Cryomech, Inc. has continued the development of two-stage 4 K pulse tube cryorefrigerators. To address the concerns of customers with vibration levels due to the stretching in the tubes, a vibration elimination apparatus reduced the maximum displacement from 25 μm to <3 μm. The PT405, 0.5 W @4 K, was modified into a model PT407, by using the CP970 Compressor Package increasing the input power from 4.7 to 7.2 kW. It provides 0.72 W @4.2 K with 34 W @55 K. A new 4 K model, PT410, has been developed with cooling capacities of 0.83 W @4.2 K with 38 W @45 K for 8.0 kW power input. The PT410 is driven by one CP970 Compressor Package. With changing the flow impedances, the PT410 also can provide 1.05 W @4.2 K and simultaneously 28 W @56 K again for 8.0 kW of input power.
Skip Nav Destination
Article navigation
10 May 2002
ADVANCES IN CRYOGENIC ENGINEERING: Proceedings of the Cryogenic Engineering Conference - CEC
16-20 July 2001
Madison, Wisconsin (USA)
Research Article|
May 10 2002
Development of 4 K pulse tube cryorefrigerators at cryomech Available to Purchase
P. E. Gifford
P. E. Gifford
Cryomech, Inc., Syracuse, New York 13211
Search for other works by this author on:
C. Wang
Cryomech, Inc., Syracuse, New York 13211
P. E. Gifford
Cryomech, Inc., Syracuse, New York 13211
AIP Conf. Proc. 613, 641–648 (2002)
Citation
C. Wang, P. E. Gifford; Development of 4 K pulse tube cryorefrigerators at cryomech. AIP Conf. Proc. 10 May 2002; 613 (1): 641–648. https://doi.org/10.1063/1.1472076
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Citing articles via
The implementation of reflective assessment using Gibbs’ reflective cycle in assessing students’ writing skill
Lala Nurlatifah, Pupung Purnawarman, et al.
Effect of coupling agent type on the self-cleaning and anti-reflective behaviour of advance nanocoating for PV panels application
Taha Tareq Mohammed, Hadia Kadhim Judran, et al.
Design of a 100 MW solar power plant on wetland in Bangladesh
Apu Kowsar, Sumon Chandra Debnath, et al.
Related Content
Small Helium Liquefiers Using 4 K Pulse Tube Cryocoolers
AIP Conf. Proc. (April 2006)
A vibration free closed-cycle 1 K cryostat with a 4 K pulse tube cryocooler
AIP Conf. Proc. (January 2014)
Cryogen‐Free Dilution Refrigerator Precooled by a Pulse‐Tube Refrigerator with Non‐Magnetic Regenerator
AIP Conf. Proc. (April 2006)
PERFORMANCE IMPROVEMENT OF A SINGLE STAGE GM CRYOCOOLER AT 25 K
AIP Conf. Proc. (March 2008)
High capacity 30 K remote helium cooling loop
AIP Conf. Proc. (January 2014)