The radio frequency ion source NIO1, jointly developed by Consorzio RFX and INFN-LNL, will generate a 60kV-135mA hydrogen negative ion beam, composed of 9 beamlets over an area of about 40 × 40 mm2. This experiment will operate in continuous mode and in conditions similar to those foreseen for the larger ion sources of the Neutral Beam Injectors for ITER. The modular design of NIO1 is convenient to address the several still open important issues related to beam extraction, optics, and performance optimization. To this purpose a set of diagnostics is being implemented. Electric and water cooling plant related measurements will allow monitoring current, pressure, flow, and temperature. The plasma in the source will be characterized by emission spectroscopy, cavity ring-down and laser absorption spectroscopy. The accelerated beam will be analyzed with a fast emittance scanner, its intensity profile and divergence with beam emission spectroscopy and visible tomography. The power distribution of the beam on the calorimeter will be monitored by thermocouples and by an infrared camera. This contribution presents the implementation and initial operation of some of these diagnostics in the commissioning phase of the experiment, in particular the cooling water calorimetry and emission spectroscopy.
Skip Nav Destination
,
,
,
,
,
,
Article navigation
8 April 2015
FOURTH INTERNATIONAL SYMPOSIUM ON NEGATIVE IONS, BEAMS AND SOURCES (NIBS 2014)
6–10 October 2014
Garching, Germany
Research Article|
April 08 2015
NIO1 diagnostics Free
B. Zaniol;
B. Zaniol
a)
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
Search for other works by this author on:
M. Barbisan;
M. Barbisan
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
Search for other works by this author on:
M. Cavenago;
M. Cavenago
2
INFN-LNL
, v.le dell'Università 2, I-35020, Legnaro (PD) Italy
Search for other works by this author on:
M. De Muri;
M. De Muri
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
2
INFN-LNL
, v.le dell'Università 2, I-35020, Legnaro (PD) Italy
Search for other works by this author on:
A. Mimo;
A. Mimo
3
Università degli Studi di Padova
, Via 8 Febbraio 2, I-35122 Padova, Italy
Search for other works by this author on:
R. Pasqualotto;
R. Pasqualotto
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
Search for other works by this author on:
G. Serianni
G. Serianni
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
Search for other works by this author on:
B. Zaniol
1,a)
M. Barbisan
1
M. Cavenago
2
M. De Muri
1,2
A. Mimo
3
R. Pasqualotto
1
G. Serianni
1
1
Consorzio RFX
, Corso Stati Uniti 4, I-35127 Padova, Italy
2
INFN-LNL
, v.le dell'Università 2, I-35020, Legnaro (PD) Italy
3
Università degli Studi di Padova
, Via 8 Febbraio 2, I-35122 Padova, Italy
a)
Corresponding author: [email protected].
AIP Conf. Proc. 1655, 060010 (2015)
Citation
B. Zaniol, M. Barbisan, M. Cavenago, M. De Muri, A. Mimo, R. Pasqualotto, G. Serianni; NIO1 diagnostics. AIP Conf. Proc. 8 April 2015; 1655 (1): 060010. https://doi.org/10.1063/1.4916479
Download citation file:
Citing articles via
The implementation of reflective assessment using Gibbs’ reflective cycle in assessing students’ writing skill
Lala Nurlatifah, Pupung Purnawarman, et al.
Effect of coupling agent type on the self-cleaning and anti-reflective behaviour of advance nanocoating for PV panels application
Taha Tareq Mohammed, Hadia Kadhim Judran, et al.
Design of a 100 MW solar power plant on wetland in Bangladesh
Apu Kowsar, Sumon Chandra Debnath, et al.
Related Content
Modeling and design of a beam emission spectroscopy diagnostic for the negative ion source NIO1
Rev. Sci. Instrum. (October 2013)
Acquisition, data retrieval, interlock and control systems for the negative ion source NIO1
AIP Conf. Proc. (August 2017)
First experiments with the negative ion source NIO1
Rev. Sci. Instrum. (November 2015)
Electron density and temperature in NIO1 RF source operated in oxygen and argon
AIP Conf. Proc. (August 2017)
A first characterization of the NIO1 particle beam by means of a diagnostic calorimeter
AIP Conf. Proc. (August 2017)