We revisit laser intensity noise in the context of stimulated Raman scattering (SRS), which has recently proved to be a key technique to provide label free images of chemical bonds in biological and medical samples. Contrary to most microscopy techniques, which detect a weak photon flux resulting from light matter interactions, SRS is a pump-probe scheme that works in the high flux regime and happens as a weak modulation (10−4–10−6) in a strong laser field. As a result, laser noise is a key issue in SRS detection. This practical tutorial provides the experimentalists with the tools required to assess the amount of noise and the ultimate SRS detection limit in a conventional lock-in-based SRS system. We first define the quantities that are relevant when discussing intensity noise and illustrate them through a conventional model of light detection by a photodiode. SRS is then introduced in its lock-in-based implementation, and the model presented is adapted in this particular case. The power spectral density, relative intensity noise (RIN), signal to noise ratio, and sensitivity of the system are derived and discussed. Two complementary methods are presented that allow measurement of the RIN and assessment of the performance of a SRS system. Such measurements are illustrated on two commercial laser systems. Finally, the consequences of noise in SRS are discussed, and future developments are suggested. The presentation is made simple enough for undergraduate students, graduate students, and newcomers in the field of stimulated Raman and more generally in pump-probe based schemes.
I. INTRODUCTION
Over the last two decades, coherent Raman imaging has evolved as a mature, label-free, imaging technique with numerous applications in biology and medicine.1 The seminal work of Zumbusch in 1999 revived coherent anti-Stokes Raman scattering (CARS)2 as a vibrational microscopic imaging modality.3 Since then, the coherent Raman imaging field experienced a second revolution in 2008 when stimulated Raman scattering (SRS)4 was also demonstrated as a powerful vibrational imaging scheme.5,6 Contrary to CARS, SRS is free of nonresonant background and scales linearly with the molecular concentration.7 These key features initiated the development of SRS imaging technologies8–17 and facilitated their successful applications in biology,18–21 chemistry,22,23 and medicine24–27 as a quantitative and label free chemical imaging modality.
Contrary to CARS, which detects faint generated photons at specific wavelengths, SRS is a pump-probe scheme that works in the high photon flux regime. It manifests itself as a weak modulation (10−4–10−6) that is transferred from an amplitude modulated (AM) laser on an unmodulated (probe) laser beam.5
Because the modulation transferred to the probe beam is weak, the laser noise and the detection of electronic noise are key components to achieve the ultimate SRS detection level.28,29 For instance, performing SRS with noisy fiber lasers requires the development of specific balanced detection schemes,15,30,31 which are at best 3 dB above the shot noise limit. Although they are keys to SRS, it appears that considerations on laser and detection noise in SRS imaging systems are still lagging behind when compared to the demonstrated technological and application advances. For instance, it is often not clear in published papers how the reported SRS detection ability compares to the shot noise limit.
This paper is intended to provide the SRS experimentalist with theoretical basics and most importantly a reliable experimental method to characterize the noise of an SRS system as compared to the ultimate shot noise. More generally, our presentation applies also to any pump probe spectroscopy scheme. Although no fundamental breakthrough is presented here, it is our understanding that the required background to master physical description and measurement of noise is often not available to physical chemists that are building and using SRS systems. The scope of this paper is to provide a tutorial on noise found in SRS systems accessible for undergraduate students, graduate students, and newcomers in the field.
We start by defining noise in the context of light intensity detection and present a conventional model used to describe such noise. The important quantities—the power spectral density (PSD), relative intensity noise (RIN), and signal to noise ratio (SNR)—are calculated for this model. We then present a typical lock-in-based SRS measurement system and describe how the previous model changes in this context. The SNR and sensitivity of such SRS system are discussed and linked to the laser RIN. Two complementary measurements of the RIN are presented that can be used to characterize and optimize the performance of a SRS system. These measurements are illustrated using two commercial laser systems commonly used for SRS imaging. We finally discuss the implications of such measurements and additional means to further increase the signal to noise ratio in SRS in both current systems and in future technological developments. The paper is complemented with a supplementary material that is intended to be used as a practical handbook to perform RIN noise measurement of a SRS laser system. For a comprehensive introduction to coherent Raman and SRS, we invite the reader to refer to the tutorial.7
II. DEFINITIONS AND LASER INTENSITY NOISE MODEL
Although various implementations of SRS have been developed, they mostly share the same working principle, which is the detection of small intensity variations in an intense laser beam. In this section, a standard model of laser intensity measurement is presented, along with the definitions of the power spectral density (PSD), relative intensity noise (RIN), and signal to noise ratio (SNR). The RIN of a laser is calculated for this model to illustrate the contribution of classical intensity noise, quantum fluctuations, and electrical noise.
A. Model
The optical intensity of a laser is modeled by a time-varying photon rate . In this tutorial, the term optical intensity is preferred to optical power in order to avoid confusion when discussing electrical power later. The laser is assumed monochromatic for simplification and is detected with a photodiode.
The electrical current I(t) at the output of this photodiode is modeled using a semiclassical approach, consistent with the detection of coherent states of light. Similar work has been done by Quinlan et al.32 and can be found in textbooks.33
B. Definitions
Let us consider a quantity A that is a function of time. In the following, definitions are made using the general quantity A although A will be replaced by photodiode current I for calculations. Similar derivations could be done using voltages, but the discussion, in particular, in terms of power, signal, and noise, would be identical.
1. DC power
2. Power Spectral Density (PSD)
3. Relative Intensity Noise (RIN)
4. Signal to Noise Ratio (SNR)
C. Calculations
1. Average and relative intensity
2. Relative Intensity Noise
The three measurement regimes and associated dominant power spectral densities and relative intensity noise.
Limitation . | Electronic . | Shot noise . | Excess (classical) noise . |
---|---|---|---|
2qI | |||
Limitation . | Electronic . | Shot noise . | Excess (classical) noise . |
---|---|---|---|
2qI | |||
As can be seen in Eq. (15), the SNR for laser intensity measurements is inversely proportional to the RIN integrated over the photodiode bandwidth. Equation (22) illustrates that the RIN depends on the frequency and average laser intensity. At low frequencies (f < 1 MHz), lasers usually feature fluctuations in intensity significantly above the shot noise . Frequently, this noise renders impossible the direct measurements of small intensity fluctuations (10−4–10−6) induced by SRS. For this reason, lock-in amplification is typically used to increase the SNR by moving the measurement toward higher frequencies.
III. LOW NOISE SRS DETECTION THROUGH LOCK-IN AMPLIFICATION
A. System and model description
Typical stimulated Raman scattering scanning microscope system. The SRS pump laser is modulated using an acousto-optic modulator (AOM). It recombines with the Stokes beam using a dichroic mirror (DM), and both are sent on the sample through a scanning microscope. The modulated beam is filtered out using an optical filter, and the intensity of the other beam is measured with a photodiode. The signal from the photodiode I is mixed with a reference signal r, filtered and amplified using a lock-in amplifier (LIA), which generated an output current Im.
Typical stimulated Raman scattering scanning microscope system. The SRS pump laser is modulated using an acousto-optic modulator (AOM). It recombines with the Stokes beam using a dichroic mirror (DM), and both are sent on the sample through a scanning microscope. The modulated beam is filtered out using an optical filter, and the intensity of the other beam is measured with a photodiode. The signal from the photodiode I is mixed with a reference signal r, filtered and amplified using a lock-in amplifier (LIA), which generated an output current Im.
The amplitude of the relative SRS gain β is typically 10−4 to 10−6. The exact expression of the proportionality factor in Eq. (25) depends on many experimental parameters, and in-depth developments can be found in the literature.7,35 Those experimental parameters may be the source of additional noise (change in pump laser intensity, laser spatial drift, etc), that should be examined individually. In most experimental settings, however, the limiting noise will be introduced by the probe laser.
B. Average current and DC power
C. Power spectral density
D. Signal to noise ratio
E. SNR optimization
The different terms of the denominator in Eq. (46) are precisely the RIN associated with electronic noise, shot noise, and laser excess noise, respectively (Table I). Using this equation, one can identify all the parameters that can maximize the SNR in a lock-in-based SRS system as follows:
Maximizing β, the relative modulation of the probe beam. This goes beyond the scope of this tutorial, but β increases linearly with the number of molecular bonds probed, as well as with the laser intensity in the AM beam [Eq. (25)]. It is important to note that β is constant with respect to the laser intensity in the probed beam and therefore does not depend on I. It does, however, depend linearly on the AM laser average intensity, and further details are discussed in Sec. V B.
Minimizing Δf, the lock-in bandwidth. In principle, an arbitrarily high SNR can be achieved by reducing Δf. In practice, the bandwidth has to be large enough to allow changes in the SRS signal. In a SRS point scanning microscope, the signal is expected to vary from one pixel to the next. In this case, Δf should match the pixel acquisition rate to allow fluctuations from one pixel to the next, while filtering out fluctuations within a pixel.
Maximizing I to minimize electrical noise and shot noise. Depending on the photodiode noise features, one can increase the probed laser beam intensity (and therefore I) to reach the regime where the electronic noise becomes negligible with respect to classical and quantum laser fluctuations. In the systems described below, this regime was reached for a few milliwatts of average intensity in the probe beam. When the SNR is limited by the laser, one of two scenarios has to be considered.
The RIN is limited by the shot noise, in which case the SNR will increase linearly with I. The SNR is then limited by how much I can be increased. This limit is set by the laser maximum output power, the photodamage threshold, or the laser excess noise.
The RIN is limited by the laser excess noise, in which case increasing I will not change the SNR. The SNR limit allowed by the laser is reached and cannot be improved by increasing the probe laser beam intensity.
In this last scenario, the laser excess noise ultimately sets the limit on the SNR in a lock-in-based SRS system. By using a modulation frequency that minimizes this excess RIN, one can achieve the optimal SNR. A proper measurement of the laser RIN is therefore necessary to quantify and optimize the performance of such a SRS system. Two experimental procedures used for laser RIN measurement are detailed in Sec. IV: (i) a complete RIN characterization (Fig. 2) for different frequencies and optical intensities and (ii) a faster RIN assessment on an existing lock-in-based SRS microscope. Both methods are illustrated on two commercial laser sources.
Schematics of the RIN measurement methods described in Sec. IV. In both cases, measurement of the average photocurrent I can be performed either directly using an oscilloscope after the photodiode or estimated by measuring the optical average power using a power meter and using the photoconvertion equation [Eq. (17)]. Proper characterization of the gain and bandwidth of the measurement electronics is key to properly estimating the laser RIN.
Schematics of the RIN measurement methods described in Sec. IV. In both cases, measurement of the average photocurrent I can be performed either directly using an oscilloscope after the photodiode or estimated by measuring the optical average power using a power meter and using the photoconvertion equation [Eq. (17)]. Proper characterization of the gain and bandwidth of the measurement electronics is key to properly estimating the laser RIN.
IV. LASER INTENSITY NOISE MEASUREMENTS
As described previously, the RIN of a photodiode output current is composed of three different terms [Eq. (22)] that come from electronic noise, quantum fluctuations, and classical fluctuations (Table I). The electronic contribution is due only to the photodiode and can be made negligible by choosing the adequate device and laser intensity. The quantum and classical fluctuations are inherent to the laser source. Its understanding and characterization is crucial to assess the performance of a SRS system. The RIN is a function of both frequency and average current provided by the photodiode. Here, we describe a way to reconstruct the full surface over a large range of frequencies (500 kHz–30 MHz) and low laser intensities (0.1 mW–50 mW) suitable for SRS imaging applications.
A. Measured laser sources
Device 1 is a Ti:sapphire femtosecond laser (Chameleon, Coherent Inc) operating at a repetition rate of 80 MHz with a pulse width duration of 150 fs and a wavelength of 800 nm. This beam acts as the (unmodulated) SRS pump beam of a lock-in-based SRS system (SRL modality), whereas the (AM) Stokes beam is provided by a tunable femtosecond optical parametric oscillator OPO (Compact OPO, APE & Coherent, Inc.) pumped by the Ti:sapphire laser. Here, only the 800 nm beam is considered as it is the one monitored by the SRS photodiode.
Device 2 is a picosecond OPO (Emerald, APE) operating at a repetition rate of 80 MHz with a pulse duration of 2 ps and a wavelength of 800 nm. This OPO is pumped by a 20 W frequency doubled ytterbium (Yb) fiber laser (Emerald engine, APE) (80 MHz rep rate, 2 ps pulse duration) that provides also the Stokes beams at 1030 nm. As explained previously, only the 800 nm beam is considered in the RIN measurement. Note that device 2 is similar to the commercially available one box SRS laser system PicoEmerald (APE) although the PicoEmerald uses a 10 W frequency doubled Yb fiber laser.
B. Method 1: Frequency and intensity dependence (Fig. 2)
To properly quantify the noise level in devices 1 and 2, fast biased photodiode (Thorlabs DET10A/M, 1 ns rise time) is placed at the position of the forward SRS detector (Fig. 1). The average photocurrent generated by the photodiode was measured with a fast oscilloscope (HP 54111D 500 MHz bandwidth, 50 Ω input). The typical photocurrent was several milliamperes. The associated shot noise PSD can be calculated using Table I and is on the order of −190 dBW/Hz which is far below the input noise (−180 dBW/Hz) of the spectrum analyzer used here (Zurich instrument HF2LI). For this reason, a filtering and amplification step was added. The photodiode was loaded with a 50 Ω resistor, and the resulting voltage was filtered with a 30 MHz low pass filter (Mini-Circuits BLP-30+) to damp the 80 MHz repetition rate and harmonics of the lasers. The signal was then amplified with a 25 dB low noise preamplifier (Ref. 153579, APE) working in the range of 100 kHz–100 MHz. The overall effect of this filtering and amplification stage was to amplify by more than 20 dB the electrical components between 500 kHz and 30 MHz (see Fig. 1 of the supplementary material). Then, the electrical power spectral density was acquired with a spectrum analyzer (Zurich instrument HF2LI). The PSD was measured for 1000 frequencies from 500 kHz to 30 MHz with a bandwidth of 1.1 kHz–70 kHz. The effect of the filter and amplifier was numerically subtracted from the obtained data.
C. Method 1: Results
For an average photocurrent of I = 5 mA, the recovered laser RIN is plotted in Fig. 3(a) for devices 1 and 2. For device 1 (Ti:sapphire laser), the measured data match the expected shot noise limit for frequencies above 2 MHz. Consequently, any SRS related modulation of the laser above this frequency can be detected down to the shot noise limit. For device 2, the measured data are within 5 dB of the shot noise limit for frequencies above 20 MHz. This corresponds to a SNR for the system—at such modulation frequencies and photocurrent—that is only a factor of 3 below the shot noise limit, meaning that a measurement requires a bandwidth 3 times smaller than that of a shot noise limited system to receive the same SNR [Eq. (15)].
(a) Relative intensity noise measured as a function of frequency for device 1 (Ti:sapphire) and 2 (fiber laser pumped OPO) with average photodiode current I = 5 mA. The solid line is the filtered data (Savitzky–Golay filter, 51 points, 1st order). (b) Relative intensity noise (RIN) as a function of the average photodiode current for device 1 (Ti:sapphire, left triangles) and 2 (fiber laser pumped OPO, right triangles). The RIN was averaged between 10 and 30 MHz. The excess noise on device 2 (dashed line) is fitted to match the measurements. The electronic noise was removed numerically.
(a) Relative intensity noise measured as a function of frequency for device 1 (Ti:sapphire) and 2 (fiber laser pumped OPO) with average photodiode current I = 5 mA. The solid line is the filtered data (Savitzky–Golay filter, 51 points, 1st order). (b) Relative intensity noise (RIN) as a function of the average photodiode current for device 1 (Ti:sapphire, left triangles) and 2 (fiber laser pumped OPO, right triangles). The RIN was averaged between 10 and 30 MHz. The excess noise on device 2 (dashed line) is fitted to match the measurements. The electronic noise was removed numerically.
Complementary to the frequency analysis, one needs to measure the RIN for different photodiode currents to characterize the optimal setup working parameters. This analysis was performed by recording the RIN as a function of frequency for photocurrents ranging from 0.5 mA to 7 mA, after which the detector saturates. This saturation comes in the form of a distorted signal from the photodiode output current visualized on the oscilloscope. The surfaces are plotted in Fig. 4 for both devices. One can visualize the contribution of the electrical noise, the shot noise, the classical noise, and the SRS signal. To better illustrate , its value was averaged from 10 to 30 MHz in the frequency domain [Fig. 3(b)].
RIN of devices 1 (left) and 2 (right) as a function of frequency ranging from 500 kHz to 30 MHz and photocurrents ranging from 0.5 mA to 7 mA. The noise floor set by the shot noise is represented by the green bottom surface. At low photocurrents, the electronic contribution to the RIN diverges; at low frequencies, the classical noise dominates. The peak shown on device 1 at 20 MHz corresponds to a SRS induced modulation at this frequency.
RIN of devices 1 (left) and 2 (right) as a function of frequency ranging from 500 kHz to 30 MHz and photocurrents ranging from 0.5 mA to 7 mA. The noise floor set by the shot noise is represented by the green bottom surface. At low photocurrents, the electronic contribution to the RIN diverges; at low frequencies, the classical noise dominates. The peak shown on device 1 at 20 MHz corresponds to a SRS induced modulation at this frequency.
Device 1 is shot noise limited around 20 MHz for all photodiode current and therefore all laser intensities, allowed by the photodiode. Device 2 shows a small excess noise of 5 dB above the shot noise for 5 mA photocurrent. As expected from the model, device 2 is shot noise limited for small enough photocurrents (<500 μA) and reaches a plateau at . The supplementary material provides a practical handbook and software Matlab script to implement Method 1.
D. Method 2: Using an existing lock-in SRS system (Fig. 2)
The gain G of the lock-in amplifier was measured experimentally using a reference signal at f0 = 20 MHz. The input signal entering the lock-in was a sinusoidal voltage of amplitude 300 μV and frequency f0. The output signal from the lock-in resulted in a DC voltage of amplitude 1.8 V. The associated gain in power was therefore measured to be G = 20 log(Vout/Vin) = 75.56 dB. The reference, lock-in input, and output signals were generated and recorded using a multipurpose system (Zurich Instrument HF2LI).
Given G and Δf and following Eq. (49), the RIN of the photodiode current I can be measured experimentally using the lock-in, by recording the variance of the lock-in output voltage. This RIN measurement is made at the lock-in frequency f0 and varying laser average intensity (and therefore I).
More specifically, by plotting the RIN as a function of I on a log-log plot, one can identify the different regimes (electronic noise, shot noise, classical noise) of limitation of the measurement.
The PSD and RIN associated with laser shot noise are given in Table I. The average photocurrent was measured through the average optical intensity using Eq. (17), knowing the quantum efficiency η = 0.8 of the photodiode at 800 nm. As a result for a laser at 800 nm wavelength, , where is expressed in Watts and I in Amperes. The RIN was measured on the two commercial laser systems described in Sec. IV A, using the lock-in voltage output standard deviation, for laser average optical intensities ranging from 0.6 to 50 mW on the photodiode (Fig. 5). The contribution of the electronic noise was removed numerically, as described in Sec. IV B.
Photodiode current RIN (expressed in dBW), for fixed frequency f = 20 MHz and increasing average laser photodiode current I (and therefore average laser intensity). The figure shows measurements for device 1 (Ti:sapphire—orange left triangles) and device 2 (fiber laser pumped OPO—blue right triangles). The shot noise (solid line) and excess laser noise (dash line) are added to provide the modeled noise for device 2 (dotted line). The electronic noise was removed numerically from all measurements but is shown (dashed-dotted line) for comparison.
Photodiode current RIN (expressed in dBW), for fixed frequency f = 20 MHz and increasing average laser photodiode current I (and therefore average laser intensity). The figure shows measurements for device 1 (Ti:sapphire—orange left triangles) and device 2 (fiber laser pumped OPO—blue right triangles). The shot noise (solid line) and excess laser noise (dash line) are added to provide the modeled noise for device 2 (dotted line). The electronic noise was removed numerically from all measurements but is shown (dashed-dotted line) for comparison.
E. Method 2: Results
The measured RIN from device 1 matches the model (dotted orange line), with no measurable classical (excess) laser noise. For photocurrents I higher than 5 mA (i.e., laser intensities on the photodiode above 10 mW), the electronic noise is negligible and the SRS measurement is shot noise limited.
The RIN measurement obtained using Method 2 (existing lock-in amplifier) is slightly discordant with the more complete study using Method 1. The excess laser RIN obtained is 2 dB below the previously measured value. This difference can be attributed to changes in the OPO alignment that introduced additional excess laser noise in the first case. It is noted that Method 2 measurement of the RIN, for device 2, presents relatively strong discrepancies with the model, in the form of fluctuations above and below the expected value. These fluctuations are also attributed to changes in the OPO alignment during the time of the measurement, which introduced up to 2 dB of fluctuations in the RIN measurement. Despite the fluctuations, the measured RIN of −161 dBc/Hz is typical of daily operation.
V. DISCUSSION
A. Consequences of RIN on SRS measurements
Excess laser RIN has a direct and major impact on the sensitivity of lock-in-based SRS measurements. As illustrated by Eq. (46), the SNR in such measurement is limited by either electronic noise, shot noise, or laser excess intensity noise. While electronic noise can usually be avoided and shot noise sets the physical limit, excess laser RIN is more difficult to control and minimize. Such excess noise in lasers can usually be explained by excess noise in the pump laser or amplified stimulated emission, which are harder to address.
Since the RIN is a function of frequency, the choice of the laser modulation frequency used in the lock-in detection is an important parameter to minimize the laser RIN and maximize the SNR. As illustrated in Fig. 3(a), the RIN can be lowered by an order of magnitude by simply changing the modulation frequency from 5 MHz to 20 MHz, therefore increasing the SNR by a factor of 10.
It is crucial to note that the SNR is limited by the laser that is used as a probe and detected by the photodiode. In the case of SRS this is particularly important since only one of the lasers used (pump or stokes) is the probe and has to exhibit a low noise. The other laser is usually amplitude modulated, and its contribution to noise is negligible. Complementary to this, once the SNR is limited by excess laser RIN, there is no advantage of increasing the laser power at the sample plane (and therefore the current of the photodiode). Since the SRS signal scales as the laser noise in this case, the SNR will not be improved by using higher laser power, as detailed in Sec. III E. Although it is important to understand how the SNR can be improved by increasing the intensity of the SRS pump or Stokes lasers, this analysis should be done considering the photodamage on the sample that usually sets the experimental limit to the laser intensities.
B. Optimal pump probe ratio
It has been demonstrated that if the SNR of a lock-in-based system is shot noise limited, and if the photodamage is linear with the intensity of both laser beams, the optimal SNR is achieved for constant photodamage when the AM beam has twice the intensity of the probe beam.36
C. Sensitivity and performance report
In the work presented here, using device 2 as the probe beam, there is no need to push the intensity above few milliwatts (on the photodiode). In this case, it is interesting to increase the optical intensity on the second (modulated) SRS laser to increase the SNR.
To reach such sensitivity with a fiber laser usually requires noise cancellation using balanced detection schemes.15,30,37 Because the sensitivity of the system also depends on its bandwidth, the proper way to evaluate and assess the performance of a SRS system with respect to laser noise is to specify:
The average photocurrent from the detector, which sets the shot noise limited RIN and SNR (Table I).
The achieved experimental RIN by either its absolute value or its distance in decibel from the shot noise limit.
D. Future developments
When imaging biological samples and other diluted species, the number of Raman active molecules in the focal spot is such that to detect a signal, one needs to stay on a pixel from several to tens of microseconds. The associated bandwidth is therefore limited to tens of kilohertz, and the fastest one can expect to perform imaging is tens of thousands of pixels per second. This acquisition rate corresponds to a few frames per second with 100 by 100 pixel images. This limit could be overcome in two ways.
First, by reducing the laser repetition frequency and increasing the pulse peak power. This would further increase the efficiency of the SRS process and render modulation at half of the repetition frequency more practical. In this case, low frequency noise would be extremely important to characterize in order to preserve the SNR. The limitation set by nonlinear photodamage will however put a limit of the amount of peak power that can be delivered on the sample.
Second and complementary to the first point, as the time spent per pixel cannot be diminished further, major developments are to be expected in spatially multiplexed SRS. Either by having multiple foci at once in the sample16 or with more robust spatial multiplexing, one can increase the amount of information collected from the sample and further increase the imaging speed. In this case, a lot of power will be necessary to enable the SRS process to take place at multiple locations. This high power can be achieved with fiber lasers or optical parametric amplifiers, but the question of noise in such system still needs to be addressed to ensure optimal—shot noise limited–SNR.
VI. CONCLUSION
This work presented a description and ways to characterize laser noise in the context of stimulated Raman scattering. The laser excess RIN [Sα(f)] was introduced as well as the shot noise RIN. The relation between RIN and the signal to noise ratio in SRS was derived [Eq. (44)], as well as the optimal ratio of pump and Stokes intensities to maximize the SNR for nonshot noise limited systems. Two methods for measuring the RIN were presented, either with a full characterization using commercially available electronic parts or with a more accessible implementation using the SRS lockin amplifier. Two laser systems were studied. One was a solid state Ti-sapphire laser which was shot noise limited around 20 MHz for laser power up to 70 mW. The second system was an OPO pumped by a 20 W fiber laser which showed excess RIN of −160 ± 1 dBc/Hz at 20 MHz. Future developments in SRS will likely involve an improvement in the noise performance of lasers and a systematic characterization of the system’s noise.
SUPPLEMENTARY MATERIAL
See the supplementary material for the additional information on how to perform measurements as described in Method 1. The file “practical_handbook” provides a practical handbook on how to setup the measurement. It contains basic troubleshooting in order to guide the reader through the procedure. The file “data_analysis.m” contains the MATLAB script used to generate the plots in the practical handbook and is intended for direct use by other experimentalists. The other files begin with either one of the following: “amplifier, amplifier_and_filter, dark, filter, LIA, pump, Stokes.” Each of this category has five files generated by the spectrum analyzer for the different components as explained in the practical handbook. “Data.csv” is the raw data in the csv format. “Graph1.png” and “Graph2.png” are the amplitude and phase plots from the spectrum analyzer. “Readme.txt” provides additional information on the spectrum analyzer parameters and csv headers. “settings.zicfg” contains the parameter file for the spectrum analyzer.
Two supplementary figures are referred to in the manuscript: “sup_fig_1” supplementary material, Fig. 1: The measured transfer function of the filter and preamplification step. “sup_fig_2” supplementary material, Fig. 2: Equivalent electronic noise power spectral density Sϵ+(f)
ACKNOWLEDGMENTS
The authors would like to acknowledge Professor Randy A. Bartels for his useful input on this work. Dr. Ingo Rimke has financial interest in APE.