A design of multichannel gated photomultiplier tube (PMT) is presented for the 960-channel neutron time-of-flight detector at the Institute of Laser Engineering of Osaka University. This is important for the fusion science and the nuclear photonics where intense hard X-rays are generated from the interaction of ultra-short laser pulse of petawatt power density with matter. The hard X-rays often overload PMTs and cause signal-induced background noises called afterpulses, making the detection of subsequent neutrons impossible. For this reason, the PMTs are coupled with an electrical time-gating (ETG) system to avoid overloading. The ETG system disables the PMT by modulating the dynode potential during the primary X-ray flash. An after-pulsing suppression technique is demonstrated by applying a reverse bias voltage between the photocathode and the first dynode. The presented multichannel scheme provides a gate response time of 80 ns, a signal cutoff ratio of 2.5 × 102, and requires reasonably low power consumption.
Skip Nav Destination
Article navigation
October 2018
Research Article|
October 01 2018
A multichannel gated neutron detector with reduced afterpulse for low-yield neutron measurements in intense hard X-ray backgrounds
Y. Abe
;
Y. Abe
a)
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
N. Nakajima;
N. Nakajima
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
Y. Sakaguchi;
Y. Sakaguchi
2
Setsunan University
, Osaka 572-8508, Japan
Search for other works by this author on:
Y. Arikawa;
Y. Arikawa
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
S. R. Mirfayzi;
S. R. Mirfayzi
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
S. Fujioka
;
S. Fujioka
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
T. Taguchi;
T. Taguchi
2
Setsunan University
, Osaka 572-8508, Japan
Search for other works by this author on:
K. Mima
;
K. Mima
3
Graduate School for the Creation of New Photonics Industries
, Shizuoka 431-1202, Japan
Search for other works by this author on:
A. Yogo;
A. Yogo
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
H. Nishimura;
H. Nishimura
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
H. Shiraga;
H. Shiraga
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
M. Nakai
M. Nakai
1
Institute of Laser Engineering, Osaka University
, Osaka 565-0871, Japan
Search for other works by this author on:
a)
Electronic mail: abe-y@ile.osaka-u.ac.jp
Note: Paper published as part of the Proceedings of the 22nd Topical Conference on High-Temperature Plasma Diagnostics, San Diego, California, April 2018.
Rev. Sci. Instrum. 89, 10I114 (2018)
Article history
Received:
May 08 2018
Accepted:
July 11 2018
Citation
Y. Abe, N. Nakajima, Y. Sakaguchi, Y. Arikawa, S. R. Mirfayzi, S. Fujioka, T. Taguchi, K. Mima, A. Yogo, H. Nishimura, H. Shiraga, M. Nakai; A multichannel gated neutron detector with reduced afterpulse for low-yield neutron measurements in intense hard X-ray backgrounds. Rev. Sci. Instrum. 1 October 2018; 89 (10): 10I114. https://doi.org/10.1063/1.5039436
Download citation file:
Sign in
Don't already have an account? Register
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Pay-Per-View Access
$40.00
Citing articles via
Learning from each other: Cross-cutting diagnostic development activities between magnetic and inertial confinement fusion (invited)
M. Gatu Johnson, D. Schlossberg, et al.
The QICK (Quantum Instrumentation Control Kit): Readout and control for qubits and detectors
Leandro Stefanazzi, Kenneth Treptow, et al.
Related Content
Flexible normally on photomultiplier gating strategy for reducing postgate artifacts
Rev. Sci. Instrum. (August 2005)
The large helical device vertical neutron camera operating in the MHz counting rate range
Rev. Sci. Instrum. (November 2018)
Geomagnetic effects on the PMT tubes of the Pierre Auger surface detectors
AIP Conference Proceedings (May 2001)
A modified electron multiplier x-ray detector for synchrotron and laser plasma sources
AIP Conference Proceedings (May 2000)
Measurement of luminescence decays: High performance at low cost
Am. J. Phys. (November 2011)