An in situ wavelength calibration system for the X-ray Imaging Crystal Spectrometer (XICS) on W7-X has been developed to provide routine calibration between plasma shots. XICS is able to determine plasma flow profiles by measuring the Doppler shift of x-ray line emission from highly charged impurity species. A novel design is described that uses an x-ray tube with a cadmium anode placed in front of the diffracting spherically bent crystal. This arrangement provides calibration lines over the full detector extent for both the Ar16+ and Ar17+/Fe24+ spectrometer channels. This calibration system can provide a relative wavelength accuracy of 3 × 10−7 Å across the full spatial extent of the detector, which corresponds to 50 m/s in the W7-X system. An absolute wavelength calibration of 1 × 10−5 Å is expected, corresponding to 1 km/s, based on the current known accuracy of the calibration wavelength and the achievable measurement of the absolute positioning of the hardware. This calibration system can be used to independently calibrate XICS systems on both stellarators and tokamaks, without the need for special plasma conditions often used for calibration, such as locked modes on tokamaks. Experimental and simulated results are shown along with expected results, and the complete design of the calibration hardware that is to be installed in the W7-X XICS system.

X-ray Imaging Crystal Spectrometers (XICSs) detect the emission of highly charged impurity ions to provide ion temperature, electron temperature, plasma flow velocity, and impurity ion density profiles. XICS systems have been used on W7-X,1 LHD,2 Alcator C-Mod,3 EAST,4 KSTAR, and other experiments along with an XICS system planned for ITER.5 The ion temperature is found from the Doppler broadening of spectral lines; the electron temperature is found from the ratio of different lines; the flow velocity is found from the Doppler shift of the lines; and the impurity ion density is found from the line intensities. In a stellarator, the radial electric field, Er, can be inferred through measurements of the perpendicular plasma flow, u.6 The measured Er profiles are important for many aspects of stellarator physics including the study of neoclassical optimization in W7-X. Existing XICS systems lack any independent wavelength calibration capability. On tokamaks, plasma flow measurements are calibrated using locked mode plasma or through reversing the plasma flow. However, on stellarators, locked mode plasmas are not available. This paper presents an independent, robust calibration design that will work for any XICS system independent of the experiment.

XICS systems use spherically bent crystals to diffract x-rays to photon counting imaging detectors. The spectral resolution of the diagnostic is produced by the crystals diffracting light of different wavelengths at different angles according to Bragg’s law. The spatial resolution comes from the spherically bent shape of the crystal. The W7-X XICS diagnostic consists of two channels utilizing two quartz crystals which measure the Ar16+ and Ar17+/Fe24+ spectra, respectively. In particular, the Ar16+ channel utilizes a Quartz (11-20) crystal with a radius of curvature of 1450 mm.

Referenced above, XICS on W7-X is used to infer the radial electric field, Er, from the perpendicular plasma flow, u. The measurement resolution of the perpendicular plasma flow, when ignoring systematic errors, is Δu = 1 km/s (ΔEr = 2 kV/m) with typical levels of argon puffing and a 10 ms integration time. This corresponds to a shift of the line on the detector by about 5 μm. Better resolution is available through additional time binning. The primary systematic errors in these measurements are due to the absolute wavelength calibration across the detector, which in this system, is equivalent to the absolute spatial alignment between the crystal and the detector. The absolute alignment can be considered in two parts: a constant shift across the detector and a tilt of the detector leading to a linear shift of the wavelength. Without an absolute wavelength reference, calibration is based on the ability to see a portion of the plasma on both sides of the magnetic axis and the assumption that the parallel velocity is negligible.

Changes in the absolute wavelength calibration are expected to occur at least on the time scale of thermal changes in the diagnostic or the surrounding environment. Only a few degrees of temperature change, similar to what can be expected from an air conditioning cycle, have been shown to cause thermal expansion of the crystal lattice structure and therefore lead to spectral line shifts on the order of the desired plasma flow measurements.7 Similarly, thermal expansion of the steel structures, that support the crystal and detector, are also on the order to potentially cause significant calibration changes. Measurements of the crystal temperature and repeated calibrations during the planned long pulse discharges can be used to compensate for any spectral line shifts due to thermal expansion.

The wavelength calibration system designed for XICS consists of a x-ray tube positioned in front of the diffracting spherical crystal in order to provide x-ray calibration lines. The specific x-ray lines produced by the chosen anode for the x-ray tube will match the system’s spectral range. This wavelength calibration design will illuminate all spatial channels on the detector simultaneously. The full spatial illumination is achieved by positioning the x-ray tube close to the spherical crystal and inside the focal points of the spherical crystal. A discussion on the modeling and experimental testing of this setup is covered later on in this paper.

The x-ray tube for this calibration system is the XRT 30 by PROTO Manufacturing, Inc. A Cadmium anode on the x-ray tube will provide the Cd and Cd lines needed to calibrate the Ar16+ and Ar17+/Fe24+ channels, respectively, for XICS. The x-ray tube will be operated up to 15 keV and 2 mA. It will be placed 100 mm from the crystals, between the crystals and the plasma. For routine operation between discharges or during long discharges, the x-ray tube is mounted on a rotating arm attached to a standard rotary feedthrough with fine motion control. This setup allows the x-ray tube to be easily rotated in and out of the line of sight of the spherical crystals.

The assembled calibration design is pictured in Fig. 1. For simplicity, the flanges, nipples, feedthroughs, cooling lines, and cabling have been hidden from rendering in Fig. 1. The vertical shaft pictured connects the rotary feedthrough to the rotating arm holding the x-ray tube in addition to a counterweight balancing the x-ray tube. The two arms extending from the upper middle portion of the vertical shaft are physical safety stopping arms that prevent the x-ray tube from being positioned such that x-rays would enter the W7-X experiment. A post fastened to the top flange is the non-rendered component that will limit the motion of the arms.

FIG. 1.

Assembled calibration design along with the placement of the x-ray tube in the crystal chamber. (a) Assembled design and (b) X-ray tube placement.

FIG. 1.

Assembled calibration design along with the placement of the x-ray tube in the crystal chamber. (a) Assembled design and (b) X-ray tube placement.

Close modal

Figure 2 shows how the calibration system will replace the top flange on the chamber that houses the crystal. On the left, the crystal chamber is shown with new addition placed on top. Progressing to the right in Fig. 2, the location of the calibration design can be seen with respect to the spherical crystal. Shown here, the x-ray tube is in the active position required for illumination of one of the detector channels. The system can be rotated to place the x-ray tube in position for the second crystal. In the stored position, the x-ray tube will be rotated behind the crystal and will not limit any current capabilities of the XICS system. Extra ports have been put on the top flange to allow the installation of a camera to ensure that the rotary system is functioning properly.

FIG. 2.

The placement of the x-ray tube with respect to the crystal chamber.

FIG. 2.

The placement of the x-ray tube with respect to the crystal chamber.

Close modal

Figure 3 shows the placement of the X-ray tube in relation to the crystal and the plasma. This top-mounted arrangement of the x-ray tube in the chamber holding the crystal allows convenient access for installation and maintenance of the wavelength calibration system.

FIG. 3.

X-ray tube placement compared to the viewing angles of XICS.

FIG. 3.

X-ray tube placement compared to the viewing angles of XICS.

Close modal

In order to validate the calibration design, X-ray Imaging Crystal Spectrometer Ray Tracing (XICS-RT) was developed as a tool to analyze and test the proposed design. XICS-RT is an object-oriented, ray tracing code written in Python that simulates the x-ray source, the spherical crystal, and the detector. The rays of a desired wavelength are generated with a random direction inside a cone defined by the spread of the source. The rays are propagated to the crystal location, and only those satisfying the Bragg condition are diffracted toward the detector. The detector counts all incident rays and outputs a Tagged Image Format (TIF) image file.

Using XICS-RT with the actual geometry of the XICS system on W7-X coupled with the physical characteristics of the x-ray source, the available configuration space for positioning the x-ray tube such that the detectors are illuminated over the full vertical extent was determined. Figures 4 and 5, respectively, show the horizontal and vertical projections of the configuration space. If the x-ray source is positioned anywhere inside this configuration envelope and directed at the spherical crystal, then the full vertical extent of the detector will be illuminated. The optimal position is along the center of the configuration envelopes.

FIG. 4.

Horizontal projection of the configuration envelope.

FIG. 4.

Horizontal projection of the configuration envelope.

Close modal
FIG. 5.

Vertical projection of the configuration envelope.

FIG. 5.

Vertical projection of the configuration envelope.

Close modal

XICS-RT was verified using an analytic formulation that was developed to understand the geometry of a spherical crystal x-ray spectrometer. Figure 6 shows this geometry. The two circles represent spheres having radii of R and R cos θB with R being the radius of curvature of the spherical crystal and θB the Bragg angle. The crystal’s width extends from C to C+ with C0 being the center of the crystal. M is the center of curvature of the crystal, and P is the location of the planar detector. The rays, which are incident on the center of the crystal and diffracted according to the Bragg angle θB, must be tangential to the sphere with radius R cos θB. In order to satisfy the Bragg condition at the center of the spherical crystal, the source must lie on the C0Q0¯ line. For practical considerations, the source is inside the Q0 position at some arbitrary Q location.

FIG. 6.

Two-dimensional projection (top down view) of the XICS geometry.

FIG. 6.

Two-dimensional projection (top down view) of the XICS geometry.

Close modal

With the source located inside the focal lengths of the crystal, the resulting image on the detector is produced from a virtual image. The location of the virtual image (not shown in Fig. 6) is at the intersection of the lines QM¯ and PC0¯. With this virtual image, a characterization of the detector image can be made by rotating all rays which satisfy the Bragg angle about the QM¯ axis.

Using the actual geometry of XICS, Fig. 7 shows the comparison of XICS-RT to the analytical formulation. In Fig. 7, the vertical extent of the detector refers to the height of the planar detector that is illuminated by an x-ray source positioned at the corresponding location. Any vertical extent above the solid horizontal line (the physical height of the detector) corresponds to the full illumination of all spatial channels on the detector. The plateau region in Fig. 7 resulted from the finite width of the detector plane; the decaying region resulted from the fixed spread of the x-ray source (chosen to match the actual x-ray tube). Good agreement was found between XICS-RT and the analytical formulation.

FIG. 7.

Comparison of XICS-RT and the analytical formulation.

FIG. 7.

Comparison of XICS-RT and the analytical formulation.

Close modal

Testing of the calibration design has been completed at the Plasma Science and Fusion Center (Cambridge, MA) using the High Resolution X-ray Crystal Imaging Spectrometer with Spatial Resolution (HIREXSR, an almost identical system to XICS).3 An x-ray tube (the same model as the one in the calibration design) with a Cadmium anode at 9.5 kV, 0.5 mA was positioned in front of the diffracting crystal at a distance of 4 cm with the angles satisfying the Bragg condition. With an exposure of 500 s (acceptable for wavelength calibrations), full illumination of the two working modules of the detector was found.

Figure 8(a) shows the resulting image from the central module of the detector produced by the Cadmium Lα1 and Cadmium Lα2 lines. Figure 8(b) is the XICS-RT produced image with the simulated x-ray source having the same spread, size, and distance from the spherical crystal as the actual experiment but only including the Cd Lα1 line in the simulation. Figure 8(c) illustrates the agreement between the two by plotting the vertically centered, binned rows from each image.

FIG. 8.

(a) Experimental and (b) simulated images of the x-ray tube with profiles compared with simulated plasma light in (c).

FIG. 8.

(a) Experimental and (b) simulated images of the x-ray tube with profiles compared with simulated plasma light in (c).

Close modal

In an attempt to identify possible calibration schemes, several indirect illumination calibration methods were investigated. These methods were motivated in order to provide a calibration source that would use the full crystal surface, mimicking the plasma. For the first test, a Cu x-ray tube was used to illuminate a thick Cadmium sheet in an attempt to excite reflection x-ray fluorescence. With a 2 mm thick Cadmium sheet placed approximately 30 cm in front of the crystal, the x-ray tube was positioned to the side of the crystal and aimed at the Cadmium sheet such that the illuminated area was in the line of sight of the full crystal. Next, transmission x-ray fluorescence was attempted with a thin 2 μm Cadmium foil, placed 2 cm from the crystal between the crystal and the Cu x-ray source. Finally, direct x-ray reflection of a Cd x-ray tube was tested using Cadmium, steel, and carbon sheets as the reflecting surfaces. All attempted methods failed to produce any detectable Cadmium lines on the detector, even after integration times of many hours. The lack of detectable x-ray emission is likely attributable to very small fluorescence yields. The indirect calibration methods were abandoned in favor of direct illumination of the spherical crystal by the x-ray tube as described in this paper.

The wavelength calibration accuracy of this design is limited by the random and systematic errors present in the system. The first random error in the wavelength calibration is the positioning of the x-ray tube at the optimal Bragg angle location which is dependent upon the rotary feedthrough. Figure 9, the results of a positioning test using XICS-RT, shows the change in position of the spectral lines (both the peak pixel location and a Gaussian fit pixel location) with a change in the angular rotation of the rotary feedthrough. Using the slope of the line fit to the Gaussian fit pixel location, the half-step of the rotary feedthrough (0.0019°) corresponds to a spectral shift of 2 × 10−7 Å (0.0005 pixels). Another random error is the ability to fit the spectral lines of the data. From the experimental testing with a 500 s exposure, the fitting error in the location of the calibration line is 1.5 × 10−7 Å (0.0004 pixels). The third possible random error is thermal expansion of the crystal or support system during the calibration exposure time. This limits the maximum integration time that can be used during the calibration and can be potentially mitigated through characterization of the thermal effects on spectral line shifts coupled with precise temperature measurements.

FIG. 9.

The impact of the x-ray tube placement on the pixel position of the calibration spectral line.

FIG. 9.

The impact of the x-ray tube placement on the pixel position of the calibration spectral line.

Close modal

The first systematic error to be quantified is the precision of the wavelength of the Cadmium lines ( and ) to be used for calibration. A literature study produces the most precise measurements to be 3.956 35 ± 0.000 04 Å and 3.738 23 ± 0.000 04 Å, respectively.8 Two additional systematic errors arise because the x-ray tube is much smaller than the plasma and much closer, meaning only a fraction of the crystal will be used to diffract the calibration x-rays (compared to the plasma light using the full crystal). First, this difference will produce a change in the shape of the spectral lines. Second, depending on the location of the x-ray tube, a spectral line shift will also occur. Both of these systematic errors will be modeled using XICS-RT. A high quality correction is expected; however, the final accuracy of this procedure will depend on our ability to accurately characterize the extent and angular spread of the x-ray source, as well as the rocking curve of the crystal. Figure 10 demonstrates XICS-RT’s ability to produce plasma spectra (the image has been rotated for this paper). This image was generated using a slab Ar16+ plasma (only the W-line simulated) at 5 keV placed at the distance of 3.5 m from the crystal. Given an absolute position accuracy of 1.5 mm (the estimated value based on available remote measuring methods for the x-ray tube location relative to the crystal center), an absolute wavelength accuracy of 1 × 10−5 Å (1 km/s) is achievable. The future addition of a collimating slit may allow an in situ calibration of the position of the x-ray tube with respect to the crystal edges, allowing a significant improvement in the absolute calibration.

FIG. 10.

XICS-RT produced image of the principle Ar16+ line of the plasma.

FIG. 10.

XICS-RT produced image of the principle Ar16+ line of the plasma.

Close modal

A wavelength calibration system using direct illumination of the spherical diffracting crystal has been developed, tested, and designed for XICS on W7-X. This system will provide a wavelength calibration across the whole detector with very high accuracy (limited mainly by integration time) but with a single remaining systematic offset. The random error in the measurement is estimated to be 3 × 10−7 Å (50 m/s) with a systematic error of 1 × 10−5 Å (1 km/s). This greatly improves the ability to measure perpendicular flow profiles to high accuracy and will allow, for the first time on W7-X, parallel flow velocities to be simultaneously measured up to the larger systematic accuracy. Future improvements of the systematic accuracy are expected from future planned improvements in the determination of the absolute location of the x-ray source and cross calibration with other diagnostics.

Research was supported by the U.S. DOE under Contract No. DE-AC02-09CH11466 with Princeton University. This work has been carried out within the framework of the EUROfusion Consortium and has received funding from the Euratom research and training programme 2014-2018 under Grant Agreement No. 633053. The views and opinions expressed herein do not necessarily reflect those of the European Commission. We thank MIT PSFC for the use of their facilities in testing the calibration schemes.

1.
N.
Pablant
 et al., in
41st EPS Conference on Plasma Physics
(
European Physical Society
,
2014
), Vol. 38F, P1.076.
2.
N. A.
Pablant
 et al., “
Layout and results from the initial operation of the high-resolution x-ray imaging crystal spectrometer on the large helical device
,”
Rev. Sci. Instrum.
83
(
8
),
083506
(
2012
).
3.
A.
Ince-Cushman
 et al., “
Spatially resolved high resolution x-ray spectroscopy for magnetically confined fusion plasmas (invited)
,”
Rev. Sci. Instrum.
79
(
10
),
10E302
(
2008
).
4.
B.
Lyu
 et al., “
Measurement of helium-like and hydrogen-like argon spectra using double-crystal x-ray spectrometers on east
,”
Rev. Sci. Instrum.
87
(
11
),
11E326
(
2016
).
5.
J.
Maddox
 et al., “
Multi-energy x-ray detector calibration for Te and impurity density (nZ) measurements of MCF plasmas
,”
Rev. Sci. Instrum.
87
(
11
),
11E320
(
2016
).
6.
N. A.
Pablant
 et al., “
Core radial electric field and transport in Wendelstein 7-X plasmas
,”
Phys. Plasmas
25
(
2
),
022508
(
2018
).
7.
L.
Delgado-Aparicio
 et al., “
Effects of thermal expansion of the crystal lattice on x-ray crystal spectrometers used for fusion research
,”
Plasma Phys. Controlled Fusion
55
(
12
),
125011
(
2013
).
8.
J. A.
Bearden
, “
X-ray wavelengths
,”
Rev. Mod. Phys.
39
,
78
(
1967
).