Most of the thermometers used for the dissemination of the International Temperature Scale of 1990 (ITS-90) to the industry are long-stem Standard Platinum Resistance Thermometers (SPRTs). In the low temperature range, they are generally used from the triple point of water (273.16 K) down to the triple point of argon (83.805 8 K). The lack of equipment for routinely calibrating thermometers below 83 K (except comparisons at the nitrogen boiling point) limits their use at lower temperatures. The LNE-Cnam has recently developed an innovative calorimeter to improve the calibration of long-stem SPRTs at the triple point of argon and to extend the calibration range down to the oxygen triple point (54.358 4 K). Results show that at the argon triple point further investigations are needed to ascertain a reduction of the calibration uncertainty, with respect to the previous system developed by the laboratory. At the oxygen triple point, uncertainties are below the millikelvin level. With this device, the LNE-Cnam acquires the possibility to improve its calibration uncertainties for long-stem SPRTs at the argon triple point, and extends its calibration capabilities down to the triple point of oxygen.
Skip Nav Destination
Article navigation
11 September 2013
TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOLUME 8: Proceedings of the Ninth International Temperature Symposium
19–23 March 2012
Los Angeles, California, USA
Research Article|
September 11 2013
Design and implementation of a dedicated calorimeter for long stem SPRTs calibrations at the argon and oxygen triple points Available to Purchase
I. Didi-Alaoui;
I. Didi-Alaoui
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
D. Fiorillo;
D. Fiorillo
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
A. Vergé;
A. Vergé
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
F. Sparasci;
F. Sparasci
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
D. Jouin;
D. Jouin
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
Y. Hermier
Y. Hermier
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Search for other works by this author on:
I. Didi-Alaoui
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
D. Fiorillo
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
A. Vergé
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
F. Sparasci
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
D. Jouin
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
Y. Hermier
Laboratoire Commun de Métrologie LNE-Cnam, 61 rue du Landy, 93210 La Plaine Saint Denis,
France
AIP Conf. Proc. 1552, 474–479 (2013)
Citation
I. Didi-Alaoui, D. Fiorillo, A. Vergé, F. Sparasci, D. Jouin, Y. Hermier; Design and implementation of a dedicated calorimeter for long stem SPRTs calibrations at the argon and oxygen triple points. AIP Conf. Proc. 11 September 2013; 1552 (1): 474–479. https://doi.org/10.1063/1.4819587
Download citation file:
Pay-Per-View Access
$40.00
Sign In
You could not be signed in. Please check your credentials and make sure you have an active account and try again.
Citing articles via
Effect of coupling agent type on the self-cleaning and anti-reflective behaviour of advance nanocoating for PV panels application
Taha Tareq Mohammed, Hadia Kadhim Judran, et al.
Design of a 100 MW solar power plant on wetland in Bangladesh
Apu Kowsar, Sumon Chandra Debnath, et al.
With synthetic data towards part recognition generalized beyond the training instances
Paul Koch, Marian Schlüter, et al.
Related Content
A new calorimeter for the simultaneous calibration of SPRTs and CSPRTs at the triple point of mercury
AIP Conf. Proc. (September 2013)
From ITS-90 to thermodynamic temperature: Hybrid CSPRT calibrations with LNE-Cnam acoustic gas thermometry
AIP Conf. Proc. (October 2024)
Development of an adiabatic calorimeter in the range 54K-273K in frame of a scientific collaboration LNE-NIS
AIP Conf. Proc. (September 2013)
Development of a quasi-adiabatic calorimeter for the determination of the water vapor pressure curve
Rev. Sci. Instrum. (July 2012)
Liquid‐Vapor Transitions in Mercury and Sodium Gas‐Controlled Heat‐Pipes
AIP Conf. Proc. (September 2003)