We introduce a low-frequency torsional oscillator suitable for studies of quantum fluids and solids. It operates at frequencies of ∼100 Hz, achieves velocities of several cm s−1, and exhibits a quality factor of . In order to reach such velocities at this relatively low frequency, the oscillator amplitude must exceed 100 μm, which would be impracticable for a conventional capacitor-driven device where the drive is applied parallel to the main motion and there are correspondingly large changes in the separation of the capacitor plates. For the different geometry of the oscillator that we now describe, however, the separations of both the drive and detect capacitor plates remain constant regardless of the amplitude of oscillation. We discuss its design and report our initial tests of its performance.
Torsional oscillators have been a valuable tool in studies of quantum fluids and solids since the very beginning. For example, they have been used to detect phase transitions in liquid 4He (Ref. 1) and helium isotopic mixtures,2 in 2D adsorbed helium films and in restricted geometries3,4 and porous media.5–8
The most direct experiment searching for the supersolid phase in bulk solid 4He also used the torsional oscillator technique:9 the onset of superfluidity in the solid helium inside a torsional oscillator would have decreased the overall moment of inertia, and hence decreased the resonant period. Bishop et al. made measurements of solid helium from 25 to 48 bar and concluded that, if there was a supersolid state, then either the supersolid fraction (the fraction of 4He atoms participating in superflow) was less than or the critical velocity (the maximum velocity of superflow without any detectable dissipation) was less than 5 μm s−1.9 Kim and Chan10,11 employed this same technique for the detection of an anomalous response in solid helium. Others soon followed with a variety of designs and experiments. These included compound torsional oscillators to study the frequency dependence of the nonclassical rotational inertia of solid 4He,12,13 a torsional oscillator in conjunction with a square cell14 for similar studies, a cell capable of measuring the torsional oscillation and shear modulus simultaneously using a piezo-transducer, under uniform rotation15 and, more recently, a compound torsional oscillator to show the existence of a 2D spatially modulated superfluid.16 In common to all of these designs was the use of capacitive drive and detect systems. For a conventional torsional oscillator, Seamus Davis's group introduced a magnetic pickup using a superconducting quantum interference device (SQUID),17 making it sensitive to much smaller displacements and velocities.
Our immediate motivation for developing the present oscillator is to study quantum turbulence (QT). It is the most general form of dissipative motion in superfluids, and it differs from its classical counterpart mainly by the inherent quantization of fluid circulation, which manifests itself as the existence of quantized vortices.18,19 These vortices are identical to each other in all respects and save position and core velocity. They not only serve as the building blocks for macroscopic quantum flows, but they also provide the only means of energy dissipation in a pure superfluid that has zero viscosity.20 It is presently understood that Kelvin wave excitations on these vortices transfer energy to progressively shorter wavelengths, until it is radiated as excess thermal quasiparticles in 3He–B or as sound excitations in 4He and Bose–Einstein condensates (BECs). In the case of superfluid 4He, the onset of QT is characterized by the critical velocity needed for self-reconnection of the Kelvin waves on remanent vortices21 attached to walls and surfaces.
At temperatures above K, vortex dynamics in 4He becomes complicated because of a significant density of viscous normal fluid component (quasiparticles). This requires the use of phenomenological, e.g., Hall-Vinen-Bekarevich-Khalatnikov (HVBK)22,23 or geometrical24–26 models to describe the QT and its decay27 when the thermal or mechanical driving force is removed. Our interest, however, centers on the simpler situation that pertains in the low temperature limit where the effect of normal fluid can be ignored. Even then, when only the superfluid component need be considered, there still remains much to be understood, and fascinating new ideas are being introduced on a regular basis, e.g., the recent identification of “fine structure” in the energy levels of model vortex rings in a cylindrical domain.28
Our main concern here is to develop the technology needed to study QT creation through the excitation of Kelvin waves. To generate waves of sufficient amplitude for reconnection to occur requires that we lower the frequency range of operation, compared to most earlier oscillators and, at the same time, that we increase the velocity range. These are seemingly opposed ambitions because a reduction in frequency at a given amplitude must also reduce the velocity. We report here a design that allows us to fulfill our goal but, in order to appreciate its unusual mode of operation, we first summarize the design and function of a traditional cryogenic oscillator.
A common design is illustrated in Figs. 1(a) and 1(b). The oscillator, shown in yellow in (b), is composed of a cylindrical body and a support rod. The body contributes to the total moment of inertia I, and the support rod provides the necessary return spring, with spring constant k. To drive the oscillator and measure its motion, two capacitors are used. Each of them is formed of a moving electrode (shown in blue) attached to the main body, and a fixed electrode (red). Each pair of electrodes is separated by a very small gap on the order of 100 m.
The capacitors are each charged to a substantial DC voltage, so that the plates strongly attract. An additional AC potential is then applied to the driving capacitor, which causes small oscillations of the moving plate and thus of the oscillator body to which it is fixed. The corresponding capacitance changes in the second transducer induce a small AC to its fixed electrode, measured by a phase-sensitive detector referenced to the AC drive. This system works well at small amplitudes but is constrained by the very narrow plate separations needed in the capacitors.
To investigate the onset of QT creation, we wish to study the oscillation of a pill-box-shaped experimental cell as sketched in Fig. 2. At low temperatures, in the absence of vortices, the coin-shaped sample of superfluid within the cell would remain stationary within the oscillating structure. However, it is to be expected that, prior to oscillation, there will exist remanent vortices21 pinned between the top and bottom plates, parallel to the axis of symmetry. The movements of the oscillator will induce Kelvin waves on these vortices and, for sufficient wave amplitude, there will be vortex reconnections, production of QT, and significant dissipation. For typical vortices of length between several hundreds of micrometers up to millimeters, the Kelvin wave resonant frequencies are hundreds of Hz. The critical velocity, vc, is proportional to the square root of the angular frequency ω, and we need to bring vc down into the measurable range. Based on dynamics of quantized vortices, the critical velocity for inducing reconnections can be estimated29 as , where m2 s−1 is the quantum of circulation in 4He. It is estimated that will be sufficient to detect the dissipation corresponding to the onset of reconnections at this critical velocity.
Thus, for frequencies near 100 Hz, critical velocities of order 20 mm s−1 are expected, corresponding to amplitudes near 30 μm at the circumference of the superfluid-filled cell. To search for convincing evidence of QT production, the critical velocity must be exceeded substantially, which is impossible using the conventional oscillator design because the capacitor plates would be liable to touch, given the large oscillation amplitude required. To solve this problem, we use a transducer geometry that moves the electrodes parallel to the motion of the torsional oscillator, so that they cannot touch even at the largest amplitudes and corresponding velocities. The oscillator drive and detection capacitors have partially overlapping electrodes. Electrostatic shear forces are therefore generated in the plane of the capacitor plates, and the capacitance changes with the area of overlap, leaving the small distance between the plates unaffected. We use the same external electrical drive and detection systems as for the conventional oscillator.30 As we demonstrate below, this kind of oscillator can achieve velocities of up to at least 300 mm s−1, giving stable and reproducible vacuum data. Its design and construction are shown in Fig. 2. As in a conventional torsional oscillator, the active elements are parallel plate capacitors charged to a high DC bias voltage, and with an AC voltage superimposed in the case of the driving electrodes. The shear forces due to the incomplete overlap of the capacitor plates tend to increase the overlap and to bring the plates toward alignment.
To make full use of this effect, and to spread the force evenly around the torsional oscillator, we developed the design illustrated in Fig. 2(a). For our particular experiments, a sample of liquid 4He is located in a pill-box-shaped epoxy cell fixed to the center of a rigid copper framework by a BeCu tube, which forms the torsion rod. This tube is attached rigidly to the top and bottom of the framework and has a small hole in its side at the center so that it also acts as the fill line for the cell. This highly symmetrical design helps to ensure that the oscillator does not respond to unwanted nuisance modes at nearby frequencies. Each electrode consists of 12 evenly spaced sectors, electrically connected together, manufactured using conventional lithographic techniques on a thin circular circuit board of radius 2 cm. The radial segments between radii r and R are distributed in a star-like fashion and connected by a small ring. Two finned electrodes of the type described above are permanently fixed to the top and bottom of the cell. Very close to them, two similar finned electrodes are positioned above and below, with deliberate angular misalignments of one half-segment-width, as illustrated in Figs. 2(b) and 2(c). They are attached to the frame and carefully adjusted to create a narrow gap d for each capacitor, one of which is used to drive the oscillator and the other to detect its motion.
Consequently, the often-used “height-times-width-over-drive” (HW/D) parameter then depends only on the total moment of inertia and is unaffected by changes in damping. In contrast to the conventional geometry, the narrow gap between the electrodes remains fixed, and the changes of capacitance arise on account of variations of the overlap angle during oscillation. With 12 electrodes, R = 20 mm, r = 7.5 mm giving L = 12.5 mm and l = 13.75 mm with a gap of m and an overlap of 6°, we expect a capacitance of 19.1 pF.
Before discussing the calibration procedure, there is one other observation that we should mention. In testing the oscillator at room temperature, we found that the copper covered conventional fiberglass printed circuit board (PCB) material known as FR4 circuit board32 forming the substrate of the electrodes is not electrically inert: over a few hours, the material gradually charges toward the voltage applied to the electrode, thereby destroying the designed geometry and potentials, and consequently reducing the signal obtained. It seems plausible that humidity (water content in the porous plastic) could be responsible for the residual conduction and, indeed, both surface and volume conduction are known properties of FR4 boards. At liquid nitrogen temperature (77 K), however, the problem did not arise, or developed so slowly as to be undetectable over the duration of an experiment. Subsequently, therefore, we ensured that the cell remained at zero voltage (grounded) to minimize spurious effects due to polarization of the FR4 for at least a day before cooling, or prior to making quick measurements at room temperature.
Further analyses have been carried out using three separate oscillators, to measure and then compare the velocity calculated from the lock-in amplifier signal with the optically measured velocity. First we used a conventional oscillator as shown in Fig. 1, as a standard for the measurement setup: having previously measured identical results for this particular oscillator from two entirely different setups, we felt confident in using it as a consistency check for our instrumentation. With the conventional oscillator, there is an almost 1:1 correspondence between the optically measured angular velocity and the angular velocity calculated from the lock-in signal, as seen in Fig. 3(a).
In testing the present torsional oscillator, we were careful to minimize any effects from the FR4 conduction (see above). It can be seen from Fig. 3(b) that the slope of the fitted line does not pass through zero; the value of the slope is 0.16. Despite minimizing the measurement time, the conduction and charging of the FR4 must inevitably have introduced errors in measurements made at room temperature, which may be another reason why the slope does not pass through zero. Our final test used an oscillator with all-metal electrodes made of brass of thickness 1.5 mm, without any FR4 at all. As can be seen in (c), the slope is 0.49, which is markedly more than the 0.16 obtained from the FR4 oscillator. The slopes of the curves give values of the detection efficiency factor ξ2. There is still a discrepancy in the actual velocity. We are currently unsure of the reason, but it could be due in part to edge effects: decreasing the capacitor gap of the detection electrode, and therefore increasing the overall capacitance, further increased ξ2, i.e., the discrepancy between the optical and electrical velocities decreased.
During the development of the torsional oscillator, we investigated the effects of different diameter torsion rods, made from both BeCu and Sterling silver. The Sterling silver rods yielded a lower maximum oscillator velocity than we expected, produced characteristics that became non-linear at about 9 mm s−1, and cracked the cell on thermal cycling. We therefore concentrated our designs on cells with BeCu torsion rods. We now present test data from two oscillators with BeCu torsion rods. Each of them was installed in a dilution refrigerator, with the fill line and the copper framework (Fig. 2) thermally anchored to the mixing chamber at ∼15 mK.
The first oscillator had a BeCu torsion rod of 1.2 mm OD and 0.9 mm ID. The capacitor plates were set as close as possible while avoiding touches; the resultant capacitances of the drive and detection electrode pairs were measured as 11.1 and 11.3 pF, respectively. Our measurements yielded an efficiency factor . The device worked well with a resonance frequency 79.105 728 Hz, quality factor , as illustrated in Fig. 4(a).
The second oscillator had a larger diameter (stiffer) BeCu torsion rod of 1.6 mm OD and 1.24 mm ID. The resultant capacitances were 11.93 and 12.04 pF for the drive and detection electrodes, respectively. It had a resonance frequency 123.185 90 Hz, and a quality factor , as illustrated in Fig. 4(b). This oscillator had an efficiency factor .
The mechanical Q's of both oscillators are relatively low compared with the conventional design which can have Q's of between 105 and 107 depending on the material used and any heat treatment for the torsion rod: apart from the lack of heat treatment of our rods, another possibly relevant factor is that the present oscillator is rigidly connected to a polymer experimental cell, and such materials are inherently lossy.
In each case, repeating the frequency sweeps over an interval of several hours showed that the data were highly reproducible, as shown in the corresponding figures. Testing the stability of the 79 Hz oscillator over 20 h showed that the frequency was stable to within 0.03 mHz and that the 123 Hz oscillator was stable within 0.1 mHz over 10 h. There was some vibrational noise coming from fridge pumping lines which has now been reduced by putting pumping lines in sand boxes. We will test the stability of our next redesigned oscillator when installed in the dilution unit. Conventional oscillators can have a frequency stability of 1 μHz down to several nano Hz in high Q oscillators.
At higher velocities, both oscillators exhibited Duffing-style non-linearities with asymmetry in the resonance curves and increases in linewidth that reduced the maximum attainable velocity. This effect was smaller for the second oscillator with the stiffer, larger diameter, BeCu torsion rod. A noticeable warming of the cell was observed at the higher drives with increases of temperature from ∼15 up to ∼110 mK for the larger diameter torsion rod; and up to ∼80 mK for the smaller one.
Amplitude sweeps in vacuum at temperature ∼100 mK for the second oscillator (1.6 mm diameter) yielded the results shown in Fig. 5(a). Here, we have plotted peak rim velocity as a function of torque on a log–log scale. The frequency fell by 7.41 mHz from 123.321 45 Hz for data at ∼100 mK and for the amplitude sweep in vacuum at 4.2 K the frequency fell by 4.18 mHz from 123.317 77 Hz. In each case, the resonance frequency did not fully recover its original value due to increase in the fridge temperature.
We conclude that the ability of our oscillator to attain velocities of at least 300 mm s−1 at frequencies ∼100 Hz means that it is well suited to studies of the generation of QT from remanent quantized vortices in superfluid 4He, where critical velocities of this order or less are to be anticipated. The decrease in sensitivity from the change in geometry, compared to traditional cryogenic oscillators, is easily compensated by the much larger velocities that can be attained. A procedure for implementing a full calibration of the device has been established. Another immediate application could be to study a 2D spatially modulated superfluid, where the critical behavior expected by Choi et al.16 was apparently absent up to velocities of 33 mm s−1: our device would facilitate measurements at much higher velocities and might enable a determination of the critical velocity for this state and help to elucidate its spectrum of excitations.
ACKNOWLEDGMENTS
We gratefully acknowledge help, advice, and valuable discussions with O. Kolosov in relation to the optical measurements that gave us an independent method of measuring the velocity at the rim of our cell. This work was supported by the Engineering and Physical Sciences Research Council, United Kingdom (Grant Nos. EP/P022197/1, EP/P024203/1, and EP/X004597/1), as well as EU H2020 European 7 Microkelvin Platform (Grant Agreement No. 824109). D.S. acknowledges support from the Czech Science Foundation under Project No. 20-13001Y.
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
Author Contributions
Anthony Guénault: Investigation (supporting); Software (lead); Writing – original draft (equal). Kalpana Devi: Investigation (supporting); Writing – review & editing (equal). Peter Vaughan Elsmere McClintock: Funding acquisition (equal); Investigation (supporting); Project administration (lead); Writing – original draft (equal); Writing – review & editing (equal). Malcolm Poole: Funding acquisition (equal); Investigation (lead); Methodology (equal); Writing – original draft (equal); Writing – review & editing (equal). Roch Schanen: Conceptualization (lead); Investigation (equal); Methodology (equal); Writing – original draft (equal); Writing – review & editing (equal). Viktor Tsepelin: Investigation (supporting); Project administration (supporting); Writing – review & editing (equal). Dmitry Zmeev: Funding acquisition (equal); Investigation (supporting); Writing – review & editing (equal). David Schmoranzer: Writing – original draft (equal); Writing – review & editing (equal). Joe Vinen: Conceptualization (equal); Funding acquisition (equal). Deepak Garg: Investigation (supporting); Writing – review & editing (equal).
DATA AVAILABILITY
The data that support the findings of this study will be openly available in the Lancaster University Publications and Research (Pure) system at https://doi.org/10.17635/lancaster/researchdata/594.