In recent years, image-scanning microscopy (ISM, also termed pixel-reassignment microscopy) has emerged as a technique that improves the resolution and signal-to-noise compared to confocal and widefield microscopy by employing a detector array at the image plane of a confocal laser scanning microscope. Here, we present a k-space analysis of coherent ISM, showing that ISM is equivalent to spotlight synthetic-aperture radar and analogous to oblique-illumination microscopy. This insight indicates that ISM can be performed with a single detector placed in the k-space of the sample, which we numerically demonstrate.

Confocal imaging is a technique that improves axial-sectioning and transverse resolution in optical microscopy compared to conventional “widefield” imaging.1,2 In widefield imaging, the entire field of view (FOV) is illuminated, and all points in the sample are simultaneously imaged to a detector (camera) plane. Confocal imaging allows an improvement in the transverse and axial resolution by illuminating the sample with a scanned focused spot and detecting the signal emerging from that focal spot with a small “confocal” detector conjugated to the illumination spot. The effective size of the confocal detector, which determines the spatial filtering and collection efficiency, is set by a small “confocal pinhole” placed in front of the detector. While the resolution improvement in confocal-microscopy increases with decreasing pinhole size,3 a smaller pinhole lowers the detection efficiency, resulting in a lower signal-to-noise ratio (SNR).

A recently introduced method termed image-scanning microscopy4–6 (ISM) or pixel-reassignment7 allows full use of all signal photons in a confocal scanning system without sacrificing transverse imaging resolution. Moreover, depending on the imaging point spread function (PSF), ISM can also provide an improvement in imaging resolution.8,9 ISM achieves this feat by using an array of detectors at the image plane instead of the conventional single detector of confocal systems. The center detector of the ISM array collects the same information as would have been collected by a confocal pinhole. However, the neighboring detectors collect light that is otherwise rejected in a confocal system. To construct the ISM image, at each illumination point, r i n, the signals from each detector position, r out, are reassigned to the midpoint between illumination and detection positions:2,4 r ISM = 1 2 ( r i n + r out ). The reassigned signals are summed over all scan positions, forming an ISM image. ISM was first implemented in incoherent microscopy modalities4,10,11 and was recently adapted to coherent imaging modalities.12–15 

Here, we analyze coherent-ISM in the spatial Fourier domain. Utilizing the projection-slice theorem,16 we show that ISM is equivalent to spotlight synthetic-aperture radar (SAR)17,18 (Fig. 1), a well-established beam-scanning imaging technique, which utilizes a single detector. As a direct result, we numerically demonstrate that ISM can be performed with a single detector and leverage the k-space analysis to highlight the close connection to oblique-illumination microscopy.19 

FIG. 1.

Coherent image-scanning-microscopy (ISM), obtained by raster-scanning a focused illumination, and its k-space interpretation as spotlight synthetic aperture radar (SAR), obtained by plane wave illumination. (a) ISM measurements can be represented by a reflection matrix R ( x i n , x out ) providing the measured signals at coordinates xout for an illumination focused at xin [where the diagonal R(x, x) is the confocal image]. For simplicity, a one-dimensional object and monochromatic illumination are considered. An ISM image, I ISM ( x ), is obtained by summing over the anti-diagonal elements of the reflection matrix, I ISM ( x ) = x i n R ( x i n , x out = 2 x x i n ), i.e., for all points satisfying x = ( x out + x i n ) / 2 (white arrows). According to the Fourier-slice theorem, this line summation (projection) is the Fourier-pair of the diagonal of the k-space reflection matrix R ̃ ( k i n , k out ) = F { R ( x i n , x out ) }, where F { } is a 2D Fourier transform. These k-space equivalent measurements, R ̃ ( k i n , k out = k i n ), represent the reflected amplitude of a plane wave at kout obtained for a plane wave illumination at kin = kout, for a set of different illumination angles. These are analogous to spotlight SAR measurements. (b) Schematic of conventional ISM measurement setup: focused illumination is raster-scanned across the object plane (x-space). At each illumination position, the reflected fields at all positions around the illumination point are measured. (c) Schematic of a setup for k-space ISM (analogous to spotlight SAR measurements): At each measurement position (green and purple beams), the object is illuminated by a tilted plane wave, k i n, and the plane wave that is reflected in the same angle, k out = k i n, is detected.

FIG. 1.

Coherent image-scanning-microscopy (ISM), obtained by raster-scanning a focused illumination, and its k-space interpretation as spotlight synthetic aperture radar (SAR), obtained by plane wave illumination. (a) ISM measurements can be represented by a reflection matrix R ( x i n , x out ) providing the measured signals at coordinates xout for an illumination focused at xin [where the diagonal R(x, x) is the confocal image]. For simplicity, a one-dimensional object and monochromatic illumination are considered. An ISM image, I ISM ( x ), is obtained by summing over the anti-diagonal elements of the reflection matrix, I ISM ( x ) = x i n R ( x i n , x out = 2 x x i n ), i.e., for all points satisfying x = ( x out + x i n ) / 2 (white arrows). According to the Fourier-slice theorem, this line summation (projection) is the Fourier-pair of the diagonal of the k-space reflection matrix R ̃ ( k i n , k out ) = F { R ( x i n , x out ) }, where F { } is a 2D Fourier transform. These k-space equivalent measurements, R ̃ ( k i n , k out = k i n ), represent the reflected amplitude of a plane wave at kout obtained for a plane wave illumination at kin = kout, for a set of different illumination angles. These are analogous to spotlight SAR measurements. (b) Schematic of conventional ISM measurement setup: focused illumination is raster-scanned across the object plane (x-space). At each illumination position, the reflected fields at all positions around the illumination point are measured. (c) Schematic of a setup for k-space ISM (analogous to spotlight SAR measurements): At each measurement position (green and purple beams), the object is illuminated by a tilted plane wave, k i n, and the plane wave that is reflected in the same angle, k out = k i n, is detected.

Close modal

We begin our analysis by considering the case of aberration-free, single-scattering imaging scenario. We represent the signals collected in coherent-ISM using the reflection matrix formalism.14,20 In this formalism, the fields that are collected at position r out at the detection plane when the illumination is focused at r i n in the object plane are given by the matrix element R ( r i n , r out ) [see coordinate notation in Fig. 1(b)]. To simplify the mathematical derivations and without loss of generality, we consider below one transverse dimension, whose coordinate is given by x.

Considering the pixel reassignment process [ x ISM = 1 2 ( x i n + x out )] for the case of identical illumination and detection PSFs, the ISM image formation at a given imaging depth, z, is given by
I ISM ( x ) = x i n R ( x i n , x out = 2 x x i n ; z ) .
(1)

Note that the general case of non-identical PSFs is analyzed in the supplementary material, Sec. C.

In this matrical representation of Eq. (1), the ISM summation can be interpreted as a sum over the anti-diagonal elements of the reflection matrix [Fig. 1(a)].

Such a line summation [Fig. 1(a), left inset] is analogous to a line projection performed in many tomographic imaging techniques, such as x-ray CT.21 Leveraging the Fourier-slice theorem,16,22 the Fourier transform of this line-projection is the main diagonal in the 2D Fourier transform of R ( x i n , x out ; z ), scaled by a factor of half [Fig. 1(a), right inset],
F { I ISM ( x ) } = I ̃ ISM ( k ) = R ̃ ( k i n = k / 2 , k out = k / 2 ; z ) ,
(2)
where F { } is the spatial Fourier transform, and kin and kout are the k-vectors. For monochromatic illumination, this k-space reflection matrix,23  R ̃ ( k i n , k out ), can be interpreted as the measured reflected plane wave with a wavevector kout when illuminating the sample with a plane wave, kin. This k-space matrix is the 2D Fourier transform of R,
R ̃ ( k i n , k out ; z ) = R ( x i n , x out ; z ) e i k i n x i n e i k out x out d x out d x i n .
(3)
The Fourier-slice equivalence of Eq. (2) can be derived by considering the rescaled diagonal of R ̃,
R ̃ ( k i n = k / 2 , k out = k / 2 ; z ) = R ( x i n , x out ; z ) e i 1 2 k x i n e i 1 2 k x out d x out d x i n = R ( x i n , x out ; z ) e i k x i n + x out 2 d x out d x i n .
(4)
Performing inverse Fourier transform yields
R ̃ ( k i n = k / 2 , k out = k / 2 ; z ) e ikx d k = R ( x i n , x out ; z ) e i k x i n + x out 2 e ikx dkd x out d x i n = R ( x i n , x out ; z ) δ ( x x i n + x out 2 ) d x out d x i n ,
(5)
which is the same anti-diagonal summation as in ISM image formation in Eq. (1).

This mathematical equivalence implies that the same ISM image can be acquired in two different forms. One, conventionally, utilizes a scanning array in the spatial domain, detecting the reflected fields from a spatially focused scanning illumination beam [Fig. 1(b)]. Alternatively, the sample can be scanned by plane waves at different angles of incidence (given by kin), and a single detector placed in the far field of the sample that detects the reflected plane wave at the same angle, i.e., the single Fourier component kout = kin. This second approach, illustrated in Fig. 1(c), is known as spotlight synthetic aperture radar (SAR) imaging.18,24

The scanning plane wave illumination of k-space ISM is closely related to oblique-illumination microscopy.19 In oblique-illumination microscopy (originally proposed by Abbe25), the sample is illuminated by different angled plane-waves, and a camera is measuring the resulting image. In the k-space, the plane wave illumination is manifested as a shift of the high frequencies angular-spectrum information of the sample into the passband of the system, allowing its detection.26 In k-space ISM, each illumination is identical to the plane wave illumination of oblique-illumination microscopy. The difference from oblique illumination microscopy is that only a single Fourier component is detected in each illumination in k-space ISM [Eq. (2)]. Similar to oblique illumination microscopy, k-space ISM measurements, thus, yield extended k-space support. The differences between the techniques are the use of a single detector in k-space ISM rather than the detector-array of oblique-illumination microscopy, which requires a larger number of illuminations, and the angular-spectrum reweighting that is required in oblique illumination microscopy.19,27

To support and demonstrate our mathematical derivation, we numerically simulated the coherent imaging of a test sample with conventional ISM [Fig. 2(a)] and k-space ISM [Fig. 2(b)]. While the two approaches differ in both their illumination scheme (focused vs plane wave) and detection scheme (array detection in the object plane vs a single detector in the k-space), the resulting images are identical and present an improvement over confocal imaging.

FIG. 2.

Numerical comparison of coherent monochromatic imaging via conventional ISM, plane wave scanning ISM, and confocal imaging. (a) In conventional ISM, a focused illumination beam raster scans the target object plane (x, y). A detector array (CAM) that is conjugated to the object plane detects the reflected fields, and the ISM image is formed by reassignment and summation of the detected signals. A confocal image is formed from the signal emerging only from the illumination point [ ( x i n , y i n ) = ( x out , y out )]. (b) Monochromatic ISM can be performed with a single detector placed at the Fourier plane of the object, by sequentially illuminating the object with tilted plane waves. (c)–(e) We display a comparison between the reconstructed amplitudes in confocal detection, an x-scanned ISM, and a k-scanned ISM. A vertical cross section (depicted by a white dashed line in the confocal image) is also displayed. This comparison shows the equivalence between x-scan ISM and k-scan ISM results, which both result in a super-resolved image compared to the conventional confocal image.

FIG. 2.

Numerical comparison of coherent monochromatic imaging via conventional ISM, plane wave scanning ISM, and confocal imaging. (a) In conventional ISM, a focused illumination beam raster scans the target object plane (x, y). A detector array (CAM) that is conjugated to the object plane detects the reflected fields, and the ISM image is formed by reassignment and summation of the detected signals. A confocal image is formed from the signal emerging only from the illumination point [ ( x i n , y i n ) = ( x out , y out )]. (b) Monochromatic ISM can be performed with a single detector placed at the Fourier plane of the object, by sequentially illuminating the object with tilted plane waves. (c)–(e) We display a comparison between the reconstructed amplitudes in confocal detection, an x-scanned ISM, and a k-scanned ISM. A vertical cross section (depicted by a white dashed line in the confocal image) is also displayed. This comparison shows the equivalence between x-scan ISM and k-scan ISM results, which both result in a super-resolved image compared to the conventional confocal image.

Close modal

The simulated conventional ISM setup [Fig. 2(a)] consists of a 4-f imaging system, where the scanning laser and the detection array (CAM) are both imaged onto the object plane. A focused illumination beam raster scans the target object plane (x, y). At each illumination position, ( x i n , y i n ), the reflected field across the image plane, ( x out , y out ), is measured. The simulated k-space ISM setup [Fig. 2(b)] is based on the Fourier transform of a point illumination (LASER) and confocal detection, from the detector plane (DET) to the object plane using a lens. The point illumination is, thus, converted to a tilted plane wave in the object plane, and the detector measures a single Fourier component, kout = kin. Observing Fig. 2(b) reveals that k-space ISM is, in fact, confocal coherent imaging performed in the k-space domain. See simulation parameters in the supplementary material, Sec. A.

In Fig. 3, we compare the two ISM processing approaches and confocal imaging using an experimental ultrasound echography dataset. The dataset represents measurements performed on an acoustic phantom with multi-plane wave transmission.28 Data were acquired using a Verasonics P4–2v probe at a center frequency of f 0 = 4.8 MHz, having 64 elements with a total aperture size of D = 19.2 mm. The imaged target is composed of five pins [Fig. 3(a), red X's] having a diameter of 0.1 mm at a depth of 60 mm. Data were post-processed with the proper phases to perform coherent compounding for either confocal, x-space ISM,14 or k-space ISM [Figs. 3(a), 3(b), 3(c), respectively]. These experimental results agree with the analytic and numerical investigations [Fig. 3(d)]. See a discussion of the data processing in the supplementary material, Sec. B.

FIG. 3.

Experimental comparison of confocal ultrasound imaging (a), conventional ISM (b), and k-space ISM (c), of five reflecting pins (marked by red X's) in an acoustic phantom. The same experimental data, acquired by plane wave sonications, were processed offline to generate the results. (a)–(c) presents the reconstructed intensity (absolute value square of the reconstructed fields). The two approaches for ISM result in practically identical images, having the expected resolution improvement over the confocal image. Both ISM images show increased sidelobes characteristic of ultrasound ISM results.14 (d) A cross section comparison of (a)–(c) along the dotted line in (a).

FIG. 3.

Experimental comparison of confocal ultrasound imaging (a), conventional ISM (b), and k-space ISM (c), of five reflecting pins (marked by red X's) in an acoustic phantom. The same experimental data, acquired by plane wave sonications, were processed offline to generate the results. (a)–(c) presents the reconstructed intensity (absolute value square of the reconstructed fields). The two approaches for ISM result in practically identical images, having the expected resolution improvement over the confocal image. Both ISM images show increased sidelobes characteristic of ultrasound ISM results.14 (d) A cross section comparison of (a)–(c) along the dotted line in (a).

Close modal

In conclusion, by leveraging the Fourier slice theorem, we show that it is possible to perform coherent ISM with a single detector located at the Fourier plane of the object. This allows the same resolution improvement gained in ISM over confocal imaging without requiring a detector array. This may be interesting for utilizing ISM in other modalities where detector arrays are not accessible. Nevertheless, in this work, we have not analyzed imaging scenarios that include multiple-scatterings or the presence of aberrations. Tackling such propagation effects is an important concern in many imaging applications. While aberrations and scattering can be analyzed using the presented mathematical framework, by considering a distorted PSF, the full analysis of this subject is left for future work. In addition, the k-space acquisition and simple (confocal-like) analysis reduce the memory requirements and may be important also for the reduction of computational burden.

We note that a single detector approach will result in a lower SNR than the detector-array approach in many practical scenarios, where the same total illumination power is used, since only a fraction of the reflected field is detected. On the contrary, the plane wave illumination reduces the power density concentrated on the sample, as compared to focused scanned illumination, and, thus, may allow higher total illumination power in the case where damage threshold of the sample is the limiting factor for the used power.

See the supplementary material for the simulation parameters and a discussion of the data processing.

This work has received funding from The Israel Science Foundation (Grant No. 1361/18) and European Research Council (ERC) Horizon 2020 research and innovation program (No. 101002406) and was supported by the Ministry of Science and Technology in Israel.

The authors have no conflicts to disclose.

Tal Sommer and Gil Weinberg contributed equally to this work.

Tal Israel Sommer: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Methodology (equal); Writing – original draft (equal); Writing – review & editing (equal). Gil Weinberg: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Methodology (equal); Writing – review & editing (equal). Ori Katz: Funding acquisition (lead); Supervision (lead); Writing – review & editing (equal).

The data that support the findings of this study are available from the corresponding author upon reasonable request.

1.
J.
Mertz
,
Introduction to Optical Microscopy
(
Cambridge University Press
,
2019
).
2.
C. J.
Sheppard
,
M.
Castello
,
G.
Tortarolo
,
T.
Deguchi
,
S. V.
Koho
,
G.
Vicidomini
, and
A.
Diaspro
, “
Pixel reassignment in image scanning microscopy: A re-evaluation
,”
J. Opt. Soc. Am. A
37
,
154
162
(
2020
).
3.
J. G.
Fujimoto
and
D.
Farkas
,
Biomedical Optical Imaging
(
Oxford University Press
,
2009
).
4.
C. B.
Müller
and
J.
Enderlein
, “
Image scanning microscopy
,”
Phys. Rev. Lett.
104
,
198101
(
2010
).
5.
E. N.
Ward
and
R.
Pal
, “
Image scanning microscopy: An overview
,”
J. Microsc.
266
,
221
228
(
2017
).
6.
R.
Tenne
,
U.
Rossman
,
B.
Rephael
,
Y.
Israel
,
A.
Krupinski-Ptaszek
,
R.
Lapkiewicz
,
Y.
Silberberg
, and
D.
Oron
, “
Super-resolution enhancement by quantum image scanning microscopy
,”
Nat. Photonics
13
,
116
122
(
2019
).
7.
C. R.
Sheppard
, “
Super-resolution in confocal imaging
,”
Optik
80
,
53
54
(
1988
).
8.
C. J.
Sheppard
,
S. B.
Mehta
, and
R.
Heintzmann
, “
Superresolution by image scanning microscopy using pixel reassignment
,”
Opt. Lett.
38
,
2889
2892
(
2013
).
9.
S.
Roth
,
C. J.
Sheppard
,
K.
Wicker
, and
R.
Heintzmann
, “
Optical photon reassignment microscopy (OPRA)
,”
Opt. Nanosc.
2
,
5
(
2013
).
10.
P. W.
Winter
,
A. G.
York
,
D.
Dalle Nogare
,
M.
Ingaramo
,
R.
Christensen
,
A.
Chitnis
,
G. H.
Patterson
, and
H.
Shroff
, “
Two-photon instant structured illumination microscopy improves the depth penetration of super-resolution imaging in thick scattering samples
,”
Optica
1
,
181
191
(
2014
).
11.
C.
Roider
,
M.
Ritsch-Marte
, and
A.
Jesacher
, “
High-resolution confocal Raman microscopy using pixel reassignment
,”
Opt. Lett.
41
,
3825
3828
(
2016
).
12.
E.
Bo
,
L.
Wang
,
J.
Xie
,
X.
Ge
, and
L.
Liu
, “
Pixel-reassigned spectral-domain optical coherence tomography
,”
IEEE Photonics J.
10
,
17665544
(
2018
).
13.
T. B.
DuBose
,
F.
LaRocca
,
S.
Farsiu
, and
J. A.
Izatt
, “
Super-resolution retinal imaging using optically reassigned scanning laser ophthalmoscopy
,”
Nat. Photonics
13
,
257
262
(
2019
).
14.
T. I.
Sommer
and
O.
Katz
, “
Pixel-reassignment in ultrasound imaging
,”
Appl. Phys. Lett.
119
,
123701
(
2021
).
15.
D.
Raanan
,
M. S.
Song
,
W. A.
Tisdale
, and
D.
Oron
, “
Super-resolved second harmonic generation imaging by coherent image scanning microscopy
,”
Appl. Phys. Lett.
120
,
071111
(
2022
).
16.
R. N.
Bracewell
, “
Strip integration in radio astronomy
,”
Aust. J. Phys.
9
,
198
217
(
1956
).
17.
W. M.
Brown
, “
Synthetic aperture radar
,”
IEEE Trans. Aerosp. Electron. Syst.
AES-3
,
217
229
(
1967
).
18.
J. C.
Kirk
, “
A discussion of digital processing in synthetic aperture radar
,”
IEEE Trans. Aerosp. Electron. Syst.
AES-11
,
326
337
(
1975
).
19.
S.
Chowdhury
,
A.-H.
Dhalla
, and
J.
Izatt
, “
Structured oblique illumination microscopy for enhanced resolution imaging of non-fluorescent, coherently scattering samples
,”
Biomed. Opt. Express
3
,
1841
1854
(
2012
).
20.
W.
Lambert
,
L. A.
Cobus
,
M.
Couade
,
M.
Fink
, and
A.
Aubry
, “
Reflection matrix approach for quantitative imaging of scattering media
,”
Phys. Rev. X
10
,
021048
(
2020
).
21.
R. M.
Mersereau
and
A. V.
Oppenheim
, “
Digital reconstruction of multidimensional signals from their projections
,”
Proc. IEEE
62
,
1319
1338
(
1974
).
22.
S.-R.
Zhao
and
H.
Halling
, “
A new Fourier method for fan beam reconstruction
,” in
IEEE Nuclear Science Symposium and Medical Imaging Conference Record
(
IEEE
,
1995
), pp.
1287
1291
.
23.
W.
Lambert
,
L. A.
Cobus
,
T.
Frappart
,
M.
Fink
, and
A.
Aubry
, “
Distortion matrix approach for ultrasound imaging of random scattering media
,”
Proc. Natl. Acad. Sci.
117
,
14645
(
2020
).
24.
M.
Soumekh
,
Synthetic Aperture Radar Signal Processing
(
Wiley
,
New York
,
1999
), Vol.
7
.
25.
E.
Abbe
, “
Beiträge zur theorie des mikroskops und der mikroskopischen wahrnehmung
,”
Arch. Mikrosk. Anat.
9
,
413
468
(
1873
).
26.
K.
Wicker
and
R.
Heintzmann
, “
Resolving a misconception about structured illumination
,”
Nat. Photonics
8
,
342
344
(
2014
).
27.
T.
Ilovitsh
,
A.
Ilovitsh
,
J.
Foiret
,
B. Z.
Fite
, and
K. W.
Ferrara
, “
Acoustical structured illumination for super-resolution ultrasound imaging
,”
Commun. Biol.
1
,
3
(
2018
).
28.
G.
Montaldo
,
M.
Tanter
,
J.
Bercoff
,
N.
Benech
, and
M.
Fink
, “
Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography
,”
IEEE Trans. Ultrason., Ferroelectr., Freq. Control
56
,
489
506
(
2009
).

Supplementary Material