
, “
Phase transitions under shock-wave loading
,”
Rev. Mod. Phys.
49
,
523
(
1977
).
2.
R. G.
McQueen
,
S. P.
Marsh
,
J. W.
Taylor
,
J. N.
Fritz
, and
W. J.
Carter
, in
High Velocity Impact Phenomena
, edited by
R.
Kinslow
(
Academic Press
,
New York
,
1970
), p.
293
.
3.
J. M.
Brown
and
R. G.
McQueen
, “
Phase transitions, Grüneisen parameter, and elasticity for shocked iron between 77 GPa and 400 GPa
,”
J. Geophys. Res.
91
,
7485
, (
1986
).
4.
M. D.
Knudson
,
M. P.
Desjarlais
, and
D. H.
Dolan
, “
Shock-wave exploration of the high-pressure phases of carbon
,”
Science
322
,
1822
(
2008
).
5.
S. J.
Turneaure
,
N.
Sinclair
, and
Y. M.
Gupta
, “
Real-time examination of atomistic mechanisms during shock-induced structural transformation in silicon
,”
Phys. Rev. Lett.
117
,
045502
(
2016
).
6.
A. E.
Gleason
,
C. A.
Bolme
,
H. J.
Lee
,
B.
Nagler
,
E.
Galtier
,
D.
Milathianaki
,
J.
Hawreliak
,
R. G.
Kraus
,
J. H.
Eggert
,
D. E.
Fratanduono
,
G. W.
Collins
,
R.
Sandberg
,
W.
Yang
, and
W. L.
Mao
, “
Ultrafast visualization of crystallization and grain growth in shock-compressed SiO2
,”
Nat. Commun.
6
,
8191
(
2015
).
7.
C. E.
Wehrenberg
,
D.
McGonegle
,
C.
Bolme
,
A.
Higginbotham
,
A.
Lazicki
,
H. J.
Lee
,
B.
Nagler
,
H.-S.
Park
,
B. A.
Remington
,
R. E.
Rudd
,
M.
Sliwa
,
M.
Suggit
,
D.
Swift
,
F.
Tavella
,
L.
Zepeda-Ruiz
, and
J. S.
Wark
, “
In situ X-ray diffraction measurement of shock-wave-driven twinning and lattice dynamics
,”
Nature
550
,
496
499
(
2017
).
8.
S. M.
Sharma
,
S. J.
Turneaure
,
J. M.
Winey
,
P. A.
Rigg
,
N.
Sinclair
,
X.
Wang
,
Y.
Toyoda
, and
Y. M.
Gupta
, “
Real-time observation of stacking faults in gold shock compressed to 150 GPa
,”
Phys. Rev. X
10
,
011010
(
2020
).
9.
S. J.
Turneaure
,
S. M.
Sharma
, and
Y. M.
Gupta
, “
Nanosecond melting and recrystallization in shock-compressed silicon
,”
Phys. Rev. Lett.
121
,
135701
(
2018
).
10.
B.
Yaakobi
,
D. D.
Meyerhofer
,
T. R.
Boehly
,
J. J.
Rehr
,
B. A.
Remington
,
P. G.
Allen
,
S. M.
Pollaine
, and
R. C.
Albers
, “
Extended x-ray absorption fine structure measurements of laser-shocked V and Ti and crystal phase transformation in Ti
,”
Phys. Rev. Lett.
92
,
095504
(
2004
).
11.
B.
Yaakobi
,
D. D.
Meyerhofer
,
T. R.
Boehly
,
J. J.
Rehr
,
B. A.
Remington
,
P. G.
Allen
,
S. M.
Pollaine
, and
R. C.
Albers
, “
Extended x-ray absorption fine structure measurements of laser shocks in Ti and V and phase transformation in Ti
,”
Phys. Plasmas
11
,
2688
(
2004
).
12.
B.
Yaakobi
,
T. R.
Boehly
,
D. D.
Meyerhofer
,
T. J. B.
Collins
,
B. A.
Remington
,
P. G.
Allen
,
S. M.
Pollaine
,
H. E.
Lorenzana
, and
J. H.
Eggert
, “
EXAFS measurement of iron bcc-to-hcp phase transformation in nanosecond-laser shocks
,”
Phys. Rev. Lett.
95
,
075501
(
2005
).
13.
B.
Yaakobi
,
T. R.
Boehly
,
D. D.
Meyerhofer
,
T. J. B.
Collins
,
B. A.
Remington
,
P. G.
Allen
,
S. M.
Pollaine
,
H. E.
Lorenzana
, and
J. H.
Eggert
, “
Extended x-ray absorption fine structure measurement of phase transformation in iron shocked by nanosecond laser
,”
Phys. Plasmas
12
,
092703
(
2005
).
14.
B.
Yaakobi
,
T. R.
Boehly
,
T. C.
Sangster
,
D. D.
Meyerhofer
,
B. A.
Remington
,
P. G.
Allen
,
S. M.
Pollaine
,
H. E.
Lorenzana
,
K. T.
Lorenz
, and
J. A.
Hawreliak
, “
Extended x-ray absorption fine structure measurements of quasi-isentropically compressed vanadium targets on the OMEGA laser
,”
Phys. Plasmas
15
,
062703
(
2008
).
15.
Y.
Ping
,
D. G.
Hicks
,
B.
Yaakobi
,
F.
Coppari
,
J.
Eggert
, and
G. W.
Collins
, “
A platform for x-ray absorption fine structure study of dynamically compressed materials above 1 Mbar
,”
Rev. Sci. Instrum.
84
,
123105
(
2013
).
16.
Y.
Ping
,
F.
Coppari
,
D. G.
Hicks
,
B.
Yaakobi
,
D. E.
Fratanduono
,
S.
Hamel
,
J. H.
Eggert
,
J. R.
Rygg
,
R. F.
Smith
,
D. C.
Swift
,
D. G.
Braun
,
T. R.
Boehly
, and
G. W.
Collins
, “
Solid iron compressed up to 560 GPa
,”
Phys. Rev. Lett.
111
,
065501
(
2013
).
17.
Y.
Ping
and
F.
Coppari
, “
Laser shock XAFS studies at OMEGA facility
,”
High Pressure Research
36
,
303
(
2016
).
18.
M.
Harmand
,
A.
Ravasio
,
S.
Mazevet
,
J.
Bouchet
,
A.
Denoeud
,
F.
Dorchies
,
Y.
Feng
,
C.
Fourment
,
E.
Galtier
,
J.
Gaudin
,
F.
Guyot
,
R.
Kodama
,
M.
Koenig
,
H. J.
Lee
,
K.
Miyanishi
,
G.
Morard
,
R.
Musella
,
B.
Nagler
,
M.
Nakatsutsumi
,
N.
Ozaki
,
V.
Recoules
,
S.
Toleikis
,
T.
Vinci
,
U.
Zastrau
,
D.
Zhu
, and
A.
Benuzzi-Mounaix
, “
X-ray absorption spectroscopy of iron at multimegabar pressures in laser shock experiments
,”
Phys. Rev. B
92
,
024108
(
2015
).
19.
P. A.
Lee
,
P. H.
Citrin
,
P.
Eisenberger
, and
B. M.
Kincaid
, “
Extended x-ray absorption fine structure—its strengths and limitations as a structural tool
,”
Rev. Mod. Phys.
53
,
769
(
1981
).
20.
J. J.
Rehr
and
R. C.
Albers
, “
Theoretical approaches to x-ray absorption fine structure
,”
Rev. Mod. Phys.
72
,
621
(
2000
).
21.
G.
Bunker
,
Introduction to XAFS: A Practical Guide to X-Ray Absorption Fine Structure Spectroscopy
(
Cambridge University Press
,
Cambridge, UK
,
2010
).
22.
M.
Newville
, “
Fundamentals of XAFS
,” in
Consortium for Advanced Radiation Sources
(
University of Chicago
,
Chicago, IL
,
2004
).
23.
B.
Yaakobi
,
F. J.
Marshall
,
T. R.
Boehly
,
R. P. J.
Town
, and
D. D.
Meyerhofer
, “
Extended x-ray absorption fine-structure experiments with a laser-imploded target as a radiation source
,”
J. Opt. Soc. Am. B
20
,
238
(
2003
).
24.
F.
Coppari
,
D. B.
Thorn
,
G. E.
Kemp
,
R. S.
Craxton
,
E. M.
Garcia
,
Y.
Ping
,
J. H.
Eggert
, and
M. B.
Schneider
, “
X-ray source development for EXAFS measurements on the National Ignition Facility
,”
Rev. Sci. Instrum.
88
,
083907
(
2017
).
25.
A.
Krygier
,
F.
Coppari
,
G. E.
Kemp
,
D. B.
Thorn
,
R. S.
Craxton
,
J. H.
Eggert
,
E. M.
Garcia
,
J. M.
McNaney
,
H.-S.
Park
,
Y.
Ping
,
B. A.
Remington
, and
M. B.
Schneider
, “
Developing a high-flux, high-energy continuum backlighter for extended x-ray absorption fine structure measurements at the National Ignition Facility
,”
Rev. Sci. Instrum.
89
,
10F114
(
2018
).
26.
S.
Pascarelli
,
O.
Mathon
,
T.
Mairs
,
I.
Kantor
,
G.
Agostini
,
C.
Strohm
,
S.
Pasternak
,
F.
Perrin
,
G.
Berruyer
,
P.
Chappelet
,
C.
Clavel
, and
M. C.
Dominguez
, “
The time-resolved and extreme-conditions XAS (TEXAS) facility at the European Synchrotron Radiation Facility: The energy-dispersive X-ray absorption spectroscopy beamline ID24
,”
J. Synchrotron Radiat.
23
,
353
(
2016
).
27.
O.
Mathon
,
F.
Occelli
,
E.
Lescoute
,
A.
Sollier
,
P.
Loubeyre
,
W.
Helsby
,
J.
Headspith
,
R.
Torchio
,
I.
Kantor
, and
S.
Pascarelli
, “
High pressure dynamic XAS studies using an energy-dispersive spectrometer
,”
High Pressure Res.
36
,
404
(
2016
).
28.
R.
Torchio
,
F.
Occelli
,
O.
Mathon
,
A.
Sollier
,
E.
Lescoute
,
L.
Videau
,
T.
Vinci
,
A.
Benuzzi-Mounaix
,
J.
Headspith
,
W.
Helsby
,
S.
Bland
,
D.
Eakins
,
D.
Chapman
,
S.
Pascarelli
, and
P.
Loubeyre
, “
Probing local and electronic structure in warm dense matter: Single pulse synchrotron x-ray absorption spectroscopy on shocked Fe
,”
Sci. Rep.
6
,
26402
(
2016
).
29.
X.
Wang
,
P.
Rigg
,
J.
Sethian
,
N.
Sinclair
,
N.
Weir
,
B.
Williams
,
J.
Zhang
,
J.
Hawreliak
,
Y.
Toyoda
,
Y.
Gupta
,
Y.
Li
,
D.
Broege
,
J.
Bromage
,
R.
Earley
,
D.
Guy
, and
J.
Zuegel
, “
The laser shock station in the dynamic compression sector. I
,”
Rev. Sci. Instrum.
90
,
053901
(
2019
).
30.
See http://physics.nist.gov/PhysRefData/XrayTrans/index.html for x-ray transition energies database.
32.
G. E.
Ice
and
C. J.
Sparks
, “
Mosaic crystal x-ray spectrometer to resolve inelastic background from anomalous scattering experiments
,”
Nucl. Instrum. Methods Phys. Res., Sect. A
291
,
110
(
1990
).
33.
A. K.
Freund
,
A.
Munkholm
, and
S.
Brennan
, “
X-ray diffraction properties of highly oriented pyrolytic graphite
,”
Proc. SPIE
2856
,
68
(
1996
).
34.
B. E.
Warren
,
X-Ray Diffraction
(
Dover Publications, Inc.
,
NY
,
1990
).
35.
M.
Sanchez del Rio
and
R. J.
Dejus
, “
Status of XOP: An x-ray optics software toolkit
,”
Proc. SPIE
5536
,
171
(
2004
).
36.
H.
Legall
,
H.
Stiel
,
V.
Arkadiev
, and
A. A.
Bjeoumikhov
, “
High spectral resolution x-ray optics with highly oriented pyrolytic graphite
,”
Opt. Express
14
(
10
),
4570
4576
(
2006
).
37.
X. H.
Yuan
,
D. C.
Carroll
,
M.
Coury
,
R. J.
Gray
,
C. M.
Brenner
,
X. X.
Lin
,
Y. T.
Li
,
M. N.
Quinn
,
O.
Tresca
,
B.
Zielbauer
,
D.
Neely
, and
P.
McKenna
, “
Spatially resolved X-ray spectroscopy using a flat HOPG crystal
,”
Nucl. Instrum. Methods Phys. Res., Sect. A
653
,
145
(
2011
).
39.
B.
Ravel
and
M.
Newville
, “
ATHENA, ARTEMIS, HEPHAESTUS: Data analysis for X-ray absorption spectroscopy using IFEFFIT
,”
J. Synchrotron Radiat.
12
,
537
(
2005
).
40.
G. G.
Li
,
F.
Bridges
, and
C. H.
Booth
, “
X-ray-absorption fine-structure standards: A comparison of experiment and theory
,”
Phys. Rev. B
52
,
6332
(
1995
).
41.
L. M.
Barker
and
R. E.
Hollenbach
, “
Laser interferometer for measuring high velocities of any reflecting surface
,”
J. Appl. Phys.
43
,
4669
(
1972
).
42.
O. T.
Strand
,
D. R.
Goosman
,
C.
Martinez
,
T. L.
Whitworth
, and
W. W.
Kuhlow
, “
Compact system for high-speed velocimetry using heterodyne techniques
,”
Rev. Sci. Instrum.
77
,
083108
(
2006
).
43.
W. H.
Gust
and
E. B.
Royce
, “
Axial yield strengths and phase-transition stresses for ⟨100⟩, ⟨110⟩, and ⟨111⟩ germanium
,”
J. Appl. Phys.
43
,
4437
(
1972
).
44.
P.
Renganathan
,
S. J.
Turneaure
,
S. M.
Sharma
, and
Y. M.
Gupta
, “
Structural transformations including melting and recrystallization during shock compression and release of germanium up to 45 GPa
,”
Phys. Rev. B
99
,
134101
(
2019
).
45.
A. S.
Cooper
, “
Precise lattice constants of germanium, aluminum, gallium arsenide, uranium, sulphur, quartz and sapphire
,”
Acta Crystallogr.
15
,
578
582
(
1962
).
46.
M.
Baldini
,
W.
Yang
,
G.
Aquilanti
,
L.
Zhang
,
Y.
Ding
,
S.
Pascarelli
, and
W. L.
Mao
, “
High-pressure EXAFS measurements of crystalline Ge using nanocrystalline diamond anvils
,”
Phys. Rev. B
84
,
014111
(
2011
).
47.
G.
Dalba
and
P.
Fornasini
, “
EXAFS Debye-Waller factor and thermal vibrations of crystals
,”
J. Synchrotron Radiat.
4
,
243
(
1997
).
48.
A.
Yoshiasa
,
T.
Nagai
,
O.
Ohtaka
,
O.
Kamishima
, and
O.
Shimomura
, “
Pressure and temperature dependence of EXAFS Debye-Waller factors in diamond-type and white-tin-type germanium
,”
J. Synchrotron Radiat.
6
,
43
(
1999
).