Diagnosing a stroke at an emergency site is an important potential clinical application for near-infrared light techniques. We have fabricated a hybrid diffuse optical device that combines diffuse optical spectroscopy and diffuse correlation spectroscopy (DCS) to monitor the total hemoglobin concentration (HbT), tissue oxygen saturation (StO2), and blood flow index (BFI) of a patient. The device used hybrid light sources consisting of three incoherent laser diodes (LDs) and one long coherent LD. Single-photon-counting avalanche photodiodes detected light reemitted from the tissue. Arm cuff experiments were performed using a healthy subject to validate device capabilities. Changes in hemodynamic parameters including the HbT, StO2, and BFI were measured continuously during the arm cuff experiments. The StO2 decreased linearly from 63% to 51% during the 3 min occlusion period. The blood flow decreased immediately by 16% when the occlusion was placed on the forearm. Overshot increases in the HbT, StO2, and BFI were all observed after occlusion release. Stroke is related to abnormal hemodynamic changes in cerebral vasculature and our device is sensitive to such changes. Thus, we believe that the hybrid device developed here may have potential clinical applications for stroke diagnosis at emergency sites.

Stroke is one of the most common causes of death and disability and results in enormous societal costs associated with rehabilitation, long-term care, and workforce loss.1 Approximately 90% of all strokes occur due to cerebral ischemia, with the remainder due to cerebral hemorrhage.2 The only approved treatment for ischemic stroke is recanalization of occluded arteries via thrombolysis with alteplase within the first hours of symptom onset.3–5 However, implementation of recanalizing therapy within this narrow therapeutic window is difficult in clinical practice because neurological examination and imaging are needed to exclude hemorrhagic stroke. In clinical reality, this treatment can be given to only a small minority (2%–5%) of stroke patients.6,7 The establishment of mobile stroke units based on x-ray computed tomography scanning in Germany and the United States has made earlier identification and treatment of ischemic stroke a reality and has the potential to substantially improve outcomes.8,9 However, compact, less expensive devices are needed for stroke diagnosis at emergency sites in order to increase the number of patients who benefit from effective treatment.

Near-infrared light systems (NIRS) may be the ideal method of overcoming the aforementioned clinical dilemma. They are sensitive to hemodynamic parameters, safe, noninvasive, cheap, and portable. There are two basic requirements for the near-infrared light techniques that can achieve the required goals. First, the techniques should output quantitative parameters that help to differentiate ischemic and hemorrhagic tissues. One common application of NIRS is measuring brain functional activity.10 However, these systems measure relative hemodynamic changes using the modified Beer-Lambert Law method, which is not suitable for diagnosing stroke. Commercial cerebral oximeters that use spatially resolved spectroscopy (SRS)11,12 can measure the absolute tissue oxygen saturation (StO2). However, their ability to measure hemodynamic parameters is limited. The frequency and time-domain methods13,14 are ideal ways of obtaining the absolute total hemoglobin concentration (HbT) and tissue oxygen saturation, which are useful in stroke diagnosis. However, these devices are complex. Second, the technique used should measure as many parameters as possible in order to provide full hemodynamic stroke information. Diffuse correlation spectroscopy (DCS) has recently been used to measure blood flow.15 DCS is a dynamic technique that analyzes rapid temporal fluctuations in light scattered by moving red blood cells and uses a diffuse correlation approach to derive local microvascular tissue blood flow. A research group from the University of Pennsylvania has combined near-infrared diffuse optical spectroscopy (NIRS-DOS) with DCS to measure tissue oxygen saturation and blood flow.16,17 However, they used separate NIRS-DOS and DCS systems with separate sources and detectors. Yu Shang et al. have proposed adding a second long coherent continuous wave laser diode to measure tissue oxygen saturation and blood flow simultaneously.18 D. Tamborini et al. have proposed using multi-color, multi-distance DCS to fit the absolute optical properties and blood flow index (BFI) using diffuse optical theory and diffuse correlation theory and then validate it using a tissue-like phantom.19 

In this study, we fabricated a hybrid diffuse optical device that combines NIR-DOS and DCS. The device used hybrid light sources consisting of three incoherent laser diode (LD) sources (690 nm, 808 nm, and 830 nm) and one long coherent LD source (785 nm). Single-photon-counting avalanche photodiodes (APDs) detected light reemitted from the tissue. Optical switches were used to switch the four laser sources. A software correlator was utilized to calculate the optical intensity and the normalized correlation value of the optical intensity simultaneously for the four wavelengths.20 The absolute HbT and StO2 were calculated using the decays of the optical intensities of the four wavelengths based on a multi-color SRS method.11,12 The absolute BFI was derived from the normalized autocorrelation value of the optical intensity of a long, coherent laser source. Thus, this system allowed the HbT, StO2, and BFI to be obtained. The entire system was compact, cheap, and easy to manufacture. Arm cuff experiments were conducted using a healthy subject to validate the device capabilities. Changes in hemodynamic parameters including the HbT, StO2, and BFI of the healthy subject were measured continuously. The results met expectations set by a previous study.20,21 Since this device was able to measure the HbT, StO2, and BFI, we believe that it may have potential clinical applications for stroke diagnosis at emergency sites.

A schematic of the hybrid diffuse optical device is shown in Fig. 1. The device combined DCS and NIRS-DOS. The DCS portion consisted of a mode-hope free, long coherence-length, continuous-wave laser (785 nm, 90 mW, DL785-100-3O, CrystaLaser Inc., USA) and four single-photon avalanche photodiode detectors (SPCM-AQRH-W4, Excelitas, Canada) whose outputs were fed to a custom-built software autocorrelator.20 The software correlator sampled the photon count and calculated both the optical intensity and its normalized autocorrelation value. The continuous-wave NIRS-DOS module consisted of three continuous-wave lasers (Wavespectrum Laser, China) at 690 nm, 808 nm, and 830 nm with optical power levels of 50 mW, 100 mW, and 100 mW, respectively. The BFI was derived from the normalized autocorrelation value at 785 nm. The normalized autocorrelation values of the other wavelengths were abandoned. The HbT and StO2 were derived from the decays of the optical intensities of all four wavelengths. Optical switches were used to switch the four measurement wavelengths (one for DCS and three for NIRS-DOS). The data acquisition frequency was 20Hz and the optical switch frequency was 0.2 Hz. Each measurement cycle required 20 s.

FIG. 1.

The device schematic.

FIG. 1.

The device schematic.

Close modal

The only hardware in the software correlator was a set of counters (PCIe-6612, NI, USA). A shielded I/O connector block (SCB-68, NI, USA) was used to connect the PCIe-6612 board to the outputs of the single-photon-counting APD. The four counters on the PCIe-6612 board were synchronized by the same sampling clock (1/Δt) at up to 5 MHz. The lag times were M=09N=078+N2MΔt and M=00N=07N2MΔt. Since the correlation bin width was always the same as the sampling clock and only the lag time increased in line with a multi-tau scheme,22 the algorithm was quite simple. The optical intensities and normalized autocorrelation values of the four channels were calculated in real-time using a computer with an integration time of 0.05 s (20 Hz).

Optical probes consisting of a multi-mode source fiber (62.5 μM/0.27 NA, Corning, USA) and three single-mode detection fibers (4.4 μM/0.13 NA, 780-HP, Nufern, USA) with source-detector separations of 1.08 cm, 1.95 cm, and 3 cm were custom built.

Arm cuff experiments were conducted to test device performance. The subject was asked to sit comfortably in a chair. Their forearm was put on their leg. A probe with embedded sources and detection fibers was secured on the forearm of the subject. An arm cuff was placed on the bicep on the same side of the probe. Black cloth was used to protect the probe and forearm from external light. After 3 min of baseline condition measurements, the blood in the arm was blocked for 3 min by inflating the blood pressure cuff to 200 mmHg. The experiment also collected post-occlusion data for 3 min. The in vivo experiment was approved by the Institutional Review Board of the Arm Medical University.

The diffuse mode of light transportation in tissue used here has semi-infinite geometry, which was suitable for the forearm experiments. Using the diffuse theory, the effective attenuation coefficient μeff is,
(1)
where μa is the absorption coefficient, μs′ is the reduced scattering coefficient, ρ is the separation between the source and detection fibers, I(ρ) is the optical intensity at ρ, and Δ is the differential operator.
Oxyhemoglobin (HbO), deoxy-hemoglobin (Hb) and water are the main chromophores in tissue for near-infrared light absorption. The total concentration of hemoglobin (CHbT=CHbO+CHb) and tissue oxygen saturation (StO2= CHbO/CHbT) can be calculated from the concentrations of HbO and Hb. Therefore, the absorption coefficient in tissue depends linearly on the chromophore concentration as
(2)
where εHbO,λi, εHbO,λi, and εwater,λi are the wavelength-dependent oxy-hemoglobin, deoxy-hemoglobin, and water extinction coefficients, as obtained from an online reference.23  λi (i=1, 2, 3, or 4) is the wavelength. γwater is the assumed percentage of water in tissue, which was always 14%24 in our study.
The reduced scattering coefficient in tissue follows an empirical power law relationship:
(3)
where b is the scaling factor and a is the scattering power, both of which are independent of λ. The unit of wavelength is nanometers. Thus, we can produce Eq. (4)
(4)
We obtained the μeff by measuring the light intensity at each separation for each wavelength. For our device, there were four wavelengths. Thus, we generated four equations that described μeff. Since the equations are nonlinear functions, we solved them via optimization. Global fitting was performed to fit experimental data over λ to minimize the cost function (χ2) and fit for HbT, StO2, a, and b,
(5)
DCS measures temporal speckle fluctuations due to motion of scattering points (red blood cells) in tissue, and can be used to estimate a microvasculature blood flow index. The dynamic motion of the medium can be determined by measuring the autocorrelation function, as faster motion of the scattering points is indicated by faster speckle fluctuations (i.e., more rapid decay of the autocorrelation function). The Green’s function solution of the diffuse correlation equation for semi-infinite boundary conditions is
(6)
where

zb=2/μs1+Reff/1Reff, Reff is the effective reflection coefficient that accounts for the index mismatch between tissue and air, k0 is the wave number of light in the medium, λ is the wavelength of light, τ is the lag time, ρ is the source-detector separation, BFI is the quantitative blood flow index, and μa and μs’ are, respectively, the absorption and reduced scattering coefficients.

DCS measures the normalized autocorrelation function (g2=<I*I(τ)>/<I>2), while the diffuse correlation equation applies to the electric field autocorrelation function. To fit the theory to the experimental data, the normalized intensity autocorrelation function must be related to the normalized electric field temporal autocorrelation (g1=G1*G1(τ)/<G1*G1>) through the Siegert relation,
(7)
where β is a constant determined primarily by the optics of the experiment and is related to the number of modes in the detected light. In most DCS experiments that use coherent, non-polarized sources and single-mode detection fibers, β is approximately 0.5.
The normalized autocorrelation of the optical intensity and the BFI derived from it did not require calibration, since they do not depend on the absolute optical intensity. Calibration was required for the effective attenuation coefficient (μeff) measurement since the detector sensitivities, detection fiber coupling coefficient, and source-detector separation measurement accuracy all affect the value of μeff. The inaccuracy associated with source-detector separation measurement was neglected in this study since it was both small and fixed. The sensitivity of an individual detector and the coupling coefficient of an individual probe make the measured optical intensity Im
(8)
where κ is the correction coefficient and Iideal is the perfect detected optical intensity without the influences of the detector sensitivity and detection fiber coupling coefficient. Upon substituting Eq. (8) into Eq. (1), we get
(9)
where C is the deviation between the ideal and measured μeff values. Intralipid phantom titration-translation experiments described in a previous study22 were conducted for the four wavelengths. The expected μa and μs’ were 0.24 cm-1 and 4.27 cm-1 at 785 nm, respectively. Since the same detector and same probe were used, the μeff could be thought of as the ideal μeff. The custom probe used for the forearm experiment was also immersed in the same intralipid phantom to obtain the measured μeff. Finally, we can determine the deviation of μeff for the four wavelengths,
(10)

The arm cuff experiment optical intensity changes at 1.08 cm, 1.95 cm, and 3 cm for the four wavelengths are shown in Fig. 2. The vertical black lines in the figure indicate when the arm cuff is inflated and deflated. The optical intensity is averaged over 1 s. For the baseline period, the optical intensities are stable and are indicated in the figure. The optical intensities all decrease and then recover during the arm occlusion period. The optical intensities stabilize after arm cuff deflation.

FIG. 2.

Optical intensity changes for four wavelengths during the arm cuff experiment.

FIG. 2.

Optical intensity changes for four wavelengths during the arm cuff experiment.

Close modal

We used the optical intensity decay over different separations to obtain the μeff, as shown in Fig. 3. The change in μeff at 690 nm was much bigger than those of the other wavelengths, since the μeff at 690 nm is more sensitive to the Hb. Note, changes in μeff and changes in the optical intensities are not consistent with each other. For example the optical intensity at 830 nm is lower during the occlusion period than during the baseline period. This indicates that attenuation increases but the actual μeff at 830 nm is lower during the occlusion period than the baseline period. Since μeff removes the same fiber coupling influence at all distances and is robust against movement artifacts, we believe that it should be used instead of the change in optical intensity to calculate the hemodynamic parameters. This is because the single-mode detection fiber used here is susceptible to movement artifacts due to its small diameter of only 4.4 μm.

FIG. 3.

Changes in the effective attenuation coefficients of the four wavelengths during the experiment.

FIG. 3.

Changes in the effective attenuation coefficients of the four wavelengths during the experiment.

Close modal

Equation (4) is non-linear equation, so we use the optimization method to solve it. The fraction of water is fixed at 14% and does not change between experiments and subjects. The initial HbT, StO2, a, and b are assumed to be 0.126 mmol/L, 67%, 0.576, and 20 respectively.24 We determine the absolute HbT and StO2 using the ‘fminsearch’ function in Matlab. For this subject, the HbT and StO2 are 0.08 mmol/L and 63%, respectively. We assume that scattering is constant and solve the linear equation using constant a and b values to identify the relative changes in HbT and StO2, as shown in Fig. 4. The HbT is nearly constant and the StO2 decreases linearly from 63% to 51% during the 3 min arm occlusion period. After occlusion, overshot increases in the HbT and StO2 occur. This is reasonable because of the overshot perfusion of blood.

FIG. 4.

HbT and StO2 changes during an arm cuff experiment.

FIG. 4.

HbT and StO2 changes during an arm cuff experiment.

Close modal

Figure 5(a) shows the normalized autocorrelation value of the optical intensity at 785 nm. The low value of β at 1.08 cm occurs because the high detected optical intensity saturates the detector.22 Obviously, the correlation curve delay is much slower during occlusion than under normal conditions. Fig. 5(b) shows the BFI as determined from the correlation values based on diffuse correlation theory with calculated μa (0.15cm-1) and μs’ (4.93cm-1) values. At the three distances noted, the BFI decreases by 24%, 15%, and 10% immediately upon occlusion of the arm. This represents a mean 16% decrease. This is reasonable since blood cell movement slows immediately. After the cuff is released, the BFI experiences an overshot increase.

FIG. 5.

(A) Comparison of the normalized autocorrelation optical intensity curves for normal and occlusion tissue. (B) Blood flow changes during the arm cuff experiment.

FIG. 5.

(A) Comparison of the normalized autocorrelation optical intensity curves for normal and occlusion tissue. (B) Blood flow changes during the arm cuff experiment.

Close modal

Unlike a traditional CW-NIR or NIR-DOS device, our device can measure blood flow directly by adding a long, coherent near-infrared laser source. The relative BFI has already been shown to indicate relative changes in blood flow.15,25 We have developed a hybrid diffuse optical device that combines NIRS-DOS with DCS to diagnose strokes at an emergency site. The HbT and StO2 are derived using the multi-color SRS method. The BFI is determined from the normalized autocorrelation value of the optical intensity of a long coherent laser source. The single-photon-counting APDs detect light reemitted from the tissue, which simplifies detector calibration and enhances detection sensitivity. The resulting system is compact, cheap, and easy to manufacture.

The arm cuff experiments showed that the device can measure the HbT, StO2, and BFI simultaneously. The relative changes in the StO2, HbT, and BFI agree well with previous studies.20,21 The quantitatively derived StO2, HbT, and BFI parameters are reasonable.12,24 The results also show that quantitative hybrid diffuse optical measurements can distinguish ischemic tissue, and therefore that the proposed technique can be used to diagnose a stroke at an emergency site. However, further improvements are needed. First, shorter wavelengths that are sensitive to deoxy-hemoglobin should be added to improve the fitting accuracy since the μeff values at 780 nm, 808 nm, and 830 nm are similar. Second, a quantitative brain reconstruction model should be developed. There is software available that may help to achieve this goal.26–28 

In conclusion, we have built a hybrid diffuse optical device that combines NIR-DOS and DCS to noninvasively and simultaneously measure the total hemoglobin, oxygen saturation, and tissue blood flow. The device was demonstrated by measuring hemodynamic parameters during an arm cuff experiment. Since strokes are related to abnormal hemodynamic changes in cerebral vasculature and our device is sensitive to such changes, portable, and potentially easy to apply without extended technical expertise, we believe that it has potential clinical applications in stroke diagnosis at emergency sites.

This work is funded by the National Key Technology R&D Program of China (2017YFC1200400, 2014BA104B01, 2014BA104B05, 2015BA101B01) and the National Natural Science Foundation of China (11604316). We thank AIP Author Services for its linguistic assistance during the preparation of this manuscript.

1.
B.
Ovbiagele
,
L. B.
Goldstein
,
R. T.
Higashida
,
V. J.
Howard
,
S. C.
Johnston
,
O. A.
Khavjou
,
D. T.
Lackland
,
J. H.
Lichtman
,
S.
Mohl
,
R. L.
Sacco
,
J. L.
Saver
, and
J. G.
Trogdon
, “
Forecasting the future of stroke in the United States: A policy statement from the American Heart Association and American Stroke Association
,”
Stroke
44
,
2361
2375
(
2013
).
2.
L. R.
Caplan
,
D. B.
Hier
, and
I.
D’Cruz
, “
Cerebral embolism in the Michael Reese Registry
,”
Stroke
14
,
530
540
(
1983
).
3.
The National Institute of Neurological Disorders and Stroke rt-PA Stroke Study Group
, “
Tissue plasminogen activator for acute ischemic stroke
,”
N Engl J Med
333
,
1581
1588
(
1995
).
4.
H. P.
Adams
, Jr.
,
G.
del Zoppo
,
M. J.
Alberts
et al, “
Guidelines for the early management of adults with ischemic stroke: A guideline from the American Heart Association/American Stroke Association Stroke Council, Clinical Cardiology Council, Cardiovascular Radiology and Intervention Council, and the Atherosclerotic Peripheral Vascular Disease and Quality of Care Outcomes in Research Interdisciplinary Working Groups: The American Academy of Neurology affirms the value of this guideline as an educational tool for neurologists
,”
Circulation
115
,
e478
e534
(
2007
).
5.
European Stroke Organisation (ESO) Executive Committee
and
ESO Writing Committee
, “
Guidelines for management of ischaemic stroke and transient ischaemic attack 2008
,”
Cerebrovasc Dis
25
,
457
507
(
2008
).
6.
G. W.
Albers
and
J. M.
Olivot
, “
Intravenous alteplase for ischaemic stroke
,”
Lancet
369
,
249
250
(
2007
).
7.
California Acute Stroke Pilot Registry (CASPR) Investigators
, “
Prioritizing interventions to improve rates of thrombolysis for ischemic stroke
,”
Neurology
64
,
654
659
(
2005
).
8.
S.
Walter
,
P.
Kostopoulos
,
A.
Haass
,
I.
Keller
,
M.
Lesmeister
,
T.
Schlechtriemen
et al, “
Diagnosis and treatment of patients with stroke in a mobile stroke unit versus in hospital: A randomised controlled trial
,”
Lancet Neurol.
11
,
397
404
(
2012
).
9.
M.
Ebinger
,
B.
Winter
,
M.
Wendt
,
J. E.
Weber
,
C.
Waldschmidt
, and
M.
Rozanski
, “
STEMO Consortium. Effect of the use of ambulance-based thrombolysis on time to thrombolysis in acute ischemic stroke: A randomized clinical trial
,”
JAMA.
311
,
1622
1631
(
2014
).
10.
F.
Scholkmann
,
S.
Kleiser
,
A. J.
Metz
,
R.
Zimmermann
,
P. J.
Mata
,
U.
Wolf
, and
M.
Wolf
, “
A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology
,”
NeuroImage
85
,
6
27
(
2014
).
11.
S.
Suzuki
,
S.
Takasaki
,
T.
Ozaki
, and
Y.
Kobayashi
, “
Tissue oxygenation monitor using NIR spatially resolved spectroscopy
,”
Proc. SPIE
3597
,
582
592
(
1999
).
12.
R. M. P.
Doornbos
,
R.
Lang
,
M. C.
Aalders
,
F. W.
Cross
, and
H. J. C. M.
Sterenborg
, “
The determination of in vivo human tissue optical properties and absolute chromophore concentrations using spatially resolved steady-state diffuse reflectance spectroscopy
,”
Phys. Med. Biol.
44
,
967
981
(
1999
).
13.
T.
Durduran
,
R.
Choe
,
W. B.
Baker
, and
A. G.
Yodh
, “
Diffuse optics for tissue monitoring and tomography
,”
Rep. Prog. Phys.
73
,
076701
(
2010
).
14.
A.
Torricelli
,
D.
Contini
,
A.
Pifferi
,
M.
Caffini
,
R.
Re
,
L.
Zucchelli
, and
L.
Spinellic
, “
Time domain functional NIRS imaging for human brain mapping
,”
NeuroImage
85
,
28
50
(
2014
).
15.
T.
Durduran
and
A. G.
Yodh
, “
Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement
,”
NeuroImage
85
,
51
63
(
2014
).
16.
T.
Durduran
,
G.
Yu
,
M. G.
Burnett
,
J. A.
Detre
,
J. H.
Greenberg
,
J.
Wang
,
C.
Zhou
, and
A. G.
Yodh
, “
Diffuse optical measurement of blood flow, tissue oxygen saturation, and metabolism in a human brain during sensorimotor cortex activation
,”
Opt. Lett.
29
,
1766
1768
(
2004
).
17.
N.
Roche-Labarbe
,
A.
Fenoglio
,
A.
Aggarwal
,
M.
Dehaes
,
S. A.
Carp
,
M. A.
Franceschini
, and
P. E.
Grant
, “
Near-infrared spectroscopy assessment of cerebral oxygen metabolism in the developing premature brain
,”
Journal of Cerebral Blood Flow & Metabolism
32
,
481
488
(
2012
).
18.
Y.
Shang
,
Y.
Zhao
,
R.
Cheng
,
L.
Dong
,
D.
Irwin
, and
G.
Yu
, “
Portable optical tissue flow oximeter based on diffuse correlation spectroscopy
,”
Opt. Lett.
34
,
3556
3558
(
2009
).
19.
D.
Tamborini
,
P.
Farzam
,
B.
Zimmermann
,
K. C.
Wu
,
D. A.
Boas
, and
M. A.
Franceschini
, “
Development and characterization of a multidistance and multiwavelength diffuse correlation spectroscopy system
,”
Neurophotonics
5
(
1
),
011015
(
2018
).
20.
D.
Wang
,
A. B.
Parthasarathy
,
W. B.
Baker
,
K.
Gannon
,
V.
Kavuri
,
T.
Ko
,
S.
Schenkel
,
Z.
Li
,
Z.
Li
,
M. T.
Mullen
,
J. A.
Detre
, and
A. G.
Yodh
, “
Fast blood flow monitoring in deep tissues with real-time software correlators
,”
Biomed. Opt. Express
7
(
3
),
776
797
(
2016
).
21.
N. B.
Hampson
and
C. A.
Piantadosi
, “
Near infrared monitoring of human skeletal muscle oxygenation during forearm ischemia
,”
Journal of Applied Physiology
64
(
6
),
2449
2457
(
1988
).
22.
D.
Wang
,
P.
Gao
,
L.
Zhu
,
Q.
Peng
,
Z.
Li
, and
J.
Zhao
, “
Optimization of detected optical intensity for measurement of diffuse correlation spectroscopy: Intralipid phantom study
,”
AIP Advances
9
,
015315
(
2019
).
24.
S. J.
Matcher
,
M.
Cope
, and
D. T.
Delpy
, “
In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and 900 nm measured with time-resolved spectroscopy
,”
Applied Optics
36
(
1
),
386
396
(
1997
).
25.
R. C.
Mesquita
,
T.
Durduran
,
G.
Yu
,
E. M.
Buckley
,
M. N.
Kim
,
C.
Zhou
,
R.
Choe
,
U.
Sunar
, and
A. G.
Yodh
, “
Direct measurement of tissue blood flow and metabolism with diffuse optics
,”
Philosophical Transactions of the Royal Society A
369
,
4358
4379
(
2011
).
26.
M.
Jermyn
,
H.
Ghadyani
,
M. A.
Mastanduno
,
W.
Turner
,
S. C.
Davis
,
H.
Dehghani
, and
B. W.
Pogue
, “
Fast segmentation and high-quality three-dimensional volume mesh creation from medical images for diffuse optical tomography
,”
J. Biomed. Opt.
18
(
8
),
086007
(
2013
).
27.
H.
Dehghani
,
M. E.
Eames
,
P. K.
Yalavarthy
,
S. C.
Davis
,
S.
Srinivasan
,
C. M.
Carpenter
,
B. W.
Pogue
, and
K. D.
Paulsen
, “
Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstruction
,”
Communications in Numerical Methods in Engineering
25
,
711
732
(
2009
).
28.
M.
Schweiger
and
S. R.
Arridge
, “
The Toast++ software suite for forward and inverse modeling in optical tomography
,”
J. Biomed. Opt.
19
(
4
),
040801
(
2014
).