Beamline P64 at PETRA III is dedicated to x-ray absorption spectroscopy experiments which require high flux, like quick extended x-ray absorption fine structure spectroscopy on the sub-second time scale, high-resolution resonant emission spectroscopy, and x-ray absorption fine-structure spectroscopy of highly diluted systems. The beamline is installed at the high-energy storage ring PETRA III. The source is a 2 m-long undulator. The beamline covers an energy-range from 4 keV to 44 keV, which was required by the user community. Mirrors can be used to reduce the intensity of higher harmonics, and to focus the beam. A conventional photo-diode and a 100-pixel high-purity Ge-detector are used in fluorescence extended x-ray absorption fine-structure spectroscopy for diluted samples.
Skip Nav Destination
Article navigation
15 January 2019
PROCEEDINGS OF THE 13TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION – SRI2018
11–15 June 2018
Taipei, Taiwan
Research Article|
January 15 2019
High-flux XAFS-beamline P64 at PETRA III
Wolfgang A. Caliebe;
Wolfgang A. Caliebe
a)
1
DESY, Notkestrasse 85
, 22603 Hamburg, Germany
a)Corresponding author: wolfgang.caliebe@desy.de
Search for other works by this author on:
Vadim Murzin;
Vadim Murzin
1
DESY, Notkestrasse 85
, 22603 Hamburg, Germany
Search for other works by this author on:
Aleksandr Kalinko;
Aleksandr Kalinko
1
DESY, Notkestrasse 85
, 22603 Hamburg, Germany
2
Paderborn University
, Warburger Str. 100, 33098 Paderborn, Germany
Search for other works by this author on:
Marcel Görlitz
Marcel Görlitz
1
DESY, Notkestrasse 85
, 22603 Hamburg, Germany
Search for other works by this author on:
a)Corresponding author: wolfgang.caliebe@desy.de
AIP Conf. Proc. 2054, 060031 (2019)
Citation
Wolfgang A. Caliebe, Vadim Murzin, Aleksandr Kalinko, Marcel Görlitz; High-flux XAFS-beamline P64 at PETRA III. AIP Conf. Proc. 15 January 2019; 2054 (1): 060031. https://doi.org/10.1063/1.5084662
Download citation file:
Citing articles via
Related Content
The quick EXAFS setup at beamline P64 at PETRA III for up to 200 spectra per second
AIP Conference Proceedings (January 2019)
The PETRA III extension
AIP Conference Proceedings (July 2016)
A beamline for bulk sample x-ray absorption spectroscopy at the high brilliance storage ring PETRA III
AIP Conference Proceedings (January 2019)
Properties of the insertion devices for PETRA III and its extension
AIP Conference Proceedings (July 2016)
Time-resolved pump and probe x-ray absorption fine structure spectroscopy at beamline P11 at PETRA III
Rev. Sci. Instrum. (May 2016)