We describe the construction of a fast field cycling device capable of sweeping a 4-order-of-magnitude range of magnetic fields, from ∼1 mT to 7 T, in under 700 ms, and which is further extendable to a 1 nT-7 T range. Central to this system is a high-speed sample shuttling mechanism between a superconducting magnet and a magnetic shield, with the capability to access arbitrary fields in between with high resolution. Our instrument serves as a versatile platform to harness the inherent dichotomy of spin dynamics on offer at low and high fields—in particular, the low anisotropy, fast spin manipulation, and rapid entanglement growth at low field as well as the long spin lifetimes, spin specific control, and efficient inductive measurement possible at high fields. Exploiting these complementary capabilities in a single device opens up applications in a host of problems in quantum control, sensing, and information storage, besides in nuclear hyperpolarization, relaxometry, and imaging. In particular, in this paper, we focus on the ability of the device to enable low-field hyperpolarization of 13C nuclei in diamond via optically pumped electronic spins associated with nitrogen vacancy defect centers.
I. INTRODUCTION
The last few decades have witnessed rapid strides in high-field superconducting magnet technology, with fields B0 > 20 T and inhomogeneities better than 1 ppb routinely available, fueling several recent advances in biomolecular nuclear magnetic resonance (NMR).1,2 In parallel, there has been a silent revolution in the development of magnetic shielding technology,3 with specialized alloys of mu-metal providing shielding factors >106 and extinguishing fields to <0.1 nT in a relatively large volume.4–6
From a physical point of view, both extremes of ultra-high and ultra-low magnetic fields provide uniquely complimentary advantages. In quantum information science, for instance, high fields provide a means to store and protect quantum information due to long spin relaxation times (T1). In particular, the electronic spin associated with the nitrogen-vacancy (NV) center in diamond7—which has emerged as a promising platform for quantum information processing,8–10 simulation,11 and metrology12,13—has a T1 approaching 10 ms at 8 T.14 In addition, high fields enable the ability to apply highly frequency selective quantum control often with <1 ppm resolution, as well as sensitivity gains in measurement, especially bulk inductive spin readout, where signal-to-noise ratio (SNR) scales favorably, .15 Ultra-low to zero fields (1 nT-100 mT), on the other hand, provide the alternative advantages of spin indistinguishability—spins even of completely different species act identically, allowing access to heteronuclear spin singlets with long lifetimes16 and the easy construction of Hamiltonian models in naturally occurring spin networks. More specifically, the low field regime is interaction dominated, where the Larmor frequencies are smaller than the inter-spin couplings. Indeed at zero field, with the absence of any field B0 that acts to truncate the inter-spin couplings, the interaction Hamiltonians for free evolution are completely isotropic, without orientational dependence.17,18 This can allow the relatively easy production of nearest-neighbor Heisenberg models in large spin networks in liquids.19 In dipolar coupled solids, this leads to fast entanglement generation since there are no disallowed transitions from energy costs set by B0.20 Finally low-field also allows the possibility of ultra-fast quantum control since there are no speed limits set by the rotating wave approximation.
In this work, we construct a device that combines and harnesses the power of both regimes in a single platform (see Fig. 1 and video at Ref. 21) capable of sweeping magnetic fields over a four order-of-magnitude dynamic range from ∼1 mT to 7 T and extendable to a 1 nT-7 T range. The device works by physically transporting (shuttling) a sample precisely and at high speed (under 700 ms) between low and high field (7 T) centers placed 830 mm apart, exploiting a high mechanical precision (50 μm) to achieve arbitrary tunable fields in the fringing field between the two centers. The sample shuttling takes place faster than the T1 times of nuclear spins in a variety of physical systems, which coupled with high resolution inductive detection at 7 T makes the system ideally suited to studying nuclear spins under different field environments. The system also provides the ability for spin manipulation at the low and high field centers and rapid in situ sample freeze to enhance T1 lifetimes.
This capability paves the way for several versatile applications of the device. In this paper, we shall particularly focus on applications for the quantum system consisting of coupled 13C nuclear spins and NV center electrons in diamond. Coupled to the optically addressable NV center qubit, 13C spins have garnered attention as forming viable nodes of a quantum information processor22–24 due to their long lifetimes and the fact that they can be rapidly and directly manipulated by the NV center.25–27 Their utility as quantum memories have been demonstrated to provide wide gains in quantum sensing, both with respect to sensing resolution28–30 and sensitivity31—especially compelling for nanoscale MRI experiments at high fields.32 Exploiting the complimentary advantages of low and high field control that our instrument offers will enable enhanced resolution gains in ancilla assisted quantum sensing. Moreover low fields serve to strongly mitigate inherent anisotropies in the system Hamiltonians. We exploited this recently, employing our instrument, to develop the first method for efficient room temperature dynamic nuclear polarization (DNP) of 13C nuclei via the optically pumped NV centers. The polarization transfer at low fields is orientation independent, allowing 13C hyperpolarization in powdered, randomly oriented, micro- and nano-diamonds,33 and the possibility of a “nanodiamond polarizer” for the optical hyperpolarization of liquids brought in contact with these high surface area diamond particles. Sample shuttling to high fields after DNP allows an efficient detection of the 13C hyperpolarization.
In this paper, we focus on the instrumental (see Fig. 1) capabilities at low and high fields enabled by our device that enabled the hyperpolarization experiments in Ref. 33. Sections II and III describe the construction and design aspects of our instrument. Section IV demonstrates our capability for low field spin manipulation, and its application for the optical hyperpolarization of 13C in diamond powder is described in Sec. V. Section VI describes a cryogenic system that delivers in situ rapid sample cooling at low field. Finally, Sec. VII briefly outlines potential experiments harnessing the power of low and high fields enabled by our field cycling device.
II. MECHANICAL FIELD CYCLING FROM LOW FIELD TO 7 T
The field cycler consists of a tower constructed over a high field (7 T) superconducting magnet with a magnetic shield positioned below it (see Fig. 1). A video showing the working of the instrument is available online.21 A fast conveyor belt actuator stage (Parker HMRB08) carries the sample in the fringing field of the magnet and into the shield, allowing a ∼1 mT–7 T field sweep (in principle, 1 nT-7 T with mu-metal shields). The sample is carried by a carbon fiber shuttling rod (Rock West composites) that is fastened rigidly on a twin carriage mount on the actuator stage [Figs. 1(a), 1(b), and 1(g)]. Carbon fiber was chosen because of its exceptional strength, low mass, and immunity to eddy-current forces,34 while the twin-carriage minimizes yaw and aids in alignment, crucial for high sample filling factors.
Figure 2 highlights the versatile control available in our system—the ability to tune the shuttler velocity and acceleration [Fig. 2(a)] and spatial position for start and stop of motion [Fig. 2(b)] and motion trajectories (see Ref. 21). Shuttling is possible upto a speed 2 m/s and acceleration 30 m/s2 over a 1600 mm travel range, with a high positional precision of 50 μm. This positional resolution is currently limited by the conveyer belt stage used for sample shuttling and, in principle, can be enhanced further to ∼1 μm using instead a screw-based actuator stage. The field control available could be exploited in systems with specific level anti-crossings (LACs); for instance, in SABRE based DNP with parahydrogen,35 where we can precisely control the rate of passage through the LACs to optimize polarization transfer efficiency (Sec. VII).36 We have characterized the shuttling time from 7 T to 8 mT to be 648 ± 2.6 ms [see Fig. 2(a) inset]. This highlights the remarkably high repeatability in our instrument, which contrasts to conventional pneumatically controlled field cyclers. To our knowledge, this is also the first time sample shuttling times have been quantified with such high precision.
A. Shuttling alignment and vibrational stability
Our system incorporates special features to maximize the sample filling factor for highly efficient inductive detection and for low field quantum control through radiofrequency or microwave excitation (Sec. IV). For minimum possible clearance to excitation coils at both fields, and low vibration associated with motion jerk, it is essential that the shuttling rod be aligned parallel to the magnet axis over the entire distance of travel. We align to better than 1mdeg through a series of design implementations. First, the entire shuttling tower (80/20 1530-S) containing the actuator, motor, and twin carriage is on an XY tunable platform [Fig. 1(a) and video at Ref. 21]. This centers the shuttling rod to the magnet bore with a precision better than 0.25 mm over the 1600 mm travel. Second, and more critically, two alignment funnel-shaped guiding stages made of soft Teflon are employed at the magnet bore [Fig. 1(f)] and NMR probe [Fig. 5(b)]. The stages vertically align the structure and provide additional points of support to greatly reduce vibration (see video at Ref. 21). The carbon fiber shuttling rod (8 mm diameter, 1.7 m length), while soft enough to be guided by the Teflon stages, is inherently less prone to vibration due to its low moment of inertia and high strength (430 GPa tensile modulus).
Figure 3 visualizes the guiding process, taken with a camera located in the NMR probe focusing on the Teflon guide above it [Fig. 5(b)]. The rod starts out slightly misaligned but is dynamically guided to be perfectly aligned, the carbon fiber malleable enough to be able to guide into place with no jerk. The funnel guide has a 45° taper with an opening of 8.077 mm, ensuring a tight fit with the 8 mm shuttling rod. This ensures shuttling with minimum clearance to the NMR coil and consequently high filling factors (see Fig. 6).
B. Sample attachment to shuttling rod
The sample is pressure-fit to the carbon fiber rod for rapid attachment and detachment (see video at Ref. 21). The lower end of the rod contains a ceramic connection for attaching the NMR tube carrying the sample [Fig. 1(c)]. It consists of a pair of soft, high temperature modulus, silicone O-rings (McMaster 1/16 fractional width, 0.254 in. OD). Remarkably, this arrangement proves resilient for fast shuttling with just the simplicity of a hand-tight pressure fit. The diameter of the NMR tube (Wilmad 8 mm OD, 0.5 mm thickness) was chosen to match that of the shuttling rod for a seamless joint through the alignment guides.
Figure 1(d) details the tube containing a sample of powdered diamond employed in optical hyperpolarization experiments. A plunger carrying a dielectric mirror is used to isolate the sample to a compact volume. The plunger is fitted with a threaded screw hole for easy fastening access, positioning, and removal. The mirror (Thorlabs BB1-E02 Broadband Dielectric Mirror, 400–750 nm) is machine ground to the inner diameter of the NMR tube and provides a double pass for the incoming laser radiation for efficient polarization of NV center electrons.
C. Synchronization and triggering
Shuttling and inductive detection are synchronized (see Fig. 4) using a high-speed pulse generator (SpinCore PulseBlaster USB 100 MHz). The servomotor is triggered to start motion for shuttling with a 24 V 10 ms pulse. This is generated by upconverting the 3.3 V pulse trigger from the pulse generator to 5 V by a TI SN74LS04N CMOS inverter, which then serves as the gate to a high voltage Metal Oxide Semiconductor Field Effect Transistor (MOSFET) switch (Williamette MHVSW-001V-036 V). The MOSFET switch relays the 24 V pulse to the servomotor that drives the belt-driven actuator stage to the desired position. Finally, the actuator returns a 24 V pulse indicating the completion of motion, which is passed through a voltage divider to trigger the NMR console (Agilent DD2) to initiate measurement (see Fig. 4). This is also used to precisely quantify the shuttling times and jitter (Fig. 2).
D. Comparison with other field cycling platforms
Let us now briefly compare our field cycling platform to previous reports. Unlike conventional field cycling platforms that are geared toward relaxometry experiments in liquids and proteins, our device seeks to harness the versatile dichotomy in spin dynamics regimes between low and high fields. Our instrument allows a large dynamic range magnetic field sweep from 1 mT to 7 T, which can be further extended to a ten order-of-magnitude range 1 nT-7 T through enhanced low field shielding, following a zero-field setup similar to the one described in Ref. 37. Note that by contrast, fast field cycling38 which uses specialized power supplies and switched-coils39,40 to rapidly switch between magnetic fields cannot easily access fields >2 T. For experiments described in Fig. 14, the ability to manipulate and store spins at higher field enables applications not accessible by pulsed field cycling.
Focusing now on field cycling devices employing sample shuttling, they come in two flavors: pneumatic shuttling and mechanical shuttling. Pneumatic devices generally provide higher shuttling speeds even though it is challenging to measure the speeds precisely; however, compressed air causes sample vibration upon motion start and stop, which adds an additional 0.1-1 s time for stabilization. Our mechanical instrument operates with high positional precision (50 μm), shuttles with a maximum velocity of 2 m/s with high repeatability (Fig. 2), and has a large thrust force capacity (295 N). While the lower speed limits it from certain tasks (e.g., protein relaxometry), the precision of control over position and velocity makes it ideal for a plethora of other applications (Sec. VII), for instance, for field sweeps through energy level anti-crossings over a wide dynamic range.
While inspired by the pioneering mechanical field cycling platform of Redfield,41 our device extends its original innovative field cycler in several directions. First, the use of a more precise actuator stage allows the ability to access fields at very high resolution and with high repeatability (Fig. 2). The fact that the low field center is far separated from the highfield center, and primarily situated at the bottom of the magnet, allows one to accessorize the field cycler with optical and microwave irradiation and cryocooling (see Sec. VI) components for spin control at low fields. This makes the field cycler as suitable for optical spin hyperpolarizaton experiments (Sec. V) involving the nitrogen vacancy center in diamond.
Moreover unlike the field cycling platform of Kiryutin et al.,42 where the entire NMR probe is shuttled, our work relies on rapid sample-only shuttling with minimal occupancy in the bore of the NMR magnet. Indeed, the shuttling assembly occupies only 8 mm in the 54 mm magnet bore. This provides a key engineering advantage of our design, and in contrast to other mechanical shuttling designs, for instance, Chou et al.,43 it would allow interfacing with magnetic inserts44 which would allow the creation of homogeneous intermediate field regions for quantum control.45 The low bore occupancy is also crucial to the integration of the shuttler with components for optical and microwave irradiation, as well as low temperature control, the confluence of which have enabled the new optical 13C hyperpolarization mechanism described in Sec. V. Finally, we note that while this manuscript was under preparation, a similar wide-range field cycling platform was recently published by Zhukov et al.,46 which also relies on shuttling into a magnetic shield. Our device, while similar in scope, differs critically in the optical and microwave access available. The shield is situated at the bottom of the magnet allowing more adaptability and the ability to interface with multiple lasers, bulky microwave amplifiers, as well as facilitating cryogenic delivery to the sample.
III. NMR PROBE COMPATIBLE WITH SHUTTLING
Modifications were made to conventional NMR probe design to accommodate fast sample shuttling (Fig. 5) through it. Figure 6 visualizes the actual shuttling process at the probe. The probe is designed hollow for shuttling to low fields below the magnet and is constructed out of 12.7 mm thick brass plates for enhanced shielding and structural rigidity. The top plate holds the tuning and matching capacitors (Voltronics AT4HV and AP14) along with the quartz tube (ID 9 mm × OD 11 mm Technical Glass Products) outside of which is fabricated a saddle shaped NMR coil. The probe can accommodate either split solenoid coils (Fig. 6) or current saddle shaped ones [similar to Fig. 7(c)]. The Teflon funnel at the top [Fig. 5(b)] allows for the rod to self align (Sec. II A). Assuming the sample fills the full height of the coil, the filling factor is ≈0.4, which is comparable to commercial probes, allowing high sensitivity inductive detection.
A. Saddle coils for maximizing filling factor
We have developed a rapid technique of fabricating saddle coils for NMR detection that are compatible with shuttling, yet optimized to the sample of interest to provide large filling factors. A descriptive video showing the fabrication method is available online.47 The coils are wrapped on a Quartz tube matched to the NMR tube being shuttled such that the sample tightly fills the entire coil volume. The coils are fabricated by cutting them out of the flexible adhesive copper foil (Venture Tape, thickness 31.75 μm) [see Fig. 7(c)] using an inexpensive vinyl cutter (Silhouette Cameo2) in 17 s and 1.5 mm cutting depth (see video at Ref. 47). The RF coil in the probe has 10.56 mm 110° windows15 and a track width of 1.25 mm with 0.75 mm spacing. After printing, excess copper around the coil was carefully removed and contact paper (Circuit StrongGrip Transfer Tape) was applied on top to maintain the coil shape (see Ref. 47). When inserted into the quartz tube, the inside of the tube was coated with water to prevent the coil from adhering while positioning to a slit for the leads to be pulled through. A heat gun was used to release the contact paper and adhere the coil to the wall of the tube. The fabricated coils for 13C NMR had an inductance of 0.28 μH and a Q factor of 150 at 75.03 MHz.
IV. LOW-FIELD CENTER
A. Construction
Our field cycling device provides a wide dynamic field range by employing a magnetically shielded low-field location that the sample can be shuttled into [Fig. 1(e) and video at Ref. 21]. Since the shields sit in the ≈300 G fringing field of the 7 T magnet, they are positioned on sliding rails to secure them against upward magnetic forces. The shield is constructed out of concentric layers of stress annealed iron (NETIC S3-6 alloy 0.062” thick, Magnetic Shield Corp.) and mu-metal (TwinLeaf MS-1), with the iron on the outside due to its high saturation and mu-metal on the inside due to its high permeability (over 106 with 4 layers). In combination, with 3 layers of iron and 4 layers of mu-metal, one can achieve a lower field center of approximately 1 nT. In practice, the low-field shielding can be customized to suit the target field desired in experiments, in particular—low fields (1-30 mT) for optical DNP experiments in diamond,33 ultra-low fields (<1 μT) for relaxometry, and even at zero fields (<1 nT) for applications exploiting isotropic Hamiltonians models6,19,48,49 and decoherence free subspaces16,50 readily available at these fields.
B. Field map
We have obtained a Bz field map over the full travel range of the sample from low to high field (Fig. 8). Measurements were performed with a sensitive longitudinal field Gaussmeter (Lakeshore HMMA-2504-VR-10) that is inserted into the hollow shuttling rod so as to be aligned centered with the bore. The fact that the superconducting magnet is unshielded leads to rather weak gradients in the fringing field, allowing one to access fields with <1 G resolution over a wide range given the 50 μm precision over sample position. This is highlighted in Fig. 9 for the excited (ESLAC ≈ 510 G) and ground (GSLAC ≈ 1020 G) state level crossings of NV centers in diamond. The weak gradients manifest as approximately linear spatial field dependencies. Indeed our instrument allows us to access the ESLAC region with a ≈114 mG resolution and sweep through it at 0.458 T/s. This would allow applications for the optical hyperpolarization of 13C nuclei and P1 centers in the vicinity of the ESLAC [Fig. 14(f)].51–53
C. Zero-field quantum control
While our focus in this paper is the use of the device in the low-field regime, our instrument also potentially allows one to achieve a zero-field37 (<1 nT) center by interfacing concentric mu-metal shields into the iron low-field volume (Fig. 10). Quantum control at zero-field can be enabled using fast DC magnetic field pulses. We have developed a pulsing circuit consisting of high power MOSFETs that allow the rapid switching of current from a 600 W power supply generating fast DC pulses for spin manipulation [see Fig. 10(c)]. The free evolution of Hamiltonian in the absence of a static magnetic field without laser and microwave irradiation is isotropic. This opens the possibility to exploit isotropic Hamiltonians and fast entanglement growth that naturally occurs at zero-field for several quantum simulation problems,19 while retaining the high detection sensitivity at high field. A detailed description of experiments in this regime will be presented elsewhere.
V. APPLICATIONS TO 13C OPTICAL HYPERPOLARIZATION IN DIAMOND
We recently employed the unique field cycling ability of our instrument, along with the ability of laser and microwave control of the NV electronic spins at low field to develop a novel method for dynamic nuclear polarization (DNP) of 13C spins in powdered diamond.33 The aim of this paper is to highlight how the complimentary advantages of low and high fields enabled by our instrument enabled this advance. Low fields (typically employed Bpol ∼ 1-30 mT) mitigate the strong orientation dependence of the NV centers, while high fields allow efficient detection of the generated hyperpolarization. Importantly, low fields allow one to invert the conventional hierarchy used in DNP experiments, entering the regime where the nuclear Larmor frequency is smaller than the electronic hyperfine interaction. This engenders hyperpolarization generation through microwave frequency sweeps of the electronic spectrum.
To be more specific, Fig. 7 describes the DNP setup in the low-field center (see video at Ref. 21). The sample is irradiated simultaneously with laser and microwaves sweeping across the inhomogeneously broadened NV center powder pattern to perform the polarization transfer [Fig. 11(a)]. The microwave excitation is produced via coils (outlined in Sec. III A) connected to a high power (20-100 W) microwave amplifier. The DNP mechanism [see Fig. 11(c)] can be interpreted as partly adiabatic traversals of Landau-Zener crossing in the rotating frame, with a model of NV electron coupled to a single 13C nucleus.33 In the low field hyperpolarization regime, the 13C nuclear Larmor frequency . In the NV sublevel ms = ±1, the eigenstates are set by the hyperfine coupling, denoted as β↑ and β↓. On the other hand, nuclear eigenstates become dominant in the ms = 0, manifold, and are denoted as α↑ and α↓. When taking second order hyperfine coupling into consideration, Larmor frequency , where ϑ is the orientation of the N-to-V axis respect to external magnetic field and Δ = 2.87 GHz is the NV zero-field splitting. Sweeping microwaves across this set of transitions leads to their sequential excitation of a set of transitions that drive the coherence hyperpolarization transfer.33,54 We demonstrated that the hyperpolarization transfer is an orientation independent hyperpolarization technique and is enabled by the relatively small inhomogeneous broadening (≈400 MHz) of the NV electronic linewidth (powder pattern) at low fields (1-30 mT). As a typical example, for 200 μm diamond particles (Element6) containing about 1 ppm of NV centers, the 13C polarization at a field of 13 mT was enhanced to the level corresponding to thermal magnetization at a field higher than 1900 T. Field cycling allows one to harness the high sensitivity of inductive measurement at 7 T [Fig. 11(b)], enabling the unambiguous inductive readout of the bulk 13C nuclear magnetization. Indeed hyperpolarization enables a measurement time gain of about five orders of magnitude at 7 T, allowing the detection of a single 200 μm particle with unit SNR in one shot.33
The high-precision and programmable field cycling ability of our setup also proves beneficial in relaxometry, in determining the factors affecting the spin lifetimes times of the 13C nuclei in diamond. Figure 12, for instance, shows a relaxation-field map T1(B) for 200 μm powdered diamond microparticles (Element6) at various fields B. In these experiments, we used ≈8 mg of the microparticles, produced by high-pressure-high-temperature (HPHT) synthesis, and they contained ≈1 ppm NV center concentration and had natural abundance (1.1%) of 13C nuclei. We first begin by hyperpolarizing the 13C nuclei opposite to the direction set by B0 = 7 T. The sample is then rapidly shuttled to the target field Brelax, allowed to wait for time Trelax, and then rapidly shuttled again to 7 T for detection. Sweeping t and fitting the decay to a monoexponential provides the lifetime T1(B). Substantial time savings, typically 4-5 orders of magnitude, are obtained via the hyperpolarization process. Figure 12(a) shows typical relaxation data at exemplary low and high fields. We measure the T1 times to be 395.7 s at B0 = 7 T and 10.19 s at Bpol = 8 mT. Since the DNP polarization is inhomogeneous, i.e., spins closer to the NV centers are more strongly polarized, spin diffusion between them also causes a signal decay at low fields, leading to a super-exponential fall in the signal. This feature slightly underestimates T1 at Bpol. The field dependent relaxation data in Fig. 12(b) show a steep dependence with field, the spin lifetimes falling rapidly at low fields, but growing to long room temperature polarization lifetimes (∼5 min) beyond 1 T. This feature arises from an increasing overlap between the 13C Larmor and electronic dipolar reservoirs associated, for instance, with P1 centers at low fields and will be addressed in detail in a future manuscript.
VI. IN SITU FAST SAMPLE FREEZE
We have interfaced our field cycling instrument with a homebuilt liquid nitrogen (LN2) cryogenic system for fast in situ freezing of the sample at the low-field center [Fig. 1(a)]. The primary aim is to increase the sample T1 lifetimes. For instance, 13C spins in pyruvate, an important molecule in the metabolic cycle and cancer detection, can exceed 55 s at 10 mT,55 but when frozen, the resulting T1 can be nearly an hour.56 A striking example of where such sample cooling would be useful is when employed for a “nanodiamond hyperpolarizer”—where polarization can be transferred from optically polarized 13C nuclei in high surface area diamond particles to external 13C in the frozen solution33—the long T1 times enabling a larger buildup of hyperpolarization.
Figure 13 describes the setup. The design incorporates a means to rapidly cool the sample in a manner that can be synchronized with the device shuttling (Fig. 4). A consistent jet of cryogenic liquid is rapidly ejected from a nozzle (see Fig. 7) on-demand which cools the sample to 77 K (liquid nitrogen temperature) within 3-4 s. Flowing N2 gas is applied to evacuate water vapor to prevent condensation. A dewar with two openings [Fig. 13(c)] stores cryogenic liquid and creates the pressure that drives the LN2 flow. The top opening is for (re-)filling by an external liquid nitrogen source, and the bottom for injecting liquid into a quartz nitrogen nozzle to generate a jet [Fig. 13(b)]. Both inlet and outlet are regulated by two solenoid valves individually (Asco 1/4 Cryogenic Solenoid Valve, 7/32 orifice diameter, 24 VDC control voltage), which can be triggered indirectly by the Pulse Blaster and respond within 1 ms to ensure synchronization (see Sec. II C). The system is designed so that laser irradiation and cryogenic cooling can occur simultaneously, and the LN2 level in the dewar is constantly replenished to maintain a constant jet pressure.
The exact manner in which the sample cooling occurs is detailed in Fig. 13(a). A fine slit in the sample tube allows an inlet to the cryogenic liquid, ensuring that it forms a cold bath around the sample. The sample itself is held in a glass bulb within the tube. The liquid remains in the NMR tube for more than 30 s with only 1 s LN2 injection, during which period the temperature in the tube remains stable even with laser irradiation. Importantly, this implementation of sample freeze is completely compatible with field cycling—the LN2 bath shuttled along with the sample at high speed and does not affect the inductive measurement at 7 T. The sample can be rapidly thawed prior to measurement by a resistive heater attached in the NMR probe to enable high resolution readout.
VII. PROPOSED EXPERIMENTS HARNESSING HIGH AND LOW MAGNETIC FIELDS
Our field cycling instrument allows one to harness the power of quantum control and measurement at low (1-30 mT) and high (7 T) fields, along with the ability to access intermediate fields with high resolution and controllable sweep rates. Indeed programming the field trajectories with a high degree of control allows versatile use of the device for several applications. In this section, we mention some potential experiments exploiting this capability. For concreteness, we focus in Fig. 14 on potential experiments in the system consisting of NV centers and 13C nuclear spins in diamond.
A. Hyperpolarized diamond MRI
The optical access and microwave control of the NV electronic spin available at low fields enable the coherent optical hyperpolarization of 13C nuclei (see Sec. V). Our instrument opens the potential to at once hyperpolarize diamond nano- and micro-particles at low field33 and image them with MRI with high sensitivity [Fig. 14(a)]. Diamond particles are non-cytotoxic and can be easily functionalized.57–60 Hence the use of hyperpolarized 13C spins in diamond particles as MRI tracers is especially compelling—the hyperpolarization providing bright MRI contrast that reports directly on the particle position, allowing their use as microfluidic flow tracers, as well as bio-sensors.
B. Relaxometry
Our device can be employed for relaxometry using conventional Boltzmann polarization [Fig. 14(b)] or low-field hyperpolarization [Fig. 14(c)]. Particularly pertinent for these applications is the ability to field cycle with high resolution, over a wide field range, and under programmable field trajectories, all of which our device can perform exceedingly well. We envision the use of this field cycling device broadly for hyperpolarized relaxometry61 of substances that are polarized via the diamond particles, with hyperpolarization providing significant savings in experimental time for the measurements. While in Fig. 12 we had considered exemplary hyperpolarized relaxometry on 13C nuclei in diamond, a more detailed exposition of the factors determining the 13C lifetimes (extracted from these data) will be presented in a forthcoming publication.
C. Quantum memories at high field
High fields enable long 13C nuclear T1’s, which in ultrapure diamond could exceed a few hours. Even for diamond microparticles containing a high density (>1 ppm) of NV centers, we have observed nuclear T1 approaching 400 s at 7 T (see Fig. 12). These long lifetimes make the 13C nuclei very attractive as ancillary quantum memories for the NV electronic qubits.30,62,63 Field cycling hence allows the dual benefit of long nuclear lifetimes at high field and technically simple electronic control and readout at low field. High fields are also associated with a concomitant increase in the nuclear coherence time, a fact that engenders their use in ancilla assisted quantum sensing protocols [Figs. 14(e) and 14(f)]. Indeed there can be a significant boost in quantum sensing resolution in such protocols, with resolution Δf ≤ 1/T2,memory.32 Such an experiment is schematically represented in Figs. 14(e) and 14(f): a quantum sensing (e.g., magnetometery) experiment is first performed via the NV electron, the states of the NV and 13C nuclei are then correlated28 and the quantum information stored at high field for T2,memory. A subsequent sensing experiment performed with the NV can then be correlated with the previous measurement, and the result readout optically via NV center flouresence [Fig. 14(e)] or inductively via the 13C magnetization [Fig. 14(f)].
D. Accessing level anti-crossings
The high precision and repeatability of our instrument, along with its ability to sweep different fields with a programmable velocity profile, enable it to access specific level anti-crossings (LACs) over an extremely wide dynamic range (1 mT–7 T) (see Fig. 9). LACs are traditionally helpful in several hyperpolarization contexts, for instance, in optically pumped NV centers at the ESLAC (∼510 G) transferring their polarization to 13C nuclei and P1 centers,51–53 and in parahydrogen based SABRE, e.g., at 6 mT to hyperpolarize 1H and at 0.1-1 μT for 13C and 15N nuclei.64 The high dynamic range allows access to low and high field LACs65 and clock states66 that have been technically hard to probe previously. The small size of the sample and confined position at the center of the fringe field guarantee the homogeneity of the field.
VIII. CONCLUSIONS AND OUTLOOK
In summary, low and high fields offer complimentary advantages that can be harnessed together in order to open interesting avenues in metrology, hyperpolarization, and quantum information science. Low fields offer access to coupling dominated Hamiltonians, with low anisotropy, and spin indistinguishability, while high fields allow spin selectivity, long lifetime and coherence times, and high SNR bulk detection. In this manuscript, we have described the construction and operation of a device that blends these two complimentary advantages on a single platform. In particular, we have constructed a novel field cycling platform capable of rapid (<700 ms) magnetic field sweeps over a wide field range in principle up to 1 nT-7 T. High positional precision and repeatability allows access to fields with high resolution. The device also allows optical and microwave spin control, and sample cooling at low fields, along with sensitive inductive readout at high field. While we have geared our device strongly to the applications in hyperpolarization of solids using optical pumping at low fields, we envision several applications in relaxometry, imaging, and quantum sensing, control and information storage that will be enabled by our instrument.
ACKNOWLEDGMENTS
We gratefully thank A. Redfield, D. Suter, T. Zens, and A. Llor for insightful conversations. We acknowledge valuable contributions from E. Granlund, J. D. Breen, S. Thangaratnam, R. Clark, S. Ebert, M. Garcia, D. Arnold, X. Cai, A. Kumar, A. Lin, G. Li, P. Raghavan, J. Wang, T. Huynh, and J. Fyson and the UC Berkeley College of Chemistry machine shop. C.A.M. acknowledges support from the National Science Foundation through Grant Nos. NSF-1309640 and NSF-1401632 and from the Research Corporation for Science Advancement through a FRED Award. The authors and UC Berkeley have filed patents on the instrument, techniques, and applications.