We have developed a Fourier transform two-dimensional infrared (2D IR) spectrometer to probe chemical reactions and biophysical processes triggered by a nanosecond temperature jump ( jump). The technical challenges for such a spectrometer involve (1) synchronization of a nanosecond -jump laser and femtosecond laser system, (2) overcoming the decreased signal-to-noise ratio from low repetition rate data acquisition, and (3) performing an interferometric measurement through a sample with a density and index of refraction that varies with time delay after the jump. The first challenge was overcome by synchronizing the two lasers to a clock derived from the Ti:sapphire oscillator, leading to timing accuracy of for delays up to . The data collection time is reduced by using undersampling with the improved signal-to-noise ratio obtained from a balanced detection scheme with a dual stripe array detector. Transient dispersed vibrational echo and 2D IR spectroscopy are applied to -methylacetamide and ubiquitin, as examples, and the spectral responses by a temperature elevation and by structural changes of the protein are compared. The synchronization of 2D IR spectroscopy with a nanosecond temperature jump without losing its sensitivity at a low repetition rate opens a new applicability of the nonlinear spectroscopy to probe a variety of molecular structure changes induced by a nanosecond perturbation.
REFERENCES
When the switch of the -jump laser is fired, we found that a significant amount of electrical noise is collected through the detector cables. Shielding the cables with grounded aluminum foil removes most of the noise but a subtraction is still needed for a DVE data collection.