Two-photon absorption (TPA) fluorescence of biomarkers has been decisive in advancing the fields of biosensing and deep-tissue in vivo imaging of live specimens. However, due to the extremely small TPA cross section and the quadratic dependence on the input photon flux, extremely high peak-intensity pulsed lasers are imperative, which can result in significant photo- and thermal damage. Previous works on entangled TPA with spontaneous parametric downconversion light sources found a linear dependence on the input photon-pair flux, but are limited by low optical powers, along with a very broad spectrum. We report that by using a high-flux squeezed light source for TPA, a fluorescence enhancement of 47 is achieved in fluorescein biomarkers as compared to classical TPA. Moreover, a polynomial behavior of the TPA rate is observed in the the laser dye 4-dicyanomethylene-2-methyl-6-(p(dimethylamino)styryl)-4H-pyran in dimethyl sulphoxide.

Two-photon absorption (TPA) microscopy (with near-infrared illumination) is the method of choice for in vivo imaging of tissues down to millimeter depths.1 It bears several advantages including intrinsic high 3D resolution due to significant TPA occurring only in close vicinity to the focal volume, reduced out-of focus bleaching, highly reduced autofluorescence, and the capability of nearly aberration-free deep-tissue focusing along with reduced absorption.2–5 Unfortunately, classical TPA is an extremely inefficient process with absorption cross sections δr on the order of 1048cm4·s/photon.6 Therefore, TPA sensing and imaging generally require the use of high-intensity pulsed lasers, to ensure the near-simultaneous presence of two photons to induce the process.7,8 However, since the excitation pulse peak power is typically 105 times the average power, samples are prone to endure significant photo- and thermal damage.9,10

In parallel, using the unique quantum energy-time entanglement characteristics of photon pairs generated by spontaneous parametric downconversion (SPDC), the entangled TPA (ETPA) rate can be vastly enhanced,6,11–14 with the absorption cross section σe for ETPA in the range of 10181022cm2. Most notably, the linear dependence of ETPA on the input photon-pair flux, which was first predicted by Gea-Banacloche15 and Javanainen and Gould,16 was also verified experimentally.6,11,13,14,17,18 It is a major advantage over the quadratic dependence of classical TPA as the need for high intensity excitation becomes obsolete. However, most current ETPA implementations with biological specimens are limited by a low flux of 107 photon pairs /s,6,11,13,14 equivalent to only 10 pW for near-infrared wavelengths. This is mostly due to loss of correlation and difficulty of tuning biphoton wavelength in the nonlinear crystals. It is also worth mentioning that a much more efficient photon pair flux generation has been demonstrated by Jechow et al. using a type-0 quasi-phase-matched periodically poled-lithium-niobate waveguide crystal.19 Their photon-pair flux can be as high as 1011 photon pairs/s.

In this work, we utilize a different quantum light source that is based on the four-wave mixing (FWM) process in an atomic 85Rb vapor cell.20–26 The setup and the respective atomic level structure are shown in Figs. 1(a) and 1(b). The medium possesses a large third-order electric susceptibility χ(3) and is pumped by a strong narrow-band continuous-wave (CW) laser at frequency ν1 (λ = 795 nm) with a typical linewidth Δν1100 kHz. Applying an additional coherent CW seed beam at frequency νp=ν1(νHF+δ), where νHF and δ are the hyperfine splitting in the electronic ground state of 85Rb and the two-photon detuning in Fig. 1(b) (see the supplementary material for further experimental details), respectively, two pump photons are converted into a pair of twin photons, namely, “probe νp” and “conjugate νc” photons, adhering to the energy conservation 2ν1=νp+νc [see the level structure in Fig. 1(b)]. The resulting “twin beams” are strongly quantum-correlated and are also referred to as (seeded) two-mode squeezed light.27 Major advantages are narrow-band probe and conjugate beams (20 MHz)26,28 along with a freely adjustable photon-pair flux between 1013 and 1016 photons/s,20,21,23–26 which is a few orders of magnitude higher than that for SPDC. Also, since FWM in atomic vapors is an nonlinear parametric process based on ground-state coherences,29 in which the main advantage arises from small two-photon detuning from real states, whereas in nonlinear crystals, there is no real state to which the signal or idler photon is close, the generation of quantum correlations with FWM in atomic vapors can be, therefore, very efficient. As can be seen from Fig. 1(c), the twin beams exhibit an intensity-difference squeezing of 6.5 dB, which is indicative of strong quantum correlations27 (see the supplementary material for further details of the squeezing measurement).

FIG. 1.

(a) Squeezed-light TPA setup in which a seeded 85Rb cell produces strong quantum-correlated twin beams via FWM. The twin beams are focused onto the sample with a 10× objective. Fluorescence is collected at an angle of 90° with a second 10× objective and fed into a photomultiplier tube (PMT). Two short-pass filters in front of the PMT exclude stray pump photons. The setup is enclosed in a light-proof box. (b) Level structure of the D1 transition of 85Rb atoms. The optical transitions are arranged in a double-Λ configuration, where νp, νc, and ν1 stand for the probe, conjugate, and pump frequencies, respectively, fulfilling νp + νc = 2ν1. The width of the excited state in the level diagram represents the Doppler broadened line. Δ is the one-photon detuning, δ is the two-photon detuning, and νHF is the hyperfine splitting in the electronic ground state of 85Rb. (c) Measured intensity-difference noise power spectrum for the squeezed twin beams (blue line) and for the standard quantum limit (red line), obtained using a radio frequency spectrum analyzer (with a resolution and video bandwidth of 300 kHz and 100 Hz, respectively). A squeezing of 6.5 dB is achieved. (d) Molecular structures of fluorescein and DCM.

FIG. 1.

(a) Squeezed-light TPA setup in which a seeded 85Rb cell produces strong quantum-correlated twin beams via FWM. The twin beams are focused onto the sample with a 10× objective. Fluorescence is collected at an angle of 90° with a second 10× objective and fed into a photomultiplier tube (PMT). Two short-pass filters in front of the PMT exclude stray pump photons. The setup is enclosed in a light-proof box. (b) Level structure of the D1 transition of 85Rb atoms. The optical transitions are arranged in a double-Λ configuration, where νp, νc, and ν1 stand for the probe, conjugate, and pump frequencies, respectively, fulfilling νp + νc = 2ν1. The width of the excited state in the level diagram represents the Doppler broadened line. Δ is the one-photon detuning, δ is the two-photon detuning, and νHF is the hyperfine splitting in the electronic ground state of 85Rb. (c) Measured intensity-difference noise power spectrum for the squeezed twin beams (blue line) and for the standard quantum limit (red line), obtained using a radio frequency spectrum analyzer (with a resolution and video bandwidth of 300 kHz and 100 Hz, respectively). A squeezing of 6.5 dB is achieved. (d) Molecular structures of fluorescein and DCM.

Close modal

In this study, we analyze and compare classical TPA and squeezed-light induced TPA (SL-TPA) fluorescence rates in fluorescein and in he laser dye 4-dicyanomethylene-2-methyl-6-(p(dimethylamino)styryl)-4H-pyran (DCM) (see the supplementary material for sample preparation). Fluorescein is one of the most frequently used biomarkers for bioimaging and biosensing.30 Its small size is very convenient for in vivo imaging applications although its relatively small classical TPA cross section generates low amounts of fluorescence.7,8 The SL-TPA setup is depicted in Fig. 1(a). A 10× objective (Thorlabs RMS10X) focuses the near-infrared excitation light onto a solution of fluorophores. Following TPA, fluorescence is collected by a second 10× objective (Thorlabs RMS10X) at an angle of 90° and guided to a photo-multiplier tube (PMT) (Thorlabs PMTSS in conjugation with a PMT transimpedance amplifier Thorlabs TIA60). Two optical low-pass filters (Thorlabs FESH0750) exclude stray pump photons. The measured PMT voltage outputs (see the inset in Fig. 2) are converted into fluorescence rates of arbitrary units (since the PMT's response to an input photon is an inverse voltage pulse, adding all the inverse voltages in a given time window can, therefore, give us a quantity that is proportional to the input photon flux up to a conversion factor, see the supplementary material for further data acquisition details). For classical TPA measurements, only the coherent pump beam is focused into the microscope objective, with the same focus spot size at the sample. Utilized powers for the twin beams were ranging from 30μW to the maximum of 8 mW.

FIG. 2.

Fluorescence rates vs excitation power. The red diamonds and the red (dashed-dotted) line show the measured values for the coherent excitation and the respective quadratic fit. The green stars are the rates for SL-TPA induced by the twin-beam excitations, and the green (dashed-dotted) line is the respective linear fit. Error bars denoting one standard deviation. Inset: raw voltage output from the PMT for coherent light (red) and squeezed light (green) excitations of 8 mW optical power. The shaded area for each curve represents one standard deviation.

FIG. 2.

Fluorescence rates vs excitation power. The red diamonds and the red (dashed-dotted) line show the measured values for the coherent excitation and the respective quadratic fit. The green stars are the rates for SL-TPA induced by the twin-beam excitations, and the green (dashed-dotted) line is the respective linear fit. Error bars denoting one standard deviation. Inset: raw voltage output from the PMT for coherent light (red) and squeezed light (green) excitations of 8 mW optical power. The shaded area for each curve represents one standard deviation.

Close modal

Measured classical TPA fluorescence rates for fluorescein, as a function of the input power, are shown by the red diamonds in Fig. 2, with error bars denoting one standard deviation. The observed quadratic power law relationship agrees well with the established literature for classical TPA, where the fluorescence signal is proportional to the square of excitation light intensity.31 Fluorescence rates induced by SL-TPA with an optical power of 8 mW (3.5 mW + 1.0 mW probe and seed; 3.5 mW conjugate) and an optical power of 4 mW (1.75 mW + 0.5 mW probe and seed; 1.75 mW conjugate) together with an optical power of 6 mW (2.6 mW + 0.8 mW probe and seed; 2.6 mW conjugate) and an optical power of 3 mW (1.3 mW + 0.4 mW probe and seed; 1.3 mW conjugate) are depicted by the green stars (although the 3 mW and 6 mW data points were taken on a different day, the trend is similar). Due to experimental constraints, 8 mW is the maximal power we are able to acquire for the squeezed light. The measured coherent rates are fitted by the quadratic function R(I)=I2×1.5mW2 (dashed-dotted red line), which represents the benchmark of the true fluorescence rate as a function of the input power. It can be observed from the figure that the signal for 8 mW of coherent excitation deviates the strongest from the fit. This fact can be attributed to background noise (e.g., electronic dark counts and spurious counts from stray ambient light) and the overall low signal to noise ratio (SNR) of the measurement (characterized by a standard deviation encompassing negative values). Following the fitted curve, the fluorescence rate for 8 mW coherent excitation is, thus, merely 9.6×101 (a.u.). For SL-TPA of 8 mW excitation power, the fluorescence rate is 4.46×103 (a.u.). This value corresponds to a striking 46.5-fold enhancement over the coherent excitation. Vice versa, increasing the coherent excitation power sevenfold to 55 mW, and thus the classical TPA rate by 47.3-fold, the measured rates for 8 mW SL-TPA and 55 mW classical TPA match, thus confirming the previous statement. Subtracting the contribution from the 1 mW coherent seed beam power, it can be argued that the measured SL-TPA enhancement is around 60-fold over 7 mW coherent excitation. Note that the seed is uncorrelated to the quantum correlated photon pairs and that 1 mW of coherent seed excitation itself will induce negligible classical TPA rates. More importantly, the measured fluorescence rate for 4 mW SL-TPA is 2.02×103 (a.u.). Subtracting the respective optical power of the seed beam, we end up with the input flux ratio of 7/3.5=2.00, which matches the measured ratio 4.46/2.02=2.21 well (within the calculated uncertainties). This is also true for the SL-TPA of 6 mW and 3 mW, which is indicative of the linear dependence on the input photon-pair flux that is expected for fluorescein in this regime. Quadratic terms, thus, do not seem to contribute here.

TPA is highly sensitive to the near-instantaneous arrival of two photons at the sample, in particular, with respect to the virtual state lifetime of the intermediate states used for the electronic transition from the ground to the final state.13,18 In ETPA, this is quantified by the entanglement time Te.6,13,32 Adjusting Te should change the measured SL-TPA enhancement. Hence, an investigation of the effect of relative temporal delay between the entangled photon pairs is conducted. The ETPA cross section σe is inversely proportional to Te and, thus, the mean group velocity delay between the entangled photon pairs. In the FWM process of the atomic 85Rb vapor, the group delay between the entangled pairs can be adjusted by changing the two-photon detuning δ of the double-Λ configuration in Fig. 1(b).28 The red, green, and blue bars in Fig. 3 show the fluorescence rates induced by 8 mW SL-TPA for the values δ = –10 MHz, –5 MHz, and 0 MHz, respectively. For each δ value, the same relative intensity-difference squeezing of 6.5 dB [see Fig. 1(c)] is maintained, such that the results are not affected by different entanglement levels. For δ=5 MHz, a relatively small delay is achieved.28 Degraded fluorescence rates for δ=10 MHz and δ = 0 MHz confirm that the SL-TPA enhancement is degraded when the relative delay between the photon pairs is tuned away from its optimal value. Further experimental details on how to change the two-photon detuning δ can be found in the supplementary material.

FIG. 3.

Fluorescence rates for 8 mW SL-TPA with three different two-photon detunings δ, shown in the atomic level structure in Fig. 1(b) as δ. Red, green, and blue bars are for δ = –10 MHz, –5 MHz, and 0 MHz, respectively. This figure demonstrates degraded SL-TPA enhancements as a function of the relative arrival times of the entangled photon pairs.

FIG. 3.

Fluorescence rates for 8 mW SL-TPA with three different two-photon detunings δ, shown in the atomic level structure in Fig. 1(b) as δ. Red, green, and blue bars are for δ = –10 MHz, –5 MHz, and 0 MHz, respectively. This figure demonstrates degraded SL-TPA enhancements as a function of the relative arrival times of the entangled photon pairs.

Close modal

In general, the ETPA rate Re as a function of the input photon-pair flux density ϕ is expected to follow the functional behavior Re(ϕ)=σeϕ+δrϕ2, where σe and δr are the cross sections for ETPA and classical TPA, respectively, and are determined by the electronic level structure of the system undergoing TPA.12,17,18,33 Both values can be determined experimentally or calculated theoretically via second-order perturbation theory for a sufficiently simple system.18 As previously established, the coincident arrival and absorption of an entangled photon pair lead to the linear dependence Re(ϕ)=σeϕ, provided that ϕ is sufficiently small.6,13,14 For sufficiently high photon-pair fluxes, TPA can be induced by uncorrelated photons from different pairs as well. The respective rate is equivalent to the classical TPA rate δrϕ2. Parity between both contributions is reached at the critical flux ϕc=σe/δr. Most previous ETPA experiments with biomarkers and low photon-pair fluxes observed only the linear dependence.6,13,14

With means of investigating SL-TPA with high and freely adjustable optical powers, we investigated its functional behavior in DCM laser dye (see the supplementary material for its preparation). DCM dyes are known for strong TPA around an excitation wavelength of 800 nm, and along with a milli molar suspension, optical powers in the μW regime are sufficient to induce enough TPA fluorescence.34,35 The measured coherent TPA rates shown as red squares in Fig. 4 agree well with a quadratic behavior, represented by the fit function R(I)=I2×0.304μW2. For SL-TPA in DCM, on the other hand, we observed a non-linear behavior of the functional form R(I)=I×7.9μW1+I2×0.59μW2. The polynomial behavior of SL-TPA in Fig. 4 implies that the photon-pair flux is already high enough to observe both linear and quadratic contributions. In fact, given the fit values, parity is already reached at Ic=7.9μW1/0.304μW2=26.0μW for the DCM solution.

FIG. 4.

Fluorescence signal of DCM vs excitation power of coherent light (red) and squeezed light (green). Coherent light fitting curve obeys a quadratic behavior, while the squeezed light fitting curve shows a polynomial behavior, indicating a high input of entangled photon-pair flux.

FIG. 4.

Fluorescence signal of DCM vs excitation power of coherent light (red) and squeezed light (green). Coherent light fitting curve obeys a quadratic behavior, while the squeezed light fitting curve shows a polynomial behavior, indicating a high input of entangled photon-pair flux.

Close modal

It is worth pointing out that the DCM laser dye solution requires much lower excitation powers than the fluorescein solution to produce appreciable TPA fluorescence rates, most probably due to a larger classical TPA cross section δr. In Fig. 4, the DCM signal at 130 μW coherent excitation (0.45×104 a.u.) roughly equals the fluorescein signal at 55 mW coherent excitation (4.43×103 a.u.). Taking into account the concentration of the two solutions (see the supplementary material for the details of sample preparation),36 we estimate that the classical TPA cross section δr of DCM is roughly 1800 times larger than that of fluorescein. However, the ETPA cross section σe of DCM is actually smaller than that of fluorescein, as demonstrated by the relatively small SL-TPA enhancements. The difference can be attributed to different electronic level structures of these two organic molecules.6 

In conclusion, this work investigates two-photon absorption fluorescence rates in fluorescein biomarkers and in DCM laser dye, induced by a coherent CW excitation light and by the bright two-mode squeezed light. For the coherent CW excitation, both fluorophores show the well-expected quadratic dependence on the input photon flux. The experimental results for fluorescein with SL-TPA, however, demonstrate a linear dependence on the input optical power, along with an ∼47-fold TPA fluorescence enhancement for 8 mW squeezed light compared to 8 mW coherent light. This can be attributed to the predominant occurrence of entangled two-photon absorption of quantum-correlated photon pairs. In addition and different from previous works using quantum states of light for TPA in fluorophores, we report that SL-TPA in DCM laser dye is governed by a polynomial behavior, which can be entirely attributed to its far greater entangled photon-pair flux, as compared to using SPDC sources. Thus, this work demonstrates that our FWM based bright two-mode squeezed light source can achieve ultra-low intensity TPA for biosensing and bioimaging and, thus, bear the potential to open up unconventional avenues for in vivo deep tissue studies of biological specimens via TPA.

See the supplementary material for complete experimental details, squeezing measurements, data acquisition, and sample preparation.

This work was supported by the Air Force Office of Scientific Research (Award No. FA-9550-18-1-0141) and the Robert A. Welch Foundation (Award No. A-1943-20180324).

We thank V. Yakovlev for discussions on spectroscopy with entangled photon pairs and for the suggestion of fluorescein biomarkers. We also thank A. Sokolov and A. Zheltikov for informative comments. A.C. acknowledges support from the Alexander von Humboldt Foundation in the framework of a Feodor Lynen Research Fellowship. F. L. acknowledges support from the Herman F. Heep and Minnie Belle Heep Texas A\&M University Endowed Fund held/administered by the Texas A\&M Foundation.

The data that support the findings of this study are available from the corresponding author upon reasonable request.

1.
F.
Helmchen
and
W.
Denk
, “
Deep tissue two-photon microscopy
,”
Nat. Methods
2
,
932
(
2005
).
2.
J. B.
Pawley
, ed.,
Handbook of Biological Confocal Microscopy
(
Plenum
,
New York
,
1995
).
3.
C.
Xu
,
W.
Zipfel
,
J. B.
Shear
,
R. M.
Williams
, and
W. W.
Webb
, “
Multiphoton fluorescence excitation: New spectral windows for biological nonlinear microscopy
,”
Proc. Natl. Acad. Sci. U. S. A.
93
,
10763
10768
(
1996
).
4.
W. R.
Zipfel
,
R. M.
Williams
, and
W. W.
Webb
, “
Nonlinear magic: Multiphoton microscopy in the biosciences
,”
Nat. Biotechnol.
21
,
1369
1377
(
2003
).
5.
M.
Drobizhev
,
N. S.
Makarov
,
S. E.
Tillo
,
T. E.
Hughes
, and
A.
Rebane
, “
Two-photon absorption properties of fluorescent proteins
,”
Nat. Methods
8
,
393
399
(
2011
).
6.
L.
Upton
,
M.
Harpham
,
O.
Suzer
,
M.
Richter
,
S.
Mukamel
, and
T.
Goodson
, “
Optically excited entangled states in organic molecules illuminate the dark
,”
J. Phys. Chem. Lett.
4
,
2046
2052
(
2013
).
7.
C.
Xu
and
W. W.
Webb
, “
Measurement of two-photon excitation cross sections of molecular fluorophores with data from 690 to 1050 nm
,”
J. Opt. Soc. Am. B
13
,
481
491
(
1996
).
8.
M. A.
Albota
,
C.
Xu
, and
W. W.
Webb
, “
Two-photon fluorescence excitation cross sections of biomolecular probes from 690 to 960 nm
,”
Appl. Opt.
37
,
7352
7356
(
1998
).
9.
M. A.
Taylor
and
W. P.
Bowen
, “
Quantum metrology and its application in biology
,”
Phys. Rep.
615
,
1
59
(
2016
).
10.
S. N.
Arkhipov
,
I.
Saytashev
, and
M.
Dantus
, “
Intravital imaging study on photodamage produced by femtosecond near-infrared laser pulses in vivo
,”
Photochem. Photobiol.
92
,
308
313
(
2016
).
11.
J. P.
Villabona-Monsalve
,
O.
Varnavski
,
B. A.
Palfey
, and
T.
Goodson
, “
Two-photon excitation of flavins and flavoproteins with classical and quantum light
,”
J. Am. Chem. Soc.
140
,
14562
14566
(
2018
).
12.
F.
Schlawin
,
K. E.
Dorfman
, and
S.
Mukamel
, “
Entangled two-photon absorption spectroscopy
,”
Acc. Chem. Res.
51
,
2207
2214
(
2018
).
13.
O.
Varnavski
,
B.
Pinsky
, and
T.
Goodson
, “
Entangled photon excited fluorescence in organic materials: An ultrafast coincidence detector
,”
J. Phys. Chem. Lett.
8
,
388
393
(
2017
).
14.
J. P.
Villabona-Monsalve
,
O.
Calderón-Losada
,
M. N.
Portela
, and
A.
Valencia
, “
Entangled two photon absorption cross section on the 808 nm region for the common dyes zinc tetraphenylporphyrin and Rhodamine B
,”
J. Phys. Chem. A
121
,
7869
7875
(
2017
).
15.
J.
Gea-Banacloche
, “
Two-photon absorption of nonclassical light
,”
Phys. Rev. Lett.
62
,
1603
1606
(
1989
).
16.
J.
Javanainen
and
P. L.
Gould
, “
Linear intensity dependence of a two-photon transition rate
,”
Phys. Rev. A
41
,
5088
5091
(
1990
).
17.
N. P.
Georgiades
,
E. S.
Polzik
,
K.
Edamatsu
,
H. J.
Kimble
, and
A. S.
Parkins
, “
Nonclassical excitation for atoms in a squeezed vacuum
,”
Phys. Rev. Lett.
75
,
3426
3429
(
1995
).
18.
H.-B.
Fei
,
B. M.
Jost
,
S.
Popescu
,
B. E. A.
Saleh
, and
M. C.
Teich
, “
Entanglement-induced two-photon transparency
,”
Phys. Rev. Lett.
78
,
1679
1682
(
1997
).
19.
A.
Jechow
,
A.
Heuer
, and
R.
Menzel
, “
High brightness, tunable biphoton source at 976 nm for quantum spectroscopy
,”
Opt. Express
16
,
13439
13449
(
2008
).
20.
M.
Dowran
,
A.
Kumar
,
B. J.
Lawrie
,
R. C.
Pooser
, and
A. M.
Marino
, “
Quantum-enhanced plasmonic sensing
,”
Optica
5
,
628
633
(
2018
).
21.
L.
Cao
,
J.
Qi
,
J.
Du
, and
J.
Jing
, “
Experimental generation of quadruple quantum-correlated beams from hot rubidium vapor by cascaded four-wave mixing using spatial multiplexing
,”
Phys. Rev. A
95
,
023803
(
2017
).
22.
F.
Hudelist
,
J.
Kong
,
C.
Liu
,
J.
Jing
,
Z. Y.
Ou
, and
W.
Zhang
, “
Quantum metrology with parametric amplifier-based photon correlation interferometers
,”
Nat. Commun.
5
,
3049
(
2014
).
23.
B. E.
Anderson
,
P.
Gupta
,
B. L.
Schmittberger
,
T.
Horrom
,
C.
Hermann-Avigliano
,
K. M.
Jones
, and
P. D.
Lett
, “
Phase sensing beyond the standard quantum limit with a variation on the Su(1,1) interferometer
,”
Optica
4
,
752
756
(
2017
).
24.
T.
Li
,
B. E.
Anderson
,
T.
Horrom
,
B. L.
Schmittberger
,
K. M.
Jones
, and
P. D.
Lett
, “
Improved measurement of two-mode quantum correlations using a phase-sensitive amplifier
,”
Opt. Express
25
,
21301
21311
(
2017
).
25.
R. C.
Pooser
and
B.
Lawrie
, “
Ultrasensitive measurement of microcantilever displacement below the shot-noise limit
,”
Optica
2
,
393
399
(
2015
).
26.
J. B.
Clark
,
R. T.
Glasser
,
Q.
Glorieux
,
U.
Vogl
,
T.
Li
,
K. M.
Jones
, and
P. D.
Lett
, “
Quantum mutual information of an entangled state propagating through a fast-light medium
,”
Nat. Photonics
8
,
515
(
2014
).
27.
C. F.
McCormick
,
A. M.
Marino
,
V.
Boyer
, and
P. D.
Lett
, “
Strong low-frequency quantum correlations from a four-wave-mixing amplifier
,”
Phys. Rev. A
78
,
043816
(
2008
).
28.
R. T.
Glasser
,
U.
Vogl
, and
P. D.
Lett
, “
Stimulated generation of superluminal light pulses via four-wave mixing
,”
Phys. Rev. Lett.
108
,
173902
(
2012
).
29.
C. F.
McCormick
,
V.
Boyer
,
E.
Arimondo
, and
P. D.
Lett
, “
Strong relative intensity squeezing by four-wave mixing in rubidium vapor
,”
Opt. Lett.
32
,
178
180
(
2007
).
30.
T.
Robertson
,
F.
Bunel
, and
M.
Roberts
, “
Fluorescein derivatives in intravital fluorescence imaging
,”
Cells
2
,
591
(
2013
).
31.
B. R.
Mollow
, “
Two-photon absorption and field correlation functions
,”
Phys. Rev.
175
,
1555
1563
(
1968
).
32.
R. D. J.
León-Montiel
,
J. C. V.
Svozilík
,
J. P.
Torres
, and
A. B.
U'Ren
, “
Temperature-controlled entangled-photon absorption spectroscopy
,”
Phys. Rev. Lett.
123
,
023601
(
2019
).
33.
M. R.
Harpham
,
Ö.
Süzer
,
C.-Q.
Ma
,
P.
Bäuerle
, and
T.
Goodson
, “
Thiophene dendrimers as entangled photon sensor materials
,”
J. Am. Chem. Soc.
131
,
973
979
(
2009
).
34.
A.
Jechow
,
M.
Seefeldt
,
H.
Kurzke
,
A.
Heuer
, and
R.
Menzel
, “
Enhanced two-photon excited fluorescence from imaging agents using true thermal light
,”
Nat. Photonics
7
,
973
(
2013
).
35.
T.
Scheul
,
C.
D'Amico
,
I.
Wang
, and
J.-C.
Vial
, “
Two-photon excitation and stimulated emission depletion by a single wavelength
,”
Opt. Express
19
,
18036
18048
(
2011
).
36.
D.
Tabakaev
,
M.
Montagnese
,
G.
Haack
,
L.
Bonacina
,
J.-P.
Wolf
,
H.
Zbinden
, and
R.
Thew
, “
Energy-time entangled two-photon molecular absorption
,” arXiv:1910.07346 (
2020
).

Supplementary Material