The response of a commercially available liquid scintillator to D(d,n)He3 neutrons and to γ radiation from several standard sources has been measured. It is found that the pulse height response to γ‐ray energy is linear in the energy range 0.5 to 2.6 Mev, and that the fast neutron detection efficiency of a disk scintillator 5 cm thick is about 30% in the neutron energy range 2 to 3.5 Mev, and varies with neutron energy E as
where E is measured in Mev.
1.
G. C.
Neilson
and
D. B.
James
,
Rev. Sci. Instr.
26
,
1018
(
1955
).
2.
Obtained from Nuclear Enterprises Ltd., 1750 Pembina Highway, Winnipeg 9, Manitoba, Canada.
3.
G. T.
Hunter
and
H. T.
Richards
,
Phys. Rev.
76
,
1445
(
1949
).
4.
Segal
,
Swartz
, and
Owen
,
Rev. Sci. Instr.
25
,
140
(
1954
).
5.
Obtained from Chromium Corporation of America, Waterbury, Connecticut.
6.
Skyrme
,
Tunnicliffe
, and
Ward
,
Rev. Sci. Instr.
23
,
204
(
1952
).