An accurate calibration method in which an ac-modulated polarizing undulator is used for polarization modulation spectroscopy such as circular dichroism (CD) and linear dichroism (LD) has been proposed and successfully applied to vacuum ultraviolet (vuv) CD and LD spectra measured at beamline BL-5B in the electron storage ring, TERAS, at AIST. This calibration method employs an undulator-modulation spectroscopic method with a multireflection polarimeter, and it uses electronic and optical elements identical to those used for the CD and LD measurements. This method regards the polarimeter as a standard sample for the CD and LD measurements in the vuv region in which a standard sample has not yet been established. The calibration factors for the CD and LD spectra are obtained over a wide range of wavelengths, from 120 to 230 nm, at TERAS BL-5B. The calibrated CD and LD spectra measured at TERAS exhibit good agreement with the standard spectra for wavelengths greater than 170 nm; the mean differences between the standard and calibrated CD and LD spectra are approximately 7% and 4%, respectively. This method enables a remarkable reduction in the experimental time, from approximately 1 h to less than 10 min that is sufficient to observe the storage-ring current dependence of the calibration factors. This method can be applied to the calibration of vuv-CD spectra measured using a conventional photoelastic modulator and for performing an accurate analysis of protein secondary structures.
Skip Nav Destination
,
,
,
Article navigation
August 2008
Research Article|
August 01 2008
Accurate and quick calibration method for polarization-modulation spectroscopy using an ac-modulated polarizing undulator Available to Purchase
Masahito Tanaka;
Masahito Tanaka
a)
1Research Institute of Instrumentation Frontier,
National Institute of Advanced Industrial Science and Technology (AIST)
, Central 2, Umezono 1-1-1, Tsukuba, Ibaraki 305-8568, Japan
Search for other works by this author on:
Kazutoshi Yagi-Watanabe;
Kazutoshi Yagi-Watanabe
1Research Institute of Instrumentation Frontier,
National Institute of Advanced Industrial Science and Technology (AIST)
, Central 2, Umezono 1-1-1, Tsukuba, Ibaraki 305-8568, Japan
Search for other works by this author on:
Fusae Kaneko;
Fusae Kaneko
1Research Institute of Instrumentation Frontier,
National Institute of Advanced Industrial Science and Technology (AIST)
, Central 2, Umezono 1-1-1, Tsukuba, Ibaraki 305-8568, Japan
Search for other works by this author on:
Kazumichi Nakagawa
Kazumichi Nakagawa
2Graduate School of Human Development and Environment,
Kobe University
, Tsurukabuto 3-11, Nada-ku, Kobe 657-8501, Japan
Search for other works by this author on:
Masahito Tanaka
1,a)
Kazutoshi Yagi-Watanabe
1
Fusae Kaneko
1
Kazumichi Nakagawa
2
1Research Institute of Instrumentation Frontier,
National Institute of Advanced Industrial Science and Technology (AIST)
, Central 2, Umezono 1-1-1, Tsukuba, Ibaraki 305-8568, Japan
2Graduate School of Human Development and Environment,
Kobe University
, Tsurukabuto 3-11, Nada-ku, Kobe 657-8501, Japan
a)
Electronic mail: [email protected].
Rev. Sci. Instrum. 79, 083102 (2008)
Article history
Received:
May 15 2008
Accepted:
July 02 2008
Citation
Masahito Tanaka, Kazutoshi Yagi-Watanabe, Fusae Kaneko, Kazumichi Nakagawa; Accurate and quick calibration method for polarization-modulation spectroscopy using an ac-modulated polarizing undulator. Rev. Sci. Instrum. 1 August 2008; 79 (8): 083102. https://doi.org/10.1063/1.2964112
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Citing articles via
Overview of the early campaign diagnostics for the SPARC tokamak (invited)
M. L. Reinke, I. Abramovic, et al.
Controlled partial gravity platform for milligravity in drop tower experiments
Kolja Joeris, Matthias Keulen, et al.
Refurbishment and commissioning of a dual-band 23/31 GHz tipping
radiometer at potential radio astronomical sites
J. Cuazoson, D. Hiriart, et al.
Related Content
Vacuum ultraviolet circular dichroism spectroscopy using an ac-modulated polarizing undulator
Rev. Sci. Instrum. (September 2005)
Recent progress in vacuum-ultraviolet polarization modulation spectroscopy using polarizing undulator at the TERAS BL5 beamline
Rev. Sci. Instrum. (December 2007)
Fabrication of a needle array using a Si gray mask for x-ray lithography
J. Vac. Sci. Technol. B (December 2007)
TELL‐TERAS, NIJI‐I, II, III, IV, and positron facilities
Rev. Sci. Instrum. (January 1992)
Polarizing undulator with crossed and retarded magnetic fields
Rev. Sci. Instrum. (July 1989)