We report an in-depth study of the long-term reproducibility and reliability of droplet dispensing in digital microfluidic devices (DMF). This involved dispensing droplets from a reservoir, measuring the volume of both the droplet and the reservoir droplet and then returning the daughter droplet to the original reservoir. The repetition of this process over the course of several hundred iterations offers, for the first time, a long-term view of droplet dispensing in DMF devices. Results indicate that the ratio between the spacer thickness and the electrode size influences the reliability of droplet dispensing. In addition, when the separation between the plates is large, the volume of the reservoir greatly affects the reproducibility in the volume of the dispensed droplets, creating “reliability regimes.” We conclude that droplet dispensing exhibits superior reliability as inter-plate device spacing is decreased, and the daughter droplet volume is most consistent when the reservoir volume matches that of the reservoir electrode.
Skip Nav Destination
Article navigation
June 2012
Research Article|
April 06 2012
Droplet dispensing in digital microfluidic devices: Assessment of long-term reproducibility
Katherine S. Elvira;
Katherine S. Elvira
1Department of Chemistry and Applied Biosciences,
Institute for Chemical and Bioengineering
, Wolfgang-Pauli Strasse, ETH Zürich, Zürich, CH 8093, Switzerland
Search for other works by this author on:
Robin Leatherbarrow;
Robin Leatherbarrow
2Department of Chemistry,
Imperial College London
, Exhibition Road, South Kensington, SW72AZ, United Kingdom
Search for other works by this author on:
Joshua Edel;
Joshua Edel
2Department of Chemistry,
Imperial College London
, Exhibition Road, South Kensington, SW72AZ, United Kingdom
Search for other works by this author on:
Andrew deMello
Andrew deMello
1Department of Chemistry and Applied Biosciences,
Institute for Chemical and Bioengineering
, Wolfgang-Pauli Strasse, ETH Zürich, Zürich, CH 8093, Switzerland
Search for other works by this author on:
Biomicrofluidics 6, 022003 (2012)
Article history
Received:
December 01 2011
Accepted:
January 29 2012
Citation
Katherine S. Elvira, Robin Leatherbarrow, Joshua Edel, Andrew deMello; Droplet dispensing in digital microfluidic devices: Assessment of long-term reproducibility. Biomicrofluidics 1 June 2012; 6 (2): 022003. https://doi.org/10.1063/1.3693592
Download citation file:
Sign in
Don't already have an account? Register
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Pay-Per-View Access
$40.00
Citing articles via
Trajectory analysis of Bacillus subtilis in micro-droplets
Yangyang Tang, Xiaolei Cao, et al.
Visualizing liquid distribution across hyphal networks with cellular resolution
Amelia J. Clark, Emily Masters-Clark, et al.
Related Content
Floating electrode electrowetting on hydrophobic dielectric with an SiO2 layer
Appl. Phys. Lett. (May 2013)
Island-ground single-plate electro-wetting on dielectric device for digital microfluidic systems
Appl. Phys. Lett. (July 2014)
Two-phase microfluidics in electrowetting displays and its effect on optical performance
Biomicrofluidics (February 2016)
Improving the dielectric properties of an electrowetting-on-dielectric microfluidic device with a low-pressure chemical vapor deposited Si3N4 dielectric layer
Biomicrofluidics (March 2015)
A study of the chemical and physical interaction between copper and polyimide
J. Vac. Sci. Technol. A (May 1989)