Monte Carlo is a method that widely used for calculation of particle transport in radiotherapy dose distribution. This study was devoted to develop a linear accelerator head geometry model (LINAC) using GEANT4 simulation for 6 MeV photon beam. The geometric model of the accelerator head consists of electron sources, target, primary collimator, flattening filter, jaws, and MLC (multi-leaf collimator). The homogeneous water phantom size of 40 × 40 × 40 cm3 was used in the simulation, 100 cm SSD (source-skin distance), and field size are 5 cm × 5 cm, 10 cm × 10 cm, 20 cm × 20 cm, and 30 cm × 30 cm. The simulation results show the peak curve energy positions are in the range 0.3 MeV – 0.4 MeV and the maximum absorbed energy is about 5.9 MeV. According to the landau curve fitting the mean energy is about 0.385 MeV. The results of GEANT4 simulation show the energy spectrum has the same pattern as an energy spectrum from several experiments of LINAC head geometry. The simulation results also show the dose distribution based on beam profile and depth profile of water phantom with different fields.
Skip Nav Destination
Article navigation
7 November 2019
THE 8TH NATIONAL PHYSICS SEMINAR 2019
29–30 June 2019
Jakarta, Indonesia
Research Article|
November 07 2019
Distributions dose analysis for 6 MV photon beams using Monte Carlo-GEANT4 simulation
Nurul Qomariyah;
Nurul Qomariyah
b)
1
Department of Physics, Faculty of Mathematics and Natural Sciences, UniversitasMataram
, Jl. Majapahit No. 62 83125 Mataram, Indonesia
Search for other works by this author on:
Rahadi Wirawan;
Rahadi Wirawan
a)
1
Department of Physics, Faculty of Mathematics and Natural Sciences, UniversitasMataram
, Jl. Majapahit No. 62 83125 Mataram, Indonesia
a)Corresponding author: rwirawan@unram.ac.id
Search for other works by this author on:
Laili Mardiana;
Laili Mardiana
1
Department of Physics, Faculty of Mathematics and Natural Sciences, UniversitasMataram
, Jl. Majapahit No. 62 83125 Mataram, Indonesia
Search for other works by this author on:
Kasnawi Al Hadi
Kasnawi Al Hadi
1
Department of Physics, Faculty of Mathematics and Natural Sciences, UniversitasMataram
, Jl. Majapahit No. 62 83125 Mataram, Indonesia
Search for other works by this author on:
a)Corresponding author: rwirawan@unram.ac.id
AIP Conf. Proc. 2169, 030006 (2019)
Citation
Nurul Qomariyah, Rahadi Wirawan, Laili Mardiana, Kasnawi Al Hadi; Distributions dose analysis for 6 MV photon beams using Monte Carlo-GEANT4 simulation. AIP Conf. Proc. 7 November 2019; 2169 (1): 030006. https://doi.org/10.1063/1.5132656
Download citation file:
Citing articles via
Related Content
Radiation shielding potential analyze of iron sand-pumice composite based on GEANT4 simulations
AIP Conference Proceedings (April 2023)
Monte Carlo modeling of the head of Varian Clinac 2100C linear accelerator operating in 18 MV utilizing GATE/GEANT4
AIP Conference Proceedings (August 2018)
Introduction to the Geant4 Simulation toolkit
AIP Conference Proceedings (May 2011)
Dose profile measurement using an imaging plate: Evaluation of filters using Monte Carlo simulation of 4 MV x-rays
Rev. Sci. Instrum. (April 2009)
GEANT4 Tuning For pCT Development
AIP Conference Proceedings (August 2011)