We have developed and operated an electronic system for the non-destructive detection and cooling of charged-particle bunches that are captured and confined in a Penning trap, together with methods for the evaluation of corresponding measurements that allow for a detailed characterization of each individual particle bunch. Once calibrated, from a single measurement of the particles’ induced electronic signal as a function of time directly upon capture, the setup and method allow for a fast determination of the initial and final absolute particle energies, of the cooling rate, and of the absolute number of particles in the bunch. We demonstrate this with highly charged ions ( ) that are injected into the Penning trap of the HILITE experiment.
I. INTRODUCTION
The method of resistive cooling of charged particles in a Penning trap is well-established and well-understood particularly for single particles1–3 but also for ensembles of particles.4,5 It is based on image currents that the particle motions induce in the conducting trap electrodes.6–8 When these currents are damped by resistive elements, the motion of the particles is damped in turn.
Most often, resistive cooling is applied to the axial particle oscillation in the trap by means of a tuned resonant circuit that is either placed across a pair of electrodes or that connects one electrode to the ground. The real part of the circuit’s impedance dissipates energy from the axial motion into the cold bath in which the circuit is embedded.9–11 This reduces the axial particle energy and, hence, the amplitude of the oscillation, until the temperature associated with that energy comes into equilibrium with the cold bath, often at liquid-helium temperature.11
The benefit of such cooling is a significant reduction in the particles’ axial velocity and spread, which facilitates well-defined confinement close to the trap center and the conditions required for a number of spectroscopic precision measurements across a wide range of frequency domains.9
For resistive cooling of large particle ensembles, it has been shown that only the center-of-mass motion is directly cooled by the interaction with the circuit, while all other motions are indirectly cooled only to the extent of their energy transfer into the center-of-mass motion.9 For the axial motion, one expects and observes a fast cooling of the center-of-mass motion with an exponential cooling rate that is proportional to the number of particles, and a de-phasing of the axial motions due to residual imperfections of the initial ensemble and the trapping fields.10 In previous measurements,10 more than 99.99% of the initial center-of-mass axial particle energy was cooled away resistively within about 50 ms. We have also previously shown that a comparatively simple model for the energy flow within the particle–circuit system can account for the observed effects during resistive cooling.10
One can create a particularly clear-cut situation when particle capture12 into the trap is done along the central trap axis (thus avoiding significant initial magnetron motion) and when the particles form a well-defined bunch that occupies a small volume in phase space, i.e., has both a small spatial and velocity spread. This translates into a narrow distribution of the axial frequencies and phases of motion of the stored particles. Then, the initial axial kinetic energy is concentrated in the axial center-of-mass motion of the incoming particle bunch. This axial center-of-mass energy can be cooled away efficiently before the bunch has a chance to de-phase into uncorrelated motions that are far less efficiently cooled.10
When the cooling circuit is chosen to connect a pair of electrodes with mirror symmetry with respect to the trap center, rather than connecting one electrode to the ground, it becomes insensitive to electronic noise due to fast switching and pulsing of electrodes as required for the dynamic capture of externally produced particles.12
In the following, we show how this can be used to determine the initial axial particle energy directly upon trapping, the final axial particle energy, the rate of cooling, and the number of particles within the bunch. Apart from a characterization of the cooling process, this may be used, e.g., for an optimization of the injection and capture process and to determine the number of stored ions which is of high interest, i.e., for the measurement of absolute cross sections in laser-ionization experiments.13
The measurements have been performed with ions from an external source that are injected into the Penning trap of the HILITE experiment.13 They complement earlier demonstrations of non-destructive characterization of particle bunches in a beamline,14,15 in the sense that the present methods find application directly inside the trap and at much lower kinetic energies and can, hence, be used to infer the properties of the confined particles directly before a subsequent measurement.
II. ION MOTION AND COOLING IN A PENNING TRAP
We consider a cylindrical Penning trap16,17 located in the center of a homogeneous magnetic field that is produced by a superconducting magnet. The central trap axis and the magnetic field axis are identical and defined to be the -axis (see also Fig. 1). Voltages applied to the trap electrodes form a quadrupolar electrostatic potential with the well along . Any individual ion in this field configuration performs a motion that is comprised of an axial oscillation about the trap center at and a radial motion that is a cyclotron orbit around the magnetic field, perturbed by a drift motion in the crossed electric and magnetic fields (magnetron motion).9
Since we are interested in the behavior of many identical particles that initially form a bunch, we adopt the center-of-mass picture, i.e., we describe the bunch by its axial center-of-mass motion and by the axial individual-ion motions relative to it. For the present conditions, a significant amount of energy is initially concentrated in the center-of-mass motion,10 i.e., before and directly upon capture, the ions form a well-defined bunch that occupies a small phase space. During confinement, the ions move with a fixed phase initially and begin to de-phase due to their initial axial energy spread and due to residual imperfections of the confining fields, which make the axial oscillation frequencies slightly dependent on the kinetic energies of the individual ions.2,5 This leads to a re-shuffling of axial energy between the center-of-mass and individual motions at a rate that is given by the width of the axial frequency distribution within the ion bunch at any time.1
It can be shown3,9 that the resistive cooling of the total energy of an ensemble of identical ions with random phases of motion occurs at the same rate as the single-ion cooling given by Eq. (3), irrespective of the number of ions .
The actual rate of cooling that is observed in an experiment depends on the initial conditions and their evolution during cooling, in particular on the distribution of axial oscillation frequencies and of the phases of axial motion. This means that in general, we expect the observed rate of cooling to change over time. We will discuss this in detail in the framework of the cooling model presented below.
III. SETUP AND PROCEDURES
The current measurements have been performed with the setup of the HILITE experiment13 located at the Helmholtz Institute Jena, Germany.
A. Overview
The setup and methods have been described in detail previously.13–15 In short, bunches of highly charged ions are produced in a commercial electron-beam ion trap (EBIT)21 and selected (with respect to their charge-to-mass ratio) by a Wien filter.22 The pure ion bunches (presently ) are then guided by a low-energy electrostatic beamline to the Penning trap of the experiment. This is located in the center of a horizontal-bore superconducting magnet that can produce static fields up to T. A system of position-sensitive non-destructive ion detectors15 and ion optics23 outside of the magnet is used to inject the ion bunches along the central trap axis, thus avoiding significant magnetron motion upon capture. Typically, 95% of the ions are injected and confined within about from the central axis.10 A pulsed drift tube24 is used to decelerate the ion bunches from an initial energy of around 2 keV to roughly 150 eV per charge and ion. The final deceleration is achieved by a floating ground potential of the whole trap25 such that each ion enters with an axial kinetic energy of about 20–30 eV per charge. This value has been established by lowering the trap potential and observation of the ion loss. The ions are dynamically captured into the trap by fast switching of the trap potential on the entrance side.12 Then, the individual trap voltages are chosen such that the trap forms a harmonic axial potential directly upon ion capture. The measurements to be discussed commence immediately upon dynamic capture.
B. Trap and electronics
The trap is a cylindrical Penning trap16,17 that consists of a central ring electrode, two pairs of correction electrodes, one pair of endcaps, and additionally one pair of electrodes for dynamic capture.12 It has the common axial symmetry and mirror symmetry with respect to the trap center, which is located in the center of the ring electrode. The endcap electrodes have a separation of 22.4 mm, and the inner diameter of the trap is 15 mm. The outer electrodes have a central hole with a diameter of 4 mm to allow for loading the trap with the externally produced ions (entry side, left) and for the axial ejection of the trap content onto an external detector with imaging capabilities (exit side, right).
The electronic circuit for detection and cooling is a normal-conducting resonant LC circuit made of high-purity copper wire wound around a toroidal polytetrafluorethylene (PTFE) core. It connects the two inner correction electrodes of the trap, rather than connecting one electrode to ground. The details and benefits of this design have been described previously.17 The LC circuit has a resonance frequency of and a quality factor of .
C. Detection: Induced current
D. Resonator impedance and signal
The quality factor of is well-suited for the present center-of-mass cooling, since the rate is still high (of the order of 100 ) and the large bandwidth of the resonator of allows the simultaneous cooling of the entire ion ensemble even in the presence of spurious energy-dependent frequency shifts during cooling.
The principal advantage of the double-electrode configuration (aside from its advantageous off-resonance behavior17) is the cancelation of any external electrical noise or disruption, which equally affects both detection electrodes and, thus, does not contribute to the total induced current. This allows the measurements to commence immediately upon capture, since the unavoidable pulse from the fast switch of the trap electrode for dynamic capture does not influence the measurement.
IV. MODEL OF RESISTIVE COOLING
A. Quantities and rate equation
B. Assumptions and validity
This equation assumes that energy from the radial motions does not contribute to the process. This is justified since for the present conditions, the axial-radial coupling due to long-range Coulomb interaction is ineffective and its time scale has been estimated to be .10 Hence, in the current experiment, this energy exchange is not observed and can be neglected.
The overall model assumes the energies and frequencies of the center-of-mass and of individual-ion motions relative to it to be well-defined quantities that allow for a statistical treatment. This is true for sufficiently large ensembles of particles ( ), as has been shown previously.1,9 For the present ion numbers in the range of several tens of thousands, we assume to fulfill this condition. We also assume that the rate of c.m.-cooling is, indeed, given by , which is justified theoretically as discussed above, and in addition has been experimentally shown over a wide range of ion numbers.10
For the application of the presented cooling model to measured curves of induced power as a function of time, we further assume that the cooling circuit’s impedance is essentially constant over the center-of-mass frequency range, which is justified at the present resonator bandwidth. As will be shown below, the center-of-mass frequency range in each individual measurement is even smaller than the bandwidth Hz of the spectrum analyzer used for the signal power measurement, which itself is much smaller than the bandwidth of the resonator of .
C. Initial cooling and quasi-equilibrium
D. Fit quantities
V. MEASUREMENTS AND RESULTS
Bunches of Ne8+ were dynamically captured in the Penning trap as described. In order to track the behavior during the resistive cooling process, snapshot spectra of the ensemble were taken for different delays after the initial capture. A selection of these spectra is shown in Fig. 5 to illustrate the general effect of cooling. They show a decay of by several orders of magnitude within less than 50 ms and a relative shift of the central frequency by about half a percent. The former results from the rapid cooling of , and the latter follows from the residual energy-dependence of the axial frequency in a non-ideal Penning trap. A Gaussian peak model was applied to each of these spectra to determine the peak power and the central frequency . This is appropriate as the peak shape is determined by the bandwidth of the spectrum analyzer of Hz. The fact that the width of each of these peaks is determined by the bandwidth of the spectrum analyzer illustrates that a frequency window of 510 Hz is enough to cover the frequency distribution of the center-of-mass mode in each individual measurement. The shifts of the central frequency result from the resistive cooling of the center-of-mass energy and the resulting motion frequency shifts in a real Penning trap.
The measured time evolution of the peak power is shown in Fig. 6. For each time, the peak power was determined by a Gaussian fit to a measured spectrum. The resulting curve of measured power as a function of time shows three main features:
Initially, in the present case within the first 20 ms, the slope of the curve increases until it becomes essentially constant. This is due to the initially large amplitudes and, thus, smaller value of the pick-up and slower cooling [see Eqs. (8) and (9)].
After that, the curve shows an exponential decay with an essentially constant rate of currently 103(3) , which is in agreement with the expected resistive center-of-mass cooling rate from Eq. (11) for ions. This ion number is in agreement with the fit result and an independent measurement, as will be discussed below.
Finally, at present after about 100 ms, the signal reaches the expected quasi-equilibrium where c.m.-cooling occurs at a much smaller rate limited by axial re-shuffling. The model for (without a quasi-equilibrium state, green curve) fails to describe the data.
Overall, the fit according to the model as expressed by Eqs. (16) to (18) describes the data well (see the orange curve in Fig. 6). In particular, for a value of the model describes the initial decay but fails to reach equilibrium (green curve in Fig. 6), supporting the assumption that is only responsible for the equilibrium level reached. The parameters , , and and the initial center-of-mass energy have been fitted using lmfit26 and the result is shown in Table I. As we have used the initial center-of-mass energy as a fitting parameter, it was necessary to solve the equations of motion using a fourth-order Runge–Kutta algorithm.27
Parameter . | Value . |
---|---|
) | 17.2(4) |
A (fA m2 kΩ−2) | 1170(90) |
B (mm2 s−1 k Ω−1) | 1030(20) |
C (mV s−1) | 81(6) |
Parameter . | Value . |
---|---|
) | 17.2(4) |
A (fA m2 kΩ−2) | 1170(90) |
B (mm2 s−1 k Ω−1) | 1030(20) |
C (mV s−1) | 81(6) |
We can now use the obtained values for the fit parameters to calculate the quantities that characterize the ion bunch and its cooling. Parameter determines the absolute scale of the measurement by establishing a relation between absolute ion energy and measured power [yielding in Eq. (18)], determines the center-of-mass cooling rate, determines the quasi-equilibrium energy, and is the initial center-of-mass energy.
A. Initial ion energy
The value of the initial axial ion energy directly upon capture into the trap is a direct result of the fit, so no further calculation is required. In the present example, its value is V per charge and ion.
B. Absolute ion number
C. Center-of-mass cooling rate
D. Equilibrium energy
Of the initial total energy per charge and ion, given by the sum of c.m. energy and uncorrelated energy, the center-of-mass component is rapidly cooled to effectively zero, leaving the axial energy in the individual (uncorrelated) motions given by Eq. (23). Hence, the amount of energy that can be cooled away efficiently is determined by the initial relation between and , which can be determined by this method. This is a pre-requisite for an optimization of the ion production, deceleration, and capture process to the end of achieving a high initial degree of energy in the axial c.m. motion ( ), which can be cooled efficiently.
VI. SUMMARY AND CONCLUSION
We have shown that it is possible to determine important properties of captured ion bunches and their resistive cooling behavior from a measured cooling curve of detected power as a function time directly upon capture in a Penning trap. Such a measurement allows for an absolute determination of the initial energy and number of ions in a captured bunch, as well as of the rate of cooling of its center-of-mass motion and of the final ion energy. One pre-requisite for this is the ability to make such measurements immediately after dynamic capture, which is possible due to a dual-electrode resonator design.
The recorded cooling curves have been analyzed with a simple model that allows the direct determination of the mentioned quantities in the given situation. The resulting ion number was independently confirmed by a non-destructive detector, which validates the model of the cooling process and also shows the correct calibration of the dual-electrode resonator.
Once the absolute scale of the measured signal, as expressed by the parameter , has been calibrated by the presented method, the characteristic quantities of the bunches can be determined on a shot-by-shot basis. This enables the optimization of the ion-capture process towards low initial energy in the uncorrelated motion, while the center-of-mass energy can be cooled within a few tens of milliseconds. The present detector and methods may quite readily be implemented in a range of Penning trap experiments that work with externally produced particle bunches, as it affords no large deviation from the common schemes for non-destructive detection and cooling.
AUTHOR DECLARATIONS
Conflict of Interest
The authors have no conflicts to disclose.
Author Contributions
Markus Kiffer: Formal analysis (lead); Investigation (equal); Visualization (lead); Writing – original draft (lead); Writing – review & editing (equal). Stefan Ringleb: Conceptualization (equal); Formal analysis (equal); Investigation (equal); Visualization (supporting); Writing – original draft (supporting); Writing – review & editing (equal). Thomas Stöhlker: Conceptualization (supporting); Funding acquisition (lead); Project administration (lead); Supervision (lead); Writing – review & editing (equal). Manuel Vogel: Conceptualization (equal); Project administration (supporting); Supervision (supporting); Visualization (supporting); Writing – original draft (supporting); Writing – review & editing (equal).