Understanding and characterizing very low-energy ( eV) background sources is a must in rare-event searches. Oscura, an experiment aiming to probe electron recoils from sub-GeV dark matter using a 10 kg skipper-CCD detector, has recently fabricated its first two batches of sensors. In this work, we present the characterization of defects/contaminants identified in the buried-channel region of these newly fabricated skipper-CCDs. These defects/contaminants produce deferred charge from trap emission in the images next to particle tracks, which can be spatially resolved due to the sub-electron resolution achieved with these sensors. Using the trap-pumping technique, we measured the energy and cross section associated with these traps in three Oscura prototype sensors from different fabrication batches which underwent different gettering methods during fabrication. Results suggest that the type of defects/contaminants is more closely linked to the fabrication batch rather than to the gettering method used. The exposure-dependent single-electron rate (SER) of one of these sensors was measured 100 m underground, yielding /pix/day at 131 K. The impact of the identified traps on the measured exposure-dependent SER is evaluated via a Monte Carlo simulation. Results suggest that the exposure-dependent SER of Oscura prototype sensors would be lower in lower background environments as expected.
I. INTRODUCTION
Since their invention in 1969, charge-coupled devices (CCDs) have been widely adopted in space and ground based astronomical surveys. They possess appealing characteristics such as a spatial resolution as low as a few m, low readout noise, and a low dark-count rate. Recently, the skipper-CCD,1,2 with enhanced sensitivity to low-energy signals, has become one of the most promising technologies for dark matter (DM) and rare-event searches. In these applications, the discovery potential is highly constrained by the one-electron background rate.3 Currently, dedicated DM skipper-CCD experiments, reporting single-electron rates (SERs) as low as /pix/day, have set the best limits on sub-GeV DM-electron interactions.4–6 One of the objectives of future experimental efforts is to further reduce this rate by up to two orders of magnitude.7
Many background sources of single-electron events (SEEs) in skipper-CCD detectors have been identified and characterized, including temperature fluctuations, radiative processes from external radiation interactions, low-energy photons from the amplifiers and clock-induced charge.8–10 However, we have recently identified another source of SEEs in the newly fabricated skipper-CCDs for Oscura,7,11 a multi-kilogram experiment aiming to probe electron recoils from sub-GeV DM. We associate this source with defects/contaminants within the CCD buried-channel that create single-electron traps with release times comparable to consecutive pixel-readout time, causing a “tail” of deferred single-electron depositions after particle tracks. In some cases, this charge can spread within the image, leading to an apparent increase in the exposure-dependent single-electron rate, which might be mistaken for the sensor’s intrinsic dark current (DC).
In this work, we perform the established trap pumping technique8,12–14 to three different Oscura prototype sensors to characterize their buried-channel single-electron traps. We measure the energy and cross section of the main trap species found in the prototype sensors and verify the effect of deferred charge from trap emission on the measured exposure-dependent SER through a Monte Carlo simulation.
II. CHARGE TRAPPING CHARACTERIZATION IN CCDS
A. Shockley–Read–Hall theory
In p-channel CCDs, charge carriers are holes for which and between 100 and 200 K,16 with being the free electron rest mass.
B. Pocket-pumping technique
The technique of pocket pumping8,12–14 has proved to be a powerful tool to spatially localize and measure the characteristic parameters of charge traps lying within the CCD charge-transfer region. This method consists of filling the traps by “uniformly” illuminating the active area of the CCD and allowing them to emit the trapped charge into their neighbor pixel multiple times. This is done by repeatedly moving the charge back and forth, between pixel phases, creating “dipole” signals relative to the flat background. The method is illustrated in Fig. 1. The sequence of states in this figure is useful to detect traps located below phases and in a three-phase device.
Within the trap pumping sequence, charge capture occurs during the state in which charge remains under the phase with the trap. Assuming a 100% probability of capture, the emission clock starts running just after charge is moved from the phase with the trap, going through the “transient” phase(s). The state in which trap emission takes place corresponds to the one in which charge from the adjacent pixel to the pixel with the trap lies in the adjacent phase to the phase with the trap. The effective time spent in this state can be considered to be a multiple integer of , which is the time spent under the “transient” phase(s). Particularly, for the pumping sequence shown in Fig. 1, the time interval spent in this state is and the probability of emission is given by Eq. (1) evaluated within this time interval. The emission clock resets after each pumping cycle, when charge passes again through the trap.
By fitting as a function of , the emission constant of an individual trap can be extracted. Furthermore, if data are taken at different temperatures, from the fit of , given by Eq. (2), the energy level and cross section of the trap can be obtained.
C. Effects of charge traps in electron-counting CCDs
Typical images from CCDs used for DM and rare-event searches are dark exposures containing tracks of different particles. In a sensor containing traps within the sensor charge-transfer region, depending on the ratio of the traps characteristic emission time and the readout time between two consecutive pixels , trapped charge from these tracks can be emitted: (1) within the pixels of the event, when ; (2) in a highly localized region in the readout direction next to the event, when ; or (3) after several pixels, when . Because of the dependence of with , i.e., Eq. (2), the “tail” of deferred charge from trap emission next to an event is expected to span more pixels at lower temperatures.
Skipper-CCDs provide a unique tool to resolve unequivocally the spatial distribution of depositions coming from emissions of single-electron traps, due to their sub-electron resolution. This is evident in Fig. 2, where dark exposure images at different temperatures from a skipper-CCD with traps within the sensor charge-transfer region are shown. As these images were taken with multiple samples per pixel, achieving sub-electron noise levels, the spatial distribution of the deferred charge next to particle tracks is resolved.
The identification and subsequent masking of pixels with deferred charge from trap emission is trivial when as the deferred charge remains near the main event. However, deferred charge from traps with cannot be easily identified because of the spatial separation of the deferred charge from the original pixel, and taking a conservative masking approach could lead to a significant loss in exposure. To minimize the span of the deferred charge, can be decreased by going to higher temperatures and/or can be increased. With skipper-CCDs, the latter can be done by increasing the number of samples per pixel. However, these approaches lead to a background increase from other temperature and/or exposure-dependent sources, which is not desirable in some cases.
III. POCKET-PUMPING MEASUREMENTS ON OSCURA SKIPPER-CCDS
A. Oscura skipper-CCDs
The newly fabricated skipper-CCDs for Oscura are 1.35 MPix p-channel CCDs with three-phase pixels and a thickness of standard 200-mm silicon wafers (725 m).11 During the Oscura R&D phase, two batches of sensors were fabricated using two different extrinsic gettering techniques.17,18 All wafers from the first batch and one half of the wafers from the second batch underwent a P ion-implantation induced gettering.19 The second half of wafers from the second batch underwent a induced gettering.19 In this work, we characterize single-electron traps from three different Oscura prototype sensors, labeled A, B, and C in Table I, from the two fabricated batches and gettering processes.
B. Data taking
The sensors were packaged at the Silicon Detector Facility in the Fermi National Accelerator Laboratory (FNAL) and installed in an experimental setup at the surface. As shown in Fig. 3, the sensor is placed inside a stainless-steel vacuum chamber and cooled using a cryocooler; the temperature is controlled with a heater located near the sensor. The Low Threshold Acquisition (LTA) electronics board,20 connected to an external computer, is used for data acquisition and CCD control. We use the pocket-pumping technique discussed in Sec. II B to localize and characterize traps in the Oscura prototype skipper-CCDs. First, using a violet LED externally controlled by an Arduino Nano, we illuminate the active area of the sensors, which is loosely covered with a Cu plate to increase uniformity in the illumination profile. The median charge per pixel after illumination lies between 1500 and 2000 .
Then, we perform a pocket-pumping sequence to probe traps below pixel phases and , such as the one illustrated in Fig. 1, including the minimization discussed in Sec. II B. We collected images with , varying from 6.6 s to 1.3 s and from 150 to 190 K. Figure 4 shows a section of the images from the pocket-pumping measurements of prototype sensor A at 150 K, for two different . The right image in this figure reveals a higher density of traps with (ms). We found a uniform spatial distribution of traps through the whole active area of the sensors.
C. Analysis and results
With the most efficient dipole-detection algorithm discussed in the Appendix, we identify dipoles and track their position in each of the images. Using sets of images from the same sensor acquired at a fixed temperature, we compute as a function of for each dipole found. We fit this curve with the function given by Eq. (5) and extract the trap emission-time constant associated with that dipole. As a quality selection criteria to the dipole intensities, we require a coefficient of determination greater than 0.7 and a relative error on below 50%. With this criteria, we reject between 2% and 20% of dipoles, depending on the dataset. From now on, we will refer as “detected traps” to those probed below pixel phases and that were found with the trap-detection algorithm, were not rejected by the selection criteria, and are not overlapped dipoles. Figure 5(left) shows the intensity as a function of of a detected trap fitted by Eq. (5). For each set of images from the same sensor at a given temperature, we build a trap map with the position and the emission-time constant of each detected trap. One of these maps, from a 50 pix 50 pix region of the active area of Oscura prototype sensor A at 150 K, is shown in Fig. 5(right).
The histograms in Fig. 6(top) show the distributions at 190 K of the detected traps for the Oscura prototype sensors A and B, which are from different fabrication batches but underwent the same gettering process (ion implantation). The histograms in Fig. 6(bottom) show the distributions at different temperatures of the detected traps for the Oscura prototype sensor C, with the gettering.
In all the distributions in Fig. 6, a primary peak can be seen, which is associated with the largest population of traps within the sensors’ buried-channel region. Also, in Fig. 6(bottom) the peaks in the distributions move toward higher values of at lower temperatures, which is expected from the dependence of with , i.e., Eq. (2). Comparing the distributions from prototype sensors A and B in Fig. 6(top), both with the ion-implantation gettering, we see a larger population of traps with s for K in the distributions from prototype sensor A, forming a secondary peak. We associate the presence of this peak with the fabrication batch as none of the distributions from sensors from the second batch, i.e., B and C, show a significant trap population at those .
For each detected trap, we plot as a function of and fit it with the function in Eq. (2). We perform a chi-squared test on the fits and rejected those with a p-value below 0.05. From each of those fits, we extract the energy and cross section associated with each trap, shown as dots in the scatterplot in Fig. 7(top). The distributions of these variables of the detected traps in each of the Oscura prototype sensors are shown in Fig. 8. The maximum value of each of these histograms and its associated error, computed as the full width at half maximum, is shown in Table II [Hist. max.].
Prototype . | Energy (eV) . | Cross section (10−15 cm2) . | ||
---|---|---|---|---|
Hist. max. . | fit . | Hist. max. . | fit . | |
A | 0.340 ± 0.006 | 0.341 ± 0.003 | 3.06 ± 1.23 | 3.49 ± 0.92 |
B | 0.302 ± 0.008 | 0.305 ± 0.006 | 0.66 ± 0.29 | 0.85 ± 0.38 |
C | 0.310 ± 0.009 | 0.313 ± 0.007 | 0.87 ± 0.46 | 0.98 ± 0.51 |
Prototype . | Energy (eV) . | Cross section (10−15 cm2) . | ||
---|---|---|---|---|
Hist. max. . | fit . | Hist. max. . | fit . | |
A | 0.340 ± 0.006 | 0.341 ± 0.003 | 3.06 ± 1.23 | 3.49 ± 0.92 |
B | 0.302 ± 0.008 | 0.305 ± 0.006 | 0.66 ± 0.29 | 0.85 ± 0.38 |
C | 0.310 ± 0.009 | 0.313 ± 0.007 | 0.87 ± 0.46 | 0.98 ± 0.51 |
Moreover, from the distributions at different temperatures associated with each sensor, i.e., histograms in Fig. 6, we plot the emission-time constants associated with the primary peaks against , with an error given by its full width at half maximum, as shown in Fig. 7 (bottom). We fit the data points with the function in Eq. (2) and extract from it the energy and cross section associated with the largest population of traps. The value of these variables and its associated error are shown in Table II [ fit] and plotted as stars in Fig. 7(top).
As can be seen from Table II, the values of the trap parameters extracted from the primary peaks of the distributions in Fig. 8 [Hist. max.] and from the fits in Fig. 7 (bottom) [ ] are mutually consistent within errors. Furthermore, the parameters from prototype sensors B and C, both from the second fabrication batch, are also mutually consistent. This suggests that the kind of defects/contaminants inducing charge traps is related to the fabrication batch. It is worth noting that while the relative errors for energies are small, below 3%, those for cross sections are significantly higher, ranging from 26% in the best case to 53% in the worst case. This uncertainty is reflected in the spread of points in the cross section axis of Fig. 7(top) which could be associated with the change in the capture probability depending on the location of the trap. For example, a trap located near the center of the charge packet could have a different capture probability than one that is near the edge, even if they belong to the same trap population,13 as local dynamics within the silicon lattice might differ.
The trap energies and cross sections reported in Table II are similar to those reported for hole traps associated with transition metals in p-type silicon,21 which are common materials used in semiconductor processing, for example: palladium (Pd), with eV and , molybdenum (Mo), with eV and , platinum (Pt), with eV and , and silver (Ag), with eV and . Although gettering techniques are implemented during the fabrication process to capture impurities, the use of the same equipment for productions involving transition metals could lead to unwanted metal contamination in the sensors.
IV. EFFECT OF 1e− TRAPS ON DC MEASUREMENTS IN SKIPPER-CCDS
Dark current (DC) is an irreducible exposure-dependent background for skipper-CCD detectors that originates from the thermal excitation of electrons from the valence band to the conduction band. As it constrains the lowest SER that can be achieved, estimating its value is important in applications where the science reach is limited by the one-electron background rate.
Using the energy and cross section associated with the largest population of traps found from the pocket-pumping measurements (Table II), we computed the contribution to DC [Eq. (6)] from the single-electron traps obtaining /pix/day for 130 K (150 K); these numbers are several orders of magnitude below the expected DC, see discussion in Sec. IV A. Here, we assumed with being the average number of traps in the buried-channel region of one sensor identified with the detection algorithm in the pocket-pumping measurements before applying the selection criteria, and the effective sensor’s volume that was probed with the pocket-pumping technique; the factor 2.15 accounts for the traps in the second phase that were not probed and for a conservative 30% dipole-detection inefficiency.
A. DC measurements at surface and underground
A typical way to quantify dark current in skipper-CCDs is to acquire dark images with different exposure times, mask events within the images associated with any other source of background, compute the SER as a function of exposure time, and extract the slope, i.e., the dark single-electron rate, which represents an upper limit on the sensor’s DC; see discussion in Ref. 9. Performing these measurements underground allows us to minimize SEEs generated from external radiation interactions, which constitute a dominant background at the surface.10 Experiments using skipper-CCDs as particle detectors on the surface have reported SERs above /pix/day,23–25 while underground experiments have reached lower SER levels.4–6,26 In fact, the lowest single-electron rate ever achieved to date in a skipper-CCD is /pix/day,4 reported by the SENSEI Collaboration from measurements in their setup at the MINOS cavern at FNAL, 100 m underground. Next-generation skipper-CCD DM experiments aim to further reduce this rate. Oscura has a target SER of /pix/day.7 Although Oscura could already constrain theoretically motivated light DM models with SERs on the order of /pix/day due to its large mass, achieving a lower SER would further increase its sensitivity at low DM masses.11
In Refs. 11 and 27, we presented DC measurements performed in a dedicated setup with 2 in. of lead shield at the surface with a Oscura prototype sensor from the same wafer as prototype A; these correspond to the circles in black at 140, 150, and 160 K in Fig. 10. The same setup was moved 100 m underground, to the MINOS cavern at FNAL; see Fig. 9. In that setup, we performed DC measurements with a Oscura prototype sensor from the same wafer as prototype C, following the previously discussed method. We acquired images varying the exposure time from 0 to 150 min, with 324 samples/pix. The exposure-dependent single-electron rates were computed from images acquired at 131, 138, and 148 K; these are shown in Fig. 10 as blue circles. The lowest value achieved was /pix/day at 131 K.
In the images taken underground with the Oscura prototype sensors, the SEEs originating from deferred charge from trap emission constitute a significant background for the DC measurements. To mitigate their impact, we implemented a “bleeding zone” mask for pixels upstream in the horizontal and vertical directions of any event with more than , similar to what is done in skipper-CCD experiments searching for DM to discard events from charge-transfer inefficiencies;4,6 see Fig. 11.
To minimize the masked area of the images, we found the minimum bleeding-mask lengths in which the “tails” of deferred charge from trap emission did not impact the measured exposure-dependent SER. We measured this rate varying the horizontal (vertical) bleeding-mask length with a fixed vertical (horizontal) bleeding mask of 200 (1250) pixels, see Fig. 12.
The optimal mask length was determined as the minimum value after which the measured exposure-dependent SER becomes constant, being 1250 (250) pixels in the horizontal (vertical) direction.
B. Monte Carlo simulations of deferred charge from trap emission
We performed a Monte Carlo simulation to estimate the impact of deferred charge from single-electron trap emission on the measured exposure-dependent SER for the Oscura prototype sensors. For the simulation, we assumed that traps in the horizontal register have similar density, energy and cross section distributions as those in the vertical registers measured with the pocket-pumping technique, that the spatial distribution of traps is uniform and that sensors from the same fabrication batch have the same density and kind of traps. We generated trap maps at different temperatures with positions directly taken from the pocket-pumping measurements and emission-time constants computed using Eq. (2), considering the trap parameters ( , ) obtained from the measurements. Figure 13 shows a region of the detected trap maps from prototype sensor A measurements corresponding to two different temperatures.
The simulation is based on two sets of images taken with the Oscura prototype sensors: (1) underground, at 131 K, with exposure times between 0 and 150 min, and (2) at surface, at 150 K, with exposure times from 0 to 15 min. We simulated an “underground” (“at surface”) set of images, with each image containing the events with energy of the acquired image, a uniformly distributed exposure-independent SER of /pix and charge from a exposure-dependent SER of /pix/day, consistent with the exposure time of the acquired image.
Using these sets and the trap maps of prototype sensor C, we simulated two new sets accounting for the effects due to traps. For each event, we simulate the shifts of its constituting charge packets toward the readout amplifier. If the packet encounters a trap, a charge carrier is captured and released at a later time with a probability given by Eq. (1). For simplicity, we assume the same capture probability for all traps and estimated the carrier density in its vicinity as in Ref. 28. In the simulation, carriers released from trap emission can be recaptured by subsequent traps and re-emitted at a later time. This causes a larger spread of carriers from trap emission within the image, which is more evident in the horizontal direction. In Fig. 14 we show a dark exposure image from one of the data sets and its corresponding image generated with the simulation.
We extracted the exposure-dependent SER on the simulated sets of images following the recipe outlined in Sec. IV A, using the optimal mask length. The extracted exposure-dependent SER in the simulated images without the effects of traps matches the simulated exposure-dependent SER of /pix/day for the “underground” (“at surface”) set. However, in the simulated images accounting for deferred charge from trap emission, we extracted a exposure-dependent SER of /pix/day for the “at surface” simulated set and /pix/day for the “underground” simulated set. Both of these values are one order of magnitude larger than the simulated exposure-dependent SER.
These results show that, in sensors with traps that have emission times comparable to the readout time of consecutive pixels, depositions from trap emission can occur beyond the masked area, even with a conservative masking approach. Additionally, multi-electron events enhance trap capture. Overall, these factors can significantly impact the measured exposure-dependent SER. In fact, in the DC measurements at surface with Oscura prototype sensors, as presented in Refs. 11 and 27, the impact was minimized by increasing the image readout rate and selecting regions free of multi-electron events.
V. CONCLUSIONS
We identified single-electron traps in the newly fabricated skipper-CCDs for Oscura. These traps have emission-time constants similar to the typical readout time of consecutive pixels, producing a “tail” of deferred charge observed in the images next to particle tracks. These “tails” consist mainly of single-electron depositions and can only be spatially resolved due to the sub-electron noise that can be reached with skipper-CCDs. Otherwise, deferred charge would only manifest as an increase in overall charge-transfer inefficiency and dark counts. In this sense, skipper-CCDs continue to provide insights into the understanding of dark-count sources.
We studied the buried-channel single-electron traps in three Oscura prototype sensors from two different fabrication batches and two different gettering methods, and ion implantation. The pocket-pumping technique was used to measure the position and emission-time constants of defects/contaminants associated with these traps at different temperatures. The trap characteristic parameters’ cross section and energy level were measured by fitting the temperature dependence of the emission times associated with each individual trap and to the primary peak of the distributions. Results from both analyses are consistent. The energy and cross section associated with the largest population of traps in each sensor are shown in Table II. These parameters are consistent within sensors from the same fabrication batch. Moreover, a secondary peak associated with a trap population with s for K is only observed in the sensor from the first fabrication batch. These results suggest that the type of defects/contaminants is more closely related to the fabrication process than to the implemented gettering.
The exposure-dependent SER was measured for a Oscura prototype sensor at the MINOS cavern at FNAL, yielding /pix/day at 131 K. A procedure for finding the optimal bleeding-mask length to minimize the effect of charge traps encountered within the sensors was described. To estimate the impact of deferred charge from trap emission on exposure-dependent SER measurements, a Monte Carlo simulation of the trap capture and emission processes was implemented by using the trap parameters found from the pocket-pumping measurements. Results show that, even with a conservative masking approach, deferred charge from these traps can occur beyond the masked area and contribute to the measured exposure-dependent SER. More importantly, it provides an explanation to the rate measured underground of the Oscura prototype sensor. These results also suggest that the exposure-dependent SER of these sensors might be lower in lower background environments.
ACKNOWLEDGMENTS
This work was supported by the resources of the Fermi National Accelerator Laboratory (FNAL), managed by Fermi Research Alliance, LLC (FRA), and the Lawrence Berkeley National Laboratory (LBNL), acting under Contract Nos. DE-AC02-07CH11359 and DE-AC02-05CH11231 with the U.S. Department of Energy, respectively.
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
Author Contributions
Santiago Ezequiel Perez Garcia: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Methodology (equal); Writing – original draft (equal); Writing – review & editing (equal). Brenda Aurea Cervantes Vergara: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Methodology (equal); Writing – original draft (equal); Writing – review & editing (equal). Juan Cruz Estrada: Funding acquisition (lead); Project administration (lead); Supervision (lead). Stephen Holland: Conceptualization (supporting); Supervision (supporting); Writing – review & editing (equal). Dario Pablo Rodrigues Ferreira Maltez: Supervision (supporting); Writing – review & editing (equal). Javier Tiffenberg: Supervision (supporting); Writing – review & editing (equal).
DATA AVAILABILITY
The data that support the findings of this study are available from the corresponding authors upon reasonable request.
APPENDIX: DIPOLE-DETECTION ALGORITHMS
Algorithms designed to detect dipole signals against a flat background typically flag pixels with intensities that exceed or fall below a certain threshold established by the flat field signal. However, detecting dipoles becomes challenging when the dipole density increases or if the background is not flat. In this work, two different algorithms (A and B) were tested. Algorithm A subtracts the median of each row and computes the “local” standard deviation within a small window of pixels. The pixel intensity threshold is defined as a multiple of the local standard deviation. This code flags consecutive pixels if their absolute intensity is above the threshold and if one is positive and the other is negative. Algorithm B subtracts the median of each row and each column. It then asks for two consecutive pixels to be one positive and one negative, computes the dipole amplitude, scales it by a factor between 0 and 1 accounting for symmetry, and asks for the scaled amplitude to be above a certain threshold.
To select detection threshold values that yield the best dipole identification in an image with a high density of traps, a Monte Carlo simulation was made generating images with known numbers and positions of dipoles. By comparing the dipoles identified by the algorithms with the simulated ones, we computed the Precision and Recall detection metrics for each code and selected the detection threshold that maximizes their performance. Figure 15 shows one of the images with simulated dipoles (left) and the Precision-Recall curve for each algorithm when varying the detection threshold (right). The algorithm B, which accounts for the dipole symmetry, was found to have the better performance.