We report the development of a battery-powered portable chemical identification device for field use consisting of an acousto-optic tunable filter (AOTF)-based Raman spectrometer with integrated data processing and analysis software. The various components and custom circuitry are integrated into a self-contained instrument by control software that runs on an embedded single-board computer (SBC), which communicates with the various instrument modules through a 48-line bidirectional TTL bus. The user interacts with the instrument via a touch-sensitive liquid crystal display unit (LCD) that provides soft buttons for user control as well as visual feedback (e.g., spectral plots, stored data, instrument settings, etc.) from the instrument. The control software manages all operational aspects of the instrument with the exception of the power management module that is run by embedded firmware. The SBC-based software includes both automated and manual library searching capabilities, permitting rapid identification of samples in the field. The use of the SBC in tandem with the LCD touchscreen for interfacing and control provides the instrument with a great deal of flexibility as its function can be customized to specific users or tasks via software modifications alone. The instrument, as currently configured, can be operated as a research-grade Raman spectrometer for scientific applications and as a “black-box” chemical identification system for field use. The instrument can acquire 198-point spectra over a spectral range of 238–1620 cm−1, perform a library search, and display the results in less than 14 s. The operating modes of the instrument are demonstrated illustrating the utility and flexibility afforded the system by the SBC–LCD control module.
Skip Nav Destination
Article navigation
June 2004
Research Article|
June 01 2004
Single-board computer based control system for a portable Raman device with integrated chemical identification Available to Purchase
Joel Mobley;
Joel Mobley
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
Brian M. Cullum;
Brian M. Cullum
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
Alan L. Wintenberg;
Alan L. Wintenberg
Monolithic Systems Group, Engineering Science and Technology Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
S. Shane Frank;
S. Shane Frank
Monolithic Systems Group, Engineering Science and Technology Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
Robert A. Maples;
Robert A. Maples
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
David L. Stokes;
David L. Stokes
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
Tuan Vo-Dinh
Tuan Vo-Dinh
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Search for other works by this author on:
Joel Mobley
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Brian M. Cullum
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Alan L. Wintenberg
Monolithic Systems Group, Engineering Science and Technology Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
S. Shane Frank
Monolithic Systems Group, Engineering Science and Technology Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Robert A. Maples
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
David L. Stokes
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Tuan Vo-Dinh
Advanced Biomedical Science and Technology Group, Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6101
Rev. Sci. Instrum. 75, 2016–2023 (2004)
Article history
Received:
June 02 2003
Accepted:
February 16 2004
Citation
Joel Mobley, Brian M. Cullum, Alan L. Wintenberg, S. Shane Frank, Robert A. Maples, David L. Stokes, Tuan Vo-Dinh; Single-board computer based control system for a portable Raman device with integrated chemical identification. Rev. Sci. Instrum. 1 June 2004; 75 (6): 2016–2023. https://doi.org/10.1063/1.1753670
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Citing articles via
An ion-imaging detector for high count rates
Kai Golibrzuch, Florian Nitz, et al.
Overview of the early campaign diagnostics for the SPARC tokamak (invited)
M. L. Reinke, I. Abramovic, et al.
Line-scan imaging for real-time phenotypic screening of C.
elegans
Aaron Au, Maximiliano Giuliani, et al.
Related Content
Development of a compact, handheld Raman instrument with no moving parts for use in field analysis
Rev. Sci. Instrum. (April 2000)
An excitation wavelength–scanning spectral imaging system for preclinical imaging
Rev. Sci. Instrum. (February 2008)
Rapid scanning fluorescence spectroscopy using an acousto‐optic tunable filter
Rev. Sci. Instrum. (November 1987)
Configurable-bandwidth imaging spectrometer based on an acousto-optic tunable filter
Rev. Sci. Instrum. (July 2006)
Versatile single-molecule multi-color excitation and detection fluorescence setup for studying biomolecular dynamics
Rev. Sci. Instrum. (November 2011)