PRISM is one of non-commercial Treatment Planning System (TPS) and is developed at the University of Washington. In Indonesia, many cancer hospitals use expensive commercial TPS. This study aims to investigate Prism TPS which been applied to the dose distribution of brachytherapy by taking into account the effect of source position and inhomogeneities. The results will be applicable for clinical Treatment Planning System.

Dose calculation has been implemented for water phantom and CT scan images of breast cancer using point source and line source. This study used point source and line source and divided into two cases. On the first case, Ir-192 seed source is located at the center of treatment volume. On the second case, the source position is gradually changed. The dose calculation of every case performed on a homogeneous and inhomogeneous phantom with dimension 20 × 20 × 20 cm3. The inhomogeneous phantom has inhomogeneities volume 2 × 2 × 2 cm3. The results of dose calculations using PRISM TPS were compared to literature data.

From the calculation of PRISM TPS, dose rates show good agreement with Plato TPS and other study as published by Ramdhani. No deviations greater than ±4% for all case. Dose calculation in inhomogeneous and homogenous cases show similar result. This results indicate that Prism TPS is good in dose calculation of brachytherapy but not sensitive for inhomogeneities. Thus, the dose calculation parameters developed in this study were found to be applicable for clinical treatment planning of brachytherapy.

1.
M.
Haekal
, Analisa Penggunaan Piranti Lunak Treatment Planning System (TPS) Prism: Suatu Studi Kasus pada Water Phantom,
Bandung
:
Institut Teknologi Bandung
,
2012
.
2.
E.
Podgorsak
,
Radiation Oncology Physics: A Handbook for Teachers and Students
,
Vienna
:
International Atomic Energy Agency
,
2005
.
3.
F. M.
Khan
,
The Physics of Radiation Therapy
4th edition,
US
:
Lippincott Williams & Wilkins
,
2010
.
4.
M. J.
Rivard
and et al, “
Brachytherapy Dose Calculation Formalism, Dataset Evaluation, and Treatment Planning System Implementation
,”
2009
.
5.
S. Z.
Ramdhani
, Peritungan Laju Dosis Serap Untuk Simulasi Terapi Kanker Serviks dengan Metode High- and-Medium Dose Rates Berdasarkan TG-43U1,
Bandung
:
ITB
,
2011
.
This content is only available via PDF.
You do not currently have access to this content.