A new type of in-vessel Penning gauge, the Wisconsin In Situ Penning (WISP) gauge, has been developed and successfully operated in the Wendelstein 7-X (W7-X) island divertor baffle and vacuum vessel. The capacity of the quantitative measurements of the neutral reservoir for light impurities, in particular, helium, is important for tokamaks as well as stellarator divertors in order to avoid fuel dilution and radiative energy loss. Penning gauges assisted by spectroscopy are a powerful tool to obtain the total neutral pressure as well as fractional neutral pressures of specific impurities. The WISP gauge is a miniaturized Penning gauge arrangement, which exploits the ambient magnetic field of magnetic confinement fusion experiments to establish the Penning discharge. Then, in situ spectroscopy is conducted to separate the fractional neutral pressures of hydrogen, helium, and possibly also other impurities. The WISP probe head was qualified using the magnetic field of the Magnetized Dusty Plasma Experiment at Auburn University between 0.25 T and 3.5 T [E. Thomas et al., J. Plasma Phys. 81, 345810206 (2015)]. The in-depth quantitative evaluation for hydrogen and helium will be shown as well as an exploration of nitrogen, argon, and neon. A power law scaling between current I and pressure p, I = f(Gas,V) · pn(Gas, B), was shown. The factor f is gas and anode potential dependent, while n is gas and magnetic field strength dependent. Pressure measurements from 0.1 mbar and down to 1 × 10−5 mbar were achieved, demonstrating a reliable operating range for relevant pressure levels in the divertor and main vessel regions in current and future fusion devices, with a time resolution of up to 1 kHz. The lowest achievable pressure measurement increases with an increase in B and can be shifted with the anode potential V. At W7-X, the WISP probe head was mounted on an immersion tube setup that passes through the cryostat and places the probe head close to the plasma. Two probe heads were positioned in different divertor pump gaps, top and bottom, and one close to the plasma on the midplane in one module. The gauges were in situ calibrated together with the ASDEX pressure gauges [G. Haas and H.-S. Bosch, Vacuum 51, 39 (1998)]. Data were taken during the entire operation phase 1.2b, and measurements were coherent with other neutral gas pressure gauges. For the spectroscopic partial pressure measurements, channels of a spectroscopic detection system based on photo-multipliers, a so-called filterscope [R. J. Colchin et al., Rev. Sci. Instrum. 74, 2068 (2003)], provided by the Oak Ridge National Lab were used.

The fusion of deuterium (D) and tritium (T) in a burning fusion plasma will continuously produce alpha particles, leading to helium (He) ash. The burn condition for D–T plasmas limits the maximum He concentration to about 10%, making a sufficient He exhaust mandatory for a reactor.1 The He exhaust can be divided into different regions, with each region requiring different diagnostics to measure the He concentration in it. The α particles are created in the plasma core from where they have to be transported into the edge and then outside the last closed flux surface (LCFS), neutralized at a divertor, pumped through the pumping gap, and eventually removed from the device through a pump duct. Charge exchange spectroscopy can be used to measure the He concentration in the plasma core,2 edge, and LCFS.3 Outside of the machine in the pump duct, several diagnostics have been developed to measure He partial pressures in a D2 environment. Even though many residual gas analyzers (RGAs) that function on the principle of quadrupole mass spectrometers (QMSs) are struggling to resolve the mass difference between He at 4.003 amu and D2 at 4.028 amu, there are some commercial RGAs that claim to be capable of doing so. However, they are sensitive to an ambient magnetic field and thus might have to be mounted further away. Another technique is to excite the neutral gas while observing the discharge spectroscopically and using the line intensities for partial pressure measurements. Optical observations have been used on ASDEX-type hot cathode ion gauges,4 while a comparison of different electron sources to excite the neutral gas has found that the best line emitting light source is a Penning discharge.5 Penning discharges6 in combination with a spectroscopic optical observation have been used successfully at TEXTOR,7 DIII-D,8 ASDEX-U, and JET.9 However, all these techniques have to be mounted on the outer periphery of the device to avoid interference with the devices’ magnetic field. Due to this, these techniques commonly struggle with having a response time compatible with the plasma particle and impurity lifetimes in the divertor region. However, clever vacuum engineering can mitigate this effect.10 The Wisconsin In Situ Penning (WISP) gauge was developed to fill this gap and offers a first in situ partial neutral pressure measurement in magnetic fusion devices.

This paper has two main sections: In Sec. II, the WISP probe head itself as well as an in-depth study of its behavior in different operating regimes are described. Section III describes how the probe head was implemented at Wendelstein 7-X (W7-X) and how it was compared to ASDEX-type hot cathode ion gauges.

A previous anode design study described in Ref. 11 for in situ Penning gauges was extended to develop the final probe head. The ring anode design was chosen over other geometries due to its superior performance in terms of stability and photon yield. The WISP probe head consists of a ring anode surrounded by a cathode plate. The anode and cathode are mechanically connected by a base plate where a lens is mounted and protected by a pinhole, as shown in the computer aided design (CAD) model of the probe head in Figs. 1 and 2. The anode ring has an outer diameter of 9 mm and a 3 mm × 3 mm cross section. The cathode plate is made out of a bent plate with a thickness of 1.5 mm. The anode, cathode, and base plate are all made out of stainless steel. The electrical insulation is achieved by custom boron nitride pieces and ceramic beads. The cathode plates are 17 mm apart from each other with the ring anode mounted in the center. The anode and cathode are supported from the side to keep the bottom view clear for optical observation. To create a well-defined optical observation area, a combination of a pinhole camera with a collimator lens is used. The pinhole camera serves two functions. The main purpose is to protect the lens from deposition, which is achieved by making the hole as deep as its 2 mm diameter. The lens was then moved further back to increase the screening effect. The second function is that the combination of the pinhole together with the lens enables a well-defined field of view. The lens was positioned in a way that the lines of sight go through the anode ring, and the majority of the discharge can be observed.

FIG. 1.

CAD model of the WISP probe head showing (a) front, (b) side, (c) top, and (d) isometric views.

FIG. 1.

CAD model of the WISP probe head showing (a) front, (b) side, (c) top, and (d) isometric views.

Close modal
FIG. 2.

Picture of a cross section of the WISP probe head with the lens in green, pinhole section in red, and the field of view lines in blue.

FIG. 2.

Picture of a cross section of the WISP probe head with the lens in green, pinhole section in red, and the field of view lines in blue.

Close modal

Commercial Penning gauges use permanent magnets to create their magnetic field for electron trapping. This can lead to interference when the gauge is operated close to the field of magnetic confinement machines. The WISP probe head avoids this issue by utilizing the present magnetic field of the device, and therefore, it can be placed directly inside the machine as long as the magnetic field vector is co-axial to the anode ring. A potential misalignment due to tolerances in the assembly or changes in the magnetic field is discussed in Subsection III E. A picture of the probe head on an adjustment unit to align it in the magnetic field is shown in Fig. 3. For easier installation, all connections were moved to the bottom of the base plate. In operation, a high voltage is applied to the anode to create the electric field. For noise reduction, the cathode plate is also insulated from the base plate, therefore also from the machine ground, and has its own return. An optical fiber is connected to the collimator lens via an SMA905 (SubMiniature Version A905) connector and transmits the light to the spectroscopic device of choice.

FIG. 3.

Picture of a WISP probe head on top of an adjustment unit to align it to the magnetic field vector.

FIG. 3.

Picture of a WISP probe head on top of an adjustment unit to align it to the magnetic field vector.

Close modal

The gauge has been designed and tested to resist the harsh environment of fusion devices. All materials and components are capable to be used under high vacuum conditions without outgassing. The lens, as the component with the lowest maximum temperature rating, is rated for temperatures of up to 180 °C and thus allows the gauge to be baked at the typical temperatures for vacuum conditioning of about 150 °C. The metals used have a magnetic permeability of μr < 1.01 to not perturb the local magnetic field, and thus the measurement, and to avoid damaging the gauge due to a net force in the field. The cobalt concentration was kept below 0.05% to minimize activation in a neutron environment.

The WISP probe head went through an in-depth testing at the Magnetized Dusty Plasma Experiment (MDPX) at Auburn University.12 The superconducting MDPX magnet offers the unique capability of a freely adjustable magnetic field strength between 0 T and 3.5 T. The vacuum chamber was replaced for the systematic testing described in this paper with an ultrahigh vacuum chamber based on a six-way ConFlat cross and ConFlat components. For pressure measurements and vacuum monitoring, an Agilent FRG-700 Pirani Inverted Magnetron gauge together with an RGA was mounted, away from the field, close to the pump. The actual chamber was equipped with a MKS 722 B baratron (1 × 103 mbar to 1 × 10−4 mbar) and a KJLC 354 ion gauge (6.7 × 10−2 mbar to 1.3 × 10−9 mbar), measuring a base pressure of 2.6 × 10−6 mbar. All following pressure measurements were performed with the baratron since the ion gauge could not be operated in magnetic fields and was only used for vacuum verification before the field was ramped up. The baratron was positioned in close vicinity of the WISP probe head so that dynamic effects can be neglected. A He and a H gas feed were connected to the chamber, which were controlled using mass flow controllers (MFCs). Two high voltage power supplies were used to apply the high voltage to the anode and to measure the ion current. Most measurements were taken with a voltage of 0–5 kV at a maximum of 60 mA power supply, while a voltage of 0–2 kV at a maximum of 150 mA power supply was used to determine the maximum measurable pressure. The spectroscopy was performed using a universal serial bus (USB) CCD spectrometer with an integration time of 600 ms and monitored the Hα line and four different He–I lines at 587 nm, 668 nm, 706 nm, and 728 nm based on previous studies.7 The WISP probe head was tested at eight different field strengths (Subsection III A), at seven different anode potentials (Subsection III B), and in H and He (Subsection III C). Different mixtures of H and He (Subsection III D) and a systematic misalignment of the probe head to the field to determine mounting tolerances as well as effects of changes in the magnetic field direction (Subsection III E) were tested. Due to the large amount of data taken, only a few exemplary sets will be discussed in each section. At W7-X, calibration was done at the beginning and at the end of the campaign, and the aging effects are discussed in Subsection III F.

The ion current I behavior, which is the measurement proportional to the neutral pressure p, can be approximated by a power law scaling,

I=f(Gas, V)pn(Gas, B).
(1)

However, at the lower end of the operating regime, the currents are lower in H, and higher in He, than those predicted by the scaling. The line intensities are proportional to the ion current. However, this systematic deviation was only seen at MDPX and neither at the PAX magnet13 nor in the calibration at W7-X, where the measurements were true to the scaling even at the lowest measureable pressures. It can, therefore, be concluded that this deviation is unique to the experimental setup at MDPX. Because of this, the plots in this paper show the general behavior of the probe head in different scenarios but cannot serve as a calibration function. An in situ calibration in the measuring position at the particular device is inevitable.

Current (Fig. 4) as well as line intensity (Fig. 5) measurements over neutral pressure were taken at eight different magnetic field strengths, starting at 0.25 T, 0.5 T and then going up in half Tesla steps, up to 3.5 T. Each curve represents a different magnetic field strength going from weaker to stronger from left to right. The power law scaling from Eq. (1) fits well with all adjusted R-square values above 0.95 with the exception of the 0.512 T case that has an adjusted R-squared value of 0.78 and is likely due to a systematic error in the gas inlet/pressure measurement of that dataset. At low pressures, the measured current is higher for He, and lower for H, than that predicted by the fit as mentioned in Sec. III. This behavior is most likely due to the vacuum and gas feed system at MDPX since this behavior was not seen at other vacuum test stands or at W7-X. The error in current increases relative to the measured current at lower currents due to a systematic error of the ammeter, which is constant throughout its operating range. A more sensitive ammeter could be used to increase the accuracy at lower pressures. With an increase in the field, the lower end of the operating regime shrinks. The I(p) curves are getting steeper while reaching comparable upper pressures. The data in Fig. 5 are from the same settings as those in Fig. 4 but displays the 668 nm He–I line instead of the ion current. The line intensity is proportional to the ion current shown in Fig. 4 and follows the same characteristics.

FIG. 4.

Measured ion current for pressure sweeps in He at various B field strengths with the anode at 1500 V.

FIG. 4.

Measured ion current for pressure sweeps in He at various B field strengths with the anode at 1500 V.

Close modal
FIG. 5.

B field comparison of He 668 nm line intensity measurements over pressure of the WISP probe head in He at 1500 V with an integration time Tint = 600 ms.

FIG. 5.

B field comparison of He 668 nm line intensity measurements over pressure of the WISP probe head in He at 1500 V with an integration time Tint = 600 ms.

Close modal

The exponent n(Gas, B) of the power law scaling in Eq. (1) can be approximated as Eq. (2) for H and Eq. (3) for He based on fit parameters. The adjusted R2 values are 0.88 for H [Eq. (2)] and 0.86 [Eq. (3)] for He,

n=1.9B[T]+0.85,
(2)
n=0.8B[T]+1.40.
(3)

The WISP probe head was run at seven different anode potentials: 1.25 kV, 1.5 kV, 1.75 kV, 2 kV, 3 kV, 4 kV, and 5 kV, as shown in Figs. 6 and 7. Figure 6 shows the ion current and Fig. 7 the Hα line intensity plotted over the neutral pressure in H at a constant field of 1.024 T. The pressure sweeps were conducted for each of the seven anode potentials. The pressure curves can be approximated by the power law scaling (1) starting with the curve of the lowest potential on the right, going to the one with the highest potential on the left. The adjusted R-square values are above 0.95 for all fits; however, it should be noted that at the lower end of the operating regime, the power law scaling (1) overestimates the ion current. This effect is only seen in H, and the cause for this is still under investigation. The anode potential V can be used as an easy tool to shift the operating regime. A higher potential will increase the ion current, ignite a discharge at lower pressures, and therefore shift the measurement range to the left.

FIG. 6.

Ion current over pressure of the WISP probe head in H at 1.024 T at different potentials.

FIG. 6.

Ion current over pressure of the WISP probe head in H at 1.024 T at different potentials.

Close modal
FIG. 7.

Hα intensity over pressure of the WISP probe head in H at 1.024 T at different potentials.

FIG. 7.

Hα intensity over pressure of the WISP probe head in H at 1.024 T at different potentials.

Close modal

The ion current scales linearly as shown in the following equation:

f(V)=3.6V[kV].
(4)

For high pressures, a lower potential is desirable to minimize the power that is being deposited in the probe head and prevent overheating. A lower potential will generate a lower ion current at the same pressure as a higher potential. A lower potential directly lowers the power and also lowers the ion current at a given pressure and therefore decreases the input power P = V · I significantly. Overheating of the probe head can lead to a glow of the anode, which can interfere with the spectroscopic measurements and eventually melt and therefore destroy the anode.

Like any ion gauge, the WISP gauge ion current measurement is gas dependent. At the same neutral pressure, nitrogen (N) results in a larger ion current than H, which is larger than He. Figure 8 shows a comparison of H and He measurements at different magnetic field strengths. While commercial Penning gauges are usually calibrated to nitrogen and use a calibration factor for different gases, such a factor could not be identified for the WISP gauge. One reason for this is the gas dependence of the exponent, as shown in Eqs. (2) and (3). For a certain magnetic field strength and anode potential, an approximation can be made where, depending on the case, the ion current scales between IH = 4 · IHe and IH = 8 · IHe; however, no general function between the gas species and the factor f(Gas, V) from Eq. (1) could be found.

FIG. 8.

Comparison of H and He ion current measurements over pressure of the WISP probe head at 3000 V.

FIG. 8.

Comparison of H and He ion current measurements over pressure of the WISP probe head at 3000 V.

Close modal

While an ion gauge can use a calibration factor, or like in the case of the WISP gauge that can be calibrated for each gas species, all ion gauges struggle with accurate measurements in gas mixtures. While the effect is negligible for trace amounts, it becomes more severe with an increase in the fraction of the other gases. The WISP gauge can avoid this problem by using the spectroscopic signal as a partial pressure measurement, as described in Subsection III D.

For exhaust measurements, H/D and He partial pressure measurements are the most important, but Ne, N, and Ar measurements can be important since they are used for diagnostic use14 and edge cooling.15 The possible lines for partial pressure measurements, which were actually observed in the gauge, are listed in Table I.

TABLE I.

Observed atomic spectral lines in the visible range inside the WISP head in nm. Lines are listed from strongest to weakest.

HHeNeNAr
656.1 587.6 586.8 391.9 812.8 
 667.8 640.2 469.5 751.6 
 706.5 614.3 869.3 764.7 
 728.1 650.7 524.9 773.4 
HHeNeNAr
656.1 587.6 586.8 391.9 812.8 
 667.8 640.2 469.5 751.6 
 706.5 614.3 869.3 764.7 
 728.1 650.7 524.9 773.4 

As described in Subsection III C, the gas dependence of the ion current measurement increases the error of an ion gauge measurement in gas mixtures. This problem can be avoided with the WISP gauge or other optical Penning gauges by extracting partial pressures from ratios of spectral line intensities after calibration with known gas mixtures.

As long as all present species are accounted for, adding all the partial pressures together will result in an accurate total neutral pressure measurement independent of gas mixtures.

Figure 9 shows a pressure scan of He with a constant background pressure of H to study the crosstalk between the two gas species. Due to limitations in the vacuum and H supply system, only a limited flow rate of H was available. This leads to H becoming the minority species as the He partial pressure was continuously increased. While it is less common to operate optical Penning gauges with H as the minority species, it has been used to study hydrogen coming off the wall during He wall conditioning at Tokamak à configuration variable (TCV).16 This could lead to a different operation regime, since the ionization of He is providing the electrons to sustain the discharge,

pHpImpurity=AIHαIImpurity,
(5)
A=Linear H responseImpurity crosstalk into H lineH crosstalk into impurity lineLinear impurity response.
(6)

However, no mode switching or change in the operation regime was found at W7-X, where measurements of He as a minority and majority species were achieved. A scan going from 0% He to 50% He in a H environment is discussed in Fig. 21. In Fig. 9, four He-I lines and the Hα line are plotted over the He partial pressure. All He-I line intensities increase with pressure, while the Hα line intensity stays constant within normal variations.

FIG. 9.

Hα and He-I line intensities at Tint = 600 ms over pressure of the WISP probe head at 2 kV, 1.024 T, and 3 × 10−3 mbar of H.

FIG. 9.

Hα and He-I line intensities at Tint = 600 ms over pressure of the WISP probe head at 2 kV, 1.024 T, and 3 × 10−3 mbar of H.

Close modal

For partial pressure measurements, the lines have to be chosen carefully to avoid contamination through neighboring lines, such as the 586.8 nm Ne line and the 587.6 nm He line. There can also be crosstalk from molecular hydrogen at this wavelength, but the amount at the 667.8 nm line is negligible.7 All listed He-I lines show a very similar line intensity, as shown in Fig. 9, so any other line is preferred. Figures 9 and 10 also show that the Hα signal is not affected by the He. The H2 pressure is linear to the intensity of the Hα line directly determined from the measured spectra, as shown in Figs. 6 and 7. If the crosstalk cannot be avoided by the selection of lines, it will have to be accounted for with a calibration matrix [Eq. (6)], as shown in Eq. (5).

FIG. 10.

Spectra taken at Tint = 600 ms with the WISP probe head at 2 kV, 1.024 T, and a constant H partial pressure of 3 × 10−3 mbar. (a) is taken with a He partial pressure of 2 × 10−3 mbar and (b) with 1 × 10−2 mbar.

FIG. 10.

Spectra taken at Tint = 600 ms with the WISP probe head at 2 kV, 1.024 T, and a constant H partial pressure of 3 × 10−3 mbar. (a) is taken with a He partial pressure of 2 × 10−3 mbar and (b) with 1 × 10−2 mbar.

Close modal

The spectra of two data points from the He pressure scan with a constant H background pressure of 3 × 10−3 mbar are shown in Fig. 10. Spectrum (a) on the left is taken with a He partial pressure of 2 × 10−3 mbar, and the He partial pressure was increased to 1 × 10−2 mbar for spectrum (b). The Hα line and the four He-I lines from Table I are labeled. The increase in the He-I line intensities due to the increase in the He partial pressure shows no significant effect on the Hα intensity, which stays constant. The spectra also show the absence of disturbing lines and further support that the crosstalk between H and He is negligible.

To empirically test for the crosstalk between H and He, He pressure scans were done with different constant H background pressures. Figure 11 shows that the H crosstalk into any of the He lines is negligible. Crosstalk might be present at lower pressures as Fig. 11 could indicate that the lower He partial pressures are only accessible for lower H content. However, this is counter-intuitive as a lower H partial pressure also means a lower total pressure and therefore a reduction in the overall ion current and a weaker discharge. Future work and modeling will focus on assessing this behavior at lower pressures. Based on this, any line intensity is proportional to the partial pressure of He in the majority of the operating range and can be used for partial pressure measurements, as long as there are no other species with neighboring lines present. Care must be taken at the lower end of the operating regime until the crosstalk behavior is completely understood or it has to be accounted for with Eqs. (5) and (6).

FIG. 11.

He668 line intensity at Tint = 600 ms over pressure of the WISP probe head at 2 kV and different partial pressures of H.

FIG. 11.

He668 line intensity at Tint = 600 ms over pressure of the WISP probe head at 2 kV and different partial pressures of H.

Close modal

Optical studies on commercial Penning gauges in hydrogen observed a saturation in the Dα signal, and it was also shown to be different for the atomic and the molecular contributions to the excitation of the Balmer-α lines.17 The WISP probe head operated in a He majority shows no saturation up to 1 × 10−2 mbar, as shown in Figs. 9 and 11. Future work will try to investigate if this is due to the different geometry, the optimization of the Penning operational parameters, or the atomic physics.

In magnetic confinement machines, the magnetic field topology can change during discharges due to a number of reasons, including bootstrap currents and running a different field configuration. This can lead the magnetic field vector to not be co-axially aligned to the WISP probe head anymore. To study what change in the field direction is tolerable as well as to be able to define mounting tolerances, a systematic misalignment test was conducted at 0°, 5°,10°, and 15°. Due to constraints with the vacuum system, these tests were done with a base pressure of 1 × 10−3 mbar. Helium was used to sweep the pressure above that, but because of the residual water and air in the system, these measurements are only comparable among each other. Therefore, measurements were repeated with these conditions for the 0° alignment. The four different angles were tested at all seven anode potentials described in Subsection III B. Figure 12 shows the results at 1.5 kV anode potential, showing the measured ion current during pressure curves for the four different angles. The behavior is exemplary for all other tested anode potentials. The curves of 0°, 5°, and 10° lay on top of each other, while the 15° curve shows a higher ion current. A small change in alignment of up to ±10° does not seem to have an effect on the ion current measurement, while the current at 15° increases by a factor of about 2.4. When changing the magnetic field direction relative to the probe head, the plasma column (Fig. 13) inside the ring anode follows the tilt of the field. At 15°, the plasma column is tilted so far that it touches the anode and brings that spot to a glow. This leads to secondary electron emissions, which is a likely reason for the increase in the ion current.

FIG. 12.

WISP probe head at 1.5 kV and a relative misalignment of 0°, 5°,10°, and 15° to the magnetic field vector.

FIG. 12.

WISP probe head at 1.5 kV and a relative misalignment of 0°, 5°,10°, and 15° to the magnetic field vector.

Close modal
FIG. 13.

Picture of the plasma column inside the anode ring in He. Opening of the pinhole camera can be seen between the anode and the lower cathode plate.

FIG. 13.

Picture of the plasma column inside the anode ring in He. Opening of the pinhole camera can be seen between the anode and the lower cathode plate.

Close modal

Penning gauges are known to deteriorate over time, making a re-calibration or an exchange of the anode and cathode necessary. Since the WISP probe head itself is cheap and the anode and cathode are welded in, it is more cost effective to change the entire probe head if necessary. When using a new probe head, the inside of the anode ring has some initial roughness to it that the plasma slowly polishes off. This was observed through the release of sparks in the first minutes of running a new probe head. This effect can be minimized by polishing the anode ring during manufacturing; however, it is recommended to run the probe head for 5–10 min at a relatively high pressure before calibrating it. A non-polished probe head will show a higher ion current than a polished one. If calibration is done before this pre-aging process, it will quickly be invalid. To study the effect of long term operation on the probe head, multiple in situ calibrations (Subsection IV C) were performed at W7-X throughout the operation phase 1.2b (OP1.2b).18 While the gauges were running throughout an entire runday, the pressure was only high enough to measure current during plasma discharges, but not during the intermittent times. Figure 14 shows the standard neutral gas calibration program at W7-X, in this case from the end of the campaign. The calibration procedure will be discussed in more detail in Subsection IV C. The WISP curves from ports AEI40 and AEI41 are unfiltered with the calibration factor from the very first calibration of the campaign but still agree with the pressure measured using the baratrons. The WISP gauge still shows the correct pressures and therefore was not affected by a significant aging throughout the 127 min of measurements. The noise and the offset in AEI40 at lower pressures are due to the ammeters used and will be replaced by the more sensitive ones for future campaigns.

FIG. 14.

The plot shows the pressure calibration curve that was conducted at the end of the campaign (ID: 20181010.001). The commercial baratrons serve as a reference that the other neutral gas gauges are calibrated against. In this plot, the WISP signal is calibrated to the same calibration curve that was conducted at the beginning of the campaign (ID: 20180703.507). The calibration is still valid.

FIG. 14.

The plot shows the pressure calibration curve that was conducted at the end of the campaign (ID: 20181010.001). The commercial baratrons serve as a reference that the other neutral gas gauges are calibrated against. In this plot, the WISP signal is calibrated to the same calibration curve that was conducted at the beginning of the campaign (ID: 20180703.507). The calibration is still valid.

Close modal

At W7-X, the WISP manometers were mounted at three positions in module 4. Due to the easier removal of the probe head for maintenance purposes, a plugin design, based on the design for the hot cathode ion gauges,19 was chosen: one in the top divertor pump gap (AEI41), one in the bottom divertor pump gap (AEI40), and one on the midplane (AEE41), allowing a good poloidal coverage in the module (Fig. 15).

FIG. 15.

(a) Cross section of module 4 of W7-X with the three mounting positions of the WISP gauge. (b) Cross section as a sketch with the plasma facing components and the relative positions of the WISP gauges.

FIG. 15.

(a) Cross section of module 4 of W7-X with the three mounting positions of the WISP gauge. (b) Cross section as a sketch with the plasma facing components and the relative positions of the WISP gauges.

Close modal

The plugins were mounted on DN63CF flanges and can be split into the following three main components: the actual probe head space in the front, an immersion tube that holds the probe head in place and provides the infrastructure, and an interface section that also serves as a vacuum barrier. Everything on the vacuum side is covered with stainless steel components with gaps <0.7 mm between parts to allow pumping but shield from electron cyclotron resonance heating (ECRH) stray radiation. To reduce the amount of noise in the signal of the current measurement, both the high voltage anode and the ground cathode return were completely insulated from the machine.

1. Probe head space

The probe head space is confined by a 100 mm by 58 mm stainless steel cylinder. The front plate is designed as a hole diaphragm (Fig. 16) made up of 75 0.7 mm holes that block out ECRH but allow a gas exchange. The probe head is mounted on an adjustment unit to be aligned to the magnetic field of the standard configuration, as shown in Fig. 3. The maximum directional change in the magnetic field between different configurations is 2.5° and is therefore well within the misalignment tolerances, as discussed in Subsection III E. The adjustment unit was manufactured out of copper to allow for a better heat transfer from the probe head to the immersion tube.

FIG. 16.

Picture of the hole diaphragm on the front plate of the probe head space.

FIG. 16.

Picture of the hole diaphragm on the front plate of the probe head space.

Close modal

2. Immersion tube

The immersion tube holds the probe head in its designated place and guides two copper wires and the fiber over 2.2 m from the feedthrough, through the cryostat, to the probe head space. The copper wires are insulated by ceramic tubes, ceramic pearls, and custom boron nitride covers at the screw joints. The center tube is at atmospheric pressure and could later be upgraded for active water cooling. This tube is surrounded by fins for stability with the slits that guide the fiber and wires, as shown in Fig. 17.

FIG. 17.

Picture of the inside of the immersion tube without the surrounding ECRH cover. The fins as well as the insulated copper wires can be seen from the feedthrough looking toward the probe head space.

FIG. 17.

Picture of the inside of the immersion tube without the surrounding ECRH cover. The fins as well as the insulated copper wires can be seen from the feedthrough looking toward the probe head space.

Close modal

3. Interface

The DN63CF flange serves as the vacuum barrier and has two electrical feedthroughs: one for the anode potential and one for the cathode return, which are both rated for 5 kV. A DN16CF flange was added to the SMA fiber feedthrough, as shown in Fig. 18. On the air side, the two electrical feedthroughs are connected to a save high voltage (SHV) connector.

FIG. 18.

Cross section of the interface section with the SHV and SMA connectors.

FIG. 18.

Cross section of the interface section with the SHV and SMA connectors.

Close modal

On the air side, a 1 mm fiber is attached to the DN16CF fiber feedthrough, which guides the light of the WISP discharge into a filterscope20 in the spectroscopy laboratory. The incoming beam is split into four separate beams for the four filterscope channels each gauge uses. Each channel is equipped with a narrow band pass filter (±1 nm) so that the photo-multiplier tubes will only detect the desired wavelengths. For H measurements, an Hα filter was used, and for He measurements, a 667.8 nm, a 706.5 nm, and a 728.1 nm filter was used. Therefore, each channel monitors one single spectral line and measures its intensity. While this approach allows for large light throughput, it can suffer from interference from molecular features and background at He concentrations below 1%. At such low concentrations, the neutral He emission line gets weaker and the molecular spectra of hydrogen isotopes can dominate the emission within the spectral band around the He-I line. Instead of using filtered photodetectors, another study utilized dispersive spectroscopy to avoid such interference and achieved measurements down to the 0.1% level.21 

The calibration of the WISP gauges for OP1.2b was conducted in situ due to the magnetic field dependence of the measurement principle. This allowed to correct for potential misalignment along the magnetic field vector. It was conducted multiple times for 100% hydrogen and 100% helium, once at the beginning and once at the end of the campaign. Additionally, partial pressure calibrations were conducted in 5%, 10%, 20%, and 50% helium in hydrogen during the first calibration. The calibration was conducted as increasing steps of constant pressure, starting at the background pressure of about 1 × 10−7 mbar up to 9 × 10−3 mbar in 21 logarithmic equidistant steps. Of these 21 steps, not all were realized due to an insufficient response of the gas valves; however, 17 steps starting at 1 × 10−6 mbar could reliably be realized. Constant pressure was achieved by closing all gate valves from the plasma vessel to the vacuum pumps and puffing defined amounts of gas into the plasma vessel. Each step had a flat top duration of 10 s with an additional 1 s for equilibration of gas throughout the torus system after the asymmetrical puff. The receded vacuum monitoring systems, consisting of a Penning and a Pirani gauge and partially augmented with two capacitance manometers, were used as the pressure reference. In particular, the capacitance manometers provided a high-reliable, gas species independent, total pressure reference. These were mounted throughout the machine and were averaged to create the stair curve seen in Fig. 14 that was used as a reference to calibrate against. The WISP manometers were calibrated from 5 × 10−5 mbar to 9 × 10−3 mbar; however, the upper limit was due to the limitations in the vacuum system. The calibration function was extrapolated beyond the calibration, and measurements of up to 1 × 10−1 mbar were achieved. In addition, the deviations from the power law scaling at low pressures, as discussed in Sec. III, could not be found at W7-X.

At W7-X, the WISP gauge was operated in the 2.5 T field at 2 kV anode potential and would read pressures reliably between 1 × 10−1 mbar and 1 × 10−4 mbar at a time resolution of up to 1 kHz at higher pressures. While the time resolution of the ion current measurement is not affected by the pressure, the time resolution of the line intensity measurement potentially needs adjustment for longer integration times as the line intensities get weaker with a decrease in pressure. All gauges worked very reliably throughout OP1.2b without failures. The ASDEX-type pressure gauges (APGs) at W7-X19 can measure at a lower pressure, starting at 5 × 10−7 mbar up to 4 × 10−2 mbar at a time resolution of up to 5 kHz. Both gauges have an accuracy of about 15%. The APGs are suitable for a wider pressure range and can provide a faster time resolution,26 especially at lower pressures. While the APGs with tungsten filament were not as reliable due to jxB failures, the issue was resolved with the LaB6 electron emitters in the now called crystal pressure gauge22 (CPG). The WISP gauges have their strength in higher pressures, which occur inside the divertor. Besides being capable of measuring higher pressures than the CPG, the partial pressures can be extracted from spectral line ratio measurements with the WISP gauge, which is crucial for exhaust studies.

The ASDEX pressure gauges were used for comparison with the WISP gauges as both systems were operated at W7-X. However, there are multiple other in situ pressure gauges that have been developed to be used on large fusion devices. Notably, the “In-vessel Neutral Pressure,” INP gauge, and the “Princeton Divertor eXperiment,” PDX, type Penning gauge were operated between 0.45 T and 0.6 T at NSTX23 as well as the miniature Penning gauge at Alcator C-Mod capable of millisecond time resolution at ambient magnetic fields in the range of 2–8 T.24,25Figure 19 shows pressure measurements and line integrated density of a discharge. In the first 3 s, the plasma starts up with a density ramp up, after which the density is kept constant until the last few seconds of the discharge before it spikes at the shut off. The pressure measurements are from a set of two WISP gauges in module 4 and two CPG’s in the same ports in module 5. Both diagnostics scale comparably and show similar effects. WISP in AEI41 and CPG in AEI51, which are both mounted in the top divertor pump gap, show a constant decay of the divertor pressure with a four peak ripple at 20 s, while the two gauges in the lower divertor AEI40 and AEI50 show a rather constant divertor pressure. Differences in pressure can be explained through local effects in the different modules based on asymmetries or like the strong peak in AEI51 at 3 s due to localized gas puffing close to the gauge. When considering these effects, it can be concluded that measurements from both diagnostics are comparable.

FIG. 19.

Comparison of divertor pump gap pressure measurements in module 4 (WISP) and module 5 (CPG) of program ID 20181016.018 shows a general agreement between WISP and CPG.

FIG. 19.

Comparison of divertor pump gap pressure measurements in module 4 (WISP) and module 5 (CPG) of program ID 20181016.018 shows a general agreement between WISP and CPG.

Close modal

While the time resolution of the WISP gauges is slower than that of the ASDEX-type gauges, it is fast enough to see the effect of individual pellets on the neutral pressure at 10 Hz. This effect is shown in Fig. 20, where the line averaged density is shown on the right-hand side and the H partial pressure measured by using the WISP gauge in port AEI40 on the left-hand side. The density shows the plasma start up during the first second and a steep spiked density increase starting at 3 s. Each spike is an individual pellet entering the plasma and a density decay afterward until the next pellet enters the plasma. These spikes that are formed by individual pellets can also be seen in the partial pressure measurement of the WISP gauge. While the exact time response of the WISP gauge is still under investigation, this shows that effects from pellets entering at 10 Hz can be studied.

FIG. 20.

Hydrogen partial pressure and the line averaged density over time of program ID 20181010.025 with 10 Hz pellets.

FIG. 20.

Hydrogen partial pressure and the line averaged density over time of program ID 20181010.025 with 10 Hz pellets.

Close modal

He partial pressures were measured during He gas injection experiments into a H plasma. The He partial pressure measured using the WISP gauge in port AEI40 is shown in blue on the left-hand side, and the He/(H+He) ratio from charge exchange spectroscopy is shown in yellow on the right-hand side of Fig. 21. The three He gas injections at 3 s, 3.5 s, and 4 s can be seen in both signals. Each injection is followed by a sudden jump in the He partial pressure, followed by a fast decay until a plateau or very slow decay is reached. The gas injection increases the neutral He household, seen in the sudden increase in partial pressure. The He reservoir is then depleted on a fast time scale by ionization, as more and more neutral He particles enter the plasma, until an equilibrium is reached about 200 ms after the injection. This demonstrates the He partial pressure measurement capabilities in a H background plasma at multiple He to H concentrations.

FIG. 21.

He partial pressure on the left and He concentration measured by charge exchange spectroscopy on the right over time of program ID 20181010.020 with three He gas injections at 3 s, 3.5 s, and 4 s.

FIG. 21.

He partial pressure on the left and He concentration measured by charge exchange spectroscopy on the right over time of program ID 20181010.020 with three He gas injections at 3 s, 3.5 s, and 4 s.

Close modal

The Wisconsin In Situ Penning (WISP) probe head was developed and optimized for partial pressure measurements in the divertor pressure regime of fusion devices. Operation between 0.07 T and 3.5 T with anode potentials between 1.25 kV and 5 kV has been shown. The pressure range is dependent on the anode potential and the magnetic field. The magnetic field is usually dictated by the device and can therefore be a limiting factor, especially at higher fields since the lower end of the operating regime shrinks with an increase in the field. The anode potential can be used as a tool to somewhat mitigate this effect by moving the operational range from left to right. The highest achievable pressure was 0.1 mbar, which was limited by the maximum acceptable pressure of the vacuum pumping system at MDPX. As long as the WISP probe head can withstand the deposited input power, no reason could be seen why higher pressures should not be achievable. A misalignment of ±10° has no effect on the measurement. While the ion current measurement is gas dependent, this can be avoided by the light intensity measurements for partial pressures. It was shown that there is negligible crosstalk between He and H measurements, and the line intensity is proportional to the partial pressure. The ion current as well as the light intensity curves can be described by power law functions. At MDPX, there was a deviation of the power law fit at low pressures, which could not be found at other devices. Therefore, the presented data serve as a visualization of the behavior and cannot serve as calibration functions. An in situ calibration is inevitable. The error and sensitivity of the ion current measurement at low currents could be increased by more sensitive ammeters. The WISP head was shown to be a robust component, insensitive to venting of the vacuum environment. The WISP probe can be vented without permanent damage or requiring a new calibration after venting. The diagnostic was successfully implemented at W7-X on a plugin design and measured reliably throughout the entire OP1.2b campaign. The gauges were in situ calibrated and showed general agreement with other neutral gas diagnostics. The calibration functioned followed a power law fit throughout the entire operating range without any deviations. Future work will be the inclusion of a collisional radiative model to calculate density and temperature of the Penning discharge to analytically derive a partial pressure as well as a more extensive exploration of other light impurities.

This work was funded, in part, by the Department of Energy under Grant No. DE-SC0014210. The author would like to thank Edward Thomas, Jr., Darrick Artis, and the rest of the MDPX team at Auburn University for granting access to MDPX as well as their continuous help, support, and time. The operation of the MDPX device was funded by the Department of Energy under Award No. DE-SC0019176, the NSF-DOE Partnership Program in Basic Plasma Science and Engineering (Grant No. PHY-1613087/DE-SC0016330), and the NSF-EPSCoR program (Grant No. OIA-1655280). The MDPX device was originally designed and built with funding from the NSF-MRI program (Grant No. NSF-1126067).

This work was carried out within the framework of the EUROfusion Consortium and received funding from the Euratom Research and Training Programme 2014-2018 and 2019-2020 under Grant Agreement No. 633053. The views and opinions expressed herein do not necessarily reflect those of the European Commission.

1.
D.
Reiter
,
G. H.
Wolf
, and
H.
Kever
,
Nucl. Fusion
30
,
2141
(
1990
).
2.
K.
Ida
,
M.
Yoshinuma
,
B.
Wieland
,
M.
Goto
,
Y.
Nakamura
,
M.
Kobayashi
,
I.
Murakami
, and
C.
Moon
,
Rev. Sci. Instrum.
86
,
123514
(
2015
).
3.
A.
Kappatou
,
R. M.
McDermott
,
T.
Pütterich
,
R.
Dux
,
B.
Geiger
,
R. J. E.
Jaspers
,
A. J. H.
Donné
,
E.
Viezzer
,
M.
Cavedon
 et al,
Plasma Phys. Controlled Fusion
60
,
055006
(
2018
).
4.
H.-S.
Bosch
,
G.
Haas
, and
M.
Lörcher
,
J. Nucl. Mater.
196-198
,
1074
(
1992
).
5.
K. H.
Finken
,
K. H.
Dippel
,
W. Y.
Baek
, and
A.
Hardtke
,
Rev. Sci. Instrum.
63
,
1
(
1992
).
6.
F. M.
Penning
,
Physica
3
,
873
(
1936
).
7.
T.
Denner
,
K. H.
Finken
, and
G.
Mank
,
Rev. Sci. Instrum.
67
,
3515
(
1996
).
8.
M. R.
Wade
,
D. L.
Hillis
,
J. T.
Hogan
,
M. A.
Mahdavi
,
R.
Maingi
,
W. P.
West
,
N. H.
Brooks
,
K. H.
Burrell
,
R. J.
Groebner
,
G. L.
Jackson
 et al,
Phys. Rev. Lett.
74
,
2702
(
1995
).
9.
D. L.
Hillis
,
C. C.
Klepper
,
M.
Von Hellermann
,
J.
Ehrenberg
,
K. H.
Finken
, and
G.
Mank
,
Fusion Eng. Des.
34-35
,
347
(
1997
).
10.
C. C.
Klepper
,
T. M.
Biewer
,
V. B.
Graves
,
P.
Andrew
,
P. C.
Lukens
,
C.
Marcus
,
M.
Shimada
,
S.
Hughes
,
B.
Boussier
,
D. W.
Johnson
 et al,
Fusion Eng. Des.
96-97
,
803
(
2015
).
11.
K.
Flesch
,
T.
Kremeyer
,
O.
Schmitz
,
V.
Soukhanovskii
, and
U.
Wenzel
,
Rev. Sci. Instrum.
87
,
11E529
(
2016
).
12.
E.
Thomas
,
U.
Konopka
,
D.
Artis
,
B.
Lynch
,
S.
Leblanc
,
S.
Adams
,
R.
Merlino
, and
M.
Rosenberg
,
J. Plasma Phys.
81
,
345810206
(
2015
).
13.
H.
Saitoh
,
T. S.
Pedersen
,
U.
Hergenhahn
 et al,
J. Phys.: Conf. Ser.
505
,
012045
(
2014
).
14.
O.
Schmitz
,
I. L.
Beigman
,
L. A.
Vainshtein
,
B.
Schweer
,
M.
Kantor
,
A.
Pospieszczyk
,
Y.
Xu
,
M.
Krychowiak
,
M.
Lehnen
,
U.
Samm
 et al,
Plasma Phys. Controlled Fusion
50
,
115004
(
2008
).
15.
G. W.
Pacher
,
H. D.
Pacher
,
G.
Janeschitz
,
A. S.
Kukushkin
,
V.
Kotov
, and
D.
Reiter
,
Nucl. Fusion
47
,
469
(
2007
).
16.
D.
Douai
,
T.
Goodman
,
A.
Isayama
 et al,
Nucl. Fusion
58
,
026018
(
2017
).
17.
C. C.
Klepper
,
D. L.
Hillis
,
M. R.
Wade
 et al,
Rev. Sci. Instrum.
68
,
400
(
1997
).
18.
H.-S.
Bosch
,
R.
Brakel
,
T.
Braeuer
,
V.
Bykov
,
P.
van Eeten
,
J.-H.
Feist
,
F.
Füllenbach
,
M.
Gasparotto
,
H.
Grote
,
T.
Klinger
 et al,
Nucl. Fusion
57
,
116015
(
2017
).
19.
U.
Wenzel
,
T.
Kremeyer
,
G.
Schlisio
,
M.
Marquardt
,
T. S.
Pedersen
,
O.
Schmitz
,
B.
Mackie
,
J.
Maisano-Brown
 et al,
J. Instrum.
12
,
C09008
(
2017
).
20.
R. J.
Colchin
,
D. L.
Hillis
,
R.
Maingi
,
C. C.
Klepper
, and
N. H.
Brooks
,
Rev. Sci. Instrum.
74
,
2068
(
2003
).
21.
C. C.
Klepper
,
T. M.
Biewer
,
C.
Marcus
 et al,
J. Instrum.
12
,
C10012
(
2017
).
22.
U.
Wenzel
,
T. S.
Pedersen
,
M.
Marquardt
, and
M.
Singer
,
Rev. Sci. Instrum.
89
,
033503
(
2018
).
23.
R.
Raman
,
H. W.
Kugel
,
T.
Provost
,
R.
Gernhardt
,
T. R.
Jarboe
, and
M. G.
Bell
,
Rev. Sci. Instrum.
74
,
1900
(
2003
).
24.
B.
Lipschultz
,
B.
LaBombard
,
C. S.
Pitcher
 et al,
Plasma Phys. Controlled Fusion
44
,
733
(
2002
).
25.
B.
Boivin
, Report No. PSFC/RR-15-12,
2015
.
26.
G.
Haas
and
H.-S.
Bosch
,
Vacuum
51
,
39
(
1998
).