The design and fabrication of a cost-effective vibrating sample magnetometer are explained. Improvement in the detected signal can be obtained using an operational amplifier circuit. The fabricated spiral detection coil design enhances the induced voltage obtained as shielding flux is reduced. A homemade setup is obtained by taking a woofer as an actuator and plexiglass as the vibrating medium. Lock-in amplifier analyzed the booster enhanced induced voltage signal by the principle of phase detection. Study of amplified and non amplified signals from the Nickel (99% purity) sample was done. Conversion of induced voltage to magnetization was done by calibration incorporating the coercivity and retentivity measurement. Magnetic oxide can be analyzed using this cost-effective designed Vibrating sample magnetometer. The data obtained from the VSM can conclude the successful operation of the designed magnetometer.
Skip Nav Destination
Article navigation
10 December 2020
INTERNATIONAL CONFERENCE ON SMART SUSTAINABLE MATERIALS AND TECHNOLOGIES: ICSSMT-2020
12–12 August 2020
Madurai, India
Research Article|
December 10 2020
Fabrication of economical signal amplified vibrating sample magnetometer with spiral designed detector
Aditya Dash;
Aditya Dash
a)
1
Department of Physics and Astronomy, National Institute of Technology Rourkela
, Odisha, India
Search for other works by this author on:
Rabindra Sarkar;
Rabindra Sarkar
b)
1
Department of Physics and Astronomy, National Institute of Technology Rourkela
, Odisha, India
Search for other works by this author on:
Abel Mathew;
Abel Mathew
c)
1
Department of Physics and Astronomy, National Institute of Technology Rourkela
, Odisha, India
Search for other works by this author on:
Prakash Nath Vishwakarma
Prakash Nath Vishwakarma
d)*
1
Department of Physics and Astronomy, National Institute of Technology Rourkela
, Odisha, India
d)*Corresponding Author: prakashn@nitrkl.ac.in
Search for other works by this author on:
d)*Corresponding Author: prakashn@nitrkl.ac.in
AIP Conf. Proc. 2297, 020035 (2020)
Citation
Aditya Dash, Rabindra Sarkar, Abel Mathew, Prakash Nath Vishwakarma; Fabrication of economical signal amplified vibrating sample magnetometer with spiral designed detector. AIP Conf. Proc. 10 December 2020; 2297 (1): 020035. https://doi.org/10.1063/5.0030132
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
70
Views
Citing articles via
Design of a 100 MW solar power plant on wetland in Bangladesh
Apu Kowsar, Sumon Chandra Debnath, et al.
Social mediated crisis communication model: A solution for social media crisis?
S. N. A. Hamid, N. Ahmad, et al.
The effect of a balanced diet on improving the quality of life in malignant neoplasms
Yu. N. Melikova, A. S. Kuryndina, et al.
Related Content
Development of Hi-Fi microspeakers with a woofer and a tweeter used for mobile phones
J. Appl. Phys. (May 2005)
Development of a pair of loudspeaker enclosures.
J Acoust Soc Am (October 2010)
Development of panel loudspeaker system: Design, evaluation and enhancement
J Acoust Soc Am (June 2001)
Comparison of linear and nonlinear experiments for landmine detection
J Acoust Soc Am (November 2001)
Expressively actuated percussion instruments and interfaces
J Acoust Soc Am (November 2013)