Observation of time-dependent luminescence from excited states with a wide range of lifetimes allows students to explore the connection between selection rules and transition rates. It is fairly simple to measure microsecond and longer lifetimes with equipment common to undergraduate programs, because the instrument response time of even modest bandwidth systems is insignificant on microsecond and longer time scales. The measurement of nanosecond lifetimes, however, is more challenging, because the instrument response time is comparable to the lifetimes being measured. In this case, the instrument temporal response must be deconvolved from the observed luminescence signals in order to extract the actual excited state lifetime. We describe a method for measuring nanosecond fluorescence lifetimes in the advanced undergraduate laboratory that uses real-time analog luminescence signals instead of traditional photon counting techniques. The detection electronics of this method are fairly simple, consisting of an oscilloscope monitoring the time-dependent output of an inexpensive silicon photomultiplier. We introduce a simple and transparent method for students to characterize the instrument response and deconvolve it from the observed luminescence signals, yielding measured nanosecond fluorescence lifetimes in good agreement with the corresponding literature values obtained by time-correlated single photon counting. The limitations of silicon photomultipliers for this method of measuring nanosecond lifetimes are discussed in detail. Application of this treatment to decay processes that are not single exponential is also discussed.

1.
Mark
Sulkes
and
Zoe
Sulkes
, “
Measurement of luminescence decays: High performance at low cost
,”
Am. J. Phys.
79
(
11
),
1104
1111
(
2011
).
2.
Compact nanosecond pulsed diode lasers, Thorlabs
<www.thorlabs.com>.
3.
Joseph R.
Lakowicz
,
Principles of Fluorescence Spectroscopy
, 3rd ed. (
Springer
,
New York
,
2006
), pp.
103
124
.
4.
Michael
Wahl
, Time-Correlated Single Photon Counting, PicoQuant GmbH Technical Note (
2014
).
5.
Graham R.
Fleming
,
Chemical Applications of Ultrafast Spectroscopy
(
Oxford University
,
New York
,
1986
), pp.
91
95
.
6.
Yi
Tian
,
Liping
Wei
, and
Derek
Ho
, “
A low-cost time-correlated single photon counting portable DNA analyzer
,”
Sensors
19
,
2838
2850
(
2019
).
7.
Jonathan
Bouchard
et al, “
A low-cost time-correlated single photon counting system for multiview time-domain diffuse optical tomography
,”
IEEE Trans. Instrum. Meas.
66
(
10
),
2505
2515
(
2017
).
8.

The NIM based components required for generating luminescence decay curves by TCSPC include a NIM bin with power supply, two wideband amplifiers (if detector output pulses require amplification), two constant fraction discriminators, a time-to-amplitude convertor, and a multichannel analyzer. These components can sometimes be found used at affordable prices; however, an anonymous reviewer of this work has indicated that such used legacy components are becoming hard to find.

9.

Prices obtained from four of the largest manufacturers of integrated TCSPC systems indicate that the least expensive units cost about $12,000. These systems integrate the individual components listed in Ref. 8 on a PC plug-in card, necessitating the added expense of installing an appropriate communications bus in the host PC. A smaller manufacturer (Marina Photonics, <www.photoncounting.net>) offers an integrated TCSPC system for less than $4000. We do not know how well any of these systems work with the inexpensive silicon photomultiplier used here, as the manufacturers of these systems only offer discrete avalanche photodiode modules or photomultiplier tubes as detectors. The TCSPC system that was sold by the manufacturer of the silicon photomultiplier used here has been discontinued.

10.
P. H.
Binh
,
V. D.
Trong
,
P.
Renucci
, and
X.
Marie
, “
A simple sub-nanosecond ultraviolet pulse generator with high repetition rate and peak power
,”
Rev. Sci. Instrum.
84
,
0831021
(
2013
).
11.
Martin
Rongen
and
Merlin
Schaufel
, “
Design and evaluation of a versatile picosecond light pulser
,”
J. Inst.
13
(
6
),
P060021
(
2018
).
12.
Wilfried
Uhring
,
Chantal-Virginie
Zint
, and
Jeremy
Bartringer
, “
A low-cost high-repetition-rate picosecond laser diode pulse generator
,”
Proc. SPIE
5452
,
583
590
(
2004
).
13.
V. N.
Rai
,
M.
Shukla
,
R. K.
Khardekar
, and
H. C.
Pant
, “
A picosecond optical pulse generator to calibrate the optical streak camera
,”
Rev. Sci. Instrum.
66
,
3125
3130
(
1995
).
14.
Mateusz
Żbik
and
Piotr Z.
Wieczorek
, “
Versatile subnanosecond laser diode driver
,”
Proc. SPIE
10031F
,
100311F
(
2016
).
15.
Mateusz
Żbik
and
Piotr Zbigniew
Wieczorek
, “
Charge-line dual-FET high-repetition-rate pulsed laser driver
,”
Appl. Sci.
9
(
7
),
1289
1300
(
2019
).
16.
Dušan
Ponikvar
, “
A simple subnanosecond light pulser
,”
IEEE Trans. Nucl. Sci.
59
(
6
),
3218
3220
(
2012
).
17.
ON Semiconductor
, C-series SiPM sensors, Rev. 6, Publication Order Number MICROC-SERIES/D (
2018
).
18.
SensL Technologies Ltd.
C-series low noise, fast, blue-sensitive silicon photomultipliers user manual, Rev. 2.0 (
2017
).
19.
ON Semiconductor
, Introduction to the silicon photomultiplier (SiPM), Rev. 7, Publication Order Number AND9770/D (
2018
).
20.
Sergei
Dolinsky
,
Geng
Fu
, and
Adrian
Ivan
, “
Timing resolution performance comparison of fast and standard outputs of SensL SiPM
,” in Proceedings of
2013 IEEE Nuclear Science Symposium and Medical Imaging Conference
, Seoul, South Korea (
2013
), pp.
1
6
.
21.
Reference 3, p. 43 and pp. 353–364.
22.
U.S. Department of Energy Laboratory Equipment Donation Program
<www.apps.orau.gov/ledp/>.
23.

The relationship between rise time τrise and the 3 dB bandwidth BW3 dB that is most often used, τrise = 0.35/BW3 dB, defines τrise as the time it takes the signal to rise from 10% to 90% of its peak value assuming the system can be modeled as a charging RC circuit with equivalent resistance R and equivalent capacitance C.

24.
Qiyin
Fang
et al, “
Time-domain laser-induced fluorescence spectroscopy apparatus for clinical diagnostics
,”
Rev. Sci. Instrum.
75
,
151
162
(
2004
). Fang et al. suggest (in pp. 161–162) that an analog system consisting of an oscilloscope, detector, and amplifier with rise times of 75, 230, and 230 ps, respectively, should be capable of measuring fluorescence lifetimes of a few hundred picoseconds with adequate deconvolution. Extrapolating, a system consisting of a 1 ns rise time oscilloscope and 300 ps rise time SiPM should be capable of measuring fluorescence lifetimes of a few ns. Some TCSPC practitioners anecdotally suggest that the time resolution of the system need only be smaller than the fluorescence lifetime. We have not found this claim explicitly stated in print; however, it is consistent with the above observation of Fang et al.
25.
Reference 3, pp. 55–56.
26.
Jose M.
Alvarez-Pez
,
Luis
Ballesteros
,
Eva
Talavera
, and
Juan
Yguerabide
, “
Fluorescein excited-state proton exchange reactions: Nanosecond emission kinetics and correlation with steady-state fluorescence intensity
,”
J. Phys. Chem. A
105
,
6320
6332
(
2001
).
27.

The absorber concentration c is estimated from the Beer-Lambert relationship, OD = εcL, where OD is the sample optical density, ε is the extinction coefficient of the absorber, and L is the path length through the sample.

28.
Abhinav K.
Jha
,
Herman T.
van Dam
,
Matthew A.
Kupinski
, and
Eric
Clarkson
, “
Simulating silicon photomultiplier response to scintillation light
,”
IEEE Trans. Nucl. Sci.
60
(
1
),
336
351
(
2013
).
29.

Unweighted fits are appropriate when the uncertainties of the individual points on the luminescence curves are not well characterized. However, fit parameter uncertainties reported by unweighted fitting routines can be unreliable precisely because they are estimated without knowledge of the individual point uncertainties.

30.
Richard Q.
Twiss
, “
Linear-circuit analysis and transient response
,” in
Massachusetts Institute of Technology Radiation Laboratory Series Volume 18
, edited by
George E.
Valley
, Jr.
and
Henry
Wallman
(
McGraw-Hill
,
New York
,
1948
), p.
77
.
31.
Reference 4, pp. 4–5.
32.
Reference 3, pp. 106–107.
33.

It can be shown that 0xfxxgxdx=0xfxgxxdx by a change of variables.

34.
Marcos R.
Ribas
,
Ronald P.
Steer
, and
Ricardo
Rüther
, “
Photophysical properties of new bis-perylene dyads for potential upconversion use
,”
Chem. Phys. Lett.
605–606
,
126
130
(
2014
).
35.
Érica
Torres
,
Mário Nuno
Berberan-Santos
, and
Maria João
Brites
, “
Synthesis, photophysical, and electrochemical properties of perylene dyes
,”
Dye. Pigm.
112
,
298
304
(
2015
).
36.
Lucia
Flamigni
, “
Inclusion of fluorescein and halogenated derivatives in α-, β-, and γ-cyclodextrins
,”
J. Phys. Chem.
97
,
9566
9572
(
1993
).
37.
E.
Martinenghi
et al, “
Time-resolved single-photon detection module based on silicon photomultiplier: A novel building block for time-correlated measurement systems
,”
Rev. Sci. Instrum.
87
(
7
),
073101-1
0731011-8
(
2016
).
38.
Jung Yeol
Yeom
et al, “
Fast timing silicon photomultipliers for scintillation detectors
,”
IEEE Photonics Technol. Lett.
25
(
14
),
1309
1312
(
2013
). The actual electron gain per avalanched microcell is 3 × 106 (Ref. 17). However, the data presented in Yeom et al. indicate that only about 2% of the total photogenerated charge flows to the fast output and the remainder flows to the anode.
39.

A laboratory class handout (available from the corresponding author) attributed to Montana State University Professor of Chemistry Patrik Callis describes such a numerical convolution method using Excel.

AAPT members receive access to the American Journal of Physics and The Physics Teacher as a member benefit. To learn more about this member benefit and becoming an AAPT member, visit the Joining AAPT page.