A new projection type imaging system is presented. The system can directly image the field emission site distribution on a cathode surface by making use of anode screens in the standard parallel plate configuration. The lateral spatial resolution of the imager is on the order of 1–10 μm. The imaging sensitivity to the field emission current can be better than the current sensitivity of a typical electrometer, i.e., less than 1 nA.

In situ, real-time or time-resolved imaging at micro- and nano-scales has provided a great deal of understanding of various processes that take place in materials. Many contemporary imaging concepts are intrinsically based on raster scanning. This means that the pointed focused/sharp probe, be it an electron or ion probe, or a mechanical cantilever, scans across a given area to resolve and plot an image of a static or dynamic pattern. Such images consist of pixels or voxels containing locally collected information. Examples are SEM, STEM, STM, AFM, imaging TOF SIMS, and others.

A second approach is much older and comes from photography. This is the projection method and it is entirely different from raster scanning. In this approach, a surface of interest under electromagnetic irradiation or electron bombardment is imaged at a certain distance beyond the interaction location giving rise to a significant magnification. Examples are electron diffraction in TEM and LEED, and X-ray topography. The most remarkable examples here would be the field emission microscope and field ion microscope invented by Erwin Mueller, in which electron or ion imaging of a sharp tip placed in a high electric field becomes possible with nanometer or atomic resolution without complex motorized systems, data acquisition, and post-processing methods.

In the realm of finding inexpensive, reliable, and adaptive field emission sources, beyond Spindt cathodes, to drive various devices and systems, many novel advanced materials have been studied. This research field remains extremely active and is of significant interest to this very day. Field emission sources often yield laterally non-uniform emission. This is because field emission is a non-linear process and therefore surface termination and geometrical features contribute significantly to the surface barrier formation and its lateral uniformity. Thus, in practice a packaged field emission source will always have strong and weak emission points.

To the best of our knowledge, systems to observe the lateral uniformity of field emission arrays are always raster-scanning tools that move a micron scale anode tip across a field emitter surface.1–3 In this paper, we present an electron emission projection imager system that captures lateral emitter distribution of the entire surface instantaneously.

The conceptual diagram of the imaging system and its actual appearance are shown in Fig. 1. The imager has three anodes: (1) an optically polished (1 in. dia. and 100 μm thick) disk made of yttrium aluminum garnet doped with cerium (YAG:Ce) which is coated with a Mo film of about 7-8 nm in thickness; (2) a transparent and conductive tin-doped indium oxide (ITO) film of 175 nm deposited onto a boro-aluminosilicate glass (BASG) substrate of 700 μm in thickness; (3) a polished (1 in. dia. and 100 μm thick) disk made of stainless steel (SS).

FIG. 1.

A schematic of the electron emission imager. Solid arrows represent the 3 cameras: two cameras are used to monitor parallelism between the cathode and anode surfaces upon installation and to evaluate the interelectrode gap during field emission measurements, and the third camera is used to detect light from either anode screen luminescence or light originating from the cathode surface itself (exampled below), or both.

FIG. 1.

A schematic of the electron emission imager. Solid arrows represent the 3 cameras: two cameras are used to monitor parallelism between the cathode and anode surfaces upon installation and to evaluate the interelectrode gap during field emission measurements, and the third camera is used to detect light from either anode screen luminescence or light originating from the cathode surface itself (exampled below), or both.

Close modal

The Mo/YAG:Ce and ITO/BASG anodes are imaging screens and the SS anode is to measure long-term temporal current stability. Mo/YAG:Ce and ITO/BASG are semitransparent in the visible range. Fig. 2(a) illustrates detailed UV-vis. spectra and Fig. 2(b) demonstrates transparency when anode screens covering color lines are backside illuminated.

FIG. 2.

(a) UV-vis. spectra (spectral dependence of transmittance) for the YAG:Ce (red line), Mo/YAG:Ce (green line), and ITO/BASG (blue line) anode screens. (b) Photographs that demonstrate the semitransparency of the ITO/BASG anode (top) and Mo/YAG:Ce anode (bottom).

FIG. 2.

(a) UV-vis. spectra (spectral dependence of transmittance) for the YAG:Ce (red line), Mo/YAG:Ce (green line), and ITO/BASG (blue line) anode screens. (b) Photographs that demonstrate the semitransparency of the ITO/BASG anode (top) and Mo/YAG:Ce anode (bottom).

Close modal

Depending on the specific measurement to be performed, the anodes can be interchanged by linearly moving the frame holding all three anodes, by means of an ultrahigh vacuum actuator. The distance between the sample and the anode is set using a micrometer holding the sample. Top and side view cameras outside the vacuum are used to check the parallelism between the cathode and anode and to measure the gap. A Canon DSLR camera is installed at a viewport behind the anodes to take pictures of electron emission patterns. The sample electrode is at ground and the anode frame is isolated and positively biased. The bias and current readings are enabled by a Keithley 2410 electrometer. The Keithley and the Canon are programmed such that they collect data sets of current-voltage (I-V) curves and images synchronously.

Before going into a detailed description of the imager, we would like to introduce a series of pictures taken on a carbon nanotube (CNT) sample grown by a CVD process. Fig. 3(a) illustrates the appearance of the sample in an optical micrograph, Fig. 3(b) illustrates the morphology as measured using a scanning electron microscope (SEM), Fig. 3(c) illustrates the green (550 nm) light produced by the Mo/YAG:Ce anode screen under ∼1 keV and ∼100 μA electron bombardment, and Fig. 3(d) illustrates clusters of intense red light imaged with the ITO/BASG anode screen under similar ∼1 keV and ∼100 μA conditions. The red cores already can be seen in Fig. 3(c) because the red light intensity is extremely high and the Mo/YAG:Ce screen is semitransparent in red. The red light thermally produced by the CNT due to electric current flow induced heating is a well-known effect which has been documented over the last 15 yr.4–8 Note that the Mo/YAG:Ce and ITO/BASG anode images were taken at different sample installations and different camera presets and therefore the red light patterns do not coincide.

FIG. 3.

CNT on the stainless steel substrate: (a) an optical microscopy image of the CNT sample; (b) a high magnification SEM image showing the CNT morphology; (c) emission pattern imaged with the Mo/YAG:Ce anode screen; (d) emission pattern imaged with the ITO/BASG anode screen. The field of view (the white circle diameter) in the fragments (c) and (d) is 4.4 mm.

FIG. 3.

CNT on the stainless steel substrate: (a) an optical microscopy image of the CNT sample; (b) a high magnification SEM image showing the CNT morphology; (c) emission pattern imaged with the Mo/YAG:Ce anode screen; (d) emission pattern imaged with the ITO/BASG anode screen. The field of view (the white circle diameter) in the fragments (c) and (d) is 4.4 mm.

Close modal

Along with imaging, I-V curves can be collected as demonstrated in Fig. 4(a). I-V characteristics of the CNT sample are presented as semi-logarithmic and Fowler-Nordheim plots. The visible kink between 1 and 10 μA always correlates with the onset of the red light emission from the CNTs. The abscissa is presented in units of electric field after the recorded voltage was divided by the interelectrode gap taken by the top camera (Fig. 4(b), right). As mentioned, the opaque SS anode is used to measure long-term emission current stability. An example of the results of a long 22 h run carried out for the CNT sample is shown in Fig. 4(b).

FIG. 4.

Electric characterization of the CNT sample: (a) I-V characteristics in semi-log and Fowler-Nordheim representation and (b) current time stability measurement taken over a 22 h period. The right panel demonstrates an example of electrode parallelism and the gap between the SS anode and the CNT cathode measured as 496 μm.

FIG. 4.

Electric characterization of the CNT sample: (a) I-V characteristics in semi-log and Fowler-Nordheim representation and (b) current time stability measurement taken over a 22 h period. The right panel demonstrates an example of electrode parallelism and the gap between the SS anode and the CNT cathode measured as 496 μm.

Close modal

Spatial resolution of the acquired images is limited by the optical system. For the case shown in Fig. 3(c), the resolution is 8 μm per pixel (the known sample diameter is 4400 μm and the apparent sample diameter in Fig. 3(c) is 550 pixels). Using another lens, we recently were able to successfully achieve higher optical magnifications which resulted in higher spatial resolutions of 5, 2 and 1 μm per pixel.

ITO/BASG screens were purchased from Delta Technologies. The anode screen is a 1 in.2 BASG square and 0.7 mm in thickness. An ITO film of ∼175 nm is deposited on one side of the BASG square. The ITO film resistance is between 4 and 10 Ω. The conductive ITO side faces the cathode under measurement conditions.

To establish an electric field between the sample and the isolating YAG:Ce anode and to collect the current, a metal film needs to be deposited. Molybdenum was chosen because it is dense and allows for continuous ultrathin films. In addition, the melting point of Mo is 2896 K, meaning that Mo coatings can sustain exceptionally high electrical power surface densitiesIVcm2. We specify this because originally Al coatings were used that would melt upon the first run producing bulged anode surfaces at the location where the current was collected. Mo is deposited in house by magnetron sputtering on one side of an optical-quality polished YAG:Ce disk. Base pressure in the magnetron system is <5 × 10−7 Torr. Prior to coating, the YAG:Ce disk is cleaned in situ using RF discharge plasma. Without breaking vacuum, immediately after the cleaning, the Mo coating is deposited. Ar is used as a working gas for both cleaning and sputtering at a pressure of ∼10−3 Torr.

For voltages less than 1.1 kV, provided by the Keithley 2410, the Mo coating has to be ultrathin. A standard deposition time of 60 s was found to be optimal for our application. This means that the thickness of the resulting Mo film is sufficiently small to let many electrons into YAG:Ce to produce photons, but thick enough to filter electrons such that they do not penetrate too deeply into the YAG:Ce phosphor – this would lead to charging and discharging effects in YAG:Ce. For instance, a Mo thickness corresponding to a 60 s deposition time does not comprehend applied voltages over 2 kV. The electron penetration depth becomes such that electrons start accumulating inside YAG:Ce and do not efficiently drain. This results in luminescence instabilities and may lead to strong discharge events damaging both the YAG:Ce and Mo coatings. The choice of Mo thickness is determined by the application and the voltage range to be applied between the electron emitter under study and the anode.

For a specific and quantitative choice of the Mo thickness, the Kanaya-Okayama approach9 can be used. The following formula relates the electron energy, E [keV], to the electron range, RKO [μm], the maximum possible penetration depth,

(1)

where MM is the molar mass [g/mol], Z is the atomic number, and ρ is the mass density [g/cm3]. The resulting dependence is shown in Fig. 5 and provides a precise answer for the maximum thickness of Mo. The example calculation at 1 keV yields RKO = 9.3 nm. In practice, one would use not the maximum range but rather the so-called practical range which is about 0.7 of RKO.10 This leads to a specific requirement on the Mo coating – 6.5 nm. At this Mo thickness, about 10% of electrons that were not back scattered will penetrate into YAG:Ce and produce photons.

FIG. 5.

The dependence of the electron range on the energy, calculated for Mo using the KO formula, Eq. (1).

FIG. 5.

The dependence of the electron range on the energy, calculated for Mo using the KO formula, Eq. (1).

Close modal

Now we can explain why the deposition time of 60 s works best for the 1.1 keV application by converting the deposition time into the actual thickness of the Mo coating. To measure the thickness of the Mo coating, we synthesized three test samples on optically polished Si witness coupons and depth profiled them using secondary ion mass spectrometry (SIMS). One sample was deposited for 60 s (the same time used for YAG:Ce anode fabrication), and two others were deposited for 90 s and 120 s. With SIMS, it is possible to measure the thickness. By definition, the sputtering rate (SR) is calculated by dividing the layer thickness d by the time T it took to sputter through it. On the other hand, the SR can be expressed through other experimental and fundamental material parameters: the ion current density j, the sputtering yield (i.e., the number of ejected atoms per incident ion) Y, the elementary charge e, and the atomic density of the target material (units of atoms per volume) N, or alternatively the mass density of the target material ρ and the atomic mass of the target material M. Combining everything together, we can deduce a formula for the thickness of the layer to be determined as

(2)

The test samples were depth profiled using a time-of-flight SIMS instrument in the traditional single beam mode.11 The 60° oblique 5 kV Ar+ ion beam was focused to a spot of 30 μm in diameter at a current of 80 ± 8 nA as measured by an in situ Faraday cup. When depth profiling, the beam was raster scanned over an area of 675 × 675 μm2, yielding an ion current density j=17.6±1.8μA/cm2. Both the single ion beam spot and the raster sizes were measured using the in situ optical Schwarzschild microscope that directly images the surface of the sample and its modification in real time.12 Electronic gating during the profiling experiments was set to 50%. The resulting depth profiles are illustrated in Fig. 6. We determine the time T at which the Mo film is eroded away as a moment at which the Si and Mo (main isotopes 28 and 98 atomic mass units, respectively) signals cross. With this fixed, the Mo film deposited for 60 s was eroded in 72 s, the one deposited for 90 s was eroded in 109 s, and the one deposited for 120 s was eroded in 144 s. Using Eq. (2), three thicknesses 7.5 nm (60 s deposition), 11.4 nm (90 s deposition), and 15.1 nm (120 s deposition) were calculated with j=17.6±1.8μA/cm2; M = 95.95 a.m.u. = 95.95 × 1.7 × 10−24 g; ρ = 10.28 g/cm3; e = 1.6 × 10−19 C; Y = 6.02. Y=6.02 was obtained using SRIM/TRIM,13 a Monte Carlo code, in the full cascade mode on the ensemble of 105 Ar+ ions with their energy set to 5 keV and the irradiation angle set to 60°.

FIG. 6.

Depth profiles for Mo films on Si witness coupons.

FIG. 6.

Depth profiles for Mo films on Si witness coupons.

Close modal

We stress that (1) the apparent spike for the Si signal is due to the presence of the C–O molecule on the surface (no isotopic pattern for Si was observed in a full mass spectrum recorded for the pristine surface, sputtering time = 0 s) and (2) the apparent spike for the Mo signal is due to slight metal surface oxidation that always leads to orders of magnitude higher secondary ion signals in SIMS (secondary ion formation probability is higher in metal-oxides compared to clean metals).14 

The thickness of 7.5 nm explains why the standard Mo/YAG:Ce screen is semitransparent in the visible range (Fig. 2), and thus why we observe the combination of green and red photons when imaging electron emission with the YAG:Ce screen (Fig. 3). The determined 7.5 nm thickness of the standard Mo coating, being in excellent agreement with electron range calculations, straightforwardly explains why a deposition time of 60 s is optimal for imaging with the YAG:Ce phosphor screen at applied voltages of up to 1.1 kV.

Taking it a step forward, let us estimate the photon sensitivity of the system. The first step here would be to calculate the electron to 550 nm photon conversion as

(3)

where Tevac/Mo is the transmission coefficient of electrons through the interface between vacuum and the Mo film (1 minus reflection), TeMo is the transmission coefficient of electrons through the Mo film, CYAG is the electron-to-photon conversion of all the electrons that penetrate through the Mo film and get embedded in the YAG:Ce screen. Tevac/Mo is only dependent on Z and is approximately equal to 0.7.10 TeMo is equal to 0.1 at a Mo thickness close to the practical electron range (6.5 nm at 1 keV). We consider CYAG = 0.01.15 

Assuming that 1 keV electrons provide a relatively uniform distribution of a current of 100 μA over an area of 2.2 mm in radius (Fig. 1) captured by the Canon 6D on the 500 × 500 pixels sensor area and using the resulting Ceph=7×104, we obtain that the signal per pixel is 106 photons/s with a Poisson noise of 103 and a S/N ratio 103. If the current (emission is still considered uniform) is reduced to 100 nA and further to 1 nA (the limit for the Keithley), then the signal per pixel reduces to 103 (S/N ≈ 30) and 10 (S/N ≈ 3) photons/s, respectively. Since the read noise level at ISO = 1000 (used in experiments) in the Canon 6D is between 1 and 5 e, such a low electron emission should be still observable by the camera at 1 keV. This would be the worst case scenario. In fact, we typically see that emission starts with isolated small features, meaning that even though the current is reduced by many orders of magnitude the local pixel photon density on the sensor can remain very high.

The main factor here is not the current or current density providing the Canon sensor with enough photons per pixel but the electron energy that should remain higher than 700 eV in order to keep the practical electron range no less than 5 nm. In this sense, Mo/YAG:Ce anode screens with different Mo thicknesses can be combined (up to 3 on the frame holder shown on the schematic diagram in Fig. 1) and/or a larger inter-electrode gap used to enable higher electron kinetic energy at the same electric field in order to detect small signals.

At 100 μA, 1 keV, and ISO=1,000, we can calculate the photographic equivalent of the light produced on the Mo/YAG:Ce screen, namely, 106 photons per pixel per second correspond to 1016 photons per second per m2. At the wavelength of 550 nm, this is about 1/2 lx or 1 EV. This means that imaging with the Mo/YAG:Ce anode screen at 100 μA and 1 keV is equivalent to photographing the full moon on a clear night. At lower currents, it becomes equivalent to taking images of a moonless clear sky with stars.

We note here that the best practice to take pictures in the described arrangement is as follows. The ISO has to be 1000-3000 because it drastically reduces pattern/banding noise and electronic noise and allows better discrimination of small lighting features above the background when collecting for many seconds. Second, top level cameras preserve the highest dynamic range above 10 photographic stops, with the dynamic range measured in photographic stops being calculated as log2fullwellcapacity[electrons]totalnoise[electrons]. This means that a high dynamic range and low noise at ISO of a few thousands enable simultaneous detection of lighting features of exceptionally different brightnesses, from ∼10 to 105 photons per pixel. This is an extremely useful capability when one wants to collect images through the entire I-V curve acquisition process, from tens of nA to hundreds of μA.

In Fig. 7, the vacuum diagram of the imager is presented. Evacuation process of the vacuum chamber/vessel is threefold. The first pumping stage is an oil-free diaphragm pump (ultimate pressure of 1.5 Torr). Second stage pumping uses a turbo-molecular pump: the pumping speed for N2 is 67 l/s, the compression ratio for N2 and Ar is 1011, the compression ratio for H2 is 105, and the typical ultimate pressure with no high temperature baking is 2 × 108 Torr at the backing pressure of 1.5 Torr. The third stage is an ion pump (pumping speed for N2 is 30 l/s) that is turned on at 5× 106 Torr which significantly improves the pumping speed toward 2 × 108 Torr.

FIG. 7.

Vacuum diagram of the imager.

FIG. 7.

Vacuum diagram of the imager.

Close modal

To quickly (in about 24 h) achieve the nominal working pressure of ≈3 × 109 Torr, which is satisfactory enough to perform measurements, the following procedure is established. The vacuum system does not have any single Viton or rubber O-ring connections. All vacuum components such as mechanical holders and screws are ultrasonically cleaned. Step one is to clean in a mixture of distilled water and an acid based detergent which is performed for about 30 min. Step two is to clean in a 50/50 mixture of ethanol and acetone which is performed for another 30 min. All parts are dried using dry nitrogen before installation. All parts undergo a high temperature drying during the first bake out of the system. The routine bake out procedure is as follows. The system is heated up to 140-145 °C when a pressure of 8× 108 Torr or lower is achieved through the described threefold procedure. The choice of the bake out temperature is limited by the Pfeiffer full range Pirani/cold cathode gauge in use (recommended baking temperature must not exceed 150 °C with the magnet detached). The chamber is kept at 140-145 °C for about 12 h. The measure of a successful baking procedure is if the pressure (with the gauge disconnected, calculated via the tabulated “ion current versus pressure” curve provided for the ion pump by the manufacturer) returns to 8 × 108 Torr or better at 140-145 °C. After that, the heating is turned off and the chamber naturally cools downs for another 12 h and reaches the targeted ≈3 × 109 Torr. This procedure is fully automated. The final step is to close an isolation gate valve (the one atop the turbo pump symbol in Fig. 7) and turn off the diaphragm and turbo pumps and wait until the pressure stabilizes at 3×109 Torr with only the ion pump on. Mechanical vibration free pumping is necessary because field emission systems are sensitive to vibrations resulting in cathode-anode gap fluctuations producing noisy current readings. An SRS residual gas analyzer monitors gas composition in the system and checks for any virtual or actual leaks.

The entire data acquisition process is automated using the LabVIEW programming environment. The Keithley 2410 simultaneously acts as a voltage source (active mode: control signal in/feedback voltage reading out) and a pico- to micro-ammeter (passive mode: current reading out). The vacuum reading is taken directly from a Pfeiffer controller connected to the full range Pirani/cold cathode gauge. Ion pump current/pressure is acquired using a DAQ USB-6003 by National Instruments. Electron emission images are taken by the Canon 6D equipped with a 50 mm lens and close-up lenses and extensions; the standard work distance is set at 6 cm.

Fig. 8 illustrates the algorithm to measure the current as the voltage is swept. After the voltage is increased/decreased and before the current measurement starts, there is a 1 s delay (typically dt1 = 1 s, but can be varied down to 0 s) introduced to allow time for the electrometer internal circuit to settle. After the dt1 delay, the current is sampled n times with a sampling rate 1/dt2. Typically n = 20 and dt2 = 300 ms, but both values can be varied. The current, voltage, and pressure are sampled and processed to obtain the mean value, the standard deviation, and the total measurement error. All are calculated online using well-known statistics formulas

(4)
(5)
(6)

where tp,n = 2.09 is the Student’s t-distribution coefficient for a confidence interval of 95%, n = 20 is the number of measurements at a given voltage, θ = 50 pA and 10 μV are the ultimate accuracies of the Keithley 2410 electrometer for current and voltage, respectively. All data, raw and processed, are recorded and stored into files.

FIG. 8.

Schematic diagram of current measurement algorithm.

FIG. 8.

Schematic diagram of current measurement algorithm.

Close modal

The software is designed such that the I-V curve measurement and collection of images using Mo/YAG:Ce or ITO/BASG anode screens are synced within a specified run, with the voltage ramped up and down. The entire block diagram is shown in Fig. 9. The LabVIEW program allows the specification of a voltage step to repeatedly acquire images with the Canon 6D as the voltage is swept up or down. One can specify images to be taken every 20, 50, 100 V, etc., while the voltage is being ramped. The voltage step size can be between 1 V and 1 kV.

FIG. 9.

Block diagram of the LabVIEW software for fully automatic data and image acquisition.

FIG. 9.

Block diagram of the LabVIEW software for fully automatic data and image acquisition.

Close modal

We presented a design and detailed description of an apparatus for direct imaging of the electron field emission site distribution on cathodes surfaces. We implemented a projection approach to the imaging problem, making use of anode screens in the standard parallel plate configuration. Detailed characterization and performance metrics of the imaging Mo/YAG:Ce anode screen are given, and recommendations for metal coating procedures are presented based on our experience. In particular, we have calculated the electron to photon conversion efficiency as well as measured the response and sensitivity of the optical system to the emitted current. We also estimated that the lateral spatial resolution of the imager can be on the order of 1–10 μm. As a part of the imaging apparatus, the fully automated vacuum control/monitor and data acquisition software was described.

The commissioned electron emission projection imager, in combination with microscopy and spectroscopic methods, may unveil electron emission induced processes on the micron and submicron scales.

Euclid was supported by the Office of Nuclear Physics of DOE through a Small Business Innovative Research Grant No. DE-SC 0013145. We also thank the following:

Alexander Zinovev of Materials Science Division at ANL for providing the access to SARISA, a mass spectrometer. Use of SARISA was supported by the U.S. Department of Energy, Office of Science, Materials Sciences and Engineering Division.

Kiran Kovi and Anirudha Sumant of the Center of Nanoscale Materials at ANL for providing us with the CNT sample. Use of the Center for Nanoscale Materials, an Office of Science user facility, was supported by the U.S. Department of Energy, Office of Science, Office of Basic Energy Sciences, under Contract No. DE-AC02-06CH11357.

Jiaqi Qiu and Richard Konecny of Euclid TechLabs for their kind help with some of the hardware manufacture.

SEM measurements were conducted in the Electron Microscopy Center of the Center for Nanoscale Materials at Argonne National Laboratory. Use of the Center for Nanoscale Materials, an Office of Science user facility, was supported by the U.S. Department of Energy, Office of Science, Office of Basic Energy Sciences, under Contract No. DE-AC02-06CH11357. S.S.B. was partially supported by NSF Grant Nos. PHY-1535639 and PHY-1549132 during the time of preparing the manuscript for publication.

1.
D.
Lysenkov
and
G.
Müller
,
Int. J. Nanotechnol.
2
,
239
(
2005
).
2.
D.
Lysenkov
, “
Optimization of nanostructures for field emission cathodes
,” Ph.D. thesis,
The University of Wuppertal
,
2006
.
3.
A. V.
Karabutov
,
V. D.
Frolov
,
S. M.
Pimenov
, and
V. I.
Konov
,
Diamond Relat. Mater.
8
,
763
(
1999
).
4.
S. T.
Purcell
,
P.
Vincent
,
C.
Journet
, and
V. T.
Binh
,
Phys. Rev. Lett.
88
,
105502
(
2002
).
5.
M.
Sveningsson
,
M.
Jönsson
,
O. A.
Nerushev
,
F.
Rohmund
, and
E. E. B.
Campbell
,
Appl. Phys. Lett.
81
,
1095
(
2002
).
6.
J.
Wei
,
H.
Zhu
,
D.
Wu
, and
B.
Wei
,
Appl. Phys. Lett.
84
,
4869
(
2004
).
7.
D.
Mann
,
Y. K.
Kato
,
A.
Kinkhabwala
,
E.
Pop
,
J.
Cao
,
X.
Wang
,
L.
Zhang
,
Q.
Wang
,
J.
Guo
, and
H.
Dai
,
Nat. Nanotechnol.
2
,
33
(
2007
).
8.
M.
Cahay
,
P. T.
Murray
,
T. C.
Back
,
S.
Fairchild
,
J.
Boeckl
,
J.
Bulmer
,
K. K. K.
Koziol
,
G.
Gruen
,
M.
Sparkes
,
F.
Orozco
, and
W.
O’Neill
,
Appl. Phys. Lett.
105
,
173107
(
2014
).
9.
K.
Kanaya
and
S.
Okayama
,
J. Phys. D
5
,
43
(
1972
).
10.
J. I.
Goldstein
,
D. E.
Newbury
,
P.
Echlin
,
D. C.
Joy
,
C.
Fiori
, and
E.
Lifshin
,
Scanning Electron Microscopy and X-ray Microanalysis: A Text for Biologists, Materials Scientists, and Geologists
(
Plenum Press
,
New York
,
1981
).
11.
S. V.
Baryshev
,
N. G.
Becker
,
A. V.
Zinovev
,
C. E.
Tripa
, and
I. V.
Veryovkin
,
Rapid Commun. Mass Spectrom.
27
,
2828
(
2013
).
12.
I. V.
Veryovkin
,
C.
Emil Tripa
, and
M. J.
Pellin
,
Phys. Procedia
1
,
379
(
2008
).
13.
J. F.
Ziegler
,
M. D.
Ziegler
, and
J. P.
Biersack
,
Nucl. Instrum. Methods Phys. Res., Sect. B
268
,
1818
(
2010
).
14.
A.
Benninghoven
and
A.
Mueller
,
Phys. Lett. A
40
,
169
(
1972
).
15.
P.
Schauer
and
R.
Autrata
, “
Performance of YAG: Ce single crystal screens for TEM
,” in
Proceedings of the 14th International Congress on Electron Microscopy, 1998
.