Cryogenic temperature sensors are not usually thought of as electrostatic discharge (ESD) sensitive devices. However, the most common cryogenic thermometers in use today are thermally sensitive diodes or resistors - both electronic devices in their base form. As such, they are sensitive to ESD at some level above which either catastrophic or latent damage can occur. Instituting an ESD program for safe handling and installation of the sensor is costly and it is desirable to balance the risk of ESD damage against this cost. However, this risk cannot be evaluated without specific knowledge of the ESD vulnerability of the devices in question. This work examines three types of cryogenic temperature sensors for ESD sensitivity - silicon diodes, Cernox{trade mark, serif} resistors, and wire wound platinum resistors, all manufactured by Lake Shore Cryotronics, Inc. Testing was performed per TIA/EIA FOTP129 (Human Body Model). Damage was found to occur in the silicon diode sensors at discharge levels of 1,500 V. For Cernox{trade mark, serif} temperature sensors, damage was observed at 3,500 V. The platinum temperature sensors were not damaged by ESD exposure levels of 9,900 V. At the lower damage limit, both the silicon diode and the Cernox{trade mark, serif} temperature sensors showed relatively small calibration shifts of 1 to 3 K at room temperature. The diode sensors were stable with time and thermal cycling, but the long term stability of the Cernox{trade mark, serif} sensors was degraded. Catastrophic failure occurred at higher levels of ESD exposure.
Skip Nav Destination
,
Article navigation
29 January 2014
ADVANCES IN CRYOGENIC ENGINEERING: Transactions of the Cryogenic Engineering Conference - CEC
17–21 June 2013
Anchorage, Alaska, USA
Research Article|
January 29 2014
Effects of electrostatic discharge on three cryogenic temperature sensor models Free
S. Scott Courts;
S. Scott Courts
Lake Shore Cryotronics, 575 McCorkle Blvd., Westerville, OH 43082,
USA
Search for other works by this author on:
Thomas B. Mott
Thomas B. Mott
Lake Shore Cryotronics, 575 McCorkle Blvd., Westerville, OH 43082,
USA
Search for other works by this author on:
S. Scott Courts
Thomas B. Mott
Lake Shore Cryotronics, 575 McCorkle Blvd., Westerville, OH 43082,
USA
AIP Conf. Proc. 1573, 118–125 (2014)
Citation
S. Scott Courts, Thomas B. Mott; Effects of electrostatic discharge on three cryogenic temperature sensor models. AIP Conf. Proc. 29 January 2014; 1573 (1): 118–125. https://doi.org/10.1063/1.4860691
Download citation file:
1,093
Views
Citing articles via
The implementation of reflective assessment using Gibbs’ reflective cycle in assessing students’ writing skill
Lala Nurlatifah, Pupung Purnawarman, et al.
Classification data mining with Laplacian Smoothing on Naïve Bayes method
Ananda P. Noto, Dewi R. S. Saputro
Effect of coupling agent type on the self-cleaning and anti-reflective behaviour of advance nanocoating for PV panels application
Taha Tareq Mohammed, Hadia Kadhim Judran, et al.
Related Content
A comparison of GRT and CxRT cryogenic temperature sensors over the 1 K to 27 K temperature range
AIP Conf. Proc. (October 2024)
Review of Cernox™ (Zirconium Oxy‐Nitride) Thin‐Film Resistance Temperature Sensors
AIP Conf. Proc. (September 2003)
Cernox thermometer under hydrostatic pressure for enhanced temperature accuracy
Rev. Sci. Instrum. (April 2023)
Gamma Radiation Induced Calibration Shift for Four Cryogenic Thermometer Types
AIP Conf. Proc. (June 2004)
Effects of Packaging on Thermal Resistance below 1 K for Cryogenic Temperature Sensors
AIP Conf. Proc. (June 2004)