A high resolution Beam Position Monitor (BPM) is necessary for the beam‐based alignment systems of high energy and low emittance electron linacs. Such a monitor is developed in the framework of the European CARE/SRF programme, in a close collaboration between DESY and CEA/DSM/DAPNIA. This monitor is a radiofrequency re‐entrant cavity, which can be used either at room or cryogenic temperature, in an environment where dust particle contamination has to be avoided, such as superconducting cavities in a cryomodule. A first prototype of a re‐entrant BPM has already delivered measurements at 2K. inside the first cryomodule (ACC1) on the TESLA Test Facility 2 (TTF2). The performances of this BPM are analyzed both experimentally and theoretically, and the limitations of this existing system clearly identified. A new cavity and new electronics have been designed in order to improve the position resolution down to 1 μm and the damping time down to 10 ns.
Skip Nav Destination
Article navigation
20 November 2006
BEAM INSTRUMENTATION WORKSHOP 2006: Twelfth Beam Instrumentation Workshop
1-4 May 2006
Batavia, Illinois (USA)
Research Article|
November 20 2006
High Resolution BPM for Linear Colliders
C. Simon;
C. Simon
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
S. Chel;
S. Chel
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
M. Luong;
M. Luong
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
O. Napoly;
O. Napoly
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
J. Novo;
J. Novo
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
D. Roudier;
D. Roudier
*Centre à l’Energie Atomique/Saclay, DSM/DAPNIA/SACM, F91191 Gif‐sur‐Yvette Cedex France
Search for other works by this author on:
N. Baboi;
N. Baboi
†Deutsches Elektronen Synchroton, Notkestasse 85, D‐22603 Hamburg, Germany
Search for other works by this author on:
D. Noelle;
D. Noelle
†Deutsches Elektronen Synchroton, Notkestasse 85, D‐22603 Hamburg, Germany
Search for other works by this author on:
N. Mildner;
N. Mildner
†Deutsches Elektronen Synchroton, Notkestasse 85, D‐22603 Hamburg, Germany
Search for other works by this author on:
K. Zapfe;
K. Zapfe
†Deutsches Elektronen Synchroton, Notkestasse 85, D‐22603 Hamburg, Germany
Search for other works by this author on:
N. Rouvière
N. Rouvière
§Centre National de Recherche Scientifique, IN2P3 — IPN/Orsay, F91406 Orsay Cedex France
Search for other works by this author on:
AIP Conf. Proc. 868, 488–496 (2006)
Citation
C. Simon, S. Chel, M. Luong, O. Napoly, J. Novo, D. Roudier, N. Baboi, D. Noelle, N. Mildner, K. Zapfe, N. Rouvière; High Resolution BPM for Linear Colliders. AIP Conf. Proc. 20 November 2006; 868 (1): 488–496. https://doi.org/10.1063/1.2401439
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
11
Views
Citing articles via
Inkjet- and flextrail-printing of silicon polymer-based inks for local passivating contacts
Zohreh Kiaee, Andreas Lösel, et al.
Effect of coupling agent type on the self-cleaning and anti-reflective behaviour of advance nanocoating for PV panels application
Taha Tareq Mohammed, Hadia Kadhim Judran, et al.
Students’ mathematical conceptual understanding: What happens to proficient students?
Dian Putri Novita Ningrum, Budi Usodo, et al.
Related Content
Electronics and Algorithms for HOM Based Beam Diagnostics
AIP Conference Proceedings (November 2006)
BPMs with Precise Alignment for TTF2
AIP Conference Proceedings (November 2004)
The ESS elliptical cavity cryomodules
AIP Conference Proceedings (January 2014)
The ESS spoke cavity cryomodules
AIP Conference Proceedings (January 2014)
Cryogenic test of the 4 K / 2 K insert for the ARIEL e-Linac cryomodule
AIP Conference Proceedings (January 2014)