An apparatus for realizing the 3He vapour‐pressure temperature scale has been constructed at NMIJ/AIST. The apparatus is designed to be operated for a long time by adopting a continuously operating 3He cryostat. The cryostat has an experimental access port, which is a top‐loading type. The diameter of the access port is 50 mm. An insert for 3He vapour‐pressure measurements is placed in the access port. The preliminary measurement results confirm that the cryostat system stabilizes the temperature of a 3He vapour‐pressure cell to better than 50 μK over several hours over the temperatures range of the 3He vapour‐pressure scale. The 3He vapour pressure is measured with a diaphragm differential pressure gauge, which has a silicon‐resonant sensor. The uncertainties of this system are estimated in the case of calibrating a rhodium‐iron resistance thermometer measured by an automatic DC bridge.