This paper provides a detailed description of how to construct a pulsed atomic beam source [including a fast ionization gauge (FIG) for characterization] with a unique combination of characteristics. We include technical drawings for a real-time adjustable piezo electric actuated pulsed valve capable of generating a 11 μs duration pulse of gas at a repetition rate of >5 KHz, with a shot-to-shot stability of 0.6%, and maximum densities of 1015 particles/cm3. We also include details on how to construct a FIG, with a 4 μs rise time, to measure the pulse. We report a 3D density map of a supersonic expansion of helium gas with a speed ratio S = 46 and a calculated longitudinal temperature of 0.3 K. Finally, the results of a laser ionization test are provided in order to verify the performance of the pulsed valve in a typical experimental configuration.

Pulsed valves have been used to deliver molecular beams to experiments since the early 1980s. The advantages of using a pulsed source over a continuous source stem from the decreased duty cycle, which allows for operation at significantly higher backing pressures while maintaining low vacuum chamber pressures. Therefore, gas can be delivered at far greater densities than when using continuous flow nozzles. This results in increased expansion of the molecular beam and cooling to temperatures below 1 K.1 

Three competing technologies have been used to actuate fast pulsed valves: Lorentz force, solenoid valves, and piezo electric (PZT) crystals. Lorentz force valves rely on the repulsion between two wires carrying currents in opposite directions and have provided short pulse durations (∼10–20 μs). However, they have been restricted to low repetition rates (∼10 Hz) due to heating limitations.2,3 Solenoid valves operate using a magnetic plunger inside a coil and include the Even Lavie valve, which can generate 20 μs pulses at up to 1 kHz and 180 bars backing pressure.4 However, solenoid valves generate substantial RF noise and produce strong magnetic fields which must be shielded. Additionally, shot-to-shot fluctuations of order <10% can develop as the gasket sealing the valve wears. PZT actuators are attractive because they are self-contained and easy to mount, they require modest driving voltages and currents, their mechanical properties can be easily controlled by varying the dimensions of the ceramic element, and they can be driven rapidly at high repetition rates well above 1 kHz. However, there are drawbacks. For example, the low forces generated by piezo elements (∼0.5 N) make opening against high backing pressures difficult. They are also finicky and difficult to adjust since their performance depends sensitively on the distance between the piezo and the sealing surface—which must be re-adjusted when heating the nozzle, changing backing pressures, or changing carrier gases. As a result, a design which offers all of the advantages of piezo ceramics and can be adjusted in real time from outside the vacuum chamber is very attractive. This is what we have developed and describe in this paper.

At least three unique piezo electric valve designs have been tested and published: Proch and Trickl,5 Irimia et al.,6 and Abeysekera et al.7 The history of piezo electric pulsed valves and their demonstrated performance are summarized in Table I. Only the valve used in the work of Abeysekera et al. allows for adjustment of the piezo-to-seal distance via a micrometer feed through; however, the maximum repetition rate of this design is limited to 100 Hz, and the valve is physically large. We argue that real-time adjustability is a critical feature since the valve performance is very sensitive to the piezo-to-seal distance, which needs be controlled to optimize performance (molecule throughput, pulse duration, and beam contrast) under varying operating conditions.

TABLE I.

History of piezo electrically actuated pulsed valves and demonstrated performance.

YearMin. pulse width (μs)Max. frequency (Hz)Max. pressure (bars)
Auerbach and McDiarmid8  1980 300 20 
Cross and Valentini9  1981 100 750 14 
Proch and Trickl5,10 1989 150 1000 30 
Irimia et al.6  2009 5000 20-30 
Proctor, Albert and Davis11  2010 60 1000 30 
Abeysekera et al.7  2014 20 100 
YearMin. pulse width (μs)Max. frequency (Hz)Max. pressure (bars)
Auerbach and McDiarmid8  1980 300 20 
Cross and Valentini9  1981 100 750 14 
Proch and Trickl5,10 1989 150 1000 30 
Irimia et al.6  2009 5000 20-30 
Proctor, Albert and Davis11  2010 60 1000 30 
Abeysekera et al.7  2014 20 100 

This paper presents a design, with complete detailed technical drawings, for an externally adjustable pulsed atomic beam source, and the results of an experiment to map the propagation of a 11 μs duration pulse in 3D space and time. Our design allows for adjustment of the valve while in operation via a reach rod through the vacuum chamber wall. It can also be adjusted by a remote control motor. It can produce 11 μs pulses at repetition rates of 6.2 kHz. It has been tested over several months and run continuously for 2 weeks (over 7.5 × 109 cycles). It has been operated at backing pressures of 21 bars and heated to 90 °C while running for 2 weeks. The valve and nozzle can produce atomic beams with densities as high as 2 × 1015 particles/cm3 at longitudinal temperatures as low as 0.3 K, which was calculated from a measured average speed ratio of 46. Additionally, it has shot-to-shot fluctuations of only 0.56%. In addition to the technical advantages of our design, it is simple and easy to construct and operate.

The design uses a piezo electric transducer (PZT) bender actuator—a concept which was first conceived by a German group in 1989.12 Our contribution is to mount the PZT on a movable piston, which can be adjusted via a micrometer feedthrough from behind the valve body. The ability to finely control the distance between the piezo and the o-ring without having to disassemble the valve was a central requirement of our design. The valve is constructed from five major components: valve body, piston, micrometer, face plate and nozzle, and PZT (Fig. 1). A complete set of dimensioned technical drawings of the valve components are included in the supplementary material.

FIG. 1.

Picture and cutaway of the valve. (1) Valve body. (2) Piston, which serves as the motion platform. (3) Micrometer—adjusts the piezo to o-ring distance. (4) Front cover and nozzle with a micro o-ring and 40° full opening angle cone. (5) PZT, which is the valve actuator.

FIG. 1.

Picture and cutaway of the valve. (1) Valve body. (2) Piston, which serves as the motion platform. (3) Micrometer—adjusts the piezo to o-ring distance. (4) Front cover and nozzle with a micro o-ring and 40° full opening angle cone. (5) PZT, which is the valve actuator.

Close modal

The body of the valve ① is approximately 50 mm in diameter by 40 mm in length and is manufactured from 303 stainless steel. The faceplate is secured to the body using flat head screws and is sealed with a Viton o-ring. Ports for the micrometer feed through, gas inlet, and electrical connections are 5/16-24 UNF thread with a Viton o-ring. The gas connection is 1/8 in. Swagelock, and the electrical connections are LEMO connectors.13 

At the heart of the internals is a brass piston ② which functions as the motion platform on which the piezo is mounted. Inside the piston is a spring which pushes against the valve body with approximately 62 N of pressure. The restoring force of the spring is counteracted by the micrometer feed through, which is threaded into the piston.14 Turning the micrometer counterclockwise allows the spring to move the piston forward thereby decreasing the distance between the piezo and the o-ring. This closes the valve. Turning the micrometer clockwise opens the valve. Note that this is opposite from traditional valve operation. Although the micrometer was designed to be turned via a rotary feed though, it is conceivable that a servo motor could be attached to the valve to allow it to be adjusted in remote applications.

The 80 tracks per inch (TPI) thread of the micrometer feedthrough ③ allows the user to finely adjust the piezo to o-ring spacing, while the valve is operating. Furthermore, a 7/16 in. hex nut machined into the end allows the micrometer to be turned from outside the vacuum chamber via a reach rod assembly with a standard 7/16 in. socket wrench. To ensure a leak proof operation, the micrometer is constructed with a double o-ring seal, which seals reliably even with movement. No change in vacuum pressure, due to leakage through the double o-ring seal, has been noticed when rotating the micrometer during operation. The micrometer is also machined from 303 stainless steel.

The nozzle ④ is constructed from MACOR ceramic (chosen for its electrical insulating properties) and has a 250 μm orifice which opens to a 40° full angle cone. It is glued into the faceplate using high temperature epoxy, which can withstand continuous operation at 200 °C.15 Resistance to high temperature is essential for allowing the nozzle to be heated during operation. A 40° full angle conical nozzle shape was chosen because of both physical considerations as well as ease of machinability. As compared to a sonic nozzle (180° opening angle), a 40° cone will generate a collimated beam with a higher beam density along the axis of the nozzle, and this will have the effect of slightly elevating the beam translation temperature.16 It is also reasonable to believe that it would promote clustering because of the increased confinement of the particles. The conical aspect was chosen (as opposed to other shapes like a “trumpet” nozzle) because 40° “V” shaped cutting tools are readily available. Conceivably any form of tool could be used to machine the nozzle profile.

Epoxied to the nozzle orifice is a micro Viton o-ring with a 254 μm inside diameter and cross section.17 The o-ring was installed by threading it onto a wire and then threading the wire through the nozzle. In operation, the o-ring seals against 5 mil Kapton tape (127 μm thick) on the PZT. The seal is good enough such that increasing the valve backing pressure from 1 bar to 6 bars produces no detectable change in a 10−6 Torr vacuum maintained by a 100 l/s turbo pump. A small diameter o-ring is essential for allowing the valve to be operated at high backing pressures since the maximum force available from a PZT bender actuator is typically quite low. Additionally, the small diameter o-ring minimizes effects of misalignment in tilt between the o-ring and Kapton gasket. The service life of the o-ring has been determined to be at least 7 × 109 cycles, and we have seen no evidence of loss of performance of the sealing mechanism. However, it is expected that some deterioration of the o-ring will eventually occur, especially in the presence of certain molecular samples. To replace the o-ring, the old o-ring must be popped off with an X-Acto knife, and then the old epoxy can be cleaned up by taking a skim cut on the lathe. Finally, a new o-ring can be glued on. It is recommended that this be performed no more than two times before replacing the MACOR nozzle insert entirely.

The PZT ⑤ is a parallel polled, biomorph, bender actuator constructed from 2 layers of PZT-5H material with a brass shim sandwiched between.18 A 20 μs, 200 V (max), signal is applied to the center electrode to open the valve. The active region of the PZT is 10.16 mm2, and when clamped into the valve, it has 5.38 mm of free length. This gives the PZT a series resonance frequency of 5650 Hz, which was measured by applying a sine wave and noting when the current and driving voltage were in phase. The maximum static displacement at 100 V was measured to be 20 μm using a machinists test indicator. All layers are electrically isolated from one another with Kapton tape and a sheet of mica beneath the bridge clamp.

The performance and stability of the valve while flowing helium and argon has been thoroughly tested; however, we are still in the early stages of testing with molecular samples. Recently, the valve has been tested with CH2I2 and C6H8 injected into the carrier gas and we have seen no ill effects. Concerning chemical compatibility, this design is generally not as robust against corrosive samples because the gas fills the entire valve body and piston and is in contact with the PZT as well as the rest of the materials: stainless steel, brass, Teflon, ceramic, epoxy, Kapton, and Viton. Therefore, caution must be used when choosing molecular samples (especially with molecules containing fluorine or acids). There are some design improvements which could increases the resistance of the valve to corrosives. For example, brass was chosen for the piston material because of ease of machinability and could be replaced with stainless steel. Additionally, PZT elements can be bought which have a stainless steel center reinforcing strip instead of a brass strip. These are available from Piezo Systems in Woburn Massachusetts.

Conceivably the valve could perform with the nozzle heated to higher temperatures than 90 °C since the epoxy and Viton o-ring can sustain continuous operation up to 200 °C. However, the limiting factor might be the curie temperatures of the PZT, which is made from Navy Type VI material and has a curie temperature of 220 °C. Exposure to temperatures in excess of the curie temperature will cause the material to lose its piezo electric properties. The PZT generates heat when it runs, so it is conceivable that the curie temperature could be exceeded even during modest heating of the valve body. The best strategy for heating the nozzle is to focus the heat on the faceplate and allow the PZT to run at slightly cooler temperatures due to the temperature gradient this method would produce in the valve body.

The maximum backing pressure of 21 bars was chosen because of limitations of our experimental apparatus. Testing the valve up to 21 bars never resulted in failure—breaking of the PZT or the inability to open.

The electronics, which drive the piezo, can be divided up into a power circuit and a control circuit. The power circuit consists of a totem pole amplifier which utilizes an op amp to provide the voltage gain and two transistors to provide the current gain. The circuit diagram is included in the supplementary material. The power for the circuit is provided by a 200 V (190 mA) supply and a −20 V (650 mA) supply, which are adequate because the duty cycle is only 2%. The op-amp is wired as a comparator, which means the op-amp is essentially a binary switch. When the signal from the control circuit rises above approximately 1.5 V, the output of the op-amp goes from 0 V to the maximum supplied (typically 100-125 V). The time it takes to do this is a function of the slew rate of the op-amp, and the op-amp we used takes about 5 μs to rise to this voltage. So, the resulting waveform looks like an isosceles trapezoid, which is 20 μs in duration at the base and has 5 μs ramps up and down at the leading and trailing edges.

The control circuit is an embedded controller which takes a trigger from the laser and outputs a square wave to the power amplifier with a variable valve-laser delay, variable repetition rate, and variable pulse duration. It was designed to function with our laser which operates at 1 kHz, but the programming principle can be applied to a laser of any repetition rate. Basically, the microcontroller acts as a countdown timer. For example, if we want to pulse the valve 100 Hz, after every 10th laser trigger, the controller will wait some number of μs (between 0 and a laser period) before outputting the signal to the amplifier. By varying the valve-laser delay we can walk the laser shot through varying gas pulse densities in order to map the entire temporal width of the pulse. The microchip is programed in C and controlled via a laptop running PuTTY SSH client and a USB to serial converter. The user can select the valve-laser delay, the repetition rate of the valve, and the driving pulse duration.

In order to map the density of the pulse, a fast ionization gauge (FIG) was built, which was conceptually based on a 1977 design by Gentry and Giese.2 The FIG works according to the same physical principles as a Bayard-Alpert hot filament pressure gauge; however, because of FIG’s small diameter grid, the measurement is relatively localized in space as well as in time. A CAD model of the FIG, and dimensioned drawings, is available in the supplementary material.

The major components of the FIG are a filament, grid, collector, and amplifier. The filament is helically wound from 127 μm tungsten wire, and applying 15 V and approximately 1 amp to the filament causes it to become yellow hot and emit electrons. The grid is helically wound from 254 μm tungsten wire and held at a 130 V potential which accelerates the electrons emitted by the filament. Down the center of the grid is a tungsten collector where ions are collected and then amplified via a transimpedance amplifier.

The FIG electronics are installed inside the vacuum chamber and plugged directly into the collector. During a pulse, the current output by the collector is on the order of tens of nA, which is dropped across a 750 kΩ resistor and then amplified 25 times by an op amp. The resulting output is sent to an oscilloscope and ranges from tens of mV to volts.

The FIG was tested by comparing its output to the chamber pressure measured by a factory ion gauge. It was determined that, under DC operation, the FIG responds linearly throughout the 10−5 and 10−4 Torr pressure ranges. Additionally, the AC response of the FIG was tested by constructing a pulsed electron source from an old vacuum tube. It was determined that the rise time of the FIG is on the order of 4 μs, which is sufficiently fast for this application.

The valve was mounted inside a vacuum chamber, pumped by a turbo pump with an effective pumping speed of 100 l/s, and the FIG was attached to a three dimensional motion feedthrough so that it could be moved around the inside of the chamber. In the coordinate system used, the origin is located at the nozzle, the z dimension is the direction the gas pulse is propagating, the x dimension is transverse to the beam propagation in the horizontal plane, and the y dimension is transverse in the vertical plane. This experimental setup was used to measure valve’s performance in three regimes of its operating envelope: ① moderate frequency and backing pressure (1 kHz and 7 bars), ② low frequency and high backing pressure (10 Hz at 21 bars), ③ and high frequency and low backing pressure (6.2 kHz at 3.5 bars). The most detailed analysis was performed on the data collected at 1 kHz and 7 bars. Therefore, the following analysis will focus on these results while drawing comparisons with the results obtained in each test.

The valve was backed with 7 bars of helium and pulsed at 1 kHz, with a 20 μs duration driving signal. The chamber pressure increased from 1 × 10−5 to 9.5 × 10−5 Torr. Measurements of the gas density as a function of time were recorded at 36 points in space by moving the FIG around the inside of the chamber. The 36 points were all in the x-z plane and divided into 4 groups, each of which had approximately equal radius from the nozzle. The z axis is along the direction of the beam propagation, and the x dimension is the horizontal transverse displacement. Since the nozzle is a cone, it is assumed that the beam is rotationally symmetric in the x-y plane. The ion signal from each of these locations was used to reconstruct the propagation of a single gas pulse.

Figure 2 shows normalized ion yield as a function of time since the valve trigger with the FIG located on the axis of the nozzle and 1.6 cm from the valve orifice. This pulse duration is 11.5 μs at FWHM and has a rise time of 8 μs. The slightly longer fall time is probably due to the pump taking a finite amount of time to clear the helium gas from the chamber. The pulse-to-background ratio is 65, which can be interpreted as the increase in peak pulse density over the background density.

FIG. 2.

Plot of normalized ion yield versus time since the signal was sent to the piezo to open. The measurement was taken with the FIG at 0° and 1.6 cm away from the nozzle and the valve running at 1 kHz and 7 bars backing pressure. The FWHM of the pulse immediately after exiting the nozzle is 11.5 μs.

FIG. 2.

Plot of normalized ion yield versus time since the signal was sent to the piezo to open. The measurement was taken with the FIG at 0° and 1.6 cm away from the nozzle and the valve running at 1 kHz and 7 bars backing pressure. The FWHM of the pulse immediately after exiting the nozzle is 11.5 μs.

Close modal

By plotting the maximum ion signal recorded at each FIG position at an instant time, still frames of the pulse propagating were constructed. Figure 3 starts 48 μs after the valve is triggered, and each frame is separated by 2 μs. The heat map is indicative of gas density with the brightest colors representing the most dense regions. The x and z axes are in cm, and the nozzle is located at the origin of each plot.

FIG. 3.

Fast ionization gauge signal in volts as a function of z (gas pulse propagation direction) and x (transverse dimension) at snapshots in time. These snapshots begin 48 μs after the valve is triggered and are each separated by 2 μs. The heat map denotes the density of helium gas, which was plotted using the peak of the ion signal, at the specified time, measured by the fast ionization gauge placed at 36 locations in the x-z plane. The nozzle is located at the origin of each plot, and the x and z axes are in centimeters. The gas density falls off like r−1.2, and the angular spread of the beam is 20° (full angle).

FIG. 3.

Fast ionization gauge signal in volts as a function of z (gas pulse propagation direction) and x (transverse dimension) at snapshots in time. These snapshots begin 48 μs after the valve is triggered and are each separated by 2 μs. The heat map denotes the density of helium gas, which was plotted using the peak of the ion signal, at the specified time, measured by the fast ionization gauge placed at 36 locations in the x-z plane. The nozzle is located at the origin of each plot, and the x and z axes are in centimeters. The gas density falls off like r−1.2, and the angular spread of the beam is 20° (full angle).

Close modal

The density of the beam is proportional to r−1.2, which provides a high density on axis even for significant distances from the nozzle, and is consistent with density simulations from pulsed sources performed by Even.19 The angular distribution of the beam is 20° (full angle). As a comparison, when the pressure behind the nozzle was increased to 21 bars the angular width of the beam decreased to 16°. This is expected, given that higher backing pressures produce colder beams in which the molecules have a tighter speed distribution.

It was estimated that each pulse contains approximately 2 × 1014 atoms. This was calculated from the rise in chamber pressure when the valve was activated, considering the effective speed of the turbo pump. The peak density can be estimated using the ideal gas law N/V = p/KBT by first calculating the density of the background and then multiplying this by the pulse-height-to-background ratio recorded by the FIG. Where p is the chamber pressure with the valve turned off (1 × 10−5 Torr), T is the room temperature, and the pulse-height-to-background ratio is 65. The peak density was estimated to be 2 × 1014 atoms/cm3. When the backing pressure was increased to 21 bars and the valve was adjusted to achieve the maximum FIG signal, the magnitude of the FIG signal increased by a factor of 10. Therefore, the maximum density can be approximated as 2 × 1015 atoms/cm3.

The longitudinal dispersion of the pulse was used to infer the speed ratio and temperature of the beam. Considering only points along the axis of nozzle, lines were fit to the FIG-to-valve distance versus time of arrival for different gas densities (interpolated from the fraction of the pulse peak height), and from the slopes of these line, a gaussian distribution of speeds was constructed (Fig. 4). For 7 bars of backing pressure, the most probable speed was 1960 ± 130 m/s. This value is within 1σ of the accepted valve of 1856 m/s.20 Our uncertainty in speed stems from our distance measurement error. Repeated measurements of the same radius result in errors of about ±1 mm, which is consistent with speed errors on the order of ±100 m/s.

FIG. 4.

Plot of velocity, parallel the direction of pulse propagation, for different densities of gas when the valve is backed with 7 bars versus 21 bars of helium. From the spread in velocity, a speed ratio of 21 and 46 was calculated, which corresponds to a temperature of 1.7 K and 0.3 K, respectively.

FIG. 4.

Plot of velocity, parallel the direction of pulse propagation, for different densities of gas when the valve is backed with 7 bars versus 21 bars of helium. From the spread in velocity, a speed ratio of 21 and 46 was calculated, which corresponds to a temperature of 1.7 K and 0.3 K, respectively.

Close modal

According to the formula given by Hillenkamp, Keinan, and Even, the speed ratio S = 2μ/σ, where μ is the most probable velocity and σ is the standard deviation. This yielded a speed ratio S = 21. From this, the average longitudinal temperature of the beam was estimated to be 1.7 K using the following formula and room temperature as To,1 

(1)

At 21 bars backing pressure, the speed ratio increases to S = 46. This corresponds to a longitudinal beam temperature of 0.3 K. This is consistent with atomic beam speed ratio and temperature as a function of backing pressure published by Hillenkamp, Keinan, and Even.1 

As for a temperature prediction of sample molecules injected into the atomic beam, this is a complex question. However, since the mechanical properties of our valve are similar to the properties of the Amsterdam Piezo Valve (APV), it is reasonable to believe that our valve would perform similarly. In one publication, the developers of the APV reported on the results of a supersonic expansion of 0.1% CD3I in 6 bars of neon. They measured a speed ratio of 135 within the densest part of the beam (their experimental setup enabled them to probe slices of the beam) and concluded that the longitudinal temperature of CD3I was 294 mK.6 We believe our valve would perform similarly given the same experimental conditions.

The shortest pulse duration was recorded when the valve was running at 6.2 kHz with 3.5 bars backing pressure, and the FIG was located on the axis of the nozzle, 1.6 cm away from the valve. Under these conditions the chamber pressure increased from 6 × 10−6 to 5 × 10−4 Torr. This pulse measured 10.6 μs FWHM. The pulse-to-background ratio is only about 4, which is a result of the limited pumping speed of this chamber. This result is similar to that published by Irimia et al.,6 where they also measured low pulse-to-background ratios at 5 kHz. For comparison purposes, when the backing pressure was increased to 21 bars, the duration was observed to increase to 15 μs and the pulse-to-background ratio also increased to 148.

The shot-to-shot stability of the valve was characterized by constructing a histogram of the area beneath the ion signal curve for 1000 pulses. Data recorded from these pulses at 1 kHz and 7 bars backing pressure yielded σ/μ = 0.56%, indicating that shot-to-shot fluctuations are less than 0.6%. The shot-to-shot waveform is unimodal (all the pulses look like the waveform depicted in Fig. 2), and the 0.6% variation mentioned is due to fluctuations in the amplitude of the gas pulse. Although we did not measure the shot-to-shot stability in quantitative detail in the long term, qualitatively the valve generated the same pulse shape, duration, and amplitude (unless we purposely changed a parameter, e.g., backing pressure, micrometer adjustment, and temperature) over many weeks of testing.

An experiment was performed using a femtosecond laser to ionize a skimmed argon beam. The ion yield was amplified via a micro channel plate, and the laser-valve timing delay was varied in order to map the temporal duration of the pulse. Additionally, the transverse extent of the pulse was measured by adjusting the focus of the laser.

The valve was backed with 7 bars of argon and pulsed at 1 kHz. The valve adjustment was set using the micrometer so that the pressure in primary chamber, which was pumped by a 250 l/s turbo pump (100 l/s effective at the chamber), remained in the high 10−5 Torr, while the valve was operating. Approximately 2 cm from the pulsed valve nozzle was a molecular skimmer (beam dynamics) with an orifice of 200 μm. Although it is well known that pulsed beams require skimmers with entrance diameters on the order of mm to prevent substantial beam attenuation, a 200 μm skimmer was freely available and deemed acceptable for this test. The secondary chamber was pumped via a turbo pump with an effective speed of 33 l/s, and the pressure remained in the high 10−7 Torr during valve operation. The laser crossing point was located 14.5 cm from the valve nozzle.

Figure 5 shows the integrated ion yield vs. the time delay between the valve trigger and the laser shot. From a normal distribution fit to these data, it was determined that FWHM = 65 μs. This duration is consistent with the measurements made using the FIG, which showed that the FWHM of a pulse of argon will increase by about 3.7 μs/cm traveled. It corresponds to a pulse duration of 17.2 μs directly in front of the nozzle.

FIG. 5.

Integrated ion yield as a function of valve-laser delay. By fitting a normal distribution, FWHM was determined to be 65 μs. This is consistent with the longitudinal dispersion of a pulse of argon gas predicted by the FIG data.

FIG. 5.

Integrated ion yield as a function of valve-laser delay. By fitting a normal distribution, FWHM was determined to be 65 μs. This is consistent with the longitudinal dispersion of a pulse of argon gas predicted by the FIG data.

Close modal

Finally, measurements of the transverse profile of the atomic beam show that the beam is gaussian shaped with a FWHM = 650 μm and a 400 μm density ramp. This gives an angular dispersion of 0.25°, which is consistent with the expected value of 0.28° calculated using the nozzle-skimmer geometry.

We developed and tested a PZT actuated pulsed valve which can generate 11 μs pulses of gas at a maximum demonstrated frequency of 6.2 kHz and peak densities of 1015 particles/cm3. The valve has also demonstrated prolonged operation at 21 bars backing pressure and temperatures of 90 °C. A unique feature of this design is that the PZT bender actuator is mounted to a motion platform, which we control via a rotary feedthrough from outside the vacuum chamber. This allows for real-time adjustment of the distance between the piezo and the sealing o-ring. This is essential for overcoming some of the shortcomings of piezo valves, such as sensitivity to temperature, changes in backing pressure, and carrier gas type.

Additionally, a fast ionization gauge was developed, based on an earlier design by Gentry and Giese, in order to map the gas density within the pulse in 3D. We measured an 11 μs duration pulse of helium, supersonically expanded from 7 bars into 10−5 Torr vacuum from a 40° conical nozzle, and determined that the beam contains approximately 2 × 1014 particles/pulse, a divergence of 20°, and a density which falls off like r−1.2. The calculated speed ratio of this beam is 21, which corresponds to a longitudinal temperature of 1.7 K. Increasing the backing pressure to 21 bars resulted in a speed ratio of 46 and a corresponding beam temperature of 0.3 K. Finally, a laser ionization test was conducted with argon as the carrier gas, which verified the results of our experiments using the FIG. In the supplementary material, we present a complete set of technical drawings which would allow the reader to construct a pulsed valve and a FIG in order to develop their own pulsed atomic beam source.

Since it is difficult to consider this pulsed valve without bringing to mind the commercially available Amsterdam Piezo Valve (APV) and Even Lavie Valve (EL), we have offered some comparisons between the performance of our valve versus these industry standards throughout this paper. In summary, our valve is more closely related to the APV in operating principle (both use a PZT bender actuator) than the EL valve, and they are more similar in performance as well. The APV valve does generate about 1 order of magnitude more atoms per pulse than our valve, which is likely due to the greater deflection of their higher performance PZT. But, in fact, without any modifications to this valve, one could install a PZT with similar material properties to the PZT used in the APV (like PSI-5H4E, available from Piezo Systems, with a d31 constant of −320 pm/V), and presumably match the number of atoms per pulse of the APV since the performance of our valve can be optimized in real time. Furthermore, the micro o-ring in our design reduces the force on the PZT thereby facilitating operation at higher backing pressures, which combined with a stainless steel reinforced PZT could enable PZT valves to approach solenoid operated valves in backing pressure and beam densities while maintaining their 10 μs pulse durations and >5 kHz repetition rates.

See supplementary material for builders’ notes and a complete set of plans to build the valve, fast ionization gauge, and electronic circuits.

We thank Eugene Shafto and Richard Lefferts, whose support and expertise were invaluable. This research was supported by the National Science Foundation under Grant No. 1505679.

1.
M.
Hillenkamp
,
S.
Keinan
, and
U.
Even
,
J. Chem. Phys.
118
,
8699
(
2003
).
2.
W. R.
Gentry
and
C. F.
Giese
,
Rev. Sci. Instrum.
49
,
595
(
1978
).
3.
B.
Yan
,
P.
Claus
,
B.
Van Oorschot
,
L.
Gerritsen
,
A.
Eppink
,
S.
van de Meerakker
, and
D.
Parker
,
Rev. Sci. Instrum.
84
,
023102
(
2013
).
4.
U.
Even
,
J.
Jortner
,
D.
Noy
,
N.
Lavie
, and
C.
Cossart-Magos
,
J. Chem. Phys.
112
,
8068
(
2000
).
5.
D.
Proch
and
T.
Trickl
,
Rev. Sci. Instrum.
60
,
713
(
1989
).
6.
D.
Irimia
,
D.
Dobrikov
,
R.
Kortekaas
,
H.
Voet
,
D. A.
van den Ende
,
W. A.
Groen
, and
M. H.
Janssen
,
Rev. Sci. Instrum.
80
,
113303
(
2009
).
7.
C.
Abeysekera
,
B.
Joalland
,
Y.
Shi
,
A.
Kamasah
,
J. M.
Oldham
, and
A. G.
Suits
,
Rev. Sci. Instrum.
85
,
116107
(
2014
).
8.
A.
Auerbach
and
R.
McDiarmid
,
Rev. Sci. Instrum.
51
,
1273
(
1980
).
9.
J. B.
Cross
and
J. J.
Valentini
,
Rev. Sci. Instrum.
53
,
38
(
1982
).
10.
A.
Flettner
,
J.
Günther
,
M.
Mason
,
U.
Weichmann
,
R.
Düren
, and
G.
Gerber
,
Appl. Phys. B
73
,
129
(
2001
).
11.
D. L.
Proctor
,
D. R.
Albert
, and
H. F.
Davis
,
Rev. Sci. Instrum.
81
,
023106
(
2010
).
12.
D.
Gerlich
,
G.
Jerke
,
U.
Mueck
, and
U.
Person
, Internal Document Physics Department,
Freiburg University
,
Germany
,
1990
, https://www.tu-chemnitz.de/physik/ION/Technology/.
13.

The LEMO connectors have Part No. EWV.00.250.NTLPV and the plans for the adapter to mate them with the body of the valve are included in the supplementary material.

14.

The spring is distributed by LeeSpring and has a Part No. LC 093M 04 S.

15.

This epoxy was purchased from Thorlabs and has Part No. 353NDPK.

16.
K.
Luria
,
W.
Christen
, and
U.
Even
,
J. Phys. Chem. A
115
,
7362
(
2011
).
17.

The o-ring was manufactured by Precision Associates Inc. and has a Part No. 10-10 9746.

18.

The piezo element is cut down, according to the plans in the supplementary material from a stock model distributed by Steminc which has a Part No. SMBA4510T05M.

19.
20.
W.
Christen
,
T.
Krause
,
B.
Kobin
, and
K.
Rademann
,
J. Phys. Chem. A
115
,
6997
(
2011
).

Supplementary Material