With the development of the precise radiotherapy, the accuracy of radiotherapy equipment is gradually improved. The gantry, the carrier of the treatment head and various testing devices, is the most important component that determines the accuracy of the entire equipment. In this paper, the layout of the O-ring linear accelerator and the structure of the gantry are optimized to reduce the weight of the gantry by 50% and the moment of inertia by 60%. A mechanical isocenter detection method based on the laser tracker is proposed to conduct a rapid and accurate isocenter measurement for the optimized gantry. The experimental results show that the maximum deformation of the optimized gantry under load is 0.13 mm during rotation and the maximum intersection distance between the gantry axis and the treatment head axis at the isocenter position is 0.21 mm.
Skip Nav Destination
,
,
,
Article navigation
March 2021
Research Article|
March 22 2021
Optimization design of O-ring linear accelerator gantry and mechanical isocenter detection method Available to Purchase
Chaofeng Shi
;
Chaofeng Shi
a)
1
AVIC Beijing Precision Engineering Institute Aircraft Industry
, Beijing 100076, China
Search for other works by this author on:
Yanjie Jiang;
Yanjie Jiang
2
Beijing SunWise Space Technology Ltd.
, Beijing 100076, China
Search for other works by this author on:
Xi Wang;
Xi Wang
3
Line of Sight Measuring Technologies (Shanghai) Ltd.
, Beijing 100076, China
Search for other works by this author on:
Chaofeng Shi
1,a)
Yanjie Jiang
2
Xi Wang
3
Yifu Chen
1,a)
1
AVIC Beijing Precision Engineering Institute Aircraft Industry
, Beijing 100076, China
2
Beijing SunWise Space Technology Ltd.
, Beijing 100076, China
3
Line of Sight Measuring Technologies (Shanghai) Ltd.
, Beijing 100076, China
Rev. Sci. Instrum. 92, 033308 (2021)
Article history
Received:
February 15 2020
Accepted:
March 06 2021
Citation
Chaofeng Shi, Yanjie Jiang, Xi Wang, Yifu Chen; Optimization design of O-ring linear accelerator gantry and mechanical isocenter detection method. Rev. Sci. Instrum. 1 March 2021; 92 (3): 033308. https://doi.org/10.1063/5.0004907
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
354
Views
Citing articles via
Overview of the early campaign diagnostics for the SPARC tokamak (invited)
M. L. Reinke, I. Abramovic, et al.
Controlled partial gravity platform for milligravity in drop tower experiments
Kolja Joeris, Matthias Keulen, et al.
Refurbishment and commissioning of a dual-band 23/31 GHz tipping
radiometer at potential radio astronomical sites
J. Cuazoson, D. Hiriart, et al.
Related Content
The GLAST LAT Instrument Science Operations Center
AIP Conf. Proc. (July 2007)
Quantification of interfacial spin-charge conversion in hybrid devices with a metal/insulator interface
Appl. Phys. Lett. (October 2020)
An isocenter estimation tool for proton gantry alignment
Rev. Sci. Instrum. (December 2017)
Performance comparison of position and swing angle control system on a gantry crane
AIP Conf. Proc. (May 2023)
Development of a compact X-band linear accelerator system mounted on an O-arm rotating gantry for radiation therapy
Rev. Sci. Instrum. (February 2021)