We have built a high-speed, tapping mode scanning force microscope using a high frequency quartz crystal resonator. In our design, a cantilever tip was attached to the end of an optical fiber which was glued to a thickness shear mode, AT-cut quartz crystal resonator so as to vibrate in the longitudinal direction. This design allows the microscope to be operated in tapping mode with the flexibility of shear mode operation, which leads to an expected improvement of image quality. Furthermore, combining this geometry with an optical microscope leads to the possibility of commercial applications.
REFERENCES
1.
Y.
Seo
, J. H.
Park
, J. B.
Moon
, and W.
Jhe
, Appl. Phys. Lett.
77
, 4274
(2000
).2.
Y.
Seo
and W.
Jhe
, Rev. Sci. Instrum.
73
, 2057
(2002
).3.
Y.
Seo
, H.
Choe
, and W.
Jhe
, Appl. Phys. Lett.
83
, 1860
(2003
).4.
V. E.
Bottom
, Introduction to Quartz Crystal Unit Design
(Van Nostrand Reinhold
, New York, 1997
).5.
Din Ping
Tsai
and Yuan Ying
Lu
, Appl. Phys. Lett.
73
, 2724
(1998
).6.
The spring constant , where . For details, refer to
K.
Karrai
and R. D.
Grober
, Appl. Phys. Lett.
66
, 1842
(1995
).7.
W. A.
Atia
and C. C.
Davis
, Appl. Phys. Lett.
70
, 405
(1997
).8.
For AT-cut QCR, , where , , according to Ref. 4, p. 99.
9.
W. H. J.
Rensen
, N. F.
van Hulst
, A. G. T.
Ruiter
, and P. E.
West
, Appl. Phys. Lett.
75
, 1640
(1999
).© 2005 American Institute of Physics.
2005
American Institute of Physics
You do not currently have access to this content.