Recent results from simultaneous long-term stability measurements of the beam position and photon energy of the synchrotron beam at the soft X-ray beamline P04 at PETRA III (DESY, Hamburg) are presented. The data was obtained by operating the built-in, non-invasive diagnostic unit directly upstream the beamline focus in parallel with a user experiment. We demonstrate that a sophisticated data evaluation based on principal component analysis results a high positional accuracy better than a few micron and a stable photon energy measurement with an uncertainty in the low ppm regime. Even subtle effects caused by abnormalities in the electron storage ring operation (e.g. short interruptions of the electron injection) are revealed hereby. Utilizing online diagnostic data for feedback controls will enable unprecedented stability of beamline operation at synchrotrons and FEL sources in the future.
Skip Nav Destination
Article navigation
15 January 2019
PROCEEDINGS OF THE 13TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION – SRI2018
11–15 June 2018
Taipei, Taiwan
Research Article|
January 15 2019
Progress report on the XUV online diagnostic unit for the highly accurate determination of SR properties
Jens Buck;
Jens Buck
b)
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
3
IEAP, Christian-Albrechts-Universität zu Kiel
, Kiel, Germany
Search for other works by this author on:
Kai Bagschik;
Kai Bagschik
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
Search for other works by this author on:
Leif Glaser;
Leif Glaser
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
Search for other works by this author on:
Frank Scholz;
Frank Scholz
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
Search for other works by this author on:
Jörn Seltmann;
Jörn Seltmann
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
Search for other works by this author on:
Jens Viefhaus
Jens Viefhaus
a)
1
Helmholtz-Zentrum Berlin für Materialien und Energie GmbH
, Germany
2
Deutsches Elektronen-Synchrotron (DESY)
, Hamburg, Germany
a)Corresponding author: jens.viefhaus@helmholtz-berlin.de
Search for other works by this author on:
a)Corresponding author: jens.viefhaus@helmholtz-berlin.de
AIP Conf. Proc. 2054, 060057 (2019)
Citation
Jens Buck, Kai Bagschik, Leif Glaser, Frank Scholz, Jörn Seltmann, Jens Viefhaus; Progress report on the XUV online diagnostic unit for the highly accurate determination of SR properties. AIP Conf. Proc. 15 January 2019; 2054 (1): 060057. https://doi.org/10.1063/1.5084688
Download citation file:
Citing articles via
Related Content
New experimental perspectives for soft x-ray absorption spectroscopies at ultra-low temperatures below 50 mK and in high magnetic fields up to 7 T
Rev. Sci. Instrum. (April 2016)
An integrated ion trap for the photon–ion spectrometer at PETRA III
Rev. Sci. Instrum. (February 2023)
A full-field transmission x-ray microscope for time-resolved imaging of magnetic nanostructures
AIP Conference Proceedings (January 2016)
Design and test of a broadband split-and-delay unit for attosecond XUV-XUV pump-probe experiments
Rev. Sci. Instrum. (February 2016)
Generation of VUV/XUV laser light
AIP Conference Proceedings (April 1991)